MOOG Medical Devices Group

5 recalls • • Categories: Infusion Pumps

Moog is recalling Curlin 4000, 6000, and PainSmart Infusion Pumps manufactured or serviced by MMDG between March 18, 2015 and November 6, 2015 because the master pumps used to calibrate them were out of tolerance and may cause an over delivery of medications and solutions at a rate of 1.2 - 6.8% faster than programmed.

Dec 23, 2015 Infusion Pumps Nationwide View Details →

Potential for leaking between the purple ENFit connector and the white transitional stepped connector due to cracking or breaking of the purple connector.

Nov 20, 2015 Infusion Pumps View Details →

MOOG Medical Devices Group is conducting a voluntary for a limited set of serial numbers of the EnteraLite Infinity and Infinity Orange Small Volume Enteral Feeding Pumps because certain diodes on the PCB assembly may result in insufficient voltage to a portion of the pump resulting in an error 13 message and shutting the pump down entirely.

Nov 8, 2012 Infusion Pumps Nationwide View Details →

This recall is the same issue for the FDA processed recall initiated by Moog Medical Devices on 1/25/11 which was related to customers reporting unclearable Air-In-Line alarm with specific lot numbers of Curlin Administration Sets used with Curlin Ambulatory Infusion Pumps. The lots associated with this recall were inadvertently omitted from the original recall which has since been terminated by

Sep 26, 2012 Infusion Pumps Nationwide View Details →