VOCSN Multi-Function Ventilators: VOCSN+Pro (V+O+C+S+N+Pro, English), REF: PRT-00490-001; V+Pro (V+Pro, English), REF: PRT-01185-000, PRT-01185-002
Class I - DangerousWhat Should You Do?
- Check if you have this product: REF/UDI-DI/Serial Numbers: PRT-00490-001/00855573007792/123627, 121892, 119639; PRT-01185-000/00855573007877/5038421, 116700, 117933, 5038112, 5038619, 5038717, 5038953, 5039086, 5039091, 5039361, 5039499, 5039503, 5039541, 5039636, 5039938, 5040019, 5040640, 5040762, 5041060, 5041078, 5041534, 5042558, 117344, 118464, 118509, 5039387; PRT-01185-002/00850018761154/5037655
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Ventec Life Systems, Inc.
- Reason for Recall:
- Multi-Function Ventilators were serviced using incorrect parts which have the potential to cause unexpected shutdown, or when using an active circuit could result in inaccurate tidal volume monitor, not triggering on patient efforts, less inspiratory volume, less inspiratory pressure, and less PEEP delivered.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
VOCSN Multi-Function Ventilators: VOCSN+Pro (V+O+C+S+N+Pro, English), REF: PRT-00490-001; V+Pro (V+Pro, English), REF: PRT-01185-000, PRT-01185-002
Product Codes/Lot Numbers:
REF/UDI-DI/Serial Numbers: PRT-00490-001/00855573007792/123627, 121892, 119639; PRT-01185-000/00855573007877/5038421, 116700, 117933, 5038112, 5038619, 5038717, 5038953, 5039086, 5039091, 5039361, 5039499, 5039503, 5039541, 5039636, 5039938, 5040019, 5040640, 5040762, 5041060, 5041078, 5041534, 5042558, 117344, 118464, 118509, 5039387; PRT-01185-002/00850018761154/5037655
Distribution:
Distributed in: US, CA, NY, MO
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1196-2025
Related Recalls
The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions or leaks of the breathing circuit which may cause serious injury or death. The risk of injury or death is increased for ventilator dependent patients.
Due to a component failure, ventilators may unexpectedly shut down or have a loss of blower function while in patient use. The shutdown potentially prevents the ventilator from providing ventilation therapy.