Model: CT 5300; Product Code (REF): 728285; Software Versions: 5.1.0.X & 5.1.1.X;
Class I - DangerousWhat Should You Do?
- Check if you have this product: Product Code (REF): 728285; UDI-DI: 00884838113237; Serial Numbers: 800031, 800073, 800070, 35064, 35056, 35043, 35029, 800084, 800083, 800076, 800063, 800058, 800048, 800040, 800039, 800038, 800034, 800033, 800028, 800026, 800021, 800018, 800015, 800011, 800010, 800078, 800071, 35078, 35068, 35048, 35036, 800023, 800008, 800004, 800001, 35073, 35069, 35067, 35058, 35051, 35044, 35034, 35031, 35027, 35019, 35015, 35009, 35001, 800065, 35062, 35042, 35040, 35022, 35016, 35059, 800081, 800062, 800003, 35081, 35065, 35061, 35057, 35053, 35052, 35045, 35041, 35035, 35008, 800005, 35006, 800041, 35080, 800056, 35070, 35033, 35021, 800079, 800077, 800069, 800055, 800054, 800053, 800052, 800007, 800002, 35079, 35071, 35063, 35055, 35046, 35017, 35011, 35007, 35002, 800075, 800045, 800044, 800043, 800029, 800027, 800025, 800024, 800016, 800013, 800009, 800051, 800042, 800061, 800046, 800030, 800019, 800014, 800006, 800074, 800066, 35054, 35018, 800060, 800059, 800072, 35032, 35023, 35003, 800080, 800047, 35039, 35038, 35004, 76001, 35077, 35066, 35060, 35030, 35025, 35013, 35012, 35010, 35037, 35082, 35072, 35005, 35026, 800057, 800067, 800037, 800036, 800032, 800020, 35076, 35075, 35074, 35050, 35049, 35047, 35028, 35024, 35020, 35014, 800050, 800049, 800082, 800064, 800022, 800017, 800012;
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- PHILIPS MEDICAL SYSTEMS
- Reason for Recall:
- Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM (Dose Modulation), the WED (Water Equivalent Diameter) value might set itself to zero, and this may lead to an insufficient dose setting after the surview. If operator misses the insufficient dose and the WED value in User Interface and continues with the subsequent helical/Axial scans, then the obtained images will be noisy due to low dose setting. Issue 3: Due to a software failure, the ECG wave file is not saved. As a result, the cardiac offline image reconstruction fails due to the missing ECG wave file, and the user might rescan the patient. Issue 4: This issue is software fault in which the patient orientation will be changed to NULL under specific random scenarios resulting in radiation being delivered from the wrong orientation (surview scan) or in the wrong position (clinical scan) and will generate incorrect and undiagnosable surview/clinical images. Patients may be rescanned for surview and/or clinical images. Issue 5: During scanning, the preview image display is not consistent, the images are not arriving at a fixed rate, skipping images from time to time during scanning. The obtained images might not support making the clinical diagnosis and the user might decide to re-scan the patient. Issue 6: If a user presses the left and middle/right mouse buttons together or afterwards in a short time, both commands are executed and an incorrect auto ROI is created. This ROI coordinates won t be assigned as the object is empty. As a result, the threshold of contrast level will not be reached in the UI and this will prevent the system from automatically triggering the subsequent clinical scan. There is potential safety risk identified for additionally tracker shots and/or rescan of a patient when the clinical scan is not triggered. This issue only affects the manual ROI process.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Model: CT 5300; Product Code (REF): 728285; Software Versions: 5.1.0.X & 5.1.1.X;
Product Codes/Lot Numbers:
Product Code (REF): 728285; UDI-DI: 00884838113237; Serial Numbers: 800031, 800073, 800070, 35064, 35056, 35043, 35029, 800084, 800083, 800076, 800063, 800058, 800048, 800040, 800039, 800038, 800034, 800033, 800028, 800026, 800021, 800018, 800015, 800011, 800010, 800078, 800071, 35078, 35068, 35048, 35036, 800023, 800008, 800004, 800001, 35073, 35069, 35067, 35058, 35051, 35044, 35034, 35031, 35027, 35019, 35015, 35009, 35001, 800065, 35062, 35042, 35040, 35022, 35016, 35059, 800081, 800062, 800003, 35081, 35065, 35061, 35057, 35053, 35052, 35045, 35041, 35035, 35008, 800005, 35006, 800041, 35080, 800056, 35070, 35033, 35021, 800079, 800077, 800069, 800055, 800054, 800053, 800052, 800007, 800002, 35079, 35071, 35063, 35055, 35046, 35017, 35011, 35007, 35002, 800075, 800045, 800044, 800043, 800029, 800027, 800025, 800024, 800016, 800013, 800009, 800051, 800042, 800061, 800046, 800030, 800019, 800014, 800006, 800074, 800066, 35054, 35018, 800060, 800059, 800072, 35032, 35023, 35003, 800080, 800047, 35039, 35038, 35004, 76001, 35077, 35066, 35060, 35030, 35025, 35013, 35012, 35010, 35037, 35082, 35072, 35005, 35026, 800057, 800067, 800037, 800036, 800032, 800020, 35076, 35075, 35074, 35050, 35049, 35047, 35028, 35024, 35020, 35014, 800050, 800049, 800082, 800064, 800022, 800017, 800012;
Distribution:
Distributed in: AL, AR, CA, CT, FL, GA, IL, IN, KY, MD, MI, OH, PA, PR, TN, TX, UT, VA
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0376-2026
Related Recalls
Philips Spectral CT on Rails. Model Number: 728334.
PHILIPS MEDICAL SYSTEMS
Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go. 2. Potential where the message of Previous Surview Exists Select Previous Surview? <Yes> or <No> should display but does not appear. 3. After performing Surview and planning the Brain Helical acquisition by setting Brain Area DoseRight Index to increase the dose, there is the potential that the Define Head Area option in the context menu is grayed out.
Model: Incisive CT for Brazil SKD; Product Code (REF): 728146; Software Versions: 5.1.0.X & 5.1.1.X;
PHILIPS MEDICAL SYSTEMS
Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM (Dose Modulation), the WED (Water Equivalent Diameter) value might set itself to zero, and this may lead to an insufficient dose setting after the surview. If operator misses the insufficient dose and the WED value in User Interface and continues with the subsequent helical/Axial scans, then the obtained images will be noisy due to low dose setting. Issue 3: Due to a software failure, the ECG wave file is not saved. As a result, the cardiac offline image reconstruction fails due to the missing ECG wave file, and the user might rescan the patient. Issue 4: This issue is software fault in which the patient orientation will be changed to NULL under specific random scenarios resulting in radiation being delivered from the wrong orientation (surview scan) or in the wrong position (clinical scan) and will generate incorrect and undiagnosable surview/clinical images. Patients may be rescanned for surview and/or clinical images. Issue 5: During scanning, the preview image display is not consistent, the images are not arriving at a fixed rate, skipping images from time to time during scanning. The obtained images might not support making the clinical diagnosis and the user might decide to re-scan the patient. Issue 6: If a user presses the left and middle/right mouse buttons together or afterwards in a short time, both commands are executed and an incorrect auto ROI is created. This ROI coordinates won t be assigned as the object is empty. As a result, the threshold of contrast level will not be reached in the UI and this will prevent the system from automatically triggering the subsequent clinical scan. There is potential safety risk identified for additionally tracker shots and/or rescan of a patient when the clinical scan is not triggered. This issue only affects the manual ROI process.
Model: Incisive CT; Product Code (REF): (1) 728143, (2) 728144; Software Versions: 5.1.0.X & 5.1.1.X;
PHILIPS MEDICAL SYSTEMS
Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM (Dose Modulation), the WED (Water Equivalent Diameter) value might set itself to zero, and this may lead to an insufficient dose setting after the surview. If operator misses the insufficient dose and the WED value in User Interface and continues with the subsequent helical/Axial scans, then the obtained images will be noisy due to low dose setting. Issue 3: Due to a software failure, the ECG wave file is not saved. As a result, the cardiac offline image reconstruction fails due to the missing ECG wave file, and the user might rescan the patient. Issue 4: This issue is software fault in which the patient orientation will be changed to NULL under specific random scenarios resulting in radiation being delivered from the wrong orientation (surview scan) or in the wrong position (clinical scan) and will generate incorrect and undiagnosable surview/clinical images. Patients may be rescanned for surview and/or clinical images. Issue 5: During scanning, the preview image display is not consistent, the images are not arriving at a fixed rate, skipping images from time to time during scanning. The obtained images might not support making the clinical diagnosis and the user might decide to re-scan the patient. Issue 6: If a user presses the left and middle/right mouse buttons together or afterwards in a short time, both commands are executed and an incorrect auto ROI is created. This ROI coordinates won t be assigned as the object is empty. As a result, the threshold of contrast level will not be reached in the UI and this will prevent the system from automatically triggering the subsequent clinical scan. There is potential safety risk identified for additionally tracker shots and/or rescan of a patient when the clinical scan is not triggered. This issue only affects the manual ROI process.