Optetrak Offset Tibial Tray and Screws, Size 1DeltaF/1T, Alpha, Ref 208-04-13, Sterile, RX.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    All lot numbers
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Exactech, Inc.
Reason for Recall:
There is a design incompatibility between certain sizes of the offset spine stiffener screw and the constrained condylar tibial insert when used in conjunction with an offset tibial tray.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Optetrak Offset Tibial Tray and Screws, Size 1DeltaF/1T, Alpha, Ref 208-04-13, Sterile, RX.

Product Codes/Lot Numbers:

All lot numbers

Distribution:

Distributed in: AL, FL, IN, ME, MN, NJ, NY, OH, OK, VA

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0129-2018