Nova Biomedical

19 recalls • • Categories: Diagnostic Equipment, Other Medical Devices

A software bug within the Gen 2 StatStrip Hospital Meter firmware leads to transmission of erroneous glucose patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current glucose test result to the DMS. The risk of this software defect is erroneously high or low glucose results being documented in a patient's medical record, which may lead to incorrect treatment.

Nov 20, 2024 Diagnostic Equipment Nationwide View Details →

A software bug within the meter firmware leads to transmission of erroneous glucose and/or ketone patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current test result to the DMS.

Nov 20, 2024 Diagnostic Equipment Nationwide View Details →

A software bug within the meter firmware leads to transmission of erroneous glucose and/or ketone patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current test result to the DMS.

Nov 20, 2024 Diagnostic Equipment Nationwide View Details →

Potential risk of reporting erroneous creatinine patient results associated with the initial start-up of a new BUN/Creatinine Sensor Cartridge. The potential risk is within the first 2-4 hours after cartridge installation that may not be identified by traditional Quality Control practices.

Sep 18, 2023 Other Medical Devices Nationwide View Details →

Potential risk of reporting erroneous creatinine patient results associated with the initial start-up of a new BUN/Creatinine Sensor Cartridge. The potential risk is within the first 2-4 hours after cartridge installation that may not be identified by traditional Quality Control practices.

Sep 18, 2023 Other Medical Devices Nationwide View Details →

Software- scheduled A-QC analysis could initiate prior to a patient result being released by the operator and result in the pending patient result being overwritten with the A-QC test results.

Jun 19, 2017 Diagnostic Equipment Nationwide View Details →

Reports of false abnormally high glucose readings from test strips

Jul 15, 2013 Diagnostic Equipment Nationwide View Details →

Reports of false abnormally high glucose readings from test strips

Jul 15, 2013 Diagnostic Equipment Nationwide View Details →

Reports of false abnormally high glucose readings from test strips

Jul 15, 2013 Diagnostic Equipment Nationwide View Details →

Reports of false abnormally high glucose readings from test strips

Jul 15, 2013 Diagnostic Equipment Nationwide View Details →

Reports of false abnormally high glucose readings from test strips

Jul 15, 2013 Diagnostic Equipment Nationwide View Details →

Reports of false abnormally high glucose readings from test strips

Jul 15, 2013 Diagnostic Equipment Nationwide View Details →

Reports of false abnormally high glucose readings from test strips

Jul 15, 2013 Diagnostic Equipment Nationwide View Details →

Reports of false abnormally high glucose readings from test strips

Jul 15, 2013 Diagnostic Equipment Nationwide View Details →

Reports of false abnormally high glucose readings from test strips

Jul 15, 2013 Diagnostic Equipment Nationwide View Details →

Reports of false abnormally high glucose readings from test strips

Jul 15, 2013 Diagnostic Equipment Nationwide View Details →