CTL Medical Hex Driver MIS, 1/4" Square QC, 4.0 Ratcheting Ref 019.7050

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lot 7EQBB-C
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
CTL Medical Corporation
Reason for Recall:
the anti-toggle ring of the hex driver may separate from the driver shaft. If this occurs during surgery, the anti-toggle ring can be pressed into the pedicle screw head and cause separation of the screw from the housing upon tightening. if the anti-toggle ring remains in the pedicle crew head and can interfere with final tightening of the pedicle screw construct. this can lead to failure of the set crew and relies of the rod from the construct.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

CTL Medical Hex Driver MIS, 1/4" Square QC, 4.0 Ratcheting Ref 019.7050

Product Codes/Lot Numbers:

Lot 7EQBB-C

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1031-2019

Related Recalls