Horiba Instruments, Inc

17 recalls • • Categories: Other Medical Devices, Diagnostic Equipment, Infusion Pumps, Implants & Prosthetics

When the ABX Pentra Micro ALBUMIN 2 CP (ALBT2) Reagent 2 is sampled and dispensed immediately after an ABX Pentra Creatinine 120 CP (CREA3) Reagent 2 dispense, the final result of the microalbumin may be elevated more than 10%. This increase may affect serum, plasma and urine sample types.

Apr 26, 2018 Other Medical Devices Nationwide View Details →

HORIBA Medical is recalling the ABX PENTRA 400 Reagent Container because they do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm. The defect would result in a false high or false low results for HbA1c and false high results for magnes

Nov 13, 2014 Other Medical Devices Nationwide View Details →

Horiba Medical is recalling the ABX PENTRA Magnesium RTU Reagent because the on-board stability claims has been updated. Horiba has updated the ABX PENTRA Magnesium RTU package inserts by removing the 7-day on-board stability claim and replacing it with a 1-day on-board stability claim.

Aug 11, 2014 Diagnostic Equipment Nationwide View Details →

HORIBA Medical initiated this recall of all software versions of ABX PENTRA ML (Model A11A06152) due to a software bug with the potential of transmitting incorrect results to the Laboratory Information System.

Feb 14, 2014 Other Medical Devices Nationwide View Details →

HORIBA Medical is informing all ABX PENTRA MultiCal, N Control, and P Control customers running the Creatinine Rate Blank Method assay on the ABX PENTRA 400 and PC200 that the Target Values for Creatinine Rate Blank Method have been updated. This issue is only applicable to Creatinine Rate Blank Method values reported on the ABX PENTRA 400 and PC200 Chemistry Analyzers only. No other analyzers

Jan 20, 2014 Infusion Pumps Nationwide View Details →

HORIBA Medical is informing all ABX PENTRA Urine Control L/H customers running the Creatinine assay on ABX PENTRA Chemistry Analyzers that the Target Range for Creatinine has been updated. The ABX PENTRA Urine Control L/H lot specific annexes have been updated to include revised acceptable Target Ranges for Creatinine.No other analyte values are affected.

Dec 18, 2013 Diagnostic Equipment Nationwide View Details →

HORIBA Medical is informing all PENTRA 200 and PENTRA 400 customers of a potential for reporting incorrect calcium results when using the ABX PENTRA CALCIUM CP reagent.

Sep 6, 2013 Implants & Prosthetics View Details →

HORIBA Medical is informing all ABX PENTRA 400 Analyzer with I.S.E. Module customers that there are no Linearity Limit Flags for the I.S.E. assays currently programmed into the ABX PENTRA 400 Analyzer software. All other assays run on this analyzer do have Linearity Limit Flags to alert the user to values that fall below or above the validated linearity range for each assay.

Mar 20, 2013 Other Medical Devices Nationwide View Details →

Horiba Medical is recalling ABX Micros IM2 Data Management System because it may display incorrect patient test results under unanticipated operating conditions.

Feb 15, 2013 Infusion Pumps Nationwide View Details →

Horiba Medical is recalling ABX PENTRA ML because they have confirmed an issue with the ABX PENTRA ML data management and validation system. An incorrect patient result is transmitted to the Laboratory Information System. This result is from a previous run of the same patient that is stored in the ML database. This issue occurs in a very rare sequence of specific events. The affected software

Sep 20, 2012 Other Medical Devices Nationwide View Details →

Horiba Medical is recalling the ABX PENTRA AST CP because during the AST application validation on the Pentra C200 using an Audit Systems linearity kit, it was determined that two higher values for AST (expected at 800 U/L and 1000 U/L) were recorded at 280 and 199 U/L without alarm.

Oct 10, 2012 Diagnostic Equipment Nationwide View Details →

Horiba Medical is recalling the ABX PENTRA AST CP because during the AST application validation on the Pentra C200 using an Audit Systems linearity kit, it was determined that two higher values for AST (expected at 800 U/L and 1000 U/L) were recorded at 280 and 199 U/L without alarm.

Oct 10, 2012 Diagnostic Equipment Nationwide View Details →

HORIBA Medical Irvine Technical Support received reports from some customers that have experienced high recover values for Hemoglobin when using Minotrol 16 lot #MX090 (Level High). Customer communication was sent 10/01/2010.

Sep 14, 2012 Diagnostic Equipment Nationwide View Details →

This recall was initiated because some customers were experiencing an alarm during Creatinine calibration "REAG RANGE HIGH". This alarm caused the calibration to fail and prevented the user from using the Creatinine assay.

Sep 14, 2012 Diagnostic Equipment Nationwide View Details →