Discovery XR656 HD. X-Ray imaging system.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Definium Tempo (GTIN: 00195278118356) Definium Tempo Pro (GTIN: 00195278070265)
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- GE Medical Systems, LLC
- Reason for Recall:
- The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Discovery XR656 HD. X-Ray imaging system.
Product Codes/Lot Numbers:
Definium Tempo (GTIN: 00195278118356) Definium Tempo Pro (GTIN: 00195278070265)
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1045-2025
Related Recalls
Allia IGS 3 Pulse angiographic X-ray system
GE Medical Systems
GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.
Allia IGS Pulse angiographic X-ray system
GE Medical Systems
GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.
Allia Moveo angiographic X-ray system
GE Medical Systems
GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.