EnteraLite Infinity Enteral Feeding Pump identified with the following list numbers: INFKIT2 and INFORNG. The device is used with disposable PVC administration tubing sets to provide a means for the volumetric delivery of fluids used in enteral applications. The EnteraLite Infinity enteral feeding pump with disposable PVC administration tubing sets and user accessories provide a means for the volumetric delivery of fluids used in enteral applications.
Class I - DangerousWhat Should You Do?
- Check if you have this product: INFKIT2 with serial numbers: 506249031, 506264096, 506272024, 506335006, 506335145, 506336089, 506336096, 506362083, 506362086, 507018037, 507018060, 507067220, 507085140, 507087018, 507087140, 507095089, 507095171, 507121102, 507122050, 507122066, 507123149, 507123163, 507144016, 507144057, 507144215, 507144220, 507144221, 507144226, 507151002, 507155019, 507155021, 507155094, 507155127, 507155156, 507155159, 507156065, 507157048, 507157124, 507164009, 507173005, 507173019, 507173063, 507179123, 507180064, 507186018, 507186098, 507186176, 507197012, 507197024, 507197036, 507201013, 507201038, 507205054, 507206064, 507212012, 507212013, 507214065, 507214098, 507214162, 507215041, 507218016, 507219057, 507219087, 507219103, 507220106, 507221098, 507227116, 507229033, 507229039, 507229060, 507232030, 507236063, 507236203, 507236303, 507239056, 507241089, 507242130, 507242201, 507247049, 507248031, 507248087, 507248095, 507249131, 507250175, 507255025, 507257123, 507257129, 507261065, 507264041, 507264127, 507268121, 507268351, 507268374, 507269084, 507270096, 507270106, 507271157, 507275046, 507276154, 507278013, 507281028, 507281055, 507283119, 507284158, 507285186, 507289004, 507289120, 507289157, 507289165, 507291090, 507291168, 507291232, 507292108, 507292179, 507292339, 507295099, 507306069, 507311133, 507313050, 507313057, 507313059, 507313084, 507325044, 507330149, 507337080, 507337082, 507337085, 507339076, 507339095, 507344149, 507345001, 507345103, 507347150, 507348077, 507351090, 507353065, 507353120, 508018015, 508018090, 508030083, 508037024, 508037097, 508039099, 508039101, 508058001, 508058159, 508058160, 508066002, 508066030, 508066097, 508066123, 508066152, 508066153, 508066268, 508067033, 508067069, 508067092, 508067117, 508067166, 508067229, 508067231, 508070062, 508070228, 508071060, 508071132, 508071257, 508072146, 508072151, 508077022, 508099062, 508100033, 508112135, 508112159, 508114018, 508114111, 508116029, 508126038, 508131043, 508131079, 508131084, 508137099, 508137101, 508148152, 508148165, 508148216, 508148267, 508151050, 508151121, 508159063, 508165023, 508165034, 508165060, 508165231, 508166036, 508172055, 508172218, 508172260, 508172319, 508172334, 508198263, 508198264, 508199105, 508233083, 508233103, 508234066, 508234068, 508234078, 508234086, 508235009, 508235107, 508235111, 508235124, 508255070, 508255100, 508260115, 508260216, 508262074, 508263073, 508269064, 508275020, 508289015, 508289131, 508325051, 508326036, 508326058, 508340030, 508340100, 508351216, 508351237, 508351246, 508357105, 509005203, 509005218, 509008108, 509008270, 509008308, 509009120, 509015017, 509017118, 509020024, 509020140, 509021098, 509024156, 509030043, 509030055, 509030056, 509030088, 509034103, 509035272, 509054126, 509055029, 509055207, 509065195, 509065202, 509065212, 509071109, 509072019, 509072092, 509079210, 509085268, 509086084, 509086125, 509090011, 509090127, 509099077, 509100122, 509107033, 509119090, 509119120, 509120121, 509132033, 509141115, 509162073, 509183062, 509183092, 509197148, 509209048, 509231116, 509254016, 509258007, 509258203, 509274074, 509275014, 509281008, 509282068, 509285043, 509323066, 509324035, 509324059, 509344048, 509350005, 509350020, 509350025, 509350037, 509350072, 509350078, 509350084, 510015134, 510015149, 510015166, 510027107, 510027146, 510028041, 510050026, 510050094, 510063025, 510063107, 510076044, 510076045, 510088038, 510092021, 510124007, 510124012, 510128125, 510168018, 510168043, 510188065, 510193159, 510218008, 510218038, 510218051, 510218064, 510218066, 510218071, 510218084, 510218087, 510218097, 510218098, 510218106, 510218148, 510218156, 510221001, 510221003, 510221028, 510221087, 510222003, 510222005, 510222006, 510222011, 510222013, 510222016, 510222020, 510222032, 510222062, 510222112, 510222119, 510222121, 510229001, 510229002, 510229003, 510229004, 510229005, 510229006, 510229007, 510229008, 510229009, 510229010, 510229011, 510229012, 510229013, 510229014, 510229015, 510229016, 510229017, 510229018, 510229019, 510229020, 510229021, 510229022, 510229023, 510229024, 510229025, 510229026, 510229027, 510229028, 510229029, 510229030, 510229031, 510229032, 510229033, 510229034, 510229035, 510229036, 510229037, 510229038, 510229039, 510229040, 510229041, 510229042, 510229043, 510229044, 510229045, 510229046, 510229047, 510229048, 510229049, 510229050, 510229051, 510229052, 510229053, 510229054, 510229055, 510229056, 510229057, 510229058, 510229059, 510229060, 510229061, 510229062, 510229063, 510229064, 510229065, 510229066, 510229067, 510229068, 510229069, 510229070, 510229071, 510229072, 510229073, 510229074, 510229075, 510229076, 510229077, 510229079, 510229080, 510229081, 510229082, 510229083, 510229084, 510229085, 510229086, 510229087, 510229088, 510229089, 510229090, 510229091, 510229092, 510229093, 510229094, 510229095, 510229096, 510229097, 510229098, 510229099, 510229100, 510229101, 510229102, 510229103, 510229104, 510229105, 510229106, 510229107, 510229108, 510229109, 510229110, 510229111, 510229112, 510229113, 510229114, 510229115, 510229116, 510229117, 510229118, 510229119, 510229120, 510229121, 510229122, 510229123, 510229124, 510229125, 510229126, 510229127, 510229128, 510229129, 510229130, 510229131, 510229132, 510229133, 510229134, 510229135, 510229136, 510229137, 510229138, 510229139, 510229140, 510229141, 510229142, 510229143, 510229144, 510229145, 510229146, 510229147, 510229148, 510229149, 510229150, 510229151, 510231001, 510231002, 510231003, 510231004, 510231005, 510231006, 510231007, 510231008, 510231009, 510231010, 510231011, 510231012, 510231013, 510231014, 510231015, 510231016, 510231017, 510231018, 510231019, 510231020, 510231021, 510231022, 510231023, 510231024, 510231025, 510231026, 510231027, 510231028, 510231029, 510231030, 510231031, 510231032, 510231033, 510231034, 510231035, 510231036, 510231037, 510231038, 510231039, 510231040, 510231041, 510231042, 510231043, 510231044, 510231045, 510231046, 510231047, 510231048, 510231049, 510231050, 510231051, 510231052, 510231053, 510231054, 510231055, 510231056, 510231057, 510231058, 510231059, 510231060, 510231061, 510231062, 510231063, 510231064, 510231065, 510231066, 510231067, 510231068, 510231069, 510231070, 510231071, 510231072, 510231073, 510231074, 510232001, 510232002, 510232003, 510232004, 510232005, 510232006, 510232007, 510232008, 510232010, 510232011, 510232012, 510232013, 510232014, 510232015, 510232016, 510232017, 510232018, 510232019, 510232020, 510232021, 510232022, 510232023, 510232024, 510232025, 510232026, 510232027, 510232028, 510232029, 510232030, 510232031, 510232032, 510232033, 510232034, 510232035, 510232036, 510232037, 510232038, 510232039, 510232040, 510232041, 510232042, 510232043, 510232044, 510232045, 510232046, 510232047, 510232048, 510232049, 510232050, 510232051, 510232052, 510232053, 510232054, 510232055, 510232056, 510232057, 510232058, 510232059, 510232060, 510232061, 510232062, 510232063, 510232064, 510232065, 510232066, 510232067, 510232068, 510232069, 510232070, 510232071, 510232072, 510232073, 510232074, 510232075, 510232076, 510232077, 510232078, 510232079, 510232080, 510232081, 510232082, 510232083, 510232084, 510232085, 510232086, 510232087, 510232088, 510232089, 510232090, 510232091, 510232092, 510232093, 510232094, 510232095, 510232096, 510232098, 510232099, 510232100, 510232101, 510232102, 510232103, 510232104, 510232105, 510232106, 510232107, 510232108, 510232109, 510232110, 510232111, 510232112, 510232113, 510232114, 510232115, 510232116, 510232117, 510232118, 510232119, 510232120, 510232121, 510232122, 510232123, 510232124, 510232125, 510232126, 510232127, 510232128, 510232129, 510232130, 510232131, 510232132, 510232134, 510232135, 510232136, 510232137, 510232138, 510232139, 510232140, 510232141, 510232142, 510232143, 510232144, 510232145, 510232146, 510232147, 510232148, 510232149, 510232150, 510232151, 510232152, 510232153, 510232154, 510232155, 510232156, 510232157, 510232158, 510232159, 510232160, 510232161, 510232162, 510232163, 510232164, 510232165, 510232166, 510232167, 510232168, 510232169, 510232170, 510232171, 510232172, 510232173, 510232174, 510232175, 510232176, 510235001, 510235002, 510235003, 510235005, 510235007, 510235008, 510235009, 510235010, 510235011, 510235012, 510235013, 510235014, 510235015, 510235016, 510235017, 510235018, 510235019, 510235020, 510235021, 510235022, 510235023, 510235024, 510235025, 510235026, 510235027, 510235029, 510235030, 510235031, 510235032, 510235033, 510235034, 510235035, 510235036, 510235037, 510235038, 510235039, 510235040, 510235041, 510235042, 510235043, 510235044, 510235045, 510235046, 510235047, 510235048, 510235049, 510235050, 510235051, 510235052, 510235053, 510235054, 510235055, 510235056, 510235057, 510235058, 510235059, 510235060, 510235061, 510235062, 510235063, 510235064, 510235065, 510235066, 510235067, 510235068, 510235069, 510235070, 510235071, 510235072, 510235073, 510235074, 510235075, 510235076, 510235077, 510235078, 510235079, 510235080, 510235081, 510235082, 510235083, 510235084, 510235085, 510235086, 510235087, 510235088, 510235089, 510235090, 510235091, 510235092, 510235093, 510235094, 510235095, 510235096, 510235097, 510235098, 510235099, 510235100, 510235101, 510235102, 510235103, 510235104, 510235105, 510235106, 510235107, 510235108, 510235110, 510235111, 510235112, 510235113, 510235115, 510235116, 510235117, 510235118, 510235119, 510235120, 510235121, 510235122, 510235124, 510235125, 510235126, 510235127, 510235128, 510235129, 510235131, 510235132, 510235133, 510235134, 510235135, 510235136, 510235138, 510235140, 510235141, 510235142, 510235143, 510235144, 510235145, 510253001, 510253002, 510253003, 510253004, 510253005, 510253006, 510253007, 510253008, 510253009, 510253010, 510253011, 510253012, 510253014, 510253015, 510253016, 510253017, 510253018, 510253019, 510253020, 510253021, 510253022, 510253023, 510253024, 510253025, 510253026, 510253027, 510253028, 510253029, 510253030, 510253031, 510253032, 510253033, 510253034, 510253035, 510253036, 510253037, 510253038, 510253039, 510253040, 510253041, 510253042, 510253043, 510253044, 510253045, 510253046, 510253047, 510253048, 510253049, 510253050, 510253051, 510253052, 510253053, 510253054, 510253055, 510253056, 510253057, 510253058, 510253059, 510253060, 510253061, 510253063, 510253064, 510253065, 510253066, 510253067, 510253068, 510253069, 510253070, 510253071, 510253072, 510253073, 510253074, 510253075, 510253076, 510253077, 510253078, 510253079, 510253080, 510253081, 510253082, 510253083, 510253084, 510253085, 510253086, 510253087, 510253088, 510253089, 510253090, 510253091, 510253092, 510253093, 510253094, 510253095, 510253096, 510253097, 510253098, 510253099, 510253100, 510253101, 510253102, 510253103, 510253104, 510253105, 510253106, 510253107, 510253108, 510253109, 510253110, 510253111, 510253112, 510253113, 510253114, 510253115, 510253116, 510253117, 510253118, 510253119, 510253120, 510253121, 510253122, 510253123, 510253124, 510253125, 510253126, 510253127, 510253128, 510253129, 510253130, 510253131, 510253132, 510253133, 510253134, 510253135, 510253136, 510253137, 510253138, 510256001, 510256002, 510256003, 510256004, 510256005, 510256006, 510256007, 510256008, 510256009, 510256010, 510256011, 510256012, 510256013, 510256014, 510256015, 510256016, 510256017, 510256018, 510256019, 510256020, 510256021, 510256022, 510256023, 510256024, 510256025, 510256026, 510256027, 510256028, 510256029, 510256030, 510256031, 510256032, 510256033, 510256034, 510256035, 510256036, 510256037, 510256038, 510256039, 510256040, 510256041, 510256042, 510256043, 510256044, 510256045, 510256046, 510256047, 510256048, 510256049, 510256050, 510256051, 510256052, 510256053, 510256054, 510256055, 510256056, 510256057, 510256058, 510256059, 510256060, 510256061, 510256062, 510256063, 510256064, 510256065, 510256066, 510256067, 510256068, 510256069, 510256070, 510256071, 510256072, 510256073, 510256074, 510256075, 510256076, 510256077, 510256078, 510256079, 510256080, 510256081, 510256082, 510256083, 510256084, 510256085, 510256086, 510256087, 510256088, 510256089, 510256090, 510256091, 510256092, 510256093, 510256094, 510256095, 510256096, 510256097, 510256098, 510256099, 510256100, 510257001, 510257002, 510257003, 510257004, 510257005, 510257006, 510257007, 510257008, 510257010, 510257011, 510257012, 510257013, 510257014, 510257015, 510257016, 510257017, 510257018, 510257019, 510257020, 510257021, 510257022, 510257023, 510257024, 510257025, 510257026, 510257027, 510257028, 510257029, 510257030, 510257031, 510257032, 510257033, 510257034, 510257035, 510257036, 510257037, 510257038, 510257039, 510257040, 510257041, 510257042, 510257043, 510257044, 510257045, 510257046, 510257047, 510257048, 510257049, 510257050, 510257051, 510257052, 510257053, 510257054, 510257055, 510257056, 510257057, 510257058, 510257059, 510257060, 510257061, 510257062, 510257063, 510257064, 510257065, 510257066, 510257067, 510257068, 510257069, 510257070, 510257071, 510257072, 510257073, 510257074, 510257075, 510257076, 510257077, 510257078, 510257079, 510257080, 510257081, 510257082, 510257083, 510257084, 510257085, 510257086, 510257087, 510257088, 510257089, 510257090, 510257091, 510257092, 510257094, 510257095, 510257096, 510257097, 510257098, 510257099, 510257100, 510257101, 510257102, 510259001, 510259002, 510259003, 510259004, 510259005, 510259006, 510259007, 510259008, 510259009, 510259011, 510259012, 510259013, 510259014, 510259015, 510259016, 510259017, 510259018, 510259019, 510259020, 510259021, 510259022, 510259023, 510259024, 510259025, 510259026, 510259027, 510259028, 510259029, 510259030, 510259031, 510259032, 510259033, 510259034, 510259035, 510259036, 510259037, 510259038, 510259039, 510259040, 510259041, 510259042, 510259043, 510259044, 510259045, 510259046, 510259047, 510259048, 510259049, 510259050, 510259051, 510259052, 510259053, 510259054, 510259055, 510259056, 510259057, 510259058, 510259059, 510259060, 510259061, 510259062, 510259063, 510259064, 510259066, 510259067, 510259068, 510259069, 510259070, 510259071, 510259072, 510259073, 510259074, 510259075, 510259076, 510259077, 510259078, 510259079, 510259080, 510259081, 510259082, 510259083, 510259084, 510259085, 510259086, 510259087, 510259088, 510259089, 510259090, 510259091, 510259092, 510259093, 510259094, 510259095, 510259096, 510259097, 510259098, 510259099, 510259100, 510259101, 510259102, 510278001, 510278002, 510278003, 510278004, 510278005, 510278006, 510284002, 510284003, 510284004, 510284005, 510284006, 510284007, 510284008, 510284009, 510284010, 510284011, 510284012, 510284013, 510284014, 510284015, 510284016, 510284017, 510284018, 510284019, 510284020, 510284021, 510284022, 510284023, 510284024, 510284025, 510284026, 510284027, 510284028, 510284029, 510284030, 510284031, 510284032, 510284033, 510284034, 510284035, 510284036, 510284037, 510284038, 510284039, 510284040, 510284041, 510284042, 510284043, 510284044, 510284045, 510284046, 510284047, 510284048, 510284049, 510284050, 510284051, 510284052, 510284053, 510284054, 510284055, 510284056, 510284057, 510284058, 510284059, 510284060, 510284061, 510284062, 510284063, 510284064, 510284065, 510284066, 510284067, 510284068, 510284069, 510284070, 510284071, 510284072, 510284073, 510284074, 510284076, 510284077, 510284078, 510284079, 510284080, 510284081, 510284082, 510284083, 510284084, 510284085, 510284086, 510284087, 510284088, 510284089, 510284090, 510284091, 510284092, 510284093, 510284094, 510284095, 510284096, 510284097, 510284098, 510284099, 510284100, 510284101, 510284102, 510284103, 510284104, 510284105, 510284106, 510284107, 510284108, 510284109, 510284110, 510284111, 510284112, 510284113, 510284114, 510284115, 510284116, 510284117, 510284118, 510284119, 510284120, 510284121, 510284122, 510284123, 510284124, 510284125, 510284126, 510284127, 510284128, 510284129, 510284130, 510285001, 510285002, 510285003, 510285004, 510285005, 510285006, 510285007, 510285008, 510285009, 510285011, 510285012, 510285013, 510285014, 510285015, 510285016, 510285017, 510285018, 510285019, 510285020, 510285021, 510285022, 510285023, 510285024, 510285025, 510285026, 510285027, 510285028, 510285029, 510285030, 510285031, 510285032, 510285033, 510285034, 510285035, 510285036, 510285037, 510285038, 510285039, 510285040, 510285041, 510285042, 510285043, 510285044, 510285045, 510285046, 510285047, 510285048, 510285049, 510285050, 510285051, 510285052, 510285053, 510285054, 510285055, 510285056, 510285057, 510285059, 510285060, 510285061, 510285062, 510285063, 510285064, 510285065, 510285066, 510285067, 510285068, 510285069, 510285070, 510285071, 510285072, 510285073, 510285074, 510285075, 510285076, 510285077, 510285078, 510285079, 510285080, 510285081, 510285082, 510285083, 510285084, 510285085, 510285086, 510285087, 510285088, 510285089, 510285090, 510285091, 510285092, 510285093, 510285094, 510285095, 510285096, 510285098, 510285099, 510285100, 510285101, 510285102, 510285103, 510285104, 510285105, 510285106, 510285107, 510285108, 510285109, 510285110, 510285111, 510285112, 510285113, 510285114, 510285115, 510285116, 510285117, 510285118, 510285119, 510285120, 510285121, 510285122, 510285123, 510285124, 510285125, 510285126, 510285127, 510285128, 510285129, 510285130, 510285131, 510285132, 510285133, 510285134, 510285136, 510285138, 510285139, 510285140, 510285141, 510285142, 510285143, 510285144, 510285145, 510285146, 510285147, 510285148, 510285149, 510285150, 510285151, 510285152, 510285153, 510285154, 510285155, 510285156, 510285157, 510285158, 510285159, 510285160, 510285161, 510285162, 510285163, 510285164, 510285165, 510285166, 510285167, 510285168, 510285169, 510285170, 510285171, 510285172, 510285174, 510285175, 510285176, 510285177, 510285178, 510286001, 510286002, 510286003, 510286004, 510286005, 510286006, 510286007, 510286008, 510286009, 510286010, 510286011, 510286012, 510286013, 510286014, 510286015, 510286016, 510286018, 510286019, 510286020, 510286021, 510286022, 510286023, 510286024, 510286025, 510286026, 510286027, 510286028, 510286029, 510286030, 510286031, 510286033, 510286034, 510286035, 510286036, 510286037, 510286038, 510286039, 510286040, 510286041, 510286042, 510286043, 510286044, 510286045, 510286046, 510286047, 510286048, 510286049, 510286050, 510286051, 510286052, 510286053, 510286054, 510286055, 510286056, 510286057, 510286058, 510286059, 510286060, 510286061, 510286062, 510286063, 510286064, 510286065, 510286066, 510286067, 510286068, 510286069, 510286070, 510286071, 510286072, 510286073, 510286074, 510286075, 510286076, 510286077, 510286078, 510286079, 510286080, 510286081, 510286082, 510286083, 510286084, 510286085, 510286086, 510286087, 510286088, 510286089, 510286090, 510286091, 510286092, 510286093, 510286094, 510286095, 510286096, 510286098, 510286099, 510286100, 510286101, 510286102, 510286103, 510286104, 510286105, 510286106, 510286107, 510286108, 510286109, 510286110, 510286111, 510286112, 510286113, 510286114, 510286115, 510286116, 510286117, 510286118, 510286119, 510286120, 510286121, 510286122, 510286123, 510286124, 510286125, 510286126, 510286127, 510286128, 510286129, 510286130, 510286131, 510286132, 510286133, 510286134, 510286135, 510286136, 510286137, 510286138, 510286139, 510286140, 510286141, 510286142, 510286143, 510286144, 510286145, 510286146, 510286147, 510286148, 510286149, 510286150, 510286151, 510286152, 510286153, 510286154, 510286155, 510286156, 510286157, 510286158, 510286159, 510286160, 510286161, 510286162, 510286163, 510286164, 510286165, 510286166, 510286167, 510286168, 510286169, 510286170, 510286171, 510286172, 510286173, 510286174, 510286175, 510286177, 510286178, 510286179, 510286180, 510286181, 510286182, 510286183, 510286184, 510286185, 510286186, 510286187, 510286188, 510286189, 510286190, 510286192, 510286193, 510286194, 510286195, 510286196, 510286197, 510286198, 510286199, 510286200, 510286201, 510286202, 510286203, 510286204, 510286205, 510286206, 510286207, 510286208, 510286209, 510286210, 510286211, 510286212, 510286213, 510286214, 510286215, 510286216, 510286217, 510286218, 510286219, 510286220, 510286221, 510286222, 510286223, 510286224, 510286225, 510286226, 510286227, 510286228, 510286229, 510286230, 510286231, 510286232, 510286233, 510286234, 510286235, 510286237, 510286238, 510286239, 510286240, 510286241, 510286242, 510286243, 510286244, 510286245, 510286246, 510286247, 510286248, 510286249, 510286250, 510286251, 510286252, 510286253, 510286254, 510286255, 510286256, 510286257, 510286258, 510286259, 510286260, 510286261, 510286262, 510286263, 510286264, 510286265, 510286266, 510286267, 510286268, 510286269, 510286270, 510286271, 510286272, 510286273, 510286274, 510286275, 510286276, 510286277, 510286278, 510286279, 510286280, 510286281, 510286282, 510286283, 510286284, 510286285, 510286286, 510286287, 510286288, 510286289, 510286290, 510286291, 510286292, 510286293, 510286294, 510286295, 510286296, 510286297, 510286298, 510286299, 510286300, 510287001, 510287002, 510287003, 510287004, 510287005, 510287006, 510287007, 510287008, 510287009, 510287010, 510287011, 510287012, 510287013, 510287014, 510287015, 510287016, 510287017, 510287018, 510287019, 510287020, 510287021, 510287022, 510287023, 510287024, 510287025, 510287026, 510287027, 510287029, 510287030, 510287031, 510287032, 510287033, 510287034, 510287035, 510287036, 510287037, 510287038, 510287039, 510287040, 510287041, 510287042, 510287043, 510287044, 510287045, 510287046, 510287047, 510287048, 510287049, 510287050, 510287051, 510287052, 510287053, 510287054, 510287055, 510287056, 510287057, 510287058, 510287059, 510287060, 510287061, 510287062, 510287063, 510287064, 510287065, 510287066, 510287067, 510287068, 510287069, 510287070, 510287071, 510287072, 510287073, 510287074, 510287075, 510287076, 510287077, 510287078, 510287079, 510287080, 510287081, 510287082, 510287083, 510287084, 510287085, 510287086, 510287087, 510287088, 510287089, 510287090, 510287091, 510287092, 510287093, 510287094, 510287095, 510287096, 510287097, 510287098, 510287099, 510287100, 510287101, 510287102, 510287103, 510287104, 510287105, 510287106, 510287107, 510287108, 510287109, 510287110, 510287111, 510287112, 510287113, 510287114, 510287115, 510287116, 510287117, 510287118, 510287119, 510287120, 510287121, 510287122, 510287123, 510287124, 510287125, 510287126, 510287127, 510287128, 510287129, 510287130, 510287131, 510287132, 510287133, 510287134, 510287135, 510287136, 510287137, 510287138, 510287139, 510287140, 510287141, 510287142, 510287143, 510287144, 510287145, 510287146, 510287147, 510287148, 510287149, 510288001, 510288002, 510288003, 510288004, 510288005, 510288006, 510288007, 510288008, 510288009, 510288010, 510288011, 510288012, 510288013, 510288014, 510288015, 510288016, 510288017, 510288018, 510288019, 510288020, 510288021, 510288022, 510288023, 510288024, 510288025, 510288026, 510288027, 510288028, 510288029, 510288030, 510288031, 510288032, 510288033, 510288034, 510288035, 510288036, 510288037, 510288038, 510288039, 510288040, 510288041, 510288042, 510288043, 510288044, 510288046, 510288047, 510288048, 510288049, 510288050, 510288051, 510288052, 510288053, 510288054, 510288055, 510288056, 510288057, 510288058, 510288059, 510288060, 510288061, 510288062, 510288063, 510288064, 510288065, 510288066, 510288067, 510288068, 510288069, 510288070, 510288071, 510288072, 510288073, 510288074, 510288075, 510288076, 510288077, 510288078, 510288079, 510288080, 510288081, 510288082, 510288083, 510288084, 510288085, 510288086, 510288087, 510288088, 510288089, 510288090, 510288091, 510288092, 510288093, 510288094, 510288095, 510288096, 510288097, 510288098, 510288099, 510288100, 510288101, 510288102, 510288103, 510288104, 510288105, 510288106, 510288107, 510288108, 510288109, 510288110, 510288111, 510288112, 510288113, 510288114, 510288115, 510288116, 510288117, 510288118, 510288119, 510288120, 510288121, 510288122, 510288123, 510288124, 510288125, 510288126, 510288127, 510288128, 510288129, 510288130, 510288131, 510288132, 510288133, 510288134, 510288135, 510288136, 510288137, 510288138, 510288139, 510288140, 510288141, 510288142, 510288143, 510288144, 510288145, 510288146, 510288147, 510288148, 510288149, 510288150, 510288151, 510288152, 510288153, 510288154, 510288155, 510288156, 510288157, 510288158, 510288159, 510288160, 510288161, 510288162, 510288163, 510288164, 510288165, 510288166, 510288167, 510288169, 510288170, 510288171, 510288172, 510288173, 510288174, 510288175, 510288176, 510288177, 510288178, 510288179, 510288180, 510288181, 510288182, 510288183, 510288184, 510288185, 510288186, 510288187, 510288188, 510288189, 510288190, 510288191, 510288192, 510288193, 510288194, 510288195, 510288196, 510288197, 510288198, 510288199, 510288200, 510288201, 510288202, 510288203, 510288204, 510288205, 510288206, 510288207, 510288208, 510288209, 510288210, 510288211, 510288212, 510288213, 510288214, 510288215, 510288216, 510288217, 510288218, 510288219, 510288220, 510288221, 510288222, 510288223, 510288224, 510288225, 510288226, 510288227, 510288228, 510288229, 510288230, 510288231, 510288232, 510288233, 510288234, 510288235, 510288236, 510288237, 510288238, 510288239, 510288240, 510288241, 510288242, 510288243, 510288244, 510288245, 510288246, 510288247, 510288248, 510288249, 510288250, 510288251, 510288252, 510288253, 510288254, 510288255, 510288256, 510288257, 510288258, 510288259, 510288260, 510288261, 510288262, 510288263, 510288264, 510288265, 510288266, 510288267, 510288268, 510288269, 510288270, 510288271, 510288272, 510288273, 510288275, 510288276, 510288277, 510288278, 510288279, 510288280, 510288281, 510291001, 510291002, 510291003, 510291005, 510291006, 510291007, 510291008, 510291009, 510291010, 510291011, 510291012, 510291013, 510291015, 510291016, 510291017, 510291018, 510291019, 510291020, 510291021, 510291022, 510291023, 510291024, 510291025, 510291026, 510291027, 510291028, 510291029, 510291030, 510291031, 510291032, 510291033, 510291035, 510291036, 510291037, 510291038, 510291039, 510291040, 510291041, 510291042, 510291043, 510291044, 510291045, 510291046, 510291047, 510291048, 510291049, 510291050, 510291051, 510292001, 510292002, 510292003, 510292004, 510292005, 510292006, 510292007, 510292008, 510292009, 510292010, 510292011, 510292012, 510292013, 510292014, 510292015, 510292016, 510292017, 510292018, 510292019, 510292020, 510292021, 510292022, 510292023, 510292024, 510292025, 510292026, 510292027, 510292028, 510292029, 510292030, 510292031, 510292032, 510292033, 510292034, 510292035, 510292036, 510292037, 510292038, 510292039, 510292040, 510292041, 510292042, 510292043, 510292044, 510292045, 510292046, 510292047, 510292048, 510292049, 510292050, 510292051, 510292052, 510292053, 510292054, 510292055, 510292056, 510292057, 510292058, 510292059, 510292060, 510292061, 510292062, 510292063, 510292064, 510292065, 510292066, 510292067, 510292068, 510292069, 510292070, 510292071, 510292072, 510292074, 510292075, 510292076, 510292077, 510292078, 510292079, 510292080, 510292081, 510292082, 510292083, 510292084, 510292085, 510292086, 510292087, 510292088, 510292089, 510292090, 510292091, 510292092, 510292093, 510292094, 510292095, 510292096, 510292097, 510292098, 510292099, 510292100, 510292101, 510292102, 510292103, 510292104, 510292105, 510292106, 510292107, 510292108, 510292109, 510292110, 510292111, 510292112, 510292113, 510292114, 510292115, 510292116, 510292117, 510292118, 510292119, 510292120, 510292121, 510292122, 510292123, 510292124, 510292125, 510292126, 510292127, 510292128, 510292129, 510292130, 510292131, 510292132, 510292133, 510292134, 510292135, 510292136, 510292137, 510292138, 510292139, 510292140, 510292141, 510292142, 510292143, 510292144, 510292145, 510292146, 510292147, 510292148, 510292149, 510292150, 510292151, 510292152, 510292153, 510292154, 510292155, 510292156, 510292157, 510292158, 510292159, 510292160, 510292161, 510292162, 510292163, 510292164, 510292165, 510292166, 510292167, 510292168, 510292169, 510292170, 510292171, 510292172, 510293001, 510293002, 510293003, 510293004, 510293005, 510293006, 510293007, 510293008, 510293009, 510293010, 510293011, 510293012, 510293013, 510293014, 510293015, 510293016, 510293017, 510293018, 510293019, 510293020, 510293021, 510293022, 510293023, 510293025, 510293026, 510293027, 510293028, 510293029, 510293030, 510293031, 510293032, 510293033, 510293034, 510293035, 510293036, 510293037, 510293038, 510293039, 510293040, 510293041, 510293042, 510293043, 510293044, 510293045, 510293046, 510293047, 510293048, 510293049, 510293050, 510293051, 510293052, 510293053, 510293054, 510293055, 510293056, 510293057, 510293058, 510293059, 510293060, 510293061, 510293062, 510293063, 510293064, 510293065, 510293066, 510293067, 510293068, 510293069, 510293070, 510293071, 510293072, 510293073, 510293074, 510293075, 510293076, 510293077, 510293078, 510293079, 510293080, 510293081, 510293082, 510293083, 510293084, 510293085, 510293086, 510293087, 510293088, 510293089, 510293090, 510293092, 510293093, 510293094, 510293095, 510293096, 510293097, 510293098, 510293099, 510293100, 510293101, 510293102, 510293103, 510293104, 510293105, 510293106, 510293107, 510293108, 510293109, 510293110, 510293111, 510293112, 510293113, 510293114, 510293115, 510293116, 510293117, 510293118, 510293119, 510293120, 510293121, 510293122, 510293123, 510293124, 510293125, 510293126, 510293127, 510293128, 510293129, 510293130, 510293131, 510293132, 510293133, 510293134, 510293135, 510293136, 510293137, 510293138, 510293139, 510293140, 510293141, 510293142, 510293143, 510293144, 510293145, 510293146, 510293147, 510293148, 510293149, 510293150, 510293151, 510293152, 510293153, 510293154, 510293155, 510293156, 510293157, 510305001, 510305002, 510305003, 510305004, 510305005, 510305006, 510305007, 510305008, 510305009, 510305010, 510305011, 510305012, 510305013, 510305014, 510305015, 510305016, 510305017, 510305018, 510305019, 510305020, 510305021, 510305022, 510305023, 510305024, 510305025, 510305026, 510305027, 510306001, 510306002, 510306003, 510306004, 510306005, 510306006, 510306007, 510306008, 510306009, 510306010, 510306011, 510306012, 510306013, 510306014, 510306015, 510306016, 510306017, 510306018, 510306019, 510306020, 510306021, 510306022, 510306023, 510306024, 510306025, 510306026, 510306027, 510306028, 510306029, 510306030, 510306031, 510306032, 510306033, 510306034, 510306035, 510306036, 510306037, 510306038, 510306039, 510306040, 510306041, 510306042, 510306043, 510306044, 510306045, 510306046, 510306047, 510306048, 510306049, 510306050, 510306051, 510306052, 510306053, 510306054, 510306055, 510306056, 510306057, 510306058, 510306059, 510306060, 510306061, 510306062, 510306063, 510306064, 510306065, 510306066, 510306067, 510306068, 510306069, 510306070, 510306071, 510306072, 510306073, 510306074, 510306075, 510306076, 510306077, 510306078, 510306079, 510306080, 510306081, 510306082, 510306083, 510306084, 510306085, 510306086, 510306087, 510306088, 510306089, 510306090, 510306091, 510306092, 510306093, 510306094, 510306095, 510306096, 510306097, 510306098, 510306099, 510306100, 510306101, 510306102, 510306103, 510306104, 510306105, 510306106, 510306107, 510306108, 510306109, 510306110, 510306111, 510306112, 510306113, 510306114, 510306115, 510306116, 510306117, 510306118, 510306119, 510306120, 510306121, 510306122, 510306123, 510306124, 510306125, 510306126, 510306127, 510306128, 510306129, 510306130, 510306131, 510306132, 510306133, 510306134, 510306135, 510306136, 510306137, 510306138, 510306139, 510306140, 510306141, 510306142, 510306143, 510306144, 510306145, 510306146, 510306147, 510306148, 510306149, 510306150, 510306151, 510306152, 510306153, 510306154, 510306155, 510306156, 510306157, 510306158, 510306159, 510306160, 510306161, 510306162, 510306163, 510306164, 510306165, 510306166, 510306167, 510306168, 510306169, 510306170, 510306171, 510306172, 510306173, 510306174, 510306175, 510306176, 510313001, 510313002, 510313003, 510313004, 510313005, 510313006, 510313007, 510313008, 510313009, 510313010, 510313011, 510313012, 510313013, 510313014, 510313015, 510313016, 510313017, 510313018, 510313019, 510313020, 510313021, 510313022, 510313023, 510313024, 510313025, 510313026, 510313027, 510313028, 510313029, 510313030, 510313031, 510313032, 510313033, 510313034, 510313035, 510313036, 510313037, 510313038, 510313039, 510313040, 510313041, 510313042,
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- MOOG Medical Devices Group
- Reason for Recall:
- MOOG Medical Devices Group is conducting a voluntary for a limited set of serial numbers of the EnteraLite Infinity and Infinity Orange Small Volume Enteral Feeding Pumps because certain diodes on the PCB assembly may result in insufficient voltage to a portion of the pump resulting in an error 13 message and shutting the pump down entirely.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
EnteraLite Infinity Enteral Feeding Pump identified with the following list numbers: INFKIT2 and INFORNG. The device is used with disposable PVC administration tubing sets to provide a means for the volumetric delivery of fluids used in enteral applications. The EnteraLite Infinity enteral feeding pump with disposable PVC administration tubing sets and user accessories provide a means for the volumetric delivery of fluids used in enteral applications.
Product Codes/Lot Numbers:
INFKIT2 with serial numbers: 506249031, 506264096, 506272024, 506335006, 506335145, 506336089, 506336096, 506362083, 506362086, 507018037, 507018060, 507067220, 507085140, 507087018, 507087140, 507095089, 507095171, 507121102, 507122050, 507122066, 507123149, 507123163, 507144016, 507144057, 507144215, 507144220, 507144221, 507144226, 507151002, 507155019, 507155021, 507155094, 507155127, 507155156, 507155159, 507156065, 507157048, 507157124, 507164009, 507173005, 507173019, 507173063, 507179123, 507180064, 507186018, 507186098, 507186176, 507197012, 507197024, 507197036, 507201013, 507201038, 507205054, 507206064, 507212012, 507212013, 507214065, 507214098, 507214162, 507215041, 507218016, 507219057, 507219087, 507219103, 507220106, 507221098, 507227116, 507229033, 507229039, 507229060, 507232030, 507236063, 507236203, 507236303, 507239056, 507241089, 507242130, 507242201, 507247049, 507248031, 507248087, 507248095, 507249131, 507250175, 507255025, 507257123, 507257129, 507261065, 507264041, 507264127, 507268121, 507268351, 507268374, 507269084, 507270096, 507270106, 507271157, 507275046, 507276154, 507278013, 507281028, 507281055, 507283119, 507284158, 507285186, 507289004, 507289120, 507289157, 507289165, 507291090, 507291168, 507291232, 507292108, 507292179, 507292339, 507295099, 507306069, 507311133, 507313050, 507313057, 507313059, 507313084, 507325044, 507330149, 507337080, 507337082, 507337085, 507339076, 507339095, 507344149, 507345001, 507345103, 507347150, 507348077, 507351090, 507353065, 507353120, 508018015, 508018090, 508030083, 508037024, 508037097, 508039099, 508039101, 508058001, 508058159, 508058160, 508066002, 508066030, 508066097, 508066123, 508066152, 508066153, 508066268, 508067033, 508067069, 508067092, 508067117, 508067166, 508067229, 508067231, 508070062, 508070228, 508071060, 508071132, 508071257, 508072146, 508072151, 508077022, 508099062, 508100033, 508112135, 508112159, 508114018, 508114111, 508116029, 508126038, 508131043, 508131079, 508131084, 508137099, 508137101, 508148152, 508148165, 508148216, 508148267, 508151050, 508151121, 508159063, 508165023, 508165034, 508165060, 508165231, 508166036, 508172055, 508172218, 508172260, 508172319, 508172334, 508198263, 508198264, 508199105, 508233083, 508233103, 508234066, 508234068, 508234078, 508234086, 508235009, 508235107, 508235111, 508235124, 508255070, 508255100, 508260115, 508260216, 508262074, 508263073, 508269064, 508275020, 508289015, 508289131, 508325051, 508326036, 508326058, 508340030, 508340100, 508351216, 508351237, 508351246, 508357105, 509005203, 509005218, 509008108, 509008270, 509008308, 509009120, 509015017, 509017118, 509020024, 509020140, 509021098, 509024156, 509030043, 509030055, 509030056, 509030088, 509034103, 509035272, 509054126, 509055029, 509055207, 509065195, 509065202, 509065212, 509071109, 509072019, 509072092, 509079210, 509085268, 509086084, 509086125, 509090011, 509090127, 509099077, 509100122, 509107033, 509119090, 509119120, 509120121, 509132033, 509141115, 509162073, 509183062, 509183092, 509197148, 509209048, 509231116, 509254016, 509258007, 509258203, 509274074, 509275014, 509281008, 509282068, 509285043, 509323066, 509324035, 509324059, 509344048, 509350005, 509350020, 509350025, 509350037, 509350072, 509350078, 509350084, 510015134, 510015149, 510015166, 510027107, 510027146, 510028041, 510050026, 510050094, 510063025, 510063107, 510076044, 510076045, 510088038, 510092021, 510124007, 510124012, 510128125, 510168018, 510168043, 510188065, 510193159, 510218008, 510218038, 510218051, 510218064, 510218066, 510218071, 510218084, 510218087, 510218097, 510218098, 510218106, 510218148, 510218156, 510221001, 510221003, 510221028, 510221087, 510222003, 510222005, 510222006, 510222011, 510222013, 510222016, 510222020, 510222032, 510222062, 510222112, 510222119, 510222121, 510229001, 510229002, 510229003, 510229004, 510229005, 510229006, 510229007, 510229008, 510229009, 510229010, 510229011, 510229012, 510229013, 510229014, 510229015, 510229016, 510229017, 510229018, 510229019, 510229020, 510229021, 510229022, 510229023, 510229024, 510229025, 510229026, 510229027, 510229028, 510229029, 510229030, 510229031, 510229032, 510229033, 510229034, 510229035, 510229036, 510229037, 510229038, 510229039, 510229040, 510229041, 510229042, 510229043, 510229044, 510229045, 510229046, 510229047, 510229048, 510229049, 510229050, 510229051, 510229052, 510229053, 510229054, 510229055, 510229056, 510229057, 510229058, 510229059, 510229060, 510229061, 510229062, 510229063, 510229064, 510229065, 510229066, 510229067, 510229068, 510229069, 510229070, 510229071, 510229072, 510229073, 510229074, 510229075, 510229076, 510229077, 510229079, 510229080, 510229081, 510229082, 510229083, 510229084, 510229085, 510229086, 510229087, 510229088, 510229089, 510229090, 510229091, 510229092, 510229093, 510229094, 510229095, 510229096, 510229097, 510229098, 510229099, 510229100, 510229101, 510229102, 510229103, 510229104, 510229105, 510229106, 510229107, 510229108, 510229109, 510229110, 510229111, 510229112, 510229113, 510229114, 510229115, 510229116, 510229117, 510229118, 510229119, 510229120, 510229121, 510229122, 510229123, 510229124, 510229125, 510229126, 510229127, 510229128, 510229129, 510229130, 510229131, 510229132, 510229133, 510229134, 510229135, 510229136, 510229137, 510229138, 510229139, 510229140, 510229141, 510229142, 510229143, 510229144, 510229145, 510229146, 510229147, 510229148, 510229149, 510229150, 510229151, 510231001, 510231002, 510231003, 510231004, 510231005, 510231006, 510231007, 510231008, 510231009, 510231010, 510231011, 510231012, 510231013, 510231014, 510231015, 510231016, 510231017, 510231018, 510231019, 510231020, 510231021, 510231022, 510231023, 510231024, 510231025, 510231026, 510231027, 510231028, 510231029, 510231030, 510231031, 510231032, 510231033, 510231034, 510231035, 510231036, 510231037, 510231038, 510231039, 510231040, 510231041, 510231042, 510231043, 510231044, 510231045, 510231046, 510231047, 510231048, 510231049, 510231050, 510231051, 510231052, 510231053, 510231054, 510231055, 510231056, 510231057, 510231058, 510231059, 510231060, 510231061, 510231062, 510231063, 510231064, 510231065, 510231066, 510231067, 510231068, 510231069, 510231070, 510231071, 510231072, 510231073, 510231074, 510232001, 510232002, 510232003, 510232004, 510232005, 510232006, 510232007, 510232008, 510232010, 510232011, 510232012, 510232013, 510232014, 510232015, 510232016, 510232017, 510232018, 510232019, 510232020, 510232021, 510232022, 510232023, 510232024, 510232025, 510232026, 510232027, 510232028, 510232029, 510232030, 510232031, 510232032, 510232033, 510232034, 510232035, 510232036, 510232037, 510232038, 510232039, 510232040, 510232041, 510232042, 510232043, 510232044, 510232045, 510232046, 510232047, 510232048, 510232049, 510232050, 510232051, 510232052, 510232053, 510232054, 510232055, 510232056, 510232057, 510232058, 510232059, 510232060, 510232061, 510232062, 510232063, 510232064, 510232065, 510232066, 510232067, 510232068, 510232069, 510232070, 510232071, 510232072, 510232073, 510232074, 510232075, 510232076, 510232077, 510232078, 510232079, 510232080, 510232081, 510232082, 510232083, 510232084, 510232085, 510232086, 510232087, 510232088, 510232089, 510232090, 510232091, 510232092, 510232093, 510232094, 510232095, 510232096, 510232098, 510232099, 510232100, 510232101, 510232102, 510232103, 510232104, 510232105, 510232106, 510232107, 510232108, 510232109, 510232110, 510232111, 510232112, 510232113, 510232114, 510232115, 510232116, 510232117, 510232118, 510232119, 510232120, 510232121, 510232122, 510232123, 510232124, 510232125, 510232126, 510232127, 510232128, 510232129, 510232130, 510232131, 510232132, 510232134, 510232135, 510232136, 510232137, 510232138, 510232139, 510232140, 510232141, 510232142, 510232143, 510232144, 510232145, 510232146, 510232147, 510232148, 510232149, 510232150, 510232151, 510232152, 510232153, 510232154, 510232155, 510232156, 510232157, 510232158, 510232159, 510232160, 510232161, 510232162, 510232163, 510232164, 510232165, 510232166, 510232167, 510232168, 510232169, 510232170, 510232171, 510232172, 510232173, 510232174, 510232175, 510232176, 510235001, 510235002, 510235003, 510235005, 510235007, 510235008, 510235009, 510235010, 510235011, 510235012, 510235013, 510235014, 510235015, 510235016, 510235017, 510235018, 510235019, 510235020, 510235021, 510235022, 510235023, 510235024, 510235025, 510235026, 510235027, 510235029, 510235030, 510235031, 510235032, 510235033, 510235034, 510235035, 510235036, 510235037, 510235038, 510235039, 510235040, 510235041, 510235042, 510235043, 510235044, 510235045, 510235046, 510235047, 510235048, 510235049, 510235050, 510235051, 510235052, 510235053, 510235054, 510235055, 510235056, 510235057, 510235058, 510235059, 510235060, 510235061, 510235062, 510235063, 510235064, 510235065, 510235066, 510235067, 510235068, 510235069, 510235070, 510235071, 510235072, 510235073, 510235074, 510235075, 510235076, 510235077, 510235078, 510235079, 510235080, 510235081, 510235082, 510235083, 510235084, 510235085, 510235086, 510235087, 510235088, 510235089, 510235090, 510235091, 510235092, 510235093, 510235094, 510235095, 510235096, 510235097, 510235098, 510235099, 510235100, 510235101, 510235102, 510235103, 510235104, 510235105, 510235106, 510235107, 510235108, 510235110, 510235111, 510235112, 510235113, 510235115, 510235116, 510235117, 510235118, 510235119, 510235120, 510235121, 510235122, 510235124, 510235125, 510235126, 510235127, 510235128, 510235129, 510235131, 510235132, 510235133, 510235134, 510235135, 510235136, 510235138, 510235140, 510235141, 510235142, 510235143, 510235144, 510235145, 510253001, 510253002, 510253003, 510253004, 510253005, 510253006, 510253007, 510253008, 510253009, 510253010, 510253011, 510253012, 510253014, 510253015, 510253016, 510253017, 510253018, 510253019, 510253020, 510253021, 510253022, 510253023, 510253024, 510253025, 510253026, 510253027, 510253028, 510253029, 510253030, 510253031, 510253032, 510253033, 510253034, 510253035, 510253036, 510253037, 510253038, 510253039, 510253040, 510253041, 510253042, 510253043, 510253044, 510253045, 510253046, 510253047, 510253048, 510253049, 510253050, 510253051, 510253052, 510253053, 510253054, 510253055, 510253056, 510253057, 510253058, 510253059, 510253060, 510253061, 510253063, 510253064, 510253065, 510253066, 510253067, 510253068, 510253069, 510253070, 510253071, 510253072, 510253073, 510253074, 510253075, 510253076, 510253077, 510253078, 510253079, 510253080, 510253081, 510253082, 510253083, 510253084, 510253085, 510253086, 510253087, 510253088, 510253089, 510253090, 510253091, 510253092, 510253093, 510253094, 510253095, 510253096, 510253097, 510253098, 510253099, 510253100, 510253101, 510253102, 510253103, 510253104, 510253105, 510253106, 510253107, 510253108, 510253109, 510253110, 510253111, 510253112, 510253113, 510253114, 510253115, 510253116, 510253117, 510253118, 510253119, 510253120, 510253121, 510253122, 510253123, 510253124, 510253125, 510253126, 510253127, 510253128, 510253129, 510253130, 510253131, 510253132, 510253133, 510253134, 510253135, 510253136, 510253137, 510253138, 510256001, 510256002, 510256003, 510256004, 510256005, 510256006, 510256007, 510256008, 510256009, 510256010, 510256011, 510256012, 510256013, 510256014, 510256015, 510256016, 510256017, 510256018, 510256019, 510256020, 510256021, 510256022, 510256023, 510256024, 510256025, 510256026, 510256027, 510256028, 510256029, 510256030, 510256031, 510256032, 510256033, 510256034, 510256035, 510256036, 510256037, 510256038, 510256039, 510256040, 510256041, 510256042, 510256043, 510256044, 510256045, 510256046, 510256047, 510256048, 510256049, 510256050, 510256051, 510256052, 510256053, 510256054, 510256055, 510256056, 510256057, 510256058, 510256059, 510256060, 510256061, 510256062, 510256063, 510256064, 510256065, 510256066, 510256067, 510256068, 510256069, 510256070, 510256071, 510256072, 510256073, 510256074, 510256075, 510256076, 510256077, 510256078, 510256079, 510256080, 510256081, 510256082, 510256083, 510256084, 510256085, 510256086, 510256087, 510256088, 510256089, 510256090, 510256091, 510256092, 510256093, 510256094, 510256095, 510256096, 510256097, 510256098, 510256099, 510256100, 510257001, 510257002, 510257003, 510257004, 510257005, 510257006, 510257007, 510257008, 510257010, 510257011, 510257012, 510257013, 510257014, 510257015, 510257016, 510257017, 510257018, 510257019, 510257020, 510257021, 510257022, 510257023, 510257024, 510257025, 510257026, 510257027, 510257028, 510257029, 510257030, 510257031, 510257032, 510257033, 510257034, 510257035, 510257036, 510257037, 510257038, 510257039, 510257040, 510257041, 510257042, 510257043, 510257044, 510257045, 510257046, 510257047, 510257048, 510257049, 510257050, 510257051, 510257052, 510257053, 510257054, 510257055, 510257056, 510257057, 510257058, 510257059, 510257060, 510257061, 510257062, 510257063, 510257064, 510257065, 510257066, 510257067, 510257068, 510257069, 510257070, 510257071, 510257072, 510257073, 510257074, 510257075, 510257076, 510257077, 510257078, 510257079, 510257080, 510257081, 510257082, 510257083, 510257084, 510257085, 510257086, 510257087, 510257088, 510257089, 510257090, 510257091, 510257092, 510257094, 510257095, 510257096, 510257097, 510257098, 510257099, 510257100, 510257101, 510257102, 510259001, 510259002, 510259003, 510259004, 510259005, 510259006, 510259007, 510259008, 510259009, 510259011, 510259012, 510259013, 510259014, 510259015, 510259016, 510259017, 510259018, 510259019, 510259020, 510259021, 510259022, 510259023, 510259024, 510259025, 510259026, 510259027, 510259028, 510259029, 510259030, 510259031, 510259032, 510259033, 510259034, 510259035, 510259036, 510259037, 510259038, 510259039, 510259040, 510259041, 510259042, 510259043, 510259044, 510259045, 510259046, 510259047, 510259048, 510259049, 510259050, 510259051, 510259052, 510259053, 510259054, 510259055, 510259056, 510259057, 510259058, 510259059, 510259060, 510259061, 510259062, 510259063, 510259064, 510259066, 510259067, 510259068, 510259069, 510259070, 510259071, 510259072, 510259073, 510259074, 510259075, 510259076, 510259077, 510259078, 510259079, 510259080, 510259081, 510259082, 510259083, 510259084, 510259085, 510259086, 510259087, 510259088, 510259089, 510259090, 510259091, 510259092, 510259093, 510259094, 510259095, 510259096, 510259097, 510259098, 510259099, 510259100, 510259101, 510259102, 510278001, 510278002, 510278003, 510278004, 510278005, 510278006, 510284002, 510284003, 510284004, 510284005, 510284006, 510284007, 510284008, 510284009, 510284010, 510284011, 510284012, 510284013, 510284014, 510284015, 510284016, 510284017, 510284018, 510284019, 510284020, 510284021, 510284022, 510284023, 510284024, 510284025, 510284026, 510284027, 510284028, 510284029, 510284030, 510284031, 510284032, 510284033, 510284034, 510284035, 510284036, 510284037, 510284038, 510284039, 510284040, 510284041, 510284042, 510284043, 510284044, 510284045, 510284046, 510284047, 510284048, 510284049, 510284050, 510284051, 510284052, 510284053, 510284054, 510284055, 510284056, 510284057, 510284058, 510284059, 510284060, 510284061, 510284062, 510284063, 510284064, 510284065, 510284066, 510284067, 510284068, 510284069, 510284070, 510284071, 510284072, 510284073, 510284074, 510284076, 510284077, 510284078, 510284079, 510284080, 510284081, 510284082, 510284083, 510284084, 510284085, 510284086, 510284087, 510284088, 510284089, 510284090, 510284091, 510284092, 510284093, 510284094, 510284095, 510284096, 510284097, 510284098, 510284099, 510284100, 510284101, 510284102, 510284103, 510284104, 510284105, 510284106, 510284107, 510284108, 510284109, 510284110, 510284111, 510284112, 510284113, 510284114, 510284115, 510284116, 510284117, 510284118, 510284119, 510284120, 510284121, 510284122, 510284123, 510284124, 510284125, 510284126, 510284127, 510284128, 510284129, 510284130, 510285001, 510285002, 510285003, 510285004, 510285005, 510285006, 510285007, 510285008, 510285009, 510285011, 510285012, 510285013, 510285014, 510285015, 510285016, 510285017, 510285018, 510285019, 510285020, 510285021, 510285022, 510285023, 510285024, 510285025, 510285026, 510285027, 510285028, 510285029, 510285030, 510285031, 510285032, 510285033, 510285034, 510285035, 510285036, 510285037, 510285038, 510285039, 510285040, 510285041, 510285042, 510285043, 510285044, 510285045, 510285046, 510285047, 510285048, 510285049, 510285050, 510285051, 510285052, 510285053, 510285054, 510285055, 510285056, 510285057, 510285059, 510285060, 510285061, 510285062, 510285063, 510285064, 510285065, 510285066, 510285067, 510285068, 510285069, 510285070, 510285071, 510285072, 510285073, 510285074, 510285075, 510285076, 510285077, 510285078, 510285079, 510285080, 510285081, 510285082, 510285083, 510285084, 510285085, 510285086, 510285087, 510285088, 510285089, 510285090, 510285091, 510285092, 510285093, 510285094, 510285095, 510285096, 510285098, 510285099, 510285100, 510285101, 510285102, 510285103, 510285104, 510285105, 510285106, 510285107, 510285108, 510285109, 510285110, 510285111, 510285112, 510285113, 510285114, 510285115, 510285116, 510285117, 510285118, 510285119, 510285120, 510285121, 510285122, 510285123, 510285124, 510285125, 510285126, 510285127, 510285128, 510285129, 510285130, 510285131, 510285132, 510285133, 510285134, 510285136, 510285138, 510285139, 510285140, 510285141, 510285142, 510285143, 510285144, 510285145, 510285146, 510285147, 510285148, 510285149, 510285150, 510285151, 510285152, 510285153, 510285154, 510285155, 510285156, 510285157, 510285158, 510285159, 510285160, 510285161, 510285162, 510285163, 510285164, 510285165, 510285166, 510285167, 510285168, 510285169, 510285170, 510285171, 510285172, 510285174, 510285175, 510285176, 510285177, 510285178, 510286001, 510286002, 510286003, 510286004, 510286005, 510286006, 510286007, 510286008, 510286009, 510286010, 510286011, 510286012, 510286013, 510286014, 510286015, 510286016, 510286018, 510286019, 510286020, 510286021, 510286022, 510286023, 510286024, 510286025, 510286026, 510286027, 510286028, 510286029, 510286030, 510286031, 510286033, 510286034, 510286035, 510286036, 510286037, 510286038, 510286039, 510286040, 510286041, 510286042, 510286043, 510286044, 510286045, 510286046, 510286047, 510286048, 510286049, 510286050, 510286051, 510286052, 510286053, 510286054, 510286055, 510286056, 510286057, 510286058, 510286059, 510286060, 510286061, 510286062, 510286063, 510286064, 510286065, 510286066, 510286067, 510286068, 510286069, 510286070, 510286071, 510286072, 510286073, 510286074, 510286075, 510286076, 510286077, 510286078, 510286079, 510286080, 510286081, 510286082, 510286083, 510286084, 510286085, 510286086, 510286087, 510286088, 510286089, 510286090, 510286091, 510286092, 510286093, 510286094, 510286095, 510286096, 510286098, 510286099, 510286100, 510286101, 510286102, 510286103, 510286104, 510286105, 510286106, 510286107, 510286108, 510286109, 510286110, 510286111, 510286112, 510286113, 510286114, 510286115, 510286116, 510286117, 510286118, 510286119, 510286120, 510286121, 510286122, 510286123, 510286124, 510286125, 510286126, 510286127, 510286128, 510286129, 510286130, 510286131, 510286132, 510286133, 510286134, 510286135, 510286136, 510286137, 510286138, 510286139, 510286140, 510286141, 510286142, 510286143, 510286144, 510286145, 510286146, 510286147, 510286148, 510286149, 510286150, 510286151, 510286152, 510286153, 510286154, 510286155, 510286156, 510286157, 510286158, 510286159, 510286160, 510286161, 510286162, 510286163, 510286164, 510286165, 510286166, 510286167, 510286168, 510286169, 510286170, 510286171, 510286172, 510286173, 510286174, 510286175, 510286177, 510286178, 510286179, 510286180, 510286181, 510286182, 510286183, 510286184, 510286185, 510286186, 510286187, 510286188, 510286189, 510286190, 510286192, 510286193, 510286194, 510286195, 510286196, 510286197, 510286198, 510286199, 510286200, 510286201, 510286202, 510286203, 510286204, 510286205, 510286206, 510286207, 510286208, 510286209, 510286210, 510286211, 510286212, 510286213, 510286214, 510286215, 510286216, 510286217, 510286218, 510286219, 510286220, 510286221, 510286222, 510286223, 510286224, 510286225, 510286226, 510286227, 510286228, 510286229, 510286230, 510286231, 510286232, 510286233, 510286234, 510286235, 510286237, 510286238, 510286239, 510286240, 510286241, 510286242, 510286243, 510286244, 510286245, 510286246, 510286247, 510286248, 510286249, 510286250, 510286251, 510286252, 510286253, 510286254, 510286255, 510286256, 510286257, 510286258, 510286259, 510286260, 510286261, 510286262, 510286263, 510286264, 510286265, 510286266, 510286267, 510286268, 510286269, 510286270, 510286271, 510286272, 510286273, 510286274, 510286275, 510286276, 510286277, 510286278, 510286279, 510286280, 510286281, 510286282, 510286283, 510286284, 510286285, 510286286, 510286287, 510286288, 510286289, 510286290, 510286291, 510286292, 510286293, 510286294, 510286295, 510286296, 510286297, 510286298, 510286299, 510286300, 510287001, 510287002, 510287003, 510287004, 510287005, 510287006, 510287007, 510287008, 510287009, 510287010, 510287011, 510287012, 510287013, 510287014, 510287015, 510287016, 510287017, 510287018, 510287019, 510287020, 510287021, 510287022, 510287023, 510287024, 510287025, 510287026, 510287027, 510287029, 510287030, 510287031, 510287032, 510287033, 510287034, 510287035, 510287036, 510287037, 510287038, 510287039, 510287040, 510287041, 510287042, 510287043, 510287044, 510287045, 510287046, 510287047, 510287048, 510287049, 510287050, 510287051, 510287052, 510287053, 510287054, 510287055, 510287056, 510287057, 510287058, 510287059, 510287060, 510287061, 510287062, 510287063, 510287064, 510287065, 510287066, 510287067, 510287068, 510287069, 510287070, 510287071, 510287072, 510287073, 510287074, 510287075, 510287076, 510287077, 510287078, 510287079, 510287080, 510287081, 510287082, 510287083, 510287084, 510287085, 510287086, 510287087, 510287088, 510287089, 510287090, 510287091, 510287092, 510287093, 510287094, 510287095, 510287096, 510287097, 510287098, 510287099, 510287100, 510287101, 510287102, 510287103, 510287104, 510287105, 510287106, 510287107, 510287108, 510287109, 510287110, 510287111, 510287112, 510287113, 510287114, 510287115, 510287116, 510287117, 510287118, 510287119, 510287120, 510287121, 510287122, 510287123, 510287124, 510287125, 510287126, 510287127, 510287128, 510287129, 510287130, 510287131, 510287132, 510287133, 510287134, 510287135, 510287136, 510287137, 510287138, 510287139, 510287140, 510287141, 510287142, 510287143, 510287144, 510287145, 510287146, 510287147, 510287148, 510287149, 510288001, 510288002, 510288003, 510288004, 510288005, 510288006, 510288007, 510288008, 510288009, 510288010, 510288011, 510288012, 510288013, 510288014, 510288015, 510288016, 510288017, 510288018, 510288019, 510288020, 510288021, 510288022, 510288023, 510288024, 510288025, 510288026, 510288027, 510288028, 510288029, 510288030, 510288031, 510288032, 510288033, 510288034, 510288035, 510288036, 510288037, 510288038, 510288039, 510288040, 510288041, 510288042, 510288043, 510288044, 510288046, 510288047, 510288048, 510288049, 510288050, 510288051, 510288052, 510288053, 510288054, 510288055, 510288056, 510288057, 510288058, 510288059, 510288060, 510288061, 510288062, 510288063, 510288064, 510288065, 510288066, 510288067, 510288068, 510288069, 510288070, 510288071, 510288072, 510288073, 510288074, 510288075, 510288076, 510288077, 510288078, 510288079, 510288080, 510288081, 510288082, 510288083, 510288084, 510288085, 510288086, 510288087, 510288088, 510288089, 510288090, 510288091, 510288092, 510288093, 510288094, 510288095, 510288096, 510288097, 510288098, 510288099, 510288100, 510288101, 510288102, 510288103, 510288104, 510288105, 510288106, 510288107, 510288108, 510288109, 510288110, 510288111, 510288112, 510288113, 510288114, 510288115, 510288116, 510288117, 510288118, 510288119, 510288120, 510288121, 510288122, 510288123, 510288124, 510288125, 510288126, 510288127, 510288128, 510288129, 510288130, 510288131, 510288132, 510288133, 510288134, 510288135, 510288136, 510288137, 510288138, 510288139, 510288140, 510288141, 510288142, 510288143, 510288144, 510288145, 510288146, 510288147, 510288148, 510288149, 510288150, 510288151, 510288152, 510288153, 510288154, 510288155, 510288156, 510288157, 510288158, 510288159, 510288160, 510288161, 510288162, 510288163, 510288164, 510288165, 510288166, 510288167, 510288169, 510288170, 510288171, 510288172, 510288173, 510288174, 510288175, 510288176, 510288177, 510288178, 510288179, 510288180, 510288181, 510288182, 510288183, 510288184, 510288185, 510288186, 510288187, 510288188, 510288189, 510288190, 510288191, 510288192, 510288193, 510288194, 510288195, 510288196, 510288197, 510288198, 510288199, 510288200, 510288201, 510288202, 510288203, 510288204, 510288205, 510288206, 510288207, 510288208, 510288209, 510288210, 510288211, 510288212, 510288213, 510288214, 510288215, 510288216, 510288217, 510288218, 510288219, 510288220, 510288221, 510288222, 510288223, 510288224, 510288225, 510288226, 510288227, 510288228, 510288229, 510288230, 510288231, 510288232, 510288233, 510288234, 510288235, 510288236, 510288237, 510288238, 510288239, 510288240, 510288241, 510288242, 510288243, 510288244, 510288245, 510288246, 510288247, 510288248, 510288249, 510288250, 510288251, 510288252, 510288253, 510288254, 510288255, 510288256, 510288257, 510288258, 510288259, 510288260, 510288261, 510288262, 510288263, 510288264, 510288265, 510288266, 510288267, 510288268, 510288269, 510288270, 510288271, 510288272, 510288273, 510288275, 510288276, 510288277, 510288278, 510288279, 510288280, 510288281, 510291001, 510291002, 510291003, 510291005, 510291006, 510291007, 510291008, 510291009, 510291010, 510291011, 510291012, 510291013, 510291015, 510291016, 510291017, 510291018, 510291019, 510291020, 510291021, 510291022, 510291023, 510291024, 510291025, 510291026, 510291027, 510291028, 510291029, 510291030, 510291031, 510291032, 510291033, 510291035, 510291036, 510291037, 510291038, 510291039, 510291040, 510291041, 510291042, 510291043, 510291044, 510291045, 510291046, 510291047, 510291048, 510291049, 510291050, 510291051, 510292001, 510292002, 510292003, 510292004, 510292005, 510292006, 510292007, 510292008, 510292009, 510292010, 510292011, 510292012, 510292013, 510292014, 510292015, 510292016, 510292017, 510292018, 510292019, 510292020, 510292021, 510292022, 510292023, 510292024, 510292025, 510292026, 510292027, 510292028, 510292029, 510292030, 510292031, 510292032, 510292033, 510292034, 510292035, 510292036, 510292037, 510292038, 510292039, 510292040, 510292041, 510292042, 510292043, 510292044, 510292045, 510292046, 510292047, 510292048, 510292049, 510292050, 510292051, 510292052, 510292053, 510292054, 510292055, 510292056, 510292057, 510292058, 510292059, 510292060, 510292061, 510292062, 510292063, 510292064, 510292065, 510292066, 510292067, 510292068, 510292069, 510292070, 510292071, 510292072, 510292074, 510292075, 510292076, 510292077, 510292078, 510292079, 510292080, 510292081, 510292082, 510292083, 510292084, 510292085, 510292086, 510292087, 510292088, 510292089, 510292090, 510292091, 510292092, 510292093, 510292094, 510292095, 510292096, 510292097, 510292098, 510292099, 510292100, 510292101, 510292102, 510292103, 510292104, 510292105, 510292106, 510292107, 510292108, 510292109, 510292110, 510292111, 510292112, 510292113, 510292114, 510292115, 510292116, 510292117, 510292118, 510292119, 510292120, 510292121, 510292122, 510292123, 510292124, 510292125, 510292126, 510292127, 510292128, 510292129, 510292130, 510292131, 510292132, 510292133, 510292134, 510292135, 510292136, 510292137, 510292138, 510292139, 510292140, 510292141, 510292142, 510292143, 510292144, 510292145, 510292146, 510292147, 510292148, 510292149, 510292150, 510292151, 510292152, 510292153, 510292154, 510292155, 510292156, 510292157, 510292158, 510292159, 510292160, 510292161, 510292162, 510292163, 510292164, 510292165, 510292166, 510292167, 510292168, 510292169, 510292170, 510292171, 510292172, 510293001, 510293002, 510293003, 510293004, 510293005, 510293006, 510293007, 510293008, 510293009, 510293010, 510293011, 510293012, 510293013, 510293014, 510293015, 510293016, 510293017, 510293018, 510293019, 510293020, 510293021, 510293022, 510293023, 510293025, 510293026, 510293027, 510293028, 510293029, 510293030, 510293031, 510293032, 510293033, 510293034, 510293035, 510293036, 510293037, 510293038, 510293039, 510293040, 510293041, 510293042, 510293043, 510293044, 510293045, 510293046, 510293047, 510293048, 510293049, 510293050, 510293051, 510293052, 510293053, 510293054, 510293055, 510293056, 510293057, 510293058, 510293059, 510293060, 510293061, 510293062, 510293063, 510293064, 510293065, 510293066, 510293067, 510293068, 510293069, 510293070, 510293071, 510293072, 510293073, 510293074, 510293075, 510293076, 510293077, 510293078, 510293079, 510293080, 510293081, 510293082, 510293083, 510293084, 510293085, 510293086, 510293087, 510293088, 510293089, 510293090, 510293092, 510293093, 510293094, 510293095, 510293096, 510293097, 510293098, 510293099, 510293100, 510293101, 510293102, 510293103, 510293104, 510293105, 510293106, 510293107, 510293108, 510293109, 510293110, 510293111, 510293112, 510293113, 510293114, 510293115, 510293116, 510293117, 510293118, 510293119, 510293120, 510293121, 510293122, 510293123, 510293124, 510293125, 510293126, 510293127, 510293128, 510293129, 510293130, 510293131, 510293132, 510293133, 510293134, 510293135, 510293136, 510293137, 510293138, 510293139, 510293140, 510293141, 510293142, 510293143, 510293144, 510293145, 510293146, 510293147, 510293148, 510293149, 510293150, 510293151, 510293152, 510293153, 510293154, 510293155, 510293156, 510293157, 510305001, 510305002, 510305003, 510305004, 510305005, 510305006, 510305007, 510305008, 510305009, 510305010, 510305011, 510305012, 510305013, 510305014, 510305015, 510305016, 510305017, 510305018, 510305019, 510305020, 510305021, 510305022, 510305023, 510305024, 510305025, 510305026, 510305027, 510306001, 510306002, 510306003, 510306004, 510306005, 510306006, 510306007, 510306008, 510306009, 510306010, 510306011, 510306012, 510306013, 510306014, 510306015, 510306016, 510306017, 510306018, 510306019, 510306020, 510306021, 510306022, 510306023, 510306024, 510306025, 510306026, 510306027, 510306028, 510306029, 510306030, 510306031, 510306032, 510306033, 510306034, 510306035, 510306036, 510306037, 510306038, 510306039, 510306040, 510306041, 510306042, 510306043, 510306044, 510306045, 510306046, 510306047, 510306048, 510306049, 510306050, 510306051, 510306052, 510306053, 510306054, 510306055, 510306056, 510306057, 510306058, 510306059, 510306060, 510306061, 510306062, 510306063, 510306064, 510306065, 510306066, 510306067, 510306068, 510306069, 510306070, 510306071, 510306072, 510306073, 510306074, 510306075, 510306076, 510306077, 510306078, 510306079, 510306080, 510306081, 510306082, 510306083, 510306084, 510306085, 510306086, 510306087, 510306088, 510306089, 510306090, 510306091, 510306092, 510306093, 510306094, 510306095, 510306096, 510306097, 510306098, 510306099, 510306100, 510306101, 510306102, 510306103, 510306104, 510306105, 510306106, 510306107, 510306108, 510306109, 510306110, 510306111, 510306112, 510306113, 510306114, 510306115, 510306116, 510306117, 510306118, 510306119, 510306120, 510306121, 510306122, 510306123, 510306124, 510306125, 510306126, 510306127, 510306128, 510306129, 510306130, 510306131, 510306132, 510306133, 510306134, 510306135, 510306136, 510306137, 510306138, 510306139, 510306140, 510306141, 510306142, 510306143, 510306144, 510306145, 510306146, 510306147, 510306148, 510306149, 510306150, 510306151, 510306152, 510306153, 510306154, 510306155, 510306156, 510306157, 510306158, 510306159, 510306160, 510306161, 510306162, 510306163, 510306164, 510306165, 510306166, 510306167, 510306168, 510306169, 510306170, 510306171, 510306172, 510306173, 510306174, 510306175, 510306176, 510313001, 510313002, 510313003, 510313004, 510313005, 510313006, 510313007, 510313008, 510313009, 510313010, 510313011, 510313012, 510313013, 510313014, 510313015, 510313016, 510313017, 510313018, 510313019, 510313020, 510313021, 510313022, 510313023, 510313024, 510313025, 510313026, 510313027, 510313028, 510313029, 510313030, 510313031, 510313032, 510313033, 510313034, 510313035, 510313036, 510313037, 510313038, 510313039, 510313040, 510313041, 510313042,
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0294-2013
Related Recalls
Moog is recalling Curlin 4000, 6000, and PainSmart Infusion Pumps manufactured or serviced by MMDG between March 18, 2015 and November 6, 2015 because the master pumps used to calibrate them were out of tolerance and may cause an over delivery of medications and solutions at a rate of 1.2 - 6.8% faster than programmed.
Potential for leaking between the purple ENFit connector and the white transitional stepped connector due to cracking or breaking of the purple connector.
Infinity Enteralite Ambulatory Feeding Pumps, List number: INFKIT2 - Addendum to Operator's Manual Product usage: provide a means for the volumetric delivery of fluids used in Enteral applications.
MOOG Medical Devices Group
The firm is adding an addendum to the operator's manual for Infinity Enteralite Ambulatory Feeding Pumps pertaining to the Infinite Dose mode and the inability of the Pump to detect air in line.