Ventec Life Systems VOCSN Multi-Function Ventilator Model/Version/Catalog Number: V+O+C+S+N+Pro/English/PRT-00490-001; V+O+C+S+N+Pro/Japan/PRT-00490-100; V+C+Pro/English/PRT-00900-001; V+C+Pro/Japan/PRT-00900-100; V+C/English/PRT-01098-000; V+O+C+S+N/English/PRT-01100-000; V+O+C+S+N/English/PRT-01100-001;

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Catalog Number: PRT-00490-001 GTIN (Primary ID): 00855573007792/00855573007006 Serial Numbers: 114303; 114304; 113728; 113729; 113697; 114296; 113587; 113684; 113706; 113003; 112957; 112961; 112962; 112965; 112972; 112974; 112994; 112995; 112999; 113006; 113008; 113010; 113520; 113522; 113525; 113533; 113534; 113536; 113547; 113550; 113555; 113556; 113557; 113561; 113563; 113689; 113724; 114021; 114101; 114105; 114106; 114107; 114109; 114110; 114111; 114116; 114117; 114310; 114333; 114108; 112792; 112798; 114497; 114324; 114325; 114330; 114336; 114493; 114537; 113046; 113662; 113667; 113682; 113591; 113594; 113654; 113655; 113740; 114020; 114030; 114044; 114046; 114048; 114096; 114114; 114019; 114025; 114027; 114033; 114039; 114040; 114045; 114051; 114028; 113704; 113746; 113748; 113753; 113760; 113061; 113725; 113586; 113691; 114307; 114309; 113028; 113066; 113539; 113515; 113709; 114047; 112814; 112824; 112829; 112830; 112844; 112848; 112865; 112866; 112807; 112822; 112823; 112834; 112835; 112847; 112854; 112876; 122848; 113758; 114508; 114951; 114342; 114343; 114346; 114351; 114360; 114500; 114345; 114499; 114505; 112976; 113037; 113040; 113524; 114322; 114498; 114510; 114636; 114640; 114641; 114676; 114685; 114695; 114701; 114820; 114827; 114841; 114900; 114901; 114674; 114725; 114800; 114810; 114812; 114822; 114825; 113057; 113063; 112821; 112839; 112842; 112867; 112826; 112837; 112825; 112982; 112987; 113510; 113511; 113546; 113584; 113592; 113652; 113686; 113710; 113723; 113756; 113671; 113675; 113715; 114009; 114013; 114489; 114494; 113549; 113744; 114637; 114947; 112765; 112766; 112767; 112768; 112770; 112774; 112775; 112776; 112778; 112779; 112780; 112787; 112815; 112841; 112862; 112968; 112977; 113013; 113014; 113016; 113017; 113018; 113027; 113032; 113039; 113050; 113535; 113562; 113588; 113658; 113755; 114032; 114488; 114842; AARC0001; AARC002; AARC003; AARC004; AARC005; AARC006; AARC007; AARC008; AARC009; 114095; 113590; Catalog Number: PRT-00490-100 GTIN (Primary ID): 00855573007761/00855573007204 Serial Numbers: 112929; 113598; 113233; 113238; 113239; 113241; 113243; 113244; 113245; 113603; 114193; 114201; 114255; 114266; Catalog Number: PRT-00900-001 GTIN (Primary ID): 00855573007778/00855573007433 Serial Numbers: 113365; 113405; 113906; 113962; 113407; 113414; 113430; 113390; 113437; 113262; 113380; 113458; 113384; 113398; 113442; 113858; 113441; 113443; 113445; 113915; 113817; 113898; 113899; 113378; 113395; 113923; 113972; 113937; 113945; 113957; 113961; 113968; 113368; 113373; 113379; 113264; 113351; 113389; 113370; 113377; 113454; 113468; 113804; 113823; 113830; 113837; 113838; 113881; 113882; 113888; 113892; 113902; 113927; 113967; 113399; Catalog Number: PRT-00900-100 GTIN (Primary ID): 00855573007785 Serial Numbers: 113096; 113300; 112923; 113107; 112877; 110392; 113143; 113183; 113274; 113323; Catalog Number: PRT-01098-000 GTIN (Primary ID): 00855573007730 Serial Numbers: 114444; 114451; 114615; 114226; 114442; 114450; 114452; 114441; 114453; Catalog Number: PRT-01100-000 GTIN (Primary ID): 00855573007747 Serial Numbers: 114556; 114591; 114634; 114771; 114779; 114603; 114606; 114667; 114710; 114713; 114737; 114739; 114559; 114783; 114869; 114576; 114587; 114364; 114369; 114374; 114387; 114391; 114560; 114568; 114764; 114776; 114787; 114849; 114549; 114569; 114705; 114562; 114355; 114372; 114405; 114550; 114594; 114601; 114652; 114659; 114661; 114662; 114358; 114383; 114386; 114389; 114390; 114530; 114540; 114401; 114422; 114424; 114548; 114670; 114714; 114850; 114414; 114417; 114554; 114561; 114861; 114862; 114543; 114786; 114789; 114912; 114917; 114851; 114628; 114211; 114542; 114528; 114531; 114483; 114750; 114643; 114644; 114748; 114926; 114991; 114891; 114892; 114894; 114357; 114409; 114604; 114413; 114653; 114654; 114734; 114754; Catalog Number: PRT-01100-001 GTIN (Primary ID): 00855573007747 Serial Numbers: 114965; 114966; 114967;
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Ventec Life Systems, Inc.
Reason for Recall:
Due to a component failure, ventilators may unexpectedly shut down or have a loss of blower function while in patient use. The shutdown potentially prevents the ventilator from providing ventilation therapy.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Ventec Life Systems VOCSN Multi-Function Ventilator Model/Version/Catalog Number: V+O+C+S+N+Pro/English/PRT-00490-001; V+O+C+S+N+Pro/Japan/PRT-00490-100; V+C+Pro/English/PRT-00900-001; V+C+Pro/Japan/PRT-00900-100; V+C/English/PRT-01098-000; V+O+C+S+N/English/PRT-01100-000; V+O+C+S+N/English/PRT-01100-001;

Product Codes/Lot Numbers:

Catalog Number: PRT-00490-001 GTIN (Primary ID): 00855573007792/00855573007006 Serial Numbers: 114303; 114304; 113728; 113729; 113697; 114296; 113587; 113684; 113706; 113003; 112957; 112961; 112962; 112965; 112972; 112974; 112994; 112995; 112999; 113006; 113008; 113010; 113520; 113522; 113525; 113533; 113534; 113536; 113547; 113550; 113555; 113556; 113557; 113561; 113563; 113689; 113724; 114021; 114101; 114105; 114106; 114107; 114109; 114110; 114111; 114116; 114117; 114310; 114333; 114108; 112792; 112798; 114497; 114324; 114325; 114330; 114336; 114493; 114537; 113046; 113662; 113667; 113682; 113591; 113594; 113654; 113655; 113740; 114020; 114030; 114044; 114046; 114048; 114096; 114114; 114019; 114025; 114027; 114033; 114039; 114040; 114045; 114051; 114028; 113704; 113746; 113748; 113753; 113760; 113061; 113725; 113586; 113691; 114307; 114309; 113028; 113066; 113539; 113515; 113709; 114047; 112814; 112824; 112829; 112830; 112844; 112848; 112865; 112866; 112807; 112822; 112823; 112834; 112835; 112847; 112854; 112876; 122848; 113758; 114508; 114951; 114342; 114343; 114346; 114351; 114360; 114500; 114345; 114499; 114505; 112976; 113037; 113040; 113524; 114322; 114498; 114510; 114636; 114640; 114641; 114676; 114685; 114695; 114701; 114820; 114827; 114841; 114900; 114901; 114674; 114725; 114800; 114810; 114812; 114822; 114825; 113057; 113063; 112821; 112839; 112842; 112867; 112826; 112837; 112825; 112982; 112987; 113510; 113511; 113546; 113584; 113592; 113652; 113686; 113710; 113723; 113756; 113671; 113675; 113715; 114009; 114013; 114489; 114494; 113549; 113744; 114637; 114947; 112765; 112766; 112767; 112768; 112770; 112774; 112775; 112776; 112778; 112779; 112780; 112787; 112815; 112841; 112862; 112968; 112977; 113013; 113014; 113016; 113017; 113018; 113027; 113032; 113039; 113050; 113535; 113562; 113588; 113658; 113755; 114032; 114488; 114842; AARC0001; AARC002; AARC003; AARC004; AARC005; AARC006; AARC007; AARC008; AARC009; 114095; 113590; Catalog Number: PRT-00490-100 GTIN (Primary ID): 00855573007761/00855573007204 Serial Numbers: 112929; 113598; 113233; 113238; 113239; 113241; 113243; 113244; 113245; 113603; 114193; 114201; 114255; 114266; Catalog Number: PRT-00900-001 GTIN (Primary ID): 00855573007778/00855573007433 Serial Numbers: 113365; 113405; 113906; 113962; 113407; 113414; 113430; 113390; 113437; 113262; 113380; 113458; 113384; 113398; 113442; 113858; 113441; 113443; 113445; 113915; 113817; 113898; 113899; 113378; 113395; 113923; 113972; 113937; 113945; 113957; 113961; 113968; 113368; 113373; 113379; 113264; 113351; 113389; 113370; 113377; 113454; 113468; 113804; 113823; 113830; 113837; 113838; 113881; 113882; 113888; 113892; 113902; 113927; 113967; 113399; Catalog Number: PRT-00900-100 GTIN (Primary ID): 00855573007785 Serial Numbers: 113096; 113300; 112923; 113107; 112877; 110392; 113143; 113183; 113274; 113323; Catalog Number: PRT-01098-000 GTIN (Primary ID): 00855573007730 Serial Numbers: 114444; 114451; 114615; 114226; 114442; 114450; 114452; 114441; 114453; Catalog Number: PRT-01100-000 GTIN (Primary ID): 00855573007747 Serial Numbers: 114556; 114591; 114634; 114771; 114779; 114603; 114606; 114667; 114710; 114713; 114737; 114739; 114559; 114783; 114869; 114576; 114587; 114364; 114369; 114374; 114387; 114391; 114560; 114568; 114764; 114776; 114787; 114849; 114549; 114569; 114705; 114562; 114355; 114372; 114405; 114550; 114594; 114601; 114652; 114659; 114661; 114662; 114358; 114383; 114386; 114389; 114390; 114530; 114540; 114401; 114422; 114424; 114548; 114670; 114714; 114850; 114414; 114417; 114554; 114561; 114861; 114862; 114543; 114786; 114789; 114912; 114917; 114851; 114628; 114211; 114542; 114528; 114531; 114483; 114750; 114643; 114644; 114748; 114926; 114991; 114891; 114892; 114894; 114357; 114409; 114604; 114413; 114653; 114654; 114734; 114754; Catalog Number: PRT-01100-001 GTIN (Primary ID): 00855573007747 Serial Numbers: 114965; 114966; 114967;

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1902-2021

Related Recalls

Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occur that may lead to reduced below intended levels of fraction of inspired oxygen delivery prior to and during ventilation and this oxygen-enriched environment may increase fire risk

Mar 25, 2026 Other Medical Devices View Details →

Multi-Function Ventilators were serviced using incorrect parts which have the potential to cause unexpected shutdown, or when using an active circuit could result in inaccurate tidal volume monitor, not triggering on patient efforts, less inspiratory volume, less inspiratory pressure, and less PEEP delivered.

Feb 3, 2025 Patient Monitors View Details →

The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions or leaks of the breathing circuit which may cause serious injury or death. The risk of injury or death is increased for ventilator dependent patients.

Jan 9, 2024 Patient Monitors View Details →