Reprocessed Pressure Tourniquet Cuff (PTC) Tourniquet cuffs are single- or dual-bladder inflatable cuffs connected to a tourniquet system via a hose assembly. When wrapped around a limb and inflated, tourniquet cuffs apply an adequate amount of pressure on the arterial blood flow in a limb to create a bloodless surgical field. Tourniquet cuffs are available in a variety of sizes to accommodate a wide range of limb circumferences.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model Lot Lot Created Expiration 5921-018-135 2990595 5/5/2014 May-16 5921-018-235 2989898 5/5/2014 May-16 5921-018-235 2992868 5/5/2014 May-16 5921-024-235 2992501 5/5/2014 May-16 5921-030-135 2990147 5/5/2014 May-16 5921-030-135 2991575 5/5/2014 May-16 5921-030-235 2993033 5/5/2014 May-16 5921-034-135 2990152 5/5/2014 May-16 5921-034-135 2992555 5/5/2014 May-16 5921-034-235 2989674 5/5/2014 May-16 5921-034-235 2990240 5/5/2014 May-16 5921-034-235 2991095 5/5/2014 May-16 5921-034-235 2991413 5/5/2014 May-16 5921-034-235 2993046 5/5/2014 May-16 5921-034-235 2993509 5/5/2014 May-16 5921-218-235 2990217 5/5/2014 May-16 5921-218-235 2990553 5/5/2014 May-16 5921-218-235 2991743 5/5/2014 May-16 5921-218-135 2990106 5/5/2014 May-16 5921-218-135 2992540 5/5/2014 May-16 60-7070-102 2990697 5/5/2014 May-16 60-7070-103 2984111 5/5/2014 May-16 60-7070-103 2989473 5/5/2014 May-16 60-7070-103 2990110 5/5/2014 May-16 60-7070-103 2991030 5/5/2014 May-16 60-7070-103 2992398 5/5/2014 May-16 60-7070-103 2992636 5/5/2014 May-16 60-7070-105 2989822 5/5/2014 May-16 60-7070-105 2990012 5/5/2014 May-16 60-7070-105 2991271 5/5/2014 May-16 60-7070-105 2993392 5/5/2014 May-16 60-7070-104 2992643 5/5/2014 May-16 60-7070-106 2990433 5/5/2014 May-16 60-7070-106 2991817 5/5/2014 May-16 60-7070-106 2993407 5/5/2014 May-16 60-7075-004 2989867 5/5/2014 May-16 60-7075-103 2990650 5/5/2014 May-16 60-7075-103 2991432 5/5/2014 May-16 60-7075-103 2992627 5/5/2014 May-16 60-7075-105 2992282 5/5/2014 May-16 60-7075-106 2990451 5/5/2014 May-16 60-7075-106 2990958 5/5/2014 May-16 5921-018-135 2996601 5/8/2014 May-16 5921-024-135 2996808 5/8/2014 May-16 5921-030-235 2995775 5/8/2014 May-16 5921-030-235 2998064 5/8/2014 May-16 5921-034-135 2996324 5/8/2014 May-16 5921-034-235 2994193 5/8/2014 May-16 5921-034-235 2995233 5/8/2014 May-16 5921-034-235 2996975 5/8/2014 May-16 5921-218-235 2994186 5/8/2014 May-16 5921-218-235 2996503 5/8/2014 May-16 5921-218-235 2997742 5/8/2014 May-16 5921-218-135 2996590 5/8/2014 May-16 60-7070-103 2993940 5/8/2014 May-16 60-7070-103 2994661 5/8/2014 May-16 60-7070-103 2995417 5/8/2014 May-16 60-7070-103 2995747 5/8/2014 May-16 60-7070-103 2996779 5/8/2014 May-16 60-7070-103 2996896 5/8/2014 May-16 60-7070-105 2994062 5/8/2014 May-16 60-7070-105 2994707 5/8/2014 May-16 60-7070-105 2995279 5/8/2014 May-16 60-7070-105 2996094 5/8/2014 May-16 60-7070-105 2997735 5/8/2014 May-16 60-7070-105 2997902 5/8/2014 May-16 60-7070-104 2994473 5/8/2014 May-16 60-7070-104 2997034 5/8/2014 May-16 60-7070-003 2997757 5/8/2014 May-16 60-7070-006 2997496 5/8/2014 May-16 60-7070-106 2995752 5/8/2014 May-16 60-7070-106 2996232 5/8/2014 May-16 60-7070-106 2997047 5/8/2014 May-16 60-7070-107 2995995 5/8/2014 May-16 60-7075-003 2994460 5/8/2014 May-16 60-7075-006 2994782 5/8/2014 May-16 60-7075-103 2996143 5/8/2014 May-16 60-7075-103 2996981 5/8/2014 May-16 60-7075-104 2994544 5/8/2014 May-16 60-7075-106 2994879 5/8/2014 May-16 60-7075-106 2997132 5/8/2014 May-16 5921-018-235 3001050 5/12/2014 May-16 5921-024-135 3002574 5/12/2014 May-16 5921-024-235 3000098 5/12/2014 May-16 5921-030-135 2999078 5/12/2014 May-16 5921-030-135 3001370 5/12/2014 May-16 5921-030-235 2999888 5/12/2014 May-16 5921-030-235 2999944 5/12/2014 May-16 5921-030-235 3001009 5/12/2014 May-16 5921-034-135 2998628 5/12/2014 May-16 5921-034-135 2999717 5/12/2014 May-16 5921-034-135 3001084 5/12/2014 May-16 5921-034-136 3001358 5/12/2014 May-16 5921-034-235 2996307 5/12/2014 May-16 5921-034-235 2998231 5/12/2014 May-16 5921-034-235 2999013 5/12/2014 May-16 5921-034-235 3001679 5/12/2014 May-16 5921-034-235 3002934 5/12/2014 May-16 5921-034-236 3001738 5/12/2014 May-16 5921-212-235 3000108 5/12/2014 May-16 5921-218-235 3000804 5/12/2014 May-16 5921-218-235 3001508 5/12/2014 May-16 5921-044-135 3001362 5/12/2014 May-16 5921-218-135 3000316 5/12/2014 May-16 5921-218-136 3001389 5/12/2014 May-16 60-7070-103 2998671 5/12/2014 May-16 60-7070-103 2999294 5/12/2014 May-16 60-7070-103 3000363 5/12/2014 May-16 60-7070-103 3001013 5/12/2014 May-16 60-7070-103 3001622 5/12/2014 May-16 60-7070-103 3002136 5/12/2014 May-16 60-7070-105 2999021 5/12/2014 May-16 60-7070-105 2999343 5/12/2014 May-16 60-7070-105 3001633 5/12/2014 May-16 60-7070-105 3001772 5/12/2014 May-16 60-7070-105 3003024 5/12/2014 May-16 60-7070-104 2999622 5/12/2014 May-16 60-7070-104 3003302 5/12/2014 May-16 60-7070-003 3001434 5/12/2014 May-16 60-7070-004 3001680 5/12/2014 May-16 60-7070-106 2998072 5/12/2014 May-16 60-7070-106 2999336 5/12/2014 May-16 60-7070-106 3000702 5/12/2014 May-16 60-7070-106 3001646 5/12/2014 May-16 60-7070-106 3002526 5/12/2014 May-16 60-7070-107 2998980 5/12/2014 May-16 60-7075-003 2999740 5/12/2014 May-16 60-7075-004 3000067 5/12/2014 May-16 60-7075-005 3000344 5/12/2014 May-16 60-7075-006 2998732 5/12/2014 May-16 60-7075-006 2998914 5/12/2014 May-16 60-7075-102 2999633 5/12/2014 May-16 60-7075-103 2999009 5/12/2014 May-16 60-7075-104 2998778 5/12/2014 May-16 60-7075-106 2998922 5/12/2014 May-16 60-7075-106 3001391 5/12/2014 May-16 60-7075-107 3001740 5/12/2014 May-16 5921-018-135 3006954 5/14/2014 May-16 5921-018-135 3007582 5/14/2014 May-16 5921-018-235 3009132 5/14/2014 May-16 5921-024-135 2984861 5/14/2014 May-16 5921-030-135 3005777 5/14/2014 May-16 5921-030-235 3004709 5/14/2014 May-16 5921-030-235 3008589 5/14/2014 May-16 5921-030-235 3008656 5/14/2014 May-16 5921-030-235 3009386 5/14/2014 May-16 5921-034-135 3005442 5/14/2014 May-16 5921-034-235 3003387 5/14/2014 May-16 5921-034-235 3003704 5/14/2014 May-16 5921-034-235 3006724 5/14/2014 May-16 5921-034-235 3008976 5/14/2014 May-16 5921-218-235 2935644 5/14/2014 May-16 5921-218-235 3003450 5/14/2014 May-16 5921-218-235 3003893 5/14/2014 May-16 5921-218-235 3006439 5/14/2014 May-16 5921-218-235 3008090 5/14/2014 May-16 5921-218-235 3009099 5/14/2014 May-16 5921-044-235 3003840 5/14/2014 May-16 5921-218-135 3003957 5/14/2014 May-16 60-7070-103 3003935 5/14/2014 May-16 60-7070-103 3005194 5/14/2014 May-16 60-7070-103 3006603 5/14/2014 May-16 60-7070-103 3008660 5/14/2014 May-16 60-7070-105 2935323 5/14/2014 May-16 60-7070-105 3004489 5/14/2014 May-16 60-7070-105 3006091 5/14/2014 May-16 60-7070-105 3007991 5/14/2014 May-16 60-7070-105 3009113 5/14/2014 May-16 60-7070-104 3006949 5/14/2014 May-16 60-7070-104 3008429 5/14/2014 May-16 60-7070-106 3003552 5/14/2014 May-16 60-7070-106 3005488 5/14/2014 May-16 60-7070-106 3006384 5/14/2014 May-16 60-7070-106 3006487 5/14/2014 May-16 60-7070-106 3007229 5/14/2014 May-16 60-7070-106 3008969 5/14/2014 May-16 60-7070-106 3009109 5/14/2014 May-16 60-7070-106 3009291 5/14/2014 May-16 60-7070-107 3007763 5/14/2014 May-16 60-7075-005 3004539 5/14/2014 May-16 60-7075-006 3003829 5/14/2014 May-16 60-7075-103 3003048 5/14/2014 May-16 60-7075-103 3006494 5/14/2014 May-16 60-7075-104 3005765 5/14/2014 May-16 60-7075-105 3006193 5/14/2014 May-16 60-7075-106 3004573 5/14/2014 May-16 60-7075-106 3004721 5/14/2014 May-16 5921-018-135 3013487 5/16/2014 May-16 5921-024-135 3009925 5/16/2014 May-16 5921-030-135 3011981 5/16/2014 May-16 5921-030-135 3013460 5/16/2014 May-16 5921-030-235 3010399 5/16/2014 May-16 5921-030-235 3012499 5/16/2014 May-16 5921-030-235 3014359 5/16/2014 May-16 5921-034-135 3009420 5/16/2014 May-16 5921-034-135 3010671 5/16/2014 May-16 5921-034-135 3012218 5/16/2014 May-16 5921-034-136 3012819 5/16/2014 May-16 5921-034-235 3011975 5/16/2014 May-16 5921-034-235 3013688 5/16/2014 May-16 5921-034-236 3012826 5/16/2014 May-16 5921-218-235 3011952 5/16/2014 May-16 5921-218-235 3012512 5/16/2014 May-16 5921-044-135 3013456 5/16/2014 May-16 5921-218-135 3011330 5/16/2014 May-16 5921-218-135 3012531 5/16/2014 May-16 5921-218-136 3012835 5/16/2014 May-16 60-7070-102 3010017 5/16/2014 May-16 60-7070-102 3010310 5/16/2014 May-16 60-7070-103 3009412 5/16/2014 May-16 60-7070-103 3010055 5/16/2014 May-16 60-7070-103 3011561 5/16/2014 May-16 60-7070-103 3012362 5/16/2014 May-16 60-7070-103 3012549 5/16/2014 May-16 60-7070-103 3013851 5/16/2014 May-16 60-7070-105 3010352 5/16/2014 May-16 60-7070-105 3012358 5/16/2014 May-16 60-7070-105 3013661 5/16/2014 May-16 60-7070-105 3014870 5/16/2014 May-16 60-7070-104 3010980 5/16/2014 May-16 60-7070-104 3013036 5/16/2014 May-16 60-7070-104 3014578 5/16/2014 May-16 60-7070-003 3009408 5/16/2014 May-16 60-7070-003 3014051 5/16/2014 May-16 60-7070-005 3011353 5/16/2014 May-16 60-7070-006 3010029 5/16/2014 May-16 60-7070-006 3013350 5/16/2014 May-16 60-7070-106 3011705 5/16/2014 May-16 60-7070-106 3012916 5/16/2014 May-16 60-7070-106 3013062 5/16/2014 May-16 60-7070-106 3013898 5/16/2014 May-16 60-7070-107 3013674 5/16/2014 May-16 60-7075-003 3009372 5/16/2014 May-16 60-7075-003 3014584 5/16/2014 May-16 60-7075-004 3012842 5/16/2014 May-16 60-7075-006 3010784 5/16/2014 May-16 60-7075-006 3010922 5/16/2014 May-16 60-7075-104 3012900 5/16/2014 May-16 60-7075-106 3009121 5/16/2014 May-16 60-7075-106 3013439 5/16/2014 May-16 60-7075-107 3014067 5/16/2014 May-16 5921-018-235 3020828 5/20/2014 May-16 5921-024-235 3017331 5/20/2014 May-16 5921-030-135 3015291 5/20/2014 May-16 5921-030-235 3016575 5/20/2014 May-16 5921-030-235 3020317 5/20/2014 May-16 5921-034-135 3014862 5/20/2014 May-16 5921-034-135 3015372 5/20/2014 May-16 5921-034-135 3018293 5/20/2014 May-16 5921-034-135 3018728 5/20/2014 May-16 5921-034-235 3015075 5/20/2014 May-16 5921-034-235 3016568 5/20/2014 May-16 5921-034-235 3019276 5/20/2014 May-16 5921-034-235 3019754 5/20/2014 May-16 5921-034-235 3019932 5/20/2014 May-16 5921-034-235 3020551 5/20/2014 May-16 5921-218-235 3016285 5/20/2014 May-16 5921-218-235 3018110 5/20/2014 May-16 5921-218-235 3019950 5/20/2014 May-16 5921-044-235 3020863 5/20/2014 May-16 5921-218-135 3015081 5/20/2014 May-16 5921-218-135 3018303 5/20/2014 May-16 5921-218-236 3015624 5/20/2014 May-16 60-7070-103 3015068 5/20/2014 May-16 60-7070-103 3016715 5/20/2014 May-16 60-7070-103 3017019 5/20/2014 May-16 60-7070-103 3018335 5/20/2014 May-16 60-7070-103 3020034 5/20/2014 May-16 60-7070-105 3016151 5/20/2014 May-16 60-7070-105 3019501 5/20/2014 May-16 60-7070-104 3018818 5/20/2014 May-16 60-7070-003 3018666 5/20/2014 May-16 60-7070-005 3021080 5/20/2014 May-16 60-7070-106 2985131 5/20/2014 May-16 60-7070-106 3016886 5/20/2014 May-16 60-7070-106 3017706 5/20/2014 May-16 60-7070-106 3019156 5/20/2014 May-16 60-7070-107 3015630 5/20/2014 May-16 60-7070-107 3020844 5/20/2014 May-16 60-7075-006 3020892 5/20/2014 May-16 60-7080-102 3015227 5/20/2014 May-16 60-7075-103 3016445 5/20/2014 May-16 60-7075-103 3018123 5/20/2014 May-16 60-7075-105 3018264 5/20/2014 May-16 60-7075-106 3015364 5/20/2014 May-16 60-7075-106 3015682 5/20/2014 May-16 60-7075-106 3016484 5/20/2014 May-16 60-7075-106 3020598 5/20/2014 May-16 5921-024-135 3021448 5/21/2014 May-16 5921-024-235 3021476 5/21/2014 May-16 5921-024-235 3022897 5/21/2014 May-16 5921-030-135 3022504 5/21/2014 May-16 5921-030-136 3023374 5/21/2014 May-16 5921-030-235 3022074 5/21/2014 May-16 5921-030-235 3024094 5/21/2014 May-16 5921-034-135 3023841 5/21/2014 May-16 5921-034-135 3025072 5/21/2014 May-16 5921-034-235 3021219 5/21/2014 May-16 5921-034-235 3022233 5/21/2014 May-16 5921-034-235 3023943 5/21/2014 May-16 5921-218-235 3021521 5/21/2014 May-16 5921-218-235 3025174 5/21/2014 May-16 5921-044-135 3025038 5/21/2014 May-16 5921-218-135 3024209 5/21/2014 May-16 60-7070-102 3020871 5/21/2014 May-16 60-7070-103 3021430 5/21/2014 May-16 60-7070-103 3022171 5/21/2014 May-16 60-7070-103 3022377 5/21/2014 May-16 60-7070-103 3022866 5/21/2014 May-16 60-7070-103 3024382 5/21/2014 May-16 60-7070-103 3025685 5/21/2014 May-16 60-7070-105 3022019 5/21/2014 May-16 60-7070-105 3023785 5/21/2014 May-16 60-7070-105 3025119 5/21/2014 May-16 60-7070-104 3017320 5/21/2014 May-16 60-7070-104 3022694 5/21/2014 May-16 60-7070-004 3023335 5/21/2014 May-16 60-7070-005 3021344 5/21/2014 May-16 60-7070-006 3023181 5/21/2014 May-16 60-7070-106 3021226 5/21/2014 May-16 60-7070-106 3021929 5/21/2014 May-16 60-7070-106 3022850 5/21/2014 May-16 60-7070-106 3023716 5/21/2014 May-16 60-7070-106 3025029 5/21/2014 May-16 60-7070-106 3025220 5/21/2014 May-16 60-7070-107 3023264 5/21/2014 May-16 60-7075-003 3023989 5/21/2014 May-16 60-7075-005 3016649 5/21/2014 May-16 60-7075-006 3023839 5/21/2014 May-16 60-7080-102 3022591 5/21/2014 May-16 60-7075-103 3022312 5/21/2014 May-16 60-7075-103 3024056 5/21/2014 May-16 60-7075-104 3021203 5/21/2014 May-16 60-7075-106 3022354 5/21/2014 May-16 60-7075-106 3025161 5/21/2014 May-16 5921-018-135 3029394 5/27/2014 May-16 5921-024-135 3028397 5/27/2014 May-16 5921-024-235 3026224 5/27/2014 May-16 5921-024-235 3030968 5/27/2014 May-16 5921-030-135 3026211 5/27/2014 May-16 5921-030-235 3026085 5/27/2014 May-16 5921-030-235 3028129 5/27/2014 May-16 5921-030-235 3031570 5/27/2014 May-16 5921-034-135 3027531 5/27/2014 May-16 5921-034-135 3029095 5/27/2014 May-16 5921-034-135 3030896 5/27/2014 May-16 5921-034-136 3025671 5/27/2014 May-16 5921-034-136 3030590 5/27/2014 May-16 5921-034-235 3026118 5/27/2014 May-16 5921-034-235 3026783 5/27/2014 May-16 5921-034-235 3029510 5/27/2014 May-16 5921-034-235 3029864 5/27/2014 May-16 5921-034-235 3030170 5/27/2014 May-16 5921-034-236 3026540 5/27/2014 May-16 5921-212-235 3029179 5/27/2014 May-16 5921-218-235 3026461 5/27/2014 May-16 5921-218-235 3026777 5/27/2014 May-16 5921-218-235 3030923 5/27/2014 May-16 5921-218-235 3031845 5/27/2014 May-16 5921-044-235 3029523 5/27/2014 May-16 5921-218-135 3027078 5/27/2014 May-16 5921-218-135 3027391 5/27/2014 May-16 5921-218-135 3029957 5/27/2014 May-16 5921-218-135 3031440 5/27/2014 May-16 5921-218-136 3026529 5/27/2014 May-16 60-7070-103 3025914 5/27/2014 May-16 60-7070-103 3026712 5/27/2014 May-16 60-7070-103 3027111 5/27/2014 May-16 60-7070-103 3028091 5/27/2014 May-16 60-7070-103 3028996 5/27/2014 May-16 60-7070-103 3029341 5/27/2014 May-16 60-7070-103 3030542 5/27/2014 May-16 60-7070-105 3026457 5/27/2014 May-16 60-7070-105 3026720 5/27/2014 May-16 60-7070-105 3027520 5/27/2014 May-16 60-7070-105 3030428 5/27/2014 May-16 60-7070-104 3027169 5/27/2014 May-16 60-7070-104 3030947 5/27/2014 May-16 60-7070-003 3025830 5/27/2014 May-16 60-7070-005 3029838 5/27/2014 May-16 60-7070-106 3025980 5/27/2014 May-16 60-7070-106 3027370 5/27/2014 May-16 60-7070-106 3028635 5/27/2014 May-16 60-7070-106 3030162 5/27/2014 May-16 60-7070-107 3025848 5/27/2014 May-16 60-7070-107 3027786 5/27/2014 May-16 60-7070-107 3029942 5/27/2014 May-16 60-7075-003 3028402 5/27/2014 May-16 60-7075-004 3027130 5/27/2014 May-16 60-7075-006 3028321 5/27/2014 May-16 60-7075-006 3031279 5/27/2014 May-16 60-7075-102 3025784 5/27/2014 May-16 60-7075-103 3031438 5/27/2014 May-16 60-7075-104 3027124 5/27/2014 May-16 60-7075-106 3028117 5/27/2014 May-16 60-7075-106 3028313 5/27/2014 May-16 60-7075-106 3029167 5/27/2014 May-16 60-7075-106 3031543 5/27/2014 May-16 60-7075-106 3031768 5/27/2014 May-16 60-7075-107 3029146 5/27/2014 May-16 60-7075-107 3031251 5/27/2014 May-16 5921-018-135 3039220 5/29/2014 May-16 5921-018-235 3034408 5/29/2014 May-16 5921-024-135 3040193 5/29/2014 May-16 5921-024-135 3040313 5/29/2014 May-16 5921-024-235 3035743 5/29/2014 May-16 5921-024-235 3039323 5/29/2014 May-16 5921-030-135 3037252 5/29/2014 May-16 5921-030-135 3037473 5/29/2014 May-16 5921-030-136 3040908 5/29/2014 May-16 5921-030-235 3033662 5/29/2014 May-16 5921-030-235 3035198 5/29/2014 May-16 5921-030-235 3038667 5/29/2014 May-16 5921-030-235 3040384 5/29/2014 May-16 5921-034-135 3032079 5/29/2014 May-16 5921-034-135 3034262 5/29/2014 May-16 5921-034-135 3035497 5/29/2014 May-16 5921-034-135 3039824 5/29/2014 May-16 5921-034-135 3041102 5/29/2014 May-16 5921-034-235 3033187 5/29/2014 May-16 5921-034-235 3034178 5/29/2014 May-16 5921-034-235 3034301 5/29/2014 May-16 5921-034-235 3035926 5/29/2014 May-16 5921-034-235 3037366 5/29/2014 May-16 5921-034-235 3040657 5/29/2014 May-16 5921-034-235 3041092 5/29/2014 May-16 5921-212-235 3040172 5/29/2014 May-16 5921-218-235 3031911 5/29/2014 May-16 5921-218-235 3033379 5/29/2014 May-16 5921-218-235 3034386 5/29/2014 May-16 5921-218-235 3036335 5/29/2014 May-16 5921-218-235 3037642 5/29/2014 May-16 5921-218-235 3039637 5/29/2014 May-16 5921-218-235 3040418 5/29/2014 May-16 5921-218-235 3041295 5/29/2014 May-16 5921-044-135 3032739 5/29/2014 May-16 5921-044-235 3036524 5/29/2014 May-16 5921-218-135 3034669 5/29/2014 May-16 5921-218-135 3037223 5/29/2014 May-16 5921-218-135 3039126 5/29/2014 May-16 60-7070-102 3032702 5/29/2014 May-16 60-7070-103 3031924 5/29/2014 May-16 60-7070-103 3033173 5/29/2014 May-16 60-7070-103 3034372 5/29/2014 May-16 60-7070-103 3034546 5/29/2014 May-16 60-7070-103 3035484 5/29/2014 May-16 60-7070-103 3036936 5/29/2014 May-16 60-7070-103 3037441 5/29/2014 May-16 60-7070-103 3038508 5/29/2014 May-16 60-7070-103 3039671 5/29/2014 May-16 60-7070-103 3040559 5/29/2014 May-16 60-7070-103 3041501 5/29/2014 May-16 60-7070-105 3032482 5/29/2014 May-16 60-7070-105 3032794 5/29/2014 May-16 60-7070-105 3034648 5/29/2014 May-16 60-7070-105 3036217 5/29/2014 May-16 60-7070-105 3038916 5/29/2014 May-16 60-7070-105 3040268 5/29/2014 May-16 60-7070-104 3033380 5/29/2014 May-16 60-7070-104 3035758 5/29/2014 May-16 60-7070-104 3039830 5/29/2014 May-16 60-7070-003 3034994 5/29/2014 May-16 60-7070-003 3037281 5/29/2014 May-16 60-7070-006 3037275 5/29/2014 May-16 60-7070-106 3032448 5/29/2014 May-16 60-7070-106 3032453 5/29/2014 May-16 60-7070-106 3033643 5/29/2014 May-16 60-7070-106 3034859 5/29/2014 May-16 60-7070-106 3035011 5/29/2014 May-16 60-7070-106 3036943 5/29/2014 May-16 60-7070-106 3037880 5/29/2014 May-16 60-7070-106 3037977 5/29/2014 May-16 60-7070-106 3039663 5/29/2014 May-16 60-7070-106 3040571 5/29/2014 May-16 60-7070-106 3040846 5/29/2014 May-16 60-7070-106 3041662 5/29/2014 May-16 60-7070-107 3035608 5/29/2014 May-16 60-7070-107 3039310 5/29/2014 May-16 60-7070-107 3040210 5/29/2014 May-16 60-7075-003 3034885 5/29/2014 May-16 60-7075-003 3036632 5/29/2014 May-16 60-7075-004 3037384 5/29/2014 May-16 60-7075-005 3040155 5/29/2014 May-16 60-7075-006 3036181 5/29/2014 May-16 60-7080-102 3040186 5/29/2014 May-16 60-7075-102 3040138 5/29/2014 May-16 60-7075-103 3032014 5/29/2014 May-16 60-7075-103 3033529 5/29/2014 May-16 60-7075-103 3033708 5/29/2014 May-16 60-7075-103 3037402 5/29/2014 May-16 60-7075-103 3041909 5/29/2014 May-16 60-7075-104 3033796 5/29/2014 May-16 60-7075-104 3034871 5/29/2014 May-16 60-7075-104 3039231 5/29/2014 May-16 60-7075-105 3032046 5/29/2014 May-16 60-7075-106 3033966 5/29/2014 May-16 60-7075-106 3036609 5/29/2014 May-16 60-7075-106 3040223 5/29/2014 May-16 60-7075-106 3041796 5/29/2014 May-16 5921-024-235 3043588 6/3/2014 Jun-16 5921-024-235 3045890 6/3/2014 Jun-16 5921-030-135 3043934 6/3/2014 Jun-16 5921-030-135 3045900 6/3/2014 Jun-16 5921-030-135 3046794 6/3/2014 Jun-16 5921-030-136 3043041 6/3/2014 Jun-16 5921-030-235 3042974 6/3/2014 Jun-16 5921-030-235 3043638 6/3/2014 Jun-16 5921-030-235 3044175 6/3/2014 Jun-16 5921-030-235 3045370 6/3/2014 Jun-16 5921-030-235 3045729 6/3/2014 Jun-16 5921-034-135 3042990 6/3/2014 Jun-16 5921-034-135 3043463 6/3/2014 Jun-16 5921-034-135 3044201 6/3/2014 Jun-16 5921-034-135 3046079 6/3/2014 Jun-16 5921-034-235 3042339 6/3/2014 Jun-16 5921-034-235 3042942 6/3/2014 Jun-16 5921-034-235 3043831 6/3/2014 Jun-16 5921-034-235 3044802 6/3/2014 Jun-16 5921-034-235 3046482 6/3/2014 Jun-16 5921-034-236 3042114 6/3/2014 Jun-16 5921-218-235 2935491 6/3/2014 Jun-16 5921-218-235 3043104 6/3/2014 Jun-16 5921-218-235 3044943 6/3/2014 Jun-16 5921-218-235 3046469 6/3/2014 Jun-16 5921-218-135 3042146 6/3/2014 Jun-16 5921-218-135 3043883 6/3/2014 Jun-16 5921-218-135 3045835 6/3/2014 Jun-16 5921-218-136 3039628 6/3/2014 Jun-16 60-7070-103 3042568 6/3/2014 Jun-16 60-7070-103 3043574 6/3/2014 Jun-16 60-7070-103 3044240 6/3/2014 Jun-16 60-7070-103 3045410 6/3/2014 Jun-16 60-7070-103 3046521 6/3/2014 Jun-16 60-7070-105 3042179 6/3/2014 Jun-16 60-7070-105 3043519 6/3/2014 Jun-16 60-7070-105 3044255 6/3/2014 Jun-16 60-7070-105 3045218 6/3/2014 Jun-16 60-7070-105 3046261 6/3/2014 Jun-16 60-7070-105 3046509 6/3/2014 Jun-16 60-7070-104 3041488 6/3/2014 Jun-16 60-7070-104 3043629 6/3/2014 Jun-16 60-7070-104 3045633 6/3/2014 Jun-16 60-7070-003 3043136 6/3/2014 Jun-16 60-7070-004 3043002 6/3/2014 Jun-16 60-7070-005 3043266 6/3/2014 Jun-16 60-7070-006 3043073 6/3/2014 Jun-16 60-7070-006 3045916 6/3/2014 Jun-16 60-7070-106 3043416 6/3/2014 Jun-16 60-7070-106 3044065 6/3/2014 Jun-16 60-7070-106 3044530 6/3/2014 Jun-16 60-7070-106 3046067 6/3/2014 Jun-16 60-7070-107 3042961 6/3/2014 Jun-16 60-7075-006 3043767 6/3/2014 Jun-16 60-7075-102 3046668 6/3/2014 Jun-16 60-7075-103 3042133 6/3/2014 Jun-16 60-7075-103 3043162 6/3/2014 Jun-16 60-7075-103 3044272 6/3/2014 Jun-16 60-7075-105 3043820 6/3/2014 Jun-16 60-7075-105 3044054 6/3/2014 Jun-16 60-7075-106 3045764 6/3/2014 Jun-16 5921-018-135 3049884 6/5/2014 Jun-16 5921-018-235 3048533 6/5/2014 Jun-16 5921-024-135 3047998 6/5/2014 Jun-16 5921-024-235 3049907 6/5/2014 Jun-16 5921-030-135 3046972 6/5/2014 Jun-16 5921-030-135 3051965 6/5/2014 Jun-16 5921-030-235 3046862 6/5/2014 Jun-16 5921-030-235 3049515 6/5/2014 Jun-16 5921-030-235 3051418 6/5/2014 Jun-16 5921-034-135 3049589 6/5/2014 Jun-16 5921-034-135 3051460 6/5/2014 Jun-16 5921-034-235 3048258 6/5/2014 Jun-16 5921-034-235 3049267 6/5/2014 Jun-16 5921-034-235 3050694 6/5/2014 Jun-16 5921-034-235 3051070 6/5/2014 Jun-16 5921-218-235 3049279 6/5/2014 Jun-16 5921-218-235 3050976 6/5/2014 Jun-16 5921-030-236 3046870 6/5/2014 Jun-16 5921-044-135 3051748 6/5/2014 Jun-16 5921-044-235 3041834 6/5/2014 Jun-16 5921-218-135 3047491 6/5/2014 Jun-16 5921-218-135 3050504 6/5/2014 Jun-16 5921-218-135 3051920 6/5/2014 Jun-16 5921-218-136 3051949 6/5/2014 Jun-16 60-7070-103 3048113 6/5/2014 Jun-16 60-7070-103 3048859 6/5/2014 Jun-16 60-7070-103 3048871 6/5/2014 Jun-16 60-7070-103 3049527 6/5/2014 Jun-16 60-7070-103 3050251 6/5/2014 Jun-16 60-7070-103 3051487 6/5/2014 Jun-16 60-7070-105 3047518 6/5/2014 Jun-16 60-7070-105 3049205 6/5/2014 Jun-16 60-7070-105 3050222 6/5/2014 Jun-16 60-7070-105 3051425 6/5/2014 Jun-16 60-7070-105 3051851 6/5/2014 Jun-16 60-7070-104 3046843 6/5/2014 Jun-16 60-7070-104 3049660 6/5/2014 Jun-16 60-7070-104 3051908 6/5/2014 Jun-16 60-7070-003 3051004 6/5/2014 Jun-16 60-7070-006 3050589 6/5/2014 Jun-16 60-7070-007 3048514 6/5/2014 Jun-16 60-7070-106 3046606 6/5/2014 Jun-16 60-7070-106 3047103 6/5/2014 Jun-16 60-7070-106 3048790 6/5/2014 Jun-16 60-7070-106 3049747 6/5/2014 Jun-16 60-7070-106 3049969 6/5/2014 Jun-16 60-7070-106 3051186 6/5/2014 Jun-16 60-7075-003 3048372 6/5/2014 Jun-16 60-7075-006 3048449 6/5/2014 Jun-16 60-7075-103 3047671 6/5/2014 Jun-16 60-7075-103 3050781 6/5/2014 Jun-16 60-7075-104 3031562 6/5/2014 Jun-16 60-7075-104 3051912 6/5/2014 Jun-16 60-7075-105 3047142 6/5/2014 Jun-16 60-7075-105 3048317 6/5/2014 Jun-16 60-7075-105 3051657 6/5/2014 Jun-16 60-7075-106 3047541 6/5/2014 Jun-16 60-7075-106 3049710 6/5/2014 Jun-16 5921-018-135 3053778 6/9/2014 Jun-16 5921-024-135 3056944 6/9/2014 Jun-16 5921-024-235 3052099 6/9/2014 Jun-16 5921-024-235 3054704 6/9/2014 Jun-16 5921-030-135 3056391 6/9/2014 Jun-16 5921-030-235 3052121 6/9/2014 Jun-16 5921-030-235 3053220 6/9/2014 Jun-16 5921-030-235 3054727 6/9/2014 Jun-16 5921-034-135 3053811 6/9/2014 Jun-16 5921-034-135 3055012 6/9/2014 Jun-16 5921-034-135 3056389 6/9/2014 Jun-16 5921-034-235 3052132 6/9/2014 Jun-16 5921-034-235 3053462 6/9/2014 Jun-16 5921-034-235 3054478 6/9/2014 Jun-16 5921-034-236 3054840 6/9/2014 Jun-16 5921-218-235 3052145 6/9/2014 Jun-16 5921-218-235 3053103 6/9/2014 Jun-16 5921-218-235 3054942 6/9/2014 Jun-16 5921-218-135 3055008 6/9/2014 Jun-16 5921-218-236 3054497 6/9/2014 Jun-16 60-7070-102 3053914 6/9/2014 Jun-16 60-7070-103 3052150 6/9/2014 Jun-16 60-7070-103 3052546 6/9/2014 Jun-16 60-7070-103 3053541 6/9/2014 Jun-16 60-7070-103 3054855 6/9/2014 Jun-16 60-7070-103 3056651 6/9/2014 Jun-16 60-7070-105 3052166 6/9/2014 Jun-16 60-7070-105 3053064 6/9/2014 Jun-16 60-7070-105 3054356 6/9/2014 Jun-16 60-7070-105 3055096 6/9/2014 Jun-16 60-7070-104 3052412 6/9/2014 Jun-16 60-7070-104 3052477 6/9/2014 Jun-16 60-7070-104 3054373 6/9/2014 Jun-16 60-7070-005 3054838 6/9/2014 Jun-16 60-7070-106 3052178 6/9/2014 Jun-16 60-7070-106 3052837 6/9/2014 Jun-16 60-7070-106 3053980 6/9/2014 Jun-16 60-7070-106 3054677 6/9/2014 Jun-16 60-7070-106 3055873 6/9/2014 Jun-16 60-7075-003 3054961 6/9/2014 Jun-16 60-7075-103 3052505 6/9/2014 Jun-16 60-7075-103 3054919 6/9/2014 Jun-16 60-7075-105 3054387 6/9/2014 Jun-16 60-7075-106 3052431 6/9/2014 Jun-16 60-7075-106 3053882 6/9/2014 Jun-16 60-7075-106 3054988 6/9/2014 Jun-16 5921-018-135 3050464 6/11/2014 Jun-16 5921-018-235 3060341 6/11/2014 Jun-16 5921-024-235 3057461 6/11/2014 Jun-16 5921-024-235 3063218 6/11/2014 Jun-16 5921-030-135 3058578 6/11/2014 Jun-16 5921-030-235 3057276 6/11/2014 Jun-16 5921-030-235 3064504 6/11/2014 Jun-16 5921-034-135 3059307 6/11/2014 Jun-16 5921-034-135 3061081 6/11/2014 Jun-16 5921-034-135 3064589 6/11/2014 Jun-16 5921-034-235 3057118 6/11/2014 Jun-16 5921-034-235 3058131 6/11/2014 Jun-16 5921-034-235 3059693 6/11/2014 Jun-16 5921-034-235 3062833 6/11/2014 Jun-16 5921-218-235 3058917 6/11/2014 Jun-16 5921-218-235 3059664 6/11/2014 Jun-16 5921-218-235 3062867 6/11/2014 Jun-16 5921-044-235 3059934 6/11/2014 Jun-16 5921-212-135 3057141 6/11/2014 Jun-16 5921-218-135 3059311 6/11/2014 Jun-16 5921-218-135 3063717 6/11/2014 Jun-16 5921-218-135 3064035 6/11/2014 Jun-16 60-7070-103 3057501 6/11/2014 Jun-16 60-7070-103 3057670 6/11/2014 Jun-16 60-7070-103 3058566 6/11/2014 Jun-16 60-7070-103 3060048 6/11/2014 Jun-16 60-7070-103 3061837 6/11/2014 Jun-16 60-7070-103 3064042 6/11/2014 Jun-16 60-7070-103 3064687 6/11/2014 Jun-16 60-7070-103 3064775 6/11/2014 Jun-16 60-7070-105 3058288 6/11/2014 Jun-16 60-7070-105 3058532 6/11/2014 Jun-16 60-7070-105 3059655 6/11/2014 Jun-16 60-7070-105 3060990 6/11/2014 Jun-16 60-7070-105 3064639 6/11/2014 Jun-16 60-7070-104 3057643 6/11/2014 Jun-16 60-7070-104 3062898 6/11/2014 Jun-16 60-7070-106 3057490 6/11/2014 Jun-16 60-7070-106 3057586 6/11/2014 Jun-16 60-7070-106 3059209 6/11/2014 Jun-16 60-7070-106 3060133 6/11/2014 Jun-16 60-7070-106 3063202 6/11/2014 Jun-16 60-7070-106 3064591 6/11/2014 Jun-16 60-7075-003 3063959 6/11/2014 Jun-16 60-7075-103 3057658 6/11/2014 Jun-16 60-7075-103 3060179 6/11/2014 Jun-16 60-7075-103 3063646 6/11/2014 Jun-16 60-7075-105 3058889 6/11/2014 Jun-16 60-7075-105 3063963 6/11/2014 Jun-16 60-7075-106 3057426 6/11/2014 Jun-16 60-7075-106 3058891 6/11/2014 Jun-16 60-7075-106 3059722 6/11/2014 Jun-16 60-7075-106 3064781 6/11/2014 Jun-16 5921-018-135 3065910 6/13/2014 Jun-16 5921-018-235 3066249 6/13/2014 Jun-16 5921-018-235 3067699 6/13/2014 Jun-16 5921-018-235 3070406 6/13/2014 Jun-16 5921-024-135 3066212 6/13/2014 Jun-16 5921-024-235 3066709 6/13/2014 Jun-16 5921-024-235 3067862 6/13/2014 Jun-16 5921-024-236 3070614 6/13/2014 Jun-16 5921-030-135 3065002 6/13/2014 Jun-16 5921-030-135 3070475 6/13/2014 Jun-16 5921-030-235 3070021 6/13/2014 Jun-16 5921-034-135 3064907 6/13/2014 Jun-16 5921-034-135 3066161 6/13/2014 Jun-16 5921-034-135 3067827 6/13/2014 Jun-16 5921-034-135 3070370 6/13/2014 Jun-16 5921-034-135 3070985 6/13/2014 Jun-16 5921-034-136 3065722 6/13/2014 Jun-16 5921-034-235 3062487 6/13/2014 Jun-16 5921-034-235 3065400 6/13/2014 Jun-16 5921-034-235 3069374 6/13/2014 Jun-16 5921-034-236 3070755 6/13/2014 Jun-16 5921-218-235 3068529 6/13/2014 Jun-16 5921-218-235 3070888 6/13/2014 Jun-16 5921-044-135 3066265 6/13/2014 Jun-16 5921-212-135 3067804 6/13/2014 Jun-16 5921-218-135 3065561 6/13/2014 Jun-16 5921-218-135 3067668 6/13/2014 Jun-16 5921-218-135 3069065 6/13/2014 Jun-16 60-7070-102 3070148 6/13/2014 Jun-16 60-7070-103 3065143 6/13/2014 Jun-16 60-7070-103 3065376 6/13/2014 Jun-16 60-7070-103 3066671 6/13/2014 Jun-16 60-7070-103 3067550 6/13/2014 Jun-16 60-7070-103 3067584 6/13/2014 Jun-16 60-7070-103 3067836 6/13/2014 Jun-16 60-7070-103 3069123 6/13/2014 Jun-16 60-7070-103 3070729 6/13/2014 Jun-16 60-7070-105 3065087 6/13/2014 Jun-16 60-7070-105 3065865 6/13/2014 Jun-16 60-7070-105 3066830 6/13/2014 Jun-16 60-7070-105 3067476 6/13/2014 Jun-16 60-7070-105 3068950 6/13/2014 Jun-16 60-7070-104 3065975 6/13/2014 Jun-16 60-7070-104 3068398 6/13/2014 Jun-16 60-7070-003 3066895 6/13/2014 Jun-16 60-7070-006 3065714 6/13/2014 Jun-16 60-7070-006 3065780 6/13/2014 Jun-16 60-7070-006 3067371 6/13/2014 Jun-16 60-7070-106 3064897 6/13/2014 Jun-16 60-7070-106 3066637 6/13/2014 Jun-16 60-7070-106 3066732 6/13/2014 Jun-16 60-7070-106 3067606 6/13/2014 Jun-16 60-7070-106 3067929 6/13/2014 Jun-16 60-7070-106 3069860 6/13/2014 Jun-16 60-7070-106 3070890 6/13/2014 Jun-16 60-7075-003 3070963 6/13/2014 Jun-16 60-7075-006 3067363 6/13/2014 Jun-16 60-7075-006 3070583 6/13/2014 Jun-16 60-7080-102 3066905 6/13/2014 Jun-16 60-7075-103 3066616 6/13/2014 Jun-16 60-7075-103 3067741 6/13/2014 Jun-16 60-7075-103 3070959 6/13/2014 Jun-16 60-7075-105 3067469 6/13/2014 Jun-16 60-7075-106 3066824 6/13/2014 Jun-16 60-7075-106 3067794 6/13/2014 Jun-16 60-7075-106 3070066 6/13/2014 Jun-16 5921-024-135 3070410 6/16/2014 Jun-16 5921-024-235 3075460 6/16/2014 Jun-16 5921-030-135 3070740 6/16/2014 Jun-16 5921-030-135 3076267 6/16/2014 Jun-16 5921-030-235 3075350 6/16/2014 Jun-16 5921-034-135 3073317 6/16/2014 Jun-16 5921-034-135 3075769 6/16/2014 Jun-16 5921-034-235 3072399 6/16/2014 Jun-16 5921-034-235 3072489 6/16/2014 Jun-16 5921-034-235 3075101 6/16/2014 Jun-16 5921-034-235 3077237 6/16/2014 Jun-16 5921-218-235 3071096 6/16/2014 Jun-16 5921-218-235 3073166 6/16/2014 Jun-16 5921-218-235 3077263 6/16/2014 Jun-16 5921-218-235 3078767 6/16/2014 Jun-16 5921-212-135 3076813 6/16/2014 Jun-16 5921-218-135 3073956 6/16/2014 Jun-16 5921-218-135 3075908 6/16/2014 Jun-16 60-7070-103 3071168 6/16/2014 Jun-16 60-7070-103 3072462 6/16/2014 Jun-16 60-7070-103 3074636 6/16/2014 Jun-16 60-7070-103 3074881 6/16/2014 Jun-16 60-7070-103 3075494 6/16/2014 Jun-16 60-7070-103 3077179 6/16/2014 Jun-16 60-7070-105 3071110 6/16/2014 Jun-16 60-7070-105 3073624 6/16/2014 Jun-16 60-7070-105 3075418 6/16/2014 Jun-16 60-7070-105 3077271 6/16/2014 Jun-16 60-7070-104 3072135 6/16/2014 Jun-16 60-7070-104 3075879 6/16/2014 Jun-16 60-7070-003 3075172 6/16/2014 Jun-16 60-7070-006 3074092 6/16/2014 Jun-16 60-7070-006 3075027 6/16/2014 Jun-16 60-7070-106 3071229 6/16/2014 Jun-16 60-7070-106 3072040 6/16/2014 Jun-16 60-7070-106 3074059 6/16/2014 Jun-16 60-7070-106 3075563 6/16/2014 Jun-16 60-7070-106 3075626 6/16/2014 Jun-16 60-7070-106 3075795 6/16/2014 Jun-16 60-7070-106 3078087 6/16/2014 Jun-16 60-7070-107 3075062 6/16/2014 Jun-16 60-7075-006 3075299 6/16/2014 Jun-16 60-7075-006 3075402 6/16/2014 Jun-16 60-7075-103 3074966 6/16/2014 Jun-16 60-7075-103 3077817 6/16/2014 Jun-16 60-7075-105 3073227 6/16/2014 Jun-16 60-7075-105 3077829 6/16/2014 Jun-16 60-7075-106 3072254 6/16/2014 Jun-16 60-7075-106 3076280 6/16/2014 Jun-16 5921-018-235 3081082 6/19/2014 Jun-16 5921-024-135 3079098 6/19/2014 Jun-16 5921-024-235 3080650 6/19/2014 Jun-16 5921-024-235 3084833 6/19/2014 Jun-16 5921-030-135 3078883 6/19/2014 Jun-16 5921-030-135 3084155 6/19/2014 Jun-16 5921-030-235 3077801 6/19/2014 Jun-16 5921-030-235 3080106 6/19/2014 Jun-16 5921-030-235 3081727 6/19/2014 Jun-16 5921-034-135 3079149 6/19/2014 Jun-16 5921-034-136 3080904 6/19/2014 Jun-16 5921-034-235 3079288 6/19/2014 Jun-16 5921-034-235 3079578 6/19/2014 Jun-16 5921-034-235 3081356 6/19/2014 Jun-16 5921-034-235 3081903 6/19/2014 Jun-16 5921-034-235 3083448 6/19/2014 Jun-16 5921-218-235 3078780 6/19/2014 Jun-16 5921-218-235 3079702 6/19/2014 Jun-16 5921-218-2
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Stryker Sustainability Solutions
- Reason for Recall:
- Issue with the process for detection of leaking Pressure Tourniquet Cuffs (PTC). Leaking PTC devices may result in excessive blood loss requiring surgical or medical intervention or systemic leakage of local anesthetic during Bier Block requiring medical intervention.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Reprocessed Pressure Tourniquet Cuff (PTC) Tourniquet cuffs are single- or dual-bladder inflatable cuffs connected to a tourniquet system via a hose assembly. When wrapped around a limb and inflated, tourniquet cuffs apply an adequate amount of pressure on the arterial blood flow in a limb to create a bloodless surgical field. Tourniquet cuffs are available in a variety of sizes to accommodate a wide range of limb circumferences.
Product Codes/Lot Numbers:
Model Lot Lot Created Expiration 5921-018-135 2990595 5/5/2014 May-16 5921-018-235 2989898 5/5/2014 May-16 5921-018-235 2992868 5/5/2014 May-16 5921-024-235 2992501 5/5/2014 May-16 5921-030-135 2990147 5/5/2014 May-16 5921-030-135 2991575 5/5/2014 May-16 5921-030-235 2993033 5/5/2014 May-16 5921-034-135 2990152 5/5/2014 May-16 5921-034-135 2992555 5/5/2014 May-16 5921-034-235 2989674 5/5/2014 May-16 5921-034-235 2990240 5/5/2014 May-16 5921-034-235 2991095 5/5/2014 May-16 5921-034-235 2991413 5/5/2014 May-16 5921-034-235 2993046 5/5/2014 May-16 5921-034-235 2993509 5/5/2014 May-16 5921-218-235 2990217 5/5/2014 May-16 5921-218-235 2990553 5/5/2014 May-16 5921-218-235 2991743 5/5/2014 May-16 5921-218-135 2990106 5/5/2014 May-16 5921-218-135 2992540 5/5/2014 May-16 60-7070-102 2990697 5/5/2014 May-16 60-7070-103 2984111 5/5/2014 May-16 60-7070-103 2989473 5/5/2014 May-16 60-7070-103 2990110 5/5/2014 May-16 60-7070-103 2991030 5/5/2014 May-16 60-7070-103 2992398 5/5/2014 May-16 60-7070-103 2992636 5/5/2014 May-16 60-7070-105 2989822 5/5/2014 May-16 60-7070-105 2990012 5/5/2014 May-16 60-7070-105 2991271 5/5/2014 May-16 60-7070-105 2993392 5/5/2014 May-16 60-7070-104 2992643 5/5/2014 May-16 60-7070-106 2990433 5/5/2014 May-16 60-7070-106 2991817 5/5/2014 May-16 60-7070-106 2993407 5/5/2014 May-16 60-7075-004 2989867 5/5/2014 May-16 60-7075-103 2990650 5/5/2014 May-16 60-7075-103 2991432 5/5/2014 May-16 60-7075-103 2992627 5/5/2014 May-16 60-7075-105 2992282 5/5/2014 May-16 60-7075-106 2990451 5/5/2014 May-16 60-7075-106 2990958 5/5/2014 May-16 5921-018-135 2996601 5/8/2014 May-16 5921-024-135 2996808 5/8/2014 May-16 5921-030-235 2995775 5/8/2014 May-16 5921-030-235 2998064 5/8/2014 May-16 5921-034-135 2996324 5/8/2014 May-16 5921-034-235 2994193 5/8/2014 May-16 5921-034-235 2995233 5/8/2014 May-16 5921-034-235 2996975 5/8/2014 May-16 5921-218-235 2994186 5/8/2014 May-16 5921-218-235 2996503 5/8/2014 May-16 5921-218-235 2997742 5/8/2014 May-16 5921-218-135 2996590 5/8/2014 May-16 60-7070-103 2993940 5/8/2014 May-16 60-7070-103 2994661 5/8/2014 May-16 60-7070-103 2995417 5/8/2014 May-16 60-7070-103 2995747 5/8/2014 May-16 60-7070-103 2996779 5/8/2014 May-16 60-7070-103 2996896 5/8/2014 May-16 60-7070-105 2994062 5/8/2014 May-16 60-7070-105 2994707 5/8/2014 May-16 60-7070-105 2995279 5/8/2014 May-16 60-7070-105 2996094 5/8/2014 May-16 60-7070-105 2997735 5/8/2014 May-16 60-7070-105 2997902 5/8/2014 May-16 60-7070-104 2994473 5/8/2014 May-16 60-7070-104 2997034 5/8/2014 May-16 60-7070-003 2997757 5/8/2014 May-16 60-7070-006 2997496 5/8/2014 May-16 60-7070-106 2995752 5/8/2014 May-16 60-7070-106 2996232 5/8/2014 May-16 60-7070-106 2997047 5/8/2014 May-16 60-7070-107 2995995 5/8/2014 May-16 60-7075-003 2994460 5/8/2014 May-16 60-7075-006 2994782 5/8/2014 May-16 60-7075-103 2996143 5/8/2014 May-16 60-7075-103 2996981 5/8/2014 May-16 60-7075-104 2994544 5/8/2014 May-16 60-7075-106 2994879 5/8/2014 May-16 60-7075-106 2997132 5/8/2014 May-16 5921-018-235 3001050 5/12/2014 May-16 5921-024-135 3002574 5/12/2014 May-16 5921-024-235 3000098 5/12/2014 May-16 5921-030-135 2999078 5/12/2014 May-16 5921-030-135 3001370 5/12/2014 May-16 5921-030-235 2999888 5/12/2014 May-16 5921-030-235 2999944 5/12/2014 May-16 5921-030-235 3001009 5/12/2014 May-16 5921-034-135 2998628 5/12/2014 May-16 5921-034-135 2999717 5/12/2014 May-16 5921-034-135 3001084 5/12/2014 May-16 5921-034-136 3001358 5/12/2014 May-16 5921-034-235 2996307 5/12/2014 May-16 5921-034-235 2998231 5/12/2014 May-16 5921-034-235 2999013 5/12/2014 May-16 5921-034-235 3001679 5/12/2014 May-16 5921-034-235 3002934 5/12/2014 May-16 5921-034-236 3001738 5/12/2014 May-16 5921-212-235 3000108 5/12/2014 May-16 5921-218-235 3000804 5/12/2014 May-16 5921-218-235 3001508 5/12/2014 May-16 5921-044-135 3001362 5/12/2014 May-16 5921-218-135 3000316 5/12/2014 May-16 5921-218-136 3001389 5/12/2014 May-16 60-7070-103 2998671 5/12/2014 May-16 60-7070-103 2999294 5/12/2014 May-16 60-7070-103 3000363 5/12/2014 May-16 60-7070-103 3001013 5/12/2014 May-16 60-7070-103 3001622 5/12/2014 May-16 60-7070-103 3002136 5/12/2014 May-16 60-7070-105 2999021 5/12/2014 May-16 60-7070-105 2999343 5/12/2014 May-16 60-7070-105 3001633 5/12/2014 May-16 60-7070-105 3001772 5/12/2014 May-16 60-7070-105 3003024 5/12/2014 May-16 60-7070-104 2999622 5/12/2014 May-16 60-7070-104 3003302 5/12/2014 May-16 60-7070-003 3001434 5/12/2014 May-16 60-7070-004 3001680 5/12/2014 May-16 60-7070-106 2998072 5/12/2014 May-16 60-7070-106 2999336 5/12/2014 May-16 60-7070-106 3000702 5/12/2014 May-16 60-7070-106 3001646 5/12/2014 May-16 60-7070-106 3002526 5/12/2014 May-16 60-7070-107 2998980 5/12/2014 May-16 60-7075-003 2999740 5/12/2014 May-16 60-7075-004 3000067 5/12/2014 May-16 60-7075-005 3000344 5/12/2014 May-16 60-7075-006 2998732 5/12/2014 May-16 60-7075-006 2998914 5/12/2014 May-16 60-7075-102 2999633 5/12/2014 May-16 60-7075-103 2999009 5/12/2014 May-16 60-7075-104 2998778 5/12/2014 May-16 60-7075-106 2998922 5/12/2014 May-16 60-7075-106 3001391 5/12/2014 May-16 60-7075-107 3001740 5/12/2014 May-16 5921-018-135 3006954 5/14/2014 May-16 5921-018-135 3007582 5/14/2014 May-16 5921-018-235 3009132 5/14/2014 May-16 5921-024-135 2984861 5/14/2014 May-16 5921-030-135 3005777 5/14/2014 May-16 5921-030-235 3004709 5/14/2014 May-16 5921-030-235 3008589 5/14/2014 May-16 5921-030-235 3008656 5/14/2014 May-16 5921-030-235 3009386 5/14/2014 May-16 5921-034-135 3005442 5/14/2014 May-16 5921-034-235 3003387 5/14/2014 May-16 5921-034-235 3003704 5/14/2014 May-16 5921-034-235 3006724 5/14/2014 May-16 5921-034-235 3008976 5/14/2014 May-16 5921-218-235 2935644 5/14/2014 May-16 5921-218-235 3003450 5/14/2014 May-16 5921-218-235 3003893 5/14/2014 May-16 5921-218-235 3006439 5/14/2014 May-16 5921-218-235 3008090 5/14/2014 May-16 5921-218-235 3009099 5/14/2014 May-16 5921-044-235 3003840 5/14/2014 May-16 5921-218-135 3003957 5/14/2014 May-16 60-7070-103 3003935 5/14/2014 May-16 60-7070-103 3005194 5/14/2014 May-16 60-7070-103 3006603 5/14/2014 May-16 60-7070-103 3008660 5/14/2014 May-16 60-7070-105 2935323 5/14/2014 May-16 60-7070-105 3004489 5/14/2014 May-16 60-7070-105 3006091 5/14/2014 May-16 60-7070-105 3007991 5/14/2014 May-16 60-7070-105 3009113 5/14/2014 May-16 60-7070-104 3006949 5/14/2014 May-16 60-7070-104 3008429 5/14/2014 May-16 60-7070-106 3003552 5/14/2014 May-16 60-7070-106 3005488 5/14/2014 May-16 60-7070-106 3006384 5/14/2014 May-16 60-7070-106 3006487 5/14/2014 May-16 60-7070-106 3007229 5/14/2014 May-16 60-7070-106 3008969 5/14/2014 May-16 60-7070-106 3009109 5/14/2014 May-16 60-7070-106 3009291 5/14/2014 May-16 60-7070-107 3007763 5/14/2014 May-16 60-7075-005 3004539 5/14/2014 May-16 60-7075-006 3003829 5/14/2014 May-16 60-7075-103 3003048 5/14/2014 May-16 60-7075-103 3006494 5/14/2014 May-16 60-7075-104 3005765 5/14/2014 May-16 60-7075-105 3006193 5/14/2014 May-16 60-7075-106 3004573 5/14/2014 May-16 60-7075-106 3004721 5/14/2014 May-16 5921-018-135 3013487 5/16/2014 May-16 5921-024-135 3009925 5/16/2014 May-16 5921-030-135 3011981 5/16/2014 May-16 5921-030-135 3013460 5/16/2014 May-16 5921-030-235 3010399 5/16/2014 May-16 5921-030-235 3012499 5/16/2014 May-16 5921-030-235 3014359 5/16/2014 May-16 5921-034-135 3009420 5/16/2014 May-16 5921-034-135 3010671 5/16/2014 May-16 5921-034-135 3012218 5/16/2014 May-16 5921-034-136 3012819 5/16/2014 May-16 5921-034-235 3011975 5/16/2014 May-16 5921-034-235 3013688 5/16/2014 May-16 5921-034-236 3012826 5/16/2014 May-16 5921-218-235 3011952 5/16/2014 May-16 5921-218-235 3012512 5/16/2014 May-16 5921-044-135 3013456 5/16/2014 May-16 5921-218-135 3011330 5/16/2014 May-16 5921-218-135 3012531 5/16/2014 May-16 5921-218-136 3012835 5/16/2014 May-16 60-7070-102 3010017 5/16/2014 May-16 60-7070-102 3010310 5/16/2014 May-16 60-7070-103 3009412 5/16/2014 May-16 60-7070-103 3010055 5/16/2014 May-16 60-7070-103 3011561 5/16/2014 May-16 60-7070-103 3012362 5/16/2014 May-16 60-7070-103 3012549 5/16/2014 May-16 60-7070-103 3013851 5/16/2014 May-16 60-7070-105 3010352 5/16/2014 May-16 60-7070-105 3012358 5/16/2014 May-16 60-7070-105 3013661 5/16/2014 May-16 60-7070-105 3014870 5/16/2014 May-16 60-7070-104 3010980 5/16/2014 May-16 60-7070-104 3013036 5/16/2014 May-16 60-7070-104 3014578 5/16/2014 May-16 60-7070-003 3009408 5/16/2014 May-16 60-7070-003 3014051 5/16/2014 May-16 60-7070-005 3011353 5/16/2014 May-16 60-7070-006 3010029 5/16/2014 May-16 60-7070-006 3013350 5/16/2014 May-16 60-7070-106 3011705 5/16/2014 May-16 60-7070-106 3012916 5/16/2014 May-16 60-7070-106 3013062 5/16/2014 May-16 60-7070-106 3013898 5/16/2014 May-16 60-7070-107 3013674 5/16/2014 May-16 60-7075-003 3009372 5/16/2014 May-16 60-7075-003 3014584 5/16/2014 May-16 60-7075-004 3012842 5/16/2014 May-16 60-7075-006 3010784 5/16/2014 May-16 60-7075-006 3010922 5/16/2014 May-16 60-7075-104 3012900 5/16/2014 May-16 60-7075-106 3009121 5/16/2014 May-16 60-7075-106 3013439 5/16/2014 May-16 60-7075-107 3014067 5/16/2014 May-16 5921-018-235 3020828 5/20/2014 May-16 5921-024-235 3017331 5/20/2014 May-16 5921-030-135 3015291 5/20/2014 May-16 5921-030-235 3016575 5/20/2014 May-16 5921-030-235 3020317 5/20/2014 May-16 5921-034-135 3014862 5/20/2014 May-16 5921-034-135 3015372 5/20/2014 May-16 5921-034-135 3018293 5/20/2014 May-16 5921-034-135 3018728 5/20/2014 May-16 5921-034-235 3015075 5/20/2014 May-16 5921-034-235 3016568 5/20/2014 May-16 5921-034-235 3019276 5/20/2014 May-16 5921-034-235 3019754 5/20/2014 May-16 5921-034-235 3019932 5/20/2014 May-16 5921-034-235 3020551 5/20/2014 May-16 5921-218-235 3016285 5/20/2014 May-16 5921-218-235 3018110 5/20/2014 May-16 5921-218-235 3019950 5/20/2014 May-16 5921-044-235 3020863 5/20/2014 May-16 5921-218-135 3015081 5/20/2014 May-16 5921-218-135 3018303 5/20/2014 May-16 5921-218-236 3015624 5/20/2014 May-16 60-7070-103 3015068 5/20/2014 May-16 60-7070-103 3016715 5/20/2014 May-16 60-7070-103 3017019 5/20/2014 May-16 60-7070-103 3018335 5/20/2014 May-16 60-7070-103 3020034 5/20/2014 May-16 60-7070-105 3016151 5/20/2014 May-16 60-7070-105 3019501 5/20/2014 May-16 60-7070-104 3018818 5/20/2014 May-16 60-7070-003 3018666 5/20/2014 May-16 60-7070-005 3021080 5/20/2014 May-16 60-7070-106 2985131 5/20/2014 May-16 60-7070-106 3016886 5/20/2014 May-16 60-7070-106 3017706 5/20/2014 May-16 60-7070-106 3019156 5/20/2014 May-16 60-7070-107 3015630 5/20/2014 May-16 60-7070-107 3020844 5/20/2014 May-16 60-7075-006 3020892 5/20/2014 May-16 60-7080-102 3015227 5/20/2014 May-16 60-7075-103 3016445 5/20/2014 May-16 60-7075-103 3018123 5/20/2014 May-16 60-7075-105 3018264 5/20/2014 May-16 60-7075-106 3015364 5/20/2014 May-16 60-7075-106 3015682 5/20/2014 May-16 60-7075-106 3016484 5/20/2014 May-16 60-7075-106 3020598 5/20/2014 May-16 5921-024-135 3021448 5/21/2014 May-16 5921-024-235 3021476 5/21/2014 May-16 5921-024-235 3022897 5/21/2014 May-16 5921-030-135 3022504 5/21/2014 May-16 5921-030-136 3023374 5/21/2014 May-16 5921-030-235 3022074 5/21/2014 May-16 5921-030-235 3024094 5/21/2014 May-16 5921-034-135 3023841 5/21/2014 May-16 5921-034-135 3025072 5/21/2014 May-16 5921-034-235 3021219 5/21/2014 May-16 5921-034-235 3022233 5/21/2014 May-16 5921-034-235 3023943 5/21/2014 May-16 5921-218-235 3021521 5/21/2014 May-16 5921-218-235 3025174 5/21/2014 May-16 5921-044-135 3025038 5/21/2014 May-16 5921-218-135 3024209 5/21/2014 May-16 60-7070-102 3020871 5/21/2014 May-16 60-7070-103 3021430 5/21/2014 May-16 60-7070-103 3022171 5/21/2014 May-16 60-7070-103 3022377 5/21/2014 May-16 60-7070-103 3022866 5/21/2014 May-16 60-7070-103 3024382 5/21/2014 May-16 60-7070-103 3025685 5/21/2014 May-16 60-7070-105 3022019 5/21/2014 May-16 60-7070-105 3023785 5/21/2014 May-16 60-7070-105 3025119 5/21/2014 May-16 60-7070-104 3017320 5/21/2014 May-16 60-7070-104 3022694 5/21/2014 May-16 60-7070-004 3023335 5/21/2014 May-16 60-7070-005 3021344 5/21/2014 May-16 60-7070-006 3023181 5/21/2014 May-16 60-7070-106 3021226 5/21/2014 May-16 60-7070-106 3021929 5/21/2014 May-16 60-7070-106 3022850 5/21/2014 May-16 60-7070-106 3023716 5/21/2014 May-16 60-7070-106 3025029 5/21/2014 May-16 60-7070-106 3025220 5/21/2014 May-16 60-7070-107 3023264 5/21/2014 May-16 60-7075-003 3023989 5/21/2014 May-16 60-7075-005 3016649 5/21/2014 May-16 60-7075-006 3023839 5/21/2014 May-16 60-7080-102 3022591 5/21/2014 May-16 60-7075-103 3022312 5/21/2014 May-16 60-7075-103 3024056 5/21/2014 May-16 60-7075-104 3021203 5/21/2014 May-16 60-7075-106 3022354 5/21/2014 May-16 60-7075-106 3025161 5/21/2014 May-16 5921-018-135 3029394 5/27/2014 May-16 5921-024-135 3028397 5/27/2014 May-16 5921-024-235 3026224 5/27/2014 May-16 5921-024-235 3030968 5/27/2014 May-16 5921-030-135 3026211 5/27/2014 May-16 5921-030-235 3026085 5/27/2014 May-16 5921-030-235 3028129 5/27/2014 May-16 5921-030-235 3031570 5/27/2014 May-16 5921-034-135 3027531 5/27/2014 May-16 5921-034-135 3029095 5/27/2014 May-16 5921-034-135 3030896 5/27/2014 May-16 5921-034-136 3025671 5/27/2014 May-16 5921-034-136 3030590 5/27/2014 May-16 5921-034-235 3026118 5/27/2014 May-16 5921-034-235 3026783 5/27/2014 May-16 5921-034-235 3029510 5/27/2014 May-16 5921-034-235 3029864 5/27/2014 May-16 5921-034-235 3030170 5/27/2014 May-16 5921-034-236 3026540 5/27/2014 May-16 5921-212-235 3029179 5/27/2014 May-16 5921-218-235 3026461 5/27/2014 May-16 5921-218-235 3026777 5/27/2014 May-16 5921-218-235 3030923 5/27/2014 May-16 5921-218-235 3031845 5/27/2014 May-16 5921-044-235 3029523 5/27/2014 May-16 5921-218-135 3027078 5/27/2014 May-16 5921-218-135 3027391 5/27/2014 May-16 5921-218-135 3029957 5/27/2014 May-16 5921-218-135 3031440 5/27/2014 May-16 5921-218-136 3026529 5/27/2014 May-16 60-7070-103 3025914 5/27/2014 May-16 60-7070-103 3026712 5/27/2014 May-16 60-7070-103 3027111 5/27/2014 May-16 60-7070-103 3028091 5/27/2014 May-16 60-7070-103 3028996 5/27/2014 May-16 60-7070-103 3029341 5/27/2014 May-16 60-7070-103 3030542 5/27/2014 May-16 60-7070-105 3026457 5/27/2014 May-16 60-7070-105 3026720 5/27/2014 May-16 60-7070-105 3027520 5/27/2014 May-16 60-7070-105 3030428 5/27/2014 May-16 60-7070-104 3027169 5/27/2014 May-16 60-7070-104 3030947 5/27/2014 May-16 60-7070-003 3025830 5/27/2014 May-16 60-7070-005 3029838 5/27/2014 May-16 60-7070-106 3025980 5/27/2014 May-16 60-7070-106 3027370 5/27/2014 May-16 60-7070-106 3028635 5/27/2014 May-16 60-7070-106 3030162 5/27/2014 May-16 60-7070-107 3025848 5/27/2014 May-16 60-7070-107 3027786 5/27/2014 May-16 60-7070-107 3029942 5/27/2014 May-16 60-7075-003 3028402 5/27/2014 May-16 60-7075-004 3027130 5/27/2014 May-16 60-7075-006 3028321 5/27/2014 May-16 60-7075-006 3031279 5/27/2014 May-16 60-7075-102 3025784 5/27/2014 May-16 60-7075-103 3031438 5/27/2014 May-16 60-7075-104 3027124 5/27/2014 May-16 60-7075-106 3028117 5/27/2014 May-16 60-7075-106 3028313 5/27/2014 May-16 60-7075-106 3029167 5/27/2014 May-16 60-7075-106 3031543 5/27/2014 May-16 60-7075-106 3031768 5/27/2014 May-16 60-7075-107 3029146 5/27/2014 May-16 60-7075-107 3031251 5/27/2014 May-16 5921-018-135 3039220 5/29/2014 May-16 5921-018-235 3034408 5/29/2014 May-16 5921-024-135 3040193 5/29/2014 May-16 5921-024-135 3040313 5/29/2014 May-16 5921-024-235 3035743 5/29/2014 May-16 5921-024-235 3039323 5/29/2014 May-16 5921-030-135 3037252 5/29/2014 May-16 5921-030-135 3037473 5/29/2014 May-16 5921-030-136 3040908 5/29/2014 May-16 5921-030-235 3033662 5/29/2014 May-16 5921-030-235 3035198 5/29/2014 May-16 5921-030-235 3038667 5/29/2014 May-16 5921-030-235 3040384 5/29/2014 May-16 5921-034-135 3032079 5/29/2014 May-16 5921-034-135 3034262 5/29/2014 May-16 5921-034-135 3035497 5/29/2014 May-16 5921-034-135 3039824 5/29/2014 May-16 5921-034-135 3041102 5/29/2014 May-16 5921-034-235 3033187 5/29/2014 May-16 5921-034-235 3034178 5/29/2014 May-16 5921-034-235 3034301 5/29/2014 May-16 5921-034-235 3035926 5/29/2014 May-16 5921-034-235 3037366 5/29/2014 May-16 5921-034-235 3040657 5/29/2014 May-16 5921-034-235 3041092 5/29/2014 May-16 5921-212-235 3040172 5/29/2014 May-16 5921-218-235 3031911 5/29/2014 May-16 5921-218-235 3033379 5/29/2014 May-16 5921-218-235 3034386 5/29/2014 May-16 5921-218-235 3036335 5/29/2014 May-16 5921-218-235 3037642 5/29/2014 May-16 5921-218-235 3039637 5/29/2014 May-16 5921-218-235 3040418 5/29/2014 May-16 5921-218-235 3041295 5/29/2014 May-16 5921-044-135 3032739 5/29/2014 May-16 5921-044-235 3036524 5/29/2014 May-16 5921-218-135 3034669 5/29/2014 May-16 5921-218-135 3037223 5/29/2014 May-16 5921-218-135 3039126 5/29/2014 May-16 60-7070-102 3032702 5/29/2014 May-16 60-7070-103 3031924 5/29/2014 May-16 60-7070-103 3033173 5/29/2014 May-16 60-7070-103 3034372 5/29/2014 May-16 60-7070-103 3034546 5/29/2014 May-16 60-7070-103 3035484 5/29/2014 May-16 60-7070-103 3036936 5/29/2014 May-16 60-7070-103 3037441 5/29/2014 May-16 60-7070-103 3038508 5/29/2014 May-16 60-7070-103 3039671 5/29/2014 May-16 60-7070-103 3040559 5/29/2014 May-16 60-7070-103 3041501 5/29/2014 May-16 60-7070-105 3032482 5/29/2014 May-16 60-7070-105 3032794 5/29/2014 May-16 60-7070-105 3034648 5/29/2014 May-16 60-7070-105 3036217 5/29/2014 May-16 60-7070-105 3038916 5/29/2014 May-16 60-7070-105 3040268 5/29/2014 May-16 60-7070-104 3033380 5/29/2014 May-16 60-7070-104 3035758 5/29/2014 May-16 60-7070-104 3039830 5/29/2014 May-16 60-7070-003 3034994 5/29/2014 May-16 60-7070-003 3037281 5/29/2014 May-16 60-7070-006 3037275 5/29/2014 May-16 60-7070-106 3032448 5/29/2014 May-16 60-7070-106 3032453 5/29/2014 May-16 60-7070-106 3033643 5/29/2014 May-16 60-7070-106 3034859 5/29/2014 May-16 60-7070-106 3035011 5/29/2014 May-16 60-7070-106 3036943 5/29/2014 May-16 60-7070-106 3037880 5/29/2014 May-16 60-7070-106 3037977 5/29/2014 May-16 60-7070-106 3039663 5/29/2014 May-16 60-7070-106 3040571 5/29/2014 May-16 60-7070-106 3040846 5/29/2014 May-16 60-7070-106 3041662 5/29/2014 May-16 60-7070-107 3035608 5/29/2014 May-16 60-7070-107 3039310 5/29/2014 May-16 60-7070-107 3040210 5/29/2014 May-16 60-7075-003 3034885 5/29/2014 May-16 60-7075-003 3036632 5/29/2014 May-16 60-7075-004 3037384 5/29/2014 May-16 60-7075-005 3040155 5/29/2014 May-16 60-7075-006 3036181 5/29/2014 May-16 60-7080-102 3040186 5/29/2014 May-16 60-7075-102 3040138 5/29/2014 May-16 60-7075-103 3032014 5/29/2014 May-16 60-7075-103 3033529 5/29/2014 May-16 60-7075-103 3033708 5/29/2014 May-16 60-7075-103 3037402 5/29/2014 May-16 60-7075-103 3041909 5/29/2014 May-16 60-7075-104 3033796 5/29/2014 May-16 60-7075-104 3034871 5/29/2014 May-16 60-7075-104 3039231 5/29/2014 May-16 60-7075-105 3032046 5/29/2014 May-16 60-7075-106 3033966 5/29/2014 May-16 60-7075-106 3036609 5/29/2014 May-16 60-7075-106 3040223 5/29/2014 May-16 60-7075-106 3041796 5/29/2014 May-16 5921-024-235 3043588 6/3/2014 Jun-16 5921-024-235 3045890 6/3/2014 Jun-16 5921-030-135 3043934 6/3/2014 Jun-16 5921-030-135 3045900 6/3/2014 Jun-16 5921-030-135 3046794 6/3/2014 Jun-16 5921-030-136 3043041 6/3/2014 Jun-16 5921-030-235 3042974 6/3/2014 Jun-16 5921-030-235 3043638 6/3/2014 Jun-16 5921-030-235 3044175 6/3/2014 Jun-16 5921-030-235 3045370 6/3/2014 Jun-16 5921-030-235 3045729 6/3/2014 Jun-16 5921-034-135 3042990 6/3/2014 Jun-16 5921-034-135 3043463 6/3/2014 Jun-16 5921-034-135 3044201 6/3/2014 Jun-16 5921-034-135 3046079 6/3/2014 Jun-16 5921-034-235 3042339 6/3/2014 Jun-16 5921-034-235 3042942 6/3/2014 Jun-16 5921-034-235 3043831 6/3/2014 Jun-16 5921-034-235 3044802 6/3/2014 Jun-16 5921-034-235 3046482 6/3/2014 Jun-16 5921-034-236 3042114 6/3/2014 Jun-16 5921-218-235 2935491 6/3/2014 Jun-16 5921-218-235 3043104 6/3/2014 Jun-16 5921-218-235 3044943 6/3/2014 Jun-16 5921-218-235 3046469 6/3/2014 Jun-16 5921-218-135 3042146 6/3/2014 Jun-16 5921-218-135 3043883 6/3/2014 Jun-16 5921-218-135 3045835 6/3/2014 Jun-16 5921-218-136 3039628 6/3/2014 Jun-16 60-7070-103 3042568 6/3/2014 Jun-16 60-7070-103 3043574 6/3/2014 Jun-16 60-7070-103 3044240 6/3/2014 Jun-16 60-7070-103 3045410 6/3/2014 Jun-16 60-7070-103 3046521 6/3/2014 Jun-16 60-7070-105 3042179 6/3/2014 Jun-16 60-7070-105 3043519 6/3/2014 Jun-16 60-7070-105 3044255 6/3/2014 Jun-16 60-7070-105 3045218 6/3/2014 Jun-16 60-7070-105 3046261 6/3/2014 Jun-16 60-7070-105 3046509 6/3/2014 Jun-16 60-7070-104 3041488 6/3/2014 Jun-16 60-7070-104 3043629 6/3/2014 Jun-16 60-7070-104 3045633 6/3/2014 Jun-16 60-7070-003 3043136 6/3/2014 Jun-16 60-7070-004 3043002 6/3/2014 Jun-16 60-7070-005 3043266 6/3/2014 Jun-16 60-7070-006 3043073 6/3/2014 Jun-16 60-7070-006 3045916 6/3/2014 Jun-16 60-7070-106 3043416 6/3/2014 Jun-16 60-7070-106 3044065 6/3/2014 Jun-16 60-7070-106 3044530 6/3/2014 Jun-16 60-7070-106 3046067 6/3/2014 Jun-16 60-7070-107 3042961 6/3/2014 Jun-16 60-7075-006 3043767 6/3/2014 Jun-16 60-7075-102 3046668 6/3/2014 Jun-16 60-7075-103 3042133 6/3/2014 Jun-16 60-7075-103 3043162 6/3/2014 Jun-16 60-7075-103 3044272 6/3/2014 Jun-16 60-7075-105 3043820 6/3/2014 Jun-16 60-7075-105 3044054 6/3/2014 Jun-16 60-7075-106 3045764 6/3/2014 Jun-16 5921-018-135 3049884 6/5/2014 Jun-16 5921-018-235 3048533 6/5/2014 Jun-16 5921-024-135 3047998 6/5/2014 Jun-16 5921-024-235 3049907 6/5/2014 Jun-16 5921-030-135 3046972 6/5/2014 Jun-16 5921-030-135 3051965 6/5/2014 Jun-16 5921-030-235 3046862 6/5/2014 Jun-16 5921-030-235 3049515 6/5/2014 Jun-16 5921-030-235 3051418 6/5/2014 Jun-16 5921-034-135 3049589 6/5/2014 Jun-16 5921-034-135 3051460 6/5/2014 Jun-16 5921-034-235 3048258 6/5/2014 Jun-16 5921-034-235 3049267 6/5/2014 Jun-16 5921-034-235 3050694 6/5/2014 Jun-16 5921-034-235 3051070 6/5/2014 Jun-16 5921-218-235 3049279 6/5/2014 Jun-16 5921-218-235 3050976 6/5/2014 Jun-16 5921-030-236 3046870 6/5/2014 Jun-16 5921-044-135 3051748 6/5/2014 Jun-16 5921-044-235 3041834 6/5/2014 Jun-16 5921-218-135 3047491 6/5/2014 Jun-16 5921-218-135 3050504 6/5/2014 Jun-16 5921-218-135 3051920 6/5/2014 Jun-16 5921-218-136 3051949 6/5/2014 Jun-16 60-7070-103 3048113 6/5/2014 Jun-16 60-7070-103 3048859 6/5/2014 Jun-16 60-7070-103 3048871 6/5/2014 Jun-16 60-7070-103 3049527 6/5/2014 Jun-16 60-7070-103 3050251 6/5/2014 Jun-16 60-7070-103 3051487 6/5/2014 Jun-16 60-7070-105 3047518 6/5/2014 Jun-16 60-7070-105 3049205 6/5/2014 Jun-16 60-7070-105 3050222 6/5/2014 Jun-16 60-7070-105 3051425 6/5/2014 Jun-16 60-7070-105 3051851 6/5/2014 Jun-16 60-7070-104 3046843 6/5/2014 Jun-16 60-7070-104 3049660 6/5/2014 Jun-16 60-7070-104 3051908 6/5/2014 Jun-16 60-7070-003 3051004 6/5/2014 Jun-16 60-7070-006 3050589 6/5/2014 Jun-16 60-7070-007 3048514 6/5/2014 Jun-16 60-7070-106 3046606 6/5/2014 Jun-16 60-7070-106 3047103 6/5/2014 Jun-16 60-7070-106 3048790 6/5/2014 Jun-16 60-7070-106 3049747 6/5/2014 Jun-16 60-7070-106 3049969 6/5/2014 Jun-16 60-7070-106 3051186 6/5/2014 Jun-16 60-7075-003 3048372 6/5/2014 Jun-16 60-7075-006 3048449 6/5/2014 Jun-16 60-7075-103 3047671 6/5/2014 Jun-16 60-7075-103 3050781 6/5/2014 Jun-16 60-7075-104 3031562 6/5/2014 Jun-16 60-7075-104 3051912 6/5/2014 Jun-16 60-7075-105 3047142 6/5/2014 Jun-16 60-7075-105 3048317 6/5/2014 Jun-16 60-7075-105 3051657 6/5/2014 Jun-16 60-7075-106 3047541 6/5/2014 Jun-16 60-7075-106 3049710 6/5/2014 Jun-16 5921-018-135 3053778 6/9/2014 Jun-16 5921-024-135 3056944 6/9/2014 Jun-16 5921-024-235 3052099 6/9/2014 Jun-16 5921-024-235 3054704 6/9/2014 Jun-16 5921-030-135 3056391 6/9/2014 Jun-16 5921-030-235 3052121 6/9/2014 Jun-16 5921-030-235 3053220 6/9/2014 Jun-16 5921-030-235 3054727 6/9/2014 Jun-16 5921-034-135 3053811 6/9/2014 Jun-16 5921-034-135 3055012 6/9/2014 Jun-16 5921-034-135 3056389 6/9/2014 Jun-16 5921-034-235 3052132 6/9/2014 Jun-16 5921-034-235 3053462 6/9/2014 Jun-16 5921-034-235 3054478 6/9/2014 Jun-16 5921-034-236 3054840 6/9/2014 Jun-16 5921-218-235 3052145 6/9/2014 Jun-16 5921-218-235 3053103 6/9/2014 Jun-16 5921-218-235 3054942 6/9/2014 Jun-16 5921-218-135 3055008 6/9/2014 Jun-16 5921-218-236 3054497 6/9/2014 Jun-16 60-7070-102 3053914 6/9/2014 Jun-16 60-7070-103 3052150 6/9/2014 Jun-16 60-7070-103 3052546 6/9/2014 Jun-16 60-7070-103 3053541 6/9/2014 Jun-16 60-7070-103 3054855 6/9/2014 Jun-16 60-7070-103 3056651 6/9/2014 Jun-16 60-7070-105 3052166 6/9/2014 Jun-16 60-7070-105 3053064 6/9/2014 Jun-16 60-7070-105 3054356 6/9/2014 Jun-16 60-7070-105 3055096 6/9/2014 Jun-16 60-7070-104 3052412 6/9/2014 Jun-16 60-7070-104 3052477 6/9/2014 Jun-16 60-7070-104 3054373 6/9/2014 Jun-16 60-7070-005 3054838 6/9/2014 Jun-16 60-7070-106 3052178 6/9/2014 Jun-16 60-7070-106 3052837 6/9/2014 Jun-16 60-7070-106 3053980 6/9/2014 Jun-16 60-7070-106 3054677 6/9/2014 Jun-16 60-7070-106 3055873 6/9/2014 Jun-16 60-7075-003 3054961 6/9/2014 Jun-16 60-7075-103 3052505 6/9/2014 Jun-16 60-7075-103 3054919 6/9/2014 Jun-16 60-7075-105 3054387 6/9/2014 Jun-16 60-7075-106 3052431 6/9/2014 Jun-16 60-7075-106 3053882 6/9/2014 Jun-16 60-7075-106 3054988 6/9/2014 Jun-16 5921-018-135 3050464 6/11/2014 Jun-16 5921-018-235 3060341 6/11/2014 Jun-16 5921-024-235 3057461 6/11/2014 Jun-16 5921-024-235 3063218 6/11/2014 Jun-16 5921-030-135 3058578 6/11/2014 Jun-16 5921-030-235 3057276 6/11/2014 Jun-16 5921-030-235 3064504 6/11/2014 Jun-16 5921-034-135 3059307 6/11/2014 Jun-16 5921-034-135 3061081 6/11/2014 Jun-16 5921-034-135 3064589 6/11/2014 Jun-16 5921-034-235 3057118 6/11/2014 Jun-16 5921-034-235 3058131 6/11/2014 Jun-16 5921-034-235 3059693 6/11/2014 Jun-16 5921-034-235 3062833 6/11/2014 Jun-16 5921-218-235 3058917 6/11/2014 Jun-16 5921-218-235 3059664 6/11/2014 Jun-16 5921-218-235 3062867 6/11/2014 Jun-16 5921-044-235 3059934 6/11/2014 Jun-16 5921-212-135 3057141 6/11/2014 Jun-16 5921-218-135 3059311 6/11/2014 Jun-16 5921-218-135 3063717 6/11/2014 Jun-16 5921-218-135 3064035 6/11/2014 Jun-16 60-7070-103 3057501 6/11/2014 Jun-16 60-7070-103 3057670 6/11/2014 Jun-16 60-7070-103 3058566 6/11/2014 Jun-16 60-7070-103 3060048 6/11/2014 Jun-16 60-7070-103 3061837 6/11/2014 Jun-16 60-7070-103 3064042 6/11/2014 Jun-16 60-7070-103 3064687 6/11/2014 Jun-16 60-7070-103 3064775 6/11/2014 Jun-16 60-7070-105 3058288 6/11/2014 Jun-16 60-7070-105 3058532 6/11/2014 Jun-16 60-7070-105 3059655 6/11/2014 Jun-16 60-7070-105 3060990 6/11/2014 Jun-16 60-7070-105 3064639 6/11/2014 Jun-16 60-7070-104 3057643 6/11/2014 Jun-16 60-7070-104 3062898 6/11/2014 Jun-16 60-7070-106 3057490 6/11/2014 Jun-16 60-7070-106 3057586 6/11/2014 Jun-16 60-7070-106 3059209 6/11/2014 Jun-16 60-7070-106 3060133 6/11/2014 Jun-16 60-7070-106 3063202 6/11/2014 Jun-16 60-7070-106 3064591 6/11/2014 Jun-16 60-7075-003 3063959 6/11/2014 Jun-16 60-7075-103 3057658 6/11/2014 Jun-16 60-7075-103 3060179 6/11/2014 Jun-16 60-7075-103 3063646 6/11/2014 Jun-16 60-7075-105 3058889 6/11/2014 Jun-16 60-7075-105 3063963 6/11/2014 Jun-16 60-7075-106 3057426 6/11/2014 Jun-16 60-7075-106 3058891 6/11/2014 Jun-16 60-7075-106 3059722 6/11/2014 Jun-16 60-7075-106 3064781 6/11/2014 Jun-16 5921-018-135 3065910 6/13/2014 Jun-16 5921-018-235 3066249 6/13/2014 Jun-16 5921-018-235 3067699 6/13/2014 Jun-16 5921-018-235 3070406 6/13/2014 Jun-16 5921-024-135 3066212 6/13/2014 Jun-16 5921-024-235 3066709 6/13/2014 Jun-16 5921-024-235 3067862 6/13/2014 Jun-16 5921-024-236 3070614 6/13/2014 Jun-16 5921-030-135 3065002 6/13/2014 Jun-16 5921-030-135 3070475 6/13/2014 Jun-16 5921-030-235 3070021 6/13/2014 Jun-16 5921-034-135 3064907 6/13/2014 Jun-16 5921-034-135 3066161 6/13/2014 Jun-16 5921-034-135 3067827 6/13/2014 Jun-16 5921-034-135 3070370 6/13/2014 Jun-16 5921-034-135 3070985 6/13/2014 Jun-16 5921-034-136 3065722 6/13/2014 Jun-16 5921-034-235 3062487 6/13/2014 Jun-16 5921-034-235 3065400 6/13/2014 Jun-16 5921-034-235 3069374 6/13/2014 Jun-16 5921-034-236 3070755 6/13/2014 Jun-16 5921-218-235 3068529 6/13/2014 Jun-16 5921-218-235 3070888 6/13/2014 Jun-16 5921-044-135 3066265 6/13/2014 Jun-16 5921-212-135 3067804 6/13/2014 Jun-16 5921-218-135 3065561 6/13/2014 Jun-16 5921-218-135 3067668 6/13/2014 Jun-16 5921-218-135 3069065 6/13/2014 Jun-16 60-7070-102 3070148 6/13/2014 Jun-16 60-7070-103 3065143 6/13/2014 Jun-16 60-7070-103 3065376 6/13/2014 Jun-16 60-7070-103 3066671 6/13/2014 Jun-16 60-7070-103 3067550 6/13/2014 Jun-16 60-7070-103 3067584 6/13/2014 Jun-16 60-7070-103 3067836 6/13/2014 Jun-16 60-7070-103 3069123 6/13/2014 Jun-16 60-7070-103 3070729 6/13/2014 Jun-16 60-7070-105 3065087 6/13/2014 Jun-16 60-7070-105 3065865 6/13/2014 Jun-16 60-7070-105 3066830 6/13/2014 Jun-16 60-7070-105 3067476 6/13/2014 Jun-16 60-7070-105 3068950 6/13/2014 Jun-16 60-7070-104 3065975 6/13/2014 Jun-16 60-7070-104 3068398 6/13/2014 Jun-16 60-7070-003 3066895 6/13/2014 Jun-16 60-7070-006 3065714 6/13/2014 Jun-16 60-7070-006 3065780 6/13/2014 Jun-16 60-7070-006 3067371 6/13/2014 Jun-16 60-7070-106 3064897 6/13/2014 Jun-16 60-7070-106 3066637 6/13/2014 Jun-16 60-7070-106 3066732 6/13/2014 Jun-16 60-7070-106 3067606 6/13/2014 Jun-16 60-7070-106 3067929 6/13/2014 Jun-16 60-7070-106 3069860 6/13/2014 Jun-16 60-7070-106 3070890 6/13/2014 Jun-16 60-7075-003 3070963 6/13/2014 Jun-16 60-7075-006 3067363 6/13/2014 Jun-16 60-7075-006 3070583 6/13/2014 Jun-16 60-7080-102 3066905 6/13/2014 Jun-16 60-7075-103 3066616 6/13/2014 Jun-16 60-7075-103 3067741 6/13/2014 Jun-16 60-7075-103 3070959 6/13/2014 Jun-16 60-7075-105 3067469 6/13/2014 Jun-16 60-7075-106 3066824 6/13/2014 Jun-16 60-7075-106 3067794 6/13/2014 Jun-16 60-7075-106 3070066 6/13/2014 Jun-16 5921-024-135 3070410 6/16/2014 Jun-16 5921-024-235 3075460 6/16/2014 Jun-16 5921-030-135 3070740 6/16/2014 Jun-16 5921-030-135 3076267 6/16/2014 Jun-16 5921-030-235 3075350 6/16/2014 Jun-16 5921-034-135 3073317 6/16/2014 Jun-16 5921-034-135 3075769 6/16/2014 Jun-16 5921-034-235 3072399 6/16/2014 Jun-16 5921-034-235 3072489 6/16/2014 Jun-16 5921-034-235 3075101 6/16/2014 Jun-16 5921-034-235 3077237 6/16/2014 Jun-16 5921-218-235 3071096 6/16/2014 Jun-16 5921-218-235 3073166 6/16/2014 Jun-16 5921-218-235 3077263 6/16/2014 Jun-16 5921-218-235 3078767 6/16/2014 Jun-16 5921-212-135 3076813 6/16/2014 Jun-16 5921-218-135 3073956 6/16/2014 Jun-16 5921-218-135 3075908 6/16/2014 Jun-16 60-7070-103 3071168 6/16/2014 Jun-16 60-7070-103 3072462 6/16/2014 Jun-16 60-7070-103 3074636 6/16/2014 Jun-16 60-7070-103 3074881 6/16/2014 Jun-16 60-7070-103 3075494 6/16/2014 Jun-16 60-7070-103 3077179 6/16/2014 Jun-16 60-7070-105 3071110 6/16/2014 Jun-16 60-7070-105 3073624 6/16/2014 Jun-16 60-7070-105 3075418 6/16/2014 Jun-16 60-7070-105 3077271 6/16/2014 Jun-16 60-7070-104 3072135 6/16/2014 Jun-16 60-7070-104 3075879 6/16/2014 Jun-16 60-7070-003 3075172 6/16/2014 Jun-16 60-7070-006 3074092 6/16/2014 Jun-16 60-7070-006 3075027 6/16/2014 Jun-16 60-7070-106 3071229 6/16/2014 Jun-16 60-7070-106 3072040 6/16/2014 Jun-16 60-7070-106 3074059 6/16/2014 Jun-16 60-7070-106 3075563 6/16/2014 Jun-16 60-7070-106 3075626 6/16/2014 Jun-16 60-7070-106 3075795 6/16/2014 Jun-16 60-7070-106 3078087 6/16/2014 Jun-16 60-7070-107 3075062 6/16/2014 Jun-16 60-7075-006 3075299 6/16/2014 Jun-16 60-7075-006 3075402 6/16/2014 Jun-16 60-7075-103 3074966 6/16/2014 Jun-16 60-7075-103 3077817 6/16/2014 Jun-16 60-7075-105 3073227 6/16/2014 Jun-16 60-7075-105 3077829 6/16/2014 Jun-16 60-7075-106 3072254 6/16/2014 Jun-16 60-7075-106 3076280 6/16/2014 Jun-16 5921-018-235 3081082 6/19/2014 Jun-16 5921-024-135 3079098 6/19/2014 Jun-16 5921-024-235 3080650 6/19/2014 Jun-16 5921-024-235 3084833 6/19/2014 Jun-16 5921-030-135 3078883 6/19/2014 Jun-16 5921-030-135 3084155 6/19/2014 Jun-16 5921-030-235 3077801 6/19/2014 Jun-16 5921-030-235 3080106 6/19/2014 Jun-16 5921-030-235 3081727 6/19/2014 Jun-16 5921-034-135 3079149 6/19/2014 Jun-16 5921-034-136 3080904 6/19/2014 Jun-16 5921-034-235 3079288 6/19/2014 Jun-16 5921-034-235 3079578 6/19/2014 Jun-16 5921-034-235 3081356 6/19/2014 Jun-16 5921-034-235 3081903 6/19/2014 Jun-16 5921-034-235 3083448 6/19/2014 Jun-16 5921-218-235 3078780 6/19/2014 Jun-16 5921-218-235 3079702 6/19/2014 Jun-16 5921-218-2
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0324-2017
Related Recalls
HARMONIC FOCUS Shears + Adaptive Tissue Technology, 9 cm Length REF: HAR9F, Reprocessed Device for Single Use, STERILEEO, RX ONLY
Stryker Sustainability Solutions
Due to manufacturing issue, reprocessed shears, may become stuck in the "on position" resulting in continuous activation.
Disposable 5 Lead Cable and Lead Wire System, 50 inches, REF: LW-309DS50/5A
Stryker Sustainability Solutions
Disposable 5 lead cable and lead wire system package that should contain lead systems, lead telemetry systems and a dual connect cable, may in fact contain, different ECG Leads, which are not approved for reprocessing, but that may have been reprocessed, which may lead to devices not performing as intended.
REF: 60-7070-005, Zimmer, Tourniquet Cuff 30" (Blue), Luer-Lock Dual Port, single Bladder, RxOnly
Stryker Sustainability Solutions
Mislabeling; tourniquet cuff packaging labeled as dual port single bladder but contains a tourniquet cuff single port single bladder.