ABX PENTRA Urine Control L/H. The ABX PENTRA Urine Control L/H is a two-level (Low and High) quality control consisting of liquid solutions prepared from human urine with chemical additives and materials of biological origin added as required to obtain given component levels. The assigned values of the control components are given in the labeling, ensuring control of the appropriate HORIBA ABX SAS methods on the ABX PENTRA chemistry analyzers. Each control level is provided in one vial of 10 ml. The ABX PENTRA Urine Control L/H is for use in quality control by monitoring accuracy and precision.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model/Part #(s): A11A01674. Lot/Serial #(s): 1207001, 1303601, and 1300701. Shelf Life: Lot#s 1207001 & 1300701 Expire on: 06/30/2014. Lot# 1303601 Expires on: 09/30/2014.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Horiba Instruments, Inc dba Horiba Medical
- Reason for Recall:
- HORIBA Medical is informing all ABX PENTRA Urine Control L/H customers running the Creatinine assay on ABX PENTRA Chemistry Analyzers that the Target Range for Creatinine has been updated. The ABX PENTRA Urine Control L/H lot specific annexes have been updated to include revised acceptable Target Ranges for Creatinine.No other analyte values are affected.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
ABX PENTRA Urine Control L/H. The ABX PENTRA Urine Control L/H is a two-level (Low and High) quality control consisting of liquid solutions prepared from human urine with chemical additives and materials of biological origin added as required to obtain given component levels. The assigned values of the control components are given in the labeling, ensuring control of the appropriate HORIBA ABX SAS methods on the ABX PENTRA chemistry analyzers. Each control level is provided in one vial of 10 ml. The ABX PENTRA Urine Control L/H is for use in quality control by monitoring accuracy and precision.
Product Codes/Lot Numbers:
Model/Part #(s): A11A01674. Lot/Serial #(s): 1207001, 1303601, and 1300701. Shelf Life: Lot#s 1207001 & 1300701 Expire on: 06/30/2014. Lot# 1303601 Expires on: 09/30/2014.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0924-2014
Related Recalls
ABX Pentra Micro ALBUMIN 2 CP, Serum/Plasma, REF 1300032563 The device is intended for use in conjunction with certain materials to measure a variety of analytes.
Horiba Instruments, Inc
When the ABX Pentra Micro ALBUMIN 2 CP (ALBT2) Reagent 2 is sampled and dispensed immediately after an ABX Pentra Creatinine 120 CP (CREA3) Reagent 2 dispense, the final result of the microalbumin may be elevated more than 10%. This increase may affect serum, plasma and urine sample types.
10 ML REAGENT CUP, B1034626, SAP 1221034626 for use with ABX PENTRA 400 / PENTRA C400
Horiba Instruments, Inc
The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm.
15 ML REAGENT CUP, B1037307, SAP 1221037307 for use with ABX PENTRA 400 / PENTRA C400
Horiba Instruments, Inc
The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm.