Stryker Sustainability Solutions

53 recalls β€’ β€’ Categories: Other Medical Devices, Surgical Instruments, Infusion Pumps, Diagnostic Equipment, Implants & Prosthetics

Due to potential for tourniquet cuffs to detach from the bladder at the welding connection and result in the cuff not being able to maintain pressure.

Mar 12, 2026 Other Medical Devices Nationwide View Details β†’

39-inche-wide inflatable air transfer mattress packages contain 34-Inch-wide mattresses, so there may be a decreased surface area available to distribute the patient's weight across, which may increase the effort exerted by staff for a lateral patient transfer, and there may be increased friction and shearing forces when the patient is moved resulting in skin tear injury.

Jun 21, 2023 Other Medical Devices View Details β†’

Increased incidence of disposable tourniquet cuffs "not able to achieve desired pressure" and "not holding pressure" due to 1) Cuff fabric tearing during use, and 2) Pinched or unseated O-ring at the connection between the cuff, disposable fill line, and/or cuff hosing, which may lead to loss of vascular occlusion, which may lead to operative site blood loss, hemorrhage, or other complications.

Mar 1, 2023 Other Medical Devices Nationwide View Details β†’

Product was distributed without receiving regulatory clearance.

Jan 19, 2021 Implants & Prosthetics Nationwide View Details β†’

A subset of temporary intracardiac ultrasound catheters are labeled with the incorrect expiration date. They are mislabeled as having a shelf life of two years, while the actual shelf life is one year. The earliest affected product will expire in May 2020. Use of mislabeled devices may lead to hazards sterility breach, component breakdown or delay in treatment.

Mar 27, 2020 Surgical Instruments View Details β†’

Tourniquet cuffs are potentially mislabeled. If used may lead to (1) Device misuse during procedure (2) Inadequate Limb Compression Force.

Oct 9, 2019 Other Medical Devices Nationwide View Details β†’

Affected lots of reprocessed arthroscopic shavers may display the incorrect catalog number and speed settings when connected to the generator, require manual mode for use on CORE generators, or display an error message on Crossfire 1 and 2 generators which does not allow the device to be used.

Jul 27, 2018 Implants & Prosthetics Nationwide View Details β†’

Stryker s Sustainability Solutions division (SSS) has received an increase in reports indicating that an EEPROM chip error code may occur when Reprocessed 2515 NAV eco Variable Electrophysiology (EP) Catheters are used with CARTO(R) EP Navigation Systems.

Dec 17, 2018 Diagnostic Equipment View Details β†’

Reprocessed Xcel Bladeless Trocar (H12LP) Model # H12LP Reprocessed Endoscopic Trocars12x100mm - Product Usage: Physical Product Description: The Reprocessed Blunt Tip Trocar consists of a radiolucent cannula and obturator sized 12mm in diameter (refer to Figure 1). The Reprocessed Blunt Tip Trocar has an obturator that has a blunt plastic tip that gently moves aside any internal viscera that may be adjacent to the abdominal or thoracic wall. The trocar cannula consists of two seals, an outer integrated removable self-adjusting seal that may accommodate instruments ranging from 5mm to 12mm in diameter, and an internal seal. These seals minimize gas leakage when instruments are inserted or withdrawn through the seal. The stopcock valve provides attachment for gas insufflation and desufflation. The stability plug component, is specific only to the H12LP catalog number. The stability plug (adjustable plug) is placed on the outside of the trocar sleeve (cannula). The stability plug provides suture clips (suture anchors), which can be used to suture the Trocar sleeve in place. Intended Use: Reprocessed Endoscopic Trocars are indicated for use to establish a port of entry for endoscopic instruments in patients requiring minimally invasive surgical procedures. Physical Product Description: The Reprocessed Blunt Tip Trocar consists of a radiolucent cannula and obturator sized 12mm in diameter (refer to Figure 1). The Reprocessed Blunt Tip Trocar has an obturator that has a blunt plastic tip that gently moves aside any internal viscera that may be adjacent to the abdominal or thoracic wall. The trocar cannula consists of two seals, an outer integrated removable self-adjusting seal that may accommodate instruments ranging from 5mm to 12mm in diameter, and an internal seal. These seals minimize gas leakage when instruments are inserted or withdrawn through the seal. The stopcock valve provides attachment for gas insufflation and desufflation. The stability plug component, is specific only to the H12LP catalog number. The stability plug (adjustable plug) is placed on the outside of the trocar sleeve (cannula). The stability plug provides suture clips (suture anchors), which can be used to suture the Trocar sleeve in place. Intended Use: Reprocessed Endoscopic Trocars are indicated for use to establish a port of entry for endoscopic instruments in patients requiring minimally invasive surgical procedures.

Class I - Dangerous

Through investigation, it was determined that H12LP Trocars in scope of a previous H12LP recall were distributed to facilities within the distribution chain following the recall execution activities initially performed by distributor.

Jun 19, 2018 Diagnostic Equipment Nationwide View Details β†’

Stryker Sustainability Solutions has received an increase in reports indicating that the affected Reprocessed HARH36 devices may display an error code (No Instrument Uses Remaining) upon initial connection to the generator. When this error code is present, the device is not able to be used.

May 8, 2018 Implants & Prosthetics Nationwide View Details β†’

Thrombogenicity test results demonstrated that both the test devices (SSS) and control devices (OM) did not meet the internal specification of non-thrombogenic .

Mar 29, 2018 Diagnostic Equipment Nationwide View Details β†’

An EEPROM chip error code may occur when the catheters are used with CARTO EP Navigation Systems.

Jan 2, 2018 Diagnostic Equipment Nationwide View Details β†’

An EEPROM chip error code may occur when the catheters are used with CARTO EP Navigation Systems.

Jan 2, 2018 Diagnostic Equipment Nationwide View Details β†’

A breach in the sterile packaging may have occurred resulting in potentially non-sterile product.

Dec 6, 2016 Surgical Instruments Nationwide View Details β†’

A breach in the sterile packaging may have occurred resulting in potentially non-sterile product.

Dec 6, 2016 Surgical Instruments Nationwide View Details β†’

The EP Catheters may be mislabeled for French size during reprocessing.

Oct 6, 2016 Diagnostic Equipment View Details β†’

Issue with the process for detection of leaking Pressure Tourniquet Cuffs (PTC). Leaking PTC devices may result in excessive blood loss requiring surgical or medical intervention or systemic leakage of local anesthetic during Bier Block requiring medical intervention.

Sep 19, 2016 Surgical Instruments Nationwide View Details β†’

Stryker Sustainability Solutions is recalling the Fuhrman Pleural & Pneumopericardial Drainage Set due to receiving complaints that it broke off in the pleural cavity while inserting the device into the patient.

Nov 17, 2015 Surgical Instruments View Details β†’

It was reported that packaging was not properly sealed.

Oct 14, 2015 Surgical Instruments Nationwide View Details β†’

Stryker Sustainability Solutions has received reports indicating that devices from one lot (lot# 3574704) were labeled as 100mm models (CB5LT) while the devices inside of the packaging were 75mm models (CB5ST).

Jul 10, 2015 Diagnostic Equipment Nationwide View Details β†’

Stryker Sustainability Solutions is recalling Trocars because they have received reports indicating that some may fracture at the distal tip resulting in fragments at the surgical site.

Jun 1, 2015 Surgical Instruments Nationwide View Details β†’

Stryker Sustainability Solutions is recalling Trocars because they have received reports indicating that some may fracture at the distal tip resulting in fragments at the surgical site.

Jun 1, 2015 Surgical Instruments Nationwide View Details β†’

Stryker Sustainability Solutions is recalling Trocars because they have received reports indicating that some may fracture at the distal tip resulting in fragments at the surgical site.

Jun 1, 2015 Surgical Instruments Nationwide View Details β†’

Stryker Sustainability Solutions is recalling the Reprocessed Diagnostic Electrophysiology (EP) Catheters Cables due to a potential sterility breach in the package seal.

Oct 18, 2013 Diagnostic Equipment View Details β†’