Nova StatStrip Xpress Glu-Test Strips (Glucose test strips) Catalog # 42214
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot 0312349249, Expires 12/31/2014 and Lot 0312361309, Expires 06/30/2015 Expanded recall: Lot 0313021309, Expires 07/31/2015
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Nova Biomedical Corporation
- Reason for Recall:
- Glucose test strips reports no result message when tested
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Nova StatStrip Xpress Glu-Test Strips (Glucose test strips) Catalog # 42214
Product Codes/Lot Numbers:
Lot 0312349249, Expires 12/31/2014 and Lot 0312361309, Expires 06/30/2015 Expanded recall: Lot 0313021309, Expires 07/31/2015
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1153-2013
Related Recalls
StatStrip Glucose Hospital Meter System. Model/Catalog Number: 63685. Prescription use blood glucose meter for near-patient testing.
Nova Biomedical
A software bug within the Gen 2 StatStrip Hospital Meter firmware leads to transmission of erroneous glucose patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current glucose test result to the DMS. The risk of this software defect is erroneously high or low glucose results being documented in a patient's medical record, which may lead to incorrect treatment.
A software bug within the meter firmware leads to transmission of erroneous glucose and/or ketone patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current test result to the DMS.
A software bug within the meter firmware leads to transmission of erroneous glucose and/or ketone patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current test result to the DMS.