CTL Medical

6 recalls • • Categories: Other Medical Devices, Infusion Pumps

the anti-toggle ring of the hex driver may separate from the driver shaft. If this occurs during surgery, the anti-toggle ring can be pressed into the pedicle screw head and cause separation of the screw from the housing upon tightening. if the anti-toggle ring remains in the pedicle crew head and can interfere with final tightening of the pedicle screw construct. this can lead to failure of the set crew and relies of the rod from the construct.

Jul 5, 2018 Infusion Pumps View Details →

the anti-toggle ring of the hex driver may separate from the driver shaft. If this occurs during surgery, the anti-toggle ring can be pressed into the pedicle screw head and cause separation of the screw from the housing upon tightening. if the anti-toggle ring remains in the pedicle crew head and can interfere with final tightening of the pedicle screw construct. this can lead to failure of the set crew and relies of the rod from the construct.

Jul 5, 2018 Infusion Pumps View Details →