Alaris Pump Module model 8100 manufactured between November 2011 and March 2012; Alaris Pump Module serviced with LVP Mechanism Sub Assembly (P/N) 10942012 between November 2011 and March 2012; Alaris Pump module Bezel Kit Assembly (P/N) 10964559) shipped between November 2011 and March 2012. The Alaris Pump module is a large volume infusion pump offered under the Alaris System. The Alaris Pump module will deliver medication and fluids using the IV administration sets for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous or epidural.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    All Alaris Pump module model 8100 manufactured and serviced with the bezel assembly manufactured between October 2011 and November 2011; Serial Numbers - Serial 13537720 13537735 13538005 13538811 13540340 13541935 13543108 13543503 13543666 13544201 13544350 13545092 13545892 13546315 13546638 13548091 13550608 13550614 13550629 13550670 13552666 13552667 13552781 13552782 13553628 13553661 13553714 13553730 13553739 13554035 12851369 12851850 9879510 9879561 9880323 9880838 9881158 12564799 13157405 13563286 13566911 13566932 13566933 13567118 13585121 13585213 13585493 13586248 13586314 13586329 13586339 13586340 13586408 13586578 13586723 13586731 13586733 13586738 13586746 13586758 13586763 13586764 13586768 13586775 13586782 13586784 13586793 13586795 13586797 13586798 13586819 13586820 13586835 13586836 13586839 13586855 13586860 13586877 13586878 13586898 13586952 13586957 13586960 13586967 13586976 13587076 13587077 13587118 13587220 13587231 13587232 13587233 13587262 13587263 13587266 13587289 13587290 13587291 13587297 13587312 13587315 13587345 13587398 13196293 13203541 13207333 13213463 13214795 13217554 13219137 13219971 13375015 13375617 13375737 12471635 9857470 9858862 9858865 9859690 9860106 9860746 13184877 13186091 13582527 13582709 13599442 12460530 12474814 12771113 13619912 13620162 13585671 13566067 13616956 13545222 13545224 13545250 13545260 13545262 13550278 13550317 13550318 13550319 13550325 13550337 13550365 13550369 13550370 13550379 13553239 13549816 13549832 13549886 13549901 13549903 13550267 13550271 13550339 13550340 13550345 13550346 13550355 13550356 13550357 13550366 13553126 13553134 13553403 13553415 13568914 13568950 13569155 13569179 13569248 13569694 13569829 13569830 13569838 13569839 13569854 13569876 13569947 13569948 13569953 13569955 13569957 13570013 13570014 13570015 13570017 13570022 13570023 13570052 13570063 13570076 13570077 13570127 13570128 13570137 13570138 13570168 13570184 13570198 13570222 13570236 13570249 13570255 13570256 13570263 12518795 12520618 12520990 13139496 13152562 13153423 13550976 13605208 13605881 13605956 13620375 13620540 13620587 13620902 13620923 13552790 13556036 13556048 13556052 13556056 13558162 13558197 13558234 13142008 13142059 13142277 13142311 13550252 13552900 13552901 13554258 13554809 13554826 13554865 13554878 13554887 13555000 13555023 13036704 4130472 13551713 13551714 13551715 13551750 13552281 13122441 13606144 13606620 13606680 13606687 12652922 13551874 13554663 13555012 13555162 13555172 13555189 13555277 13555711 13555714 13555740 13555743 13562785 13562126 13562559 13566817 13566828 13566873 13618695 13618974 13620108 13620109 13623007 13626334 4049223 12389924 12390048 12390059 12390063 12390110 12421349 13410265 13481070 13415581 13415684 13546628 13546711 13550790 13550951 9895104 12548328 12366930 12367649 12367659 12367775 12367815 12368485 13550687 13550967 13537121 13537307 13537615 13537708 13537929 13538004 13538163 13538907 13539151 13539520 13539841 13539844 13540113 13540190 13540375 13540560 13540586 13541136 13541300 13541571 13541863 13541874 13542038 13542061 13542222 13542379 13542436 13542446 13542474 13542525 13542684 13542888 13542903 13542957 13542989 13543284 13543448 13543508 13543610 13543667 13543668 13543746 13543917 13544255 13544355 13545718 13546268 13546294 13546518 13546684 13546696 13546821 13547115 13547166 13547222 13547288 13547340 13547345 13547378 13547382 13547407 13547413 13547427 13547530 13547563 13548096 13548131 13548135 13548173 13548309 13548311 13548425 13548455 13549072 13549099 13549100 13549225 13549229 13549244 13549258 13549279 13549285 13549314 13549315 13549319 13549324 13549328 13549939 13549974 13549992 13549996 13550004 13550008 13550033 13550039 13550048 13550054 13550058 13550068 13550076 13550078 13550081 13550087 13550102 13550107 13550113 13550127 13550131 13550134 13550136 13550138 13550141 13550146 13550147 13550149 13550150 13550151 13550152 13550153 13550164 13550165 13550168 13550171 13550175 13550178 13550180 13550212 13550224 13550230 13550232 13550235 13550635 13550654 13550673 13550678 13550684 13550697 13550706 13550707 13550725 13550728 13550729 13550732 13550742 13550750 13550751 13550753 13550774 13550776 13550787 4071770 4072996 13582245 13582462 13582640 13582680 4107349 12340716 13571569 13571727 13571729 13571749 13571770 13571812 13571813 13571820 13571828 13571848 13571854 13571856 13571878 13571894 13571917 13571923 13571930 13571951 13572147 13572149 13572183 13572202 13572205 13572211 13572212 13572214 13572226 13572229 13572241 13572249 13556663 13556669 13556672 13556673 13556683 13574132 13576370 13587807 13612188 13616914 13617837 13618602 13619396 13619872 13619896 13619956 13620006 13620357 13620552 13621706 13621787 13621968 13622945 13623052 13623339 13623429 13623619 13623843 13623879 13623889 13624255 13624679 13624798 13624859 13625544 13626012 13626042 13626134 13626210 13626383 13626599 13626685 13627074 13627153 13627289 13627372 13627503 13585554 13587666 13598417 13601558 13601987 13610893 13613150 13585095 13590080 13591941 13592282 13593366 13595657 13595818 13596981 13597308 13598495 13599219 13599275 13599409 13599449 13600261 13600345 13600374 13600432 13600444 13600445 13600471 13600479 13600503 13600524 13600605 13600619 13600622 13600623 13600672 13600683 13600766 13600769 13600777 13600897 13600959 13601035 13601054 13601079 13601081 13601098 13601112 13601145 13601165 13604013 13604088 13604104 13604105 13604109 13604112 13604139 4034918 4035880 12745455 12781937 13536784 13562891 13584567 13585555 13588652 13593516 13594815 13596828 13598383 13598766 13599103 13599391 13599506 13599901 13600003 13600618 13600774 13600783 13600834 13600876 13600900 13600910 13601018 13601050 13601237 13601919 13602609 13602893 13602955 13603251 13603340 13603367 13603377 13603460 13603963 13604083 13604921 13604963 13605204 13605205 13605498 13605691 13605754 13605767 13605792 13605821 13606065 13606337 13606377 13606447 13606583 13606701 13606810 13606866 13606911 13606936 13606945 13606951 13606989 13607027 13607087 13607115 13607212 13607629 13607653 13607681 13607725 13607851 13607896 13608010 13608102 13608160 13608168 13608240 13608274 13608302 13608314 13608346 13608413 13608425 13609048 13609052 13609069 13609244 13610420 13610440 13610465 13610581 13610653 13610667 13610710 13610713 13610732 13610735 13610795 13610823 13610872 13610892 13610901 13610902 13610904 13610907 13611369 13611502 13611720 13611817 13612071 13612848 13612879 13612913 13612918 13612955 13612959 13612970 13612980 13612982 13612987 13612992 13612997 13613000 13613008 13613009 13613010 13613012 13613022 13613034 13613037 13613038 13613039 13613040 13613048 13613051 13613053 13613057 13613058 13613071 13613073 13613083 13613096 13613167 13613194 13613252 13613258 13613265 13613280 13613291 13613306 13613318 13613331 13613341 13613352 13613359 13613364 13613415 13613433 13613674 13613718 13613719 13613720 13613729 13613753 13613780 13613793 13613796 13613811 13613815 13613816 13613831 13613841 13613846 13613851 13613858 13613878 13613881 13613890 13613897 13613898 13613905 13613907 13613909 13613925 13613926 13613929 13613933 13613934 13613943 13613944 13613951 13613954 13613956 13613959 13613962 13613977 13613981 13613984 13613985 13613987 13613989 13613990 13613996 13613999 13614002 13614005 13614007 13614009 13614024 13614029 13614031 13614034 13614035 13614039 13614042 13614044 13614053 13614061 13614063 13614066 13614067 13614069 13614080 13614083 13614084 13614085 13614088 13614091 13614093 13614095 13614108 13614110 13614111 13614114 13614115 13614116 13614117 13614118 13614119 13614120 13614125 13614126 13614135 13614138 13614141 13614147 13614149 13614152 13614153 13614154 13614156 13614157 13614165 13614166 13614172 13614173 13614175 13614178 13614194 13614200 13614204 13614207 13614228 13614248 13614249 13614255 13614268 13614274 13614277 13614300 13614301 13614315 13614319 13614324 13614328 13614336 13614349 13614351 13614352 13614355 13614358 13614361 13614362 13614371 13614374 13614396 13614405 13614409 13614414 13614443 13614445 13614446 13614453 13614454 13614455 13614457 13614458 13614464 13614465 13614466 13614467 13614468 13614472 13614473 13614479 13614481 13614487 13614489 13614490 13614491 13614492 13614503 13614504 13614505 13614507 13614514 13614521 13614522 13614531 13614535 13614549 13614557 13614560 13614561 13614563 13614565 13614582 13614584 13614586 13614587 13614588 13614591 13614595 13614601 13614608 13614642 13614650 13614682 13614683 13614684 13614691 13614692 13614695 13614696 13614698 13614702 13614709 13614714 13614715 13614716 13614717 13614736 13614777 13614793 13614795 13614807 13614830 13614831 13614832 13614907 13614928 13614975 13614979 13614980 13614981 13614982 13614987 13614988 13614993 13614994 13615000 13615010 13615017 13615018 13615019 13615021 13615024 13615032 13615033 13615035 13615038 13615040 13615041 13615046 13615057 13615059 13615060 13615062 13615063 13615071 13615073 13615075 13615076 13615077 13615079 13615084 13615086 13615091 13615092 13615107 13615108 13615126 13615128 13615130 13615138 13615139 13615149 13615155 13615165 13615166 13615168 13615169 13615176 13615181 13615183 13615184 13615187 13615190 13615196 13615197 13615202 13615205 13615206 13615207 13615208 13615209 13615210 13615211 13615228 13615234 13615236 13615237 13615255 13615258 13615261 13615262 13615264 13615265 13615268 13615269 13615270 13615276 13615280 13615285 13615288 13615291 13615292 13615298 13615300 13615301 13615306 13615308 13615310 13615318 13615320 13615324 13615326 13615328 13615335 13615337 13615338 13615339 13615341 13615343 13615344 13615345 13615346 13615354 13615355 13615364 13615368 13615372 13615374 13615375 13615379 13615380 13615381 13615392 13615394 13615400 13615407 13615416 13615417 13615418 13615421 13615422 13615423 13615426 13615427 13615428 13615429 13615434 13615437 13615438 13615439 13615440 13615441 13615442 13615443 13615446 13615447 13615449 13615451 13615454 13615460 13615462 13615464 13615481 13615490 13615493 13615498 13615499 13615500 13615517 13615527 13615536 13615537 13615544 13615545 13615546 13615548 13615549 13615552 13615554 13615555 13615556 13615558 13615559 13615560 13615561 13615562 13615564 13615569 13615570 13615573 13615576 13615581 13615584 13615587 13615588 13615590 13615595 13615596 13615602 13615603 13615604 13615621 13615622 13615623 13615624 13615626 13615627 13615632 13615634 13615635 13615638 13615639 13615640 13615641 13615644 13615648 13615651 13615655 13615656 13615666 13615677 13615679 13615681 13615683 13615689 13615690 13615691 13615698 13615699 13615700 13615701 13615706 13615708 13615711 13615712 13615718 13615720 13615721 13615723 13615724 13615725 13615726 13615727 13615734 13615741 13615742 13615744 13615749 13615750 13615751 13615752 13615754 13615755 13615757 13615758 13615759 13615762 13615763 13615764 13615766 13615767 13615770 13615775 13615776 13615781 13615782 13615783 13615786 13615788 13615789 13615790 13615791 13615795 13615796 13615797 13615799 13615800 13615801 13615803 13615805 13615806 13615807 13615808 13615810 13615811 13615812 13615815 13615818 13615832 13615833 13615834 13615835 13615848 13615853 13615856 13615866 13615878 13615879 13615880 13615881 13615891 13615893 13615901 13615906 13615907 13615916 13615925 13615969 13615970 13615977 13615981 13615988 13615989 13615993 13615994 13616007 13616009 13616014 13616022 13616024 13616026 13616027 13616030 13616034 13616037 13616041 13616047 13616048 13616125 13616132 13616133 13616134 13616136 13616137 13616140 13616142 13616143 13616144 13616146 13616148 13616153 13616157 13616160 13616162 13616163 13616170 13616171 13616172 13616176 13616178 13616180 13616181 13616183 13616186 13616187 13616190 13616196 13616201 13616203 13616205 13616214 13616220 13616225 13616228 13616229 13616231 13616238 13616239 13616240 13616242 13616247 13616265 13616271 13616273 13616274 13616275 13616276 13616277 13616281 13616282 13616289 13616298 13616300 13616301 13616304 13616305 13616306 13616308 13616310 13616311 13616313 13616316 13616317 13616319 13616320 13616321 13616322 13616323 13616324 13616325 13616326 13616328 13616331 13616332 13616333 13616334 13616335 13616336 13616337 13616339 13616340 13616341 13616342 13616343 13616344 13616345 13616346 13616348 13616350 13616351 13616352 13616353 13616355 13616356 13616357 13616358 13616359 13616360 13616361 13616362 13616363 13616364 13616365 13616367 13616368 13616370 13616371 13616375 13616377 13616379 13616381 13616384 13616387 13616388 13616389 13616390 13616391 13616393 13616394 13616395 13616397 13616399 13616400 13616401 13616404 13616407 13616409 13616410 13616411 13616413 13616414 13616416 13616417 13616419 13616420 13616421 13616423 13616424 13616427 13616428 13616430 13616431 13616434 13616435 13616436 13616437 13616438 13616439 13616440 13616441 13616442 13616443 13616444 13616445 13616446 13616447 13616450 13616451 13616452 13616453 13616454 13616455 13616458 13616463 13616464 13616466 13616468 13616469 13616471 13616472 13616473 13616478 13616480 13616481 13616483 13616484 13616485 13616486 13616487 13616488 13616489 13616490 13616491 13616494 13616495 13616496 13616500 13616502 13616503 13616505 13616506 13616507 13616509 13616510 13616512 13616513 13616514 13616515 13616590 13616592 13616593 13616598 13616600 13616602 13616603 13616605 13616606 13616607 13616609 13616610 13616612 13616613 13616620 13616622 13616623 13616625 13616630 13616631 13616633 13616634 13616635 13616636 13616638 13616641 13616643 13616645 13616650 13616653 13616654 13616655 13616658 13616670 13616671 13616708 13616710 13616713 13616716 13616719 13616728 13616735 13616740 13616750 13616755 13616756 13616768 13616776 13616779 13616793 13616801 13616839 13616840 13616841 13616846 13616848 13616849 13616858 13616862 13616863 13616866 13616872 13616881 13616889 13616891 13616899 13616909 13616920 13616927 13616929 13616934 13616943 13616946 13616954 13616958 13616968 13616973 13616974 13616975 13616980 13616989 13616995 13617005 13617008 13617038 13617039 13617050 13617059 13617068 13617080 13617098 13617121 13617152 13617158 13617203 13617224 13617230 13617235 13617253 13617254 13617267 13617284 13617297 13617303 13617307 13617308 13617315 13617336 13617340 13617344 13617346 13617350 13617352 13617353 13617357 13617368 13617370 13617372 13617373 13617374 13617377 13617378 13617380 13617391 13617394 13617399 13617403 13617404 13617405 13617406 13617407 13617408 13617409 13617413 13617415 13617425 13617428 13617430 13617435 13617436 13617438 13617439 13617440 13617441 13617448 13617454 13617456 13617457 13617459 13617461 13617468 13617469 13617473 13617476 13617483 13617492 13617493 13617495 13617496 13617497 13617502 13617506 13617507 13617508 13617509 13617510 13617511 13617518 13617519 13617526 13617529 13617530 13617533 13617534 13617535 13617538 13617539 13617540 13617546 13617556 13617565 13617566 13617581 13617586 13617587 13617588 13617589 13617590 13617591 13617593 13617594 13617596 13617599 13617601 13617604 13617607 13617613 13617614 13617621 13617622 13617623 13617624 13617626 13617627 13617634 13617635 13617636 13617637 13617638 13617639 13617641 13617642 13617643 13617644 13617646 13617654 13617668 13617674 13617686 13617687 13617689 13617690 13617691 13617694 13617695 13617696 13617698 13617699 13617709 13617710 13617714 13617718 13617724 13617725 13617740 13617742 13617745 13617747 13617749 13617750 13617762 13617763 13617764 13617767 13617768 13617770 13617777 13617784 13617789 13617790 13617791 13617792 13617793 13617794 13617795 13617796 13617799 13617800 13617801 13617804 13617807 13617810 13617811 13617812 13617813 13617814 13617818 13617819 13617820 13617821 13617822 13617823 13617824 13617826 13617829 13617830 13617831 13617833 13617834 13617836 13617838 13617840 13617841 13617842 13617844 13617845 13617847 13617848 13617849 13617851 13617853 13617854 13617856 13617857 13617858 13617859 13617860 13617861 13617863 13617865 13617867 13617868 13617869 13617871 13617872 13617874 13617875 13617881 13617882 13617883 13617884 13617885 13617890 13617894 13617895 13617897 13617898 13617904 13617912 13617913 13617914 13617916 13617923 13617925 13617927 13617930 13617931 13617934 13617935 13617937 13617942 13617946 13617953 13617963 13617973 13617979 13618051 13618052 13618055 13618067 13618069 13618071 13618072 13618074 13618075 13618077 13618083 13618084 13618086 13618099 13618105 13618111 13618116 13618344 13618351 13618353 13618354 13618367 13618384 13618389 13618390 13618392 13618393 13618396 13618399 13618401 13618426 13618427 13618441 13618442 13618446 13618447 13618449 13618450 13618451 13618459 13618474 13618496 13618504 13618507 13618511 13618513 13618514 13618516 13618518 13618520 13618522 13618523 13618524 13618530 13618532 13618533 13618534 13618538 13618549 13618551 13618561 13618562 13618570 13618575 13618578 13618579 13618583 13618586 13618588 13618590 13618591 13618604 13618612 13618619 13618622 13618630 13618636 13618637 13618644 13618682 13618697 13618699 13618737 13618747 13618750 13618751 13618752 13618754 13618755 13618756 13618757 13618758 13618761 13618763 13618782 13618785 13618787 13618788 13618790 13618793 13618806 13618809 13618810 13618811 13618812 13618821 13618824 13618825 13618826 13618827 13618830 13618838 13618840 13618853 13618854 13618856 13618857 13618858 13618872 13618874 13618878 13618880 13618885 13618886 13618910 13618917 13618921 13618932 13618944 13618950 13619032 13619774 13619871 13619945 13619959 13619963 13620091 13620101 13620131 13620703 13621052 13621116 13621451 13621676 13623848 13626838 13627267 13627477 13627515 13627540 13627553 13627563 13627565 13627573 13627574 13627575 13627579 13627580 13627581 13388491 13612933 13556320 13560644 13585491 13585619 13588572 13589270 13589294 13589467 13600017 13600075 13600103 13600110 13600116 13600119 13600133 13600135 13600161 13600218 13600239 13600365 13600521 13600540 13600548 13600829 13600887 13601006 13601043 13601049 13601058 13601085 13601086 13601087 13601117 13601120 13601121 13601137 13601140 13601175 12943397 12943427 12943590 12943594 12943684 12943724 3687431 3687533 3689010 3689852 3691237 13588966 13595492 13599714 13599911 13603990 13604073 13604120 13604125 13604132 13604176 13604177 13604371 13604465 13604475 13604477 13604478 13604491 13604501 13604539 13604550 13604552 13604615 13604629 13604631 13604633 13604634 13604635 13604636 13604638 13604639 13604646 13604658 13604659 13604669 13604682 13604683 13604696 13604699 13604706 13604792 13604824 13604858 13604871 13604897 13604898 13604903 13604919 13604922 13604959 13604962 13604975 13605010 13605030 13605603 13411475 13411465 13411659 13186123 13558319 12920284 12643797 12643800 13605042 13605415 13605517 13605546 13605575 13605919 13616652 13617371 13617732 13622175 9905416 9905925 9906038 9906070 9912145 9914030 9914180 9914202 9914703 9914948 9914960 9915067 9915076 9915589 9915818 9915937 9915958 9916063 9916098 9916179 9916529 9916547 9916722 9916737 9916970 9917140 9917414 9917451 9917485 9917631 12341830 12344336 13558214 13559397 13562422 13562473 13565038 13565416 13565865 13566039 13566143 13568862 13568879 13568882 13568883 13568885 13568907 13568921 13568937 13568938 13568940 13568941 13568943 13568944 13568948 13568949 13568951 13568953 13568955 13568977 13568993 13568995 13568996 13568998 13568999 13569013 13569022 13569091 13569108 13569113 13569118 13569134 13569215 13569221 13569233 13569234 13569241 13569247 13569249 13569256 13569257 13569258 13569277 13569296 13569301 13569327 13569436 13569445 13569446 13569459 13569475 13569632 13569644 13569646 13569650 13569659 13569662 13569663 13569664 13569666 13569669 13569682 13569683 13569692 13569693 13569699 13569706 13569714 13569721 13569727 13569732 13569740 13569779 13569782 13569783 13569784 13569797 13569804 13569807 13569814 13569817 13569831 13569833 13569849 13569870 13569871 13569877 13569878 13569879 13569895 13569897 13569901 13569934 13569939 13569940 13569942 13569943 13569944 13569946 13569949 13569950 13569954 13569992 13570011 13570012 13570016 13570034 13570040 13570042 13570046 13570047 13570048 13570053 13570054 13570064 13570065 13570066 13570067 13570073 13570125 13570130 13570134 13570149 13570189 13570190 13570192 13570199 13570208 13570231 13570288 13570290 13570326 13570327 13570329 13570330 13570343 13570345 13570346 13570347 13570348 13570349 13570350 13570373 13570378 13570383 13570384 13570386 13570389 13570396 13570397 13570398 13570399 13570400 13570401 13570404 13570405 13570409 13570410 13570415 13570419 13570422 13570427 13570436 13570442 13570444 13570447 13570451 13570453 13570456 13570463 13570465 13570472 13570479 13570483 13570490 13570491 13570492 13570495 13570496 13570507 13570508 13570511 13570512 13570514 13570515 13570516 13570519 13570520 13570521 13570522 13570525 13570526 13570529 13570530 13570531 13570533 13570534 13570535 13570536 13570537 13570538 13570539 13570540 13570542 13570543 13570544 13570546 13570548 13570561 13570562 13570563 13570564 13570567 13570568 13570569 13570570 13570571 13570572 13570573 13570574 13570575 13570577 13570581 13570583 13570590 13570595 13570603 13570604 13570606 13570801 13570811 13570823 13570847 13570859 13570866 13570947 13570960 13570965 13570966 13570979 13571030 13571032 13551107 12389182 12389259 12389723 13585503 12583308 9896323 12344631 12951526 12792273 12816712 12827932 2000687 12594225 9870369 9870862 9871285 9871311 4055327 4055340 4055686 4055696 4055728 4055758 4055864 4055938 4055997 4060066 4060332 4060373 4063251 4063254 4063264 4063301 4063327 4063328 4063368 4063372 4063528 4063580 4063585 4063589 4063603 4063829 4064628 4064712 4064797 4064802 4064818 4064861 4064875 4065047 4065075 4065178 4065215 4065318 4065361 4065374 4065561 4065564 4065566 4065585 4065655 4065690 4068327 4068409 4068456 4068463 4068473 4068773 4068885 4068906 4068909 4068915 4069042 4069113 4069163 4069175 4069200 4069205 4069215 4069226 4069229 4069249 4069296 4069310 4069312 4069339 4069359 4069485 4069515 4069619 4069648 4069649 4069654 4069658 4069697 4069712 4069760 4069763 4069766 4069771 4069774 4069780 4069782 4069793 4069798 4069821 4069832 4069970 4070038 4090905 4090909 12390952 12392611 12392763 12426401 12430945 12432027 12432057 12432310 12432375 12432443 12432536 12433282 12433635 9360734 13585621 13586208 13586229 13586235 13586243 13586298 13586315 13586325 13586328 13586342 13586357 13586396 13586409 13586416 13586600 13586729 13586737 13586756 13586760 13586762 13586766 13586770 13586774 13586776 13586781 13586794 13586796 13586813 13586815 13586816 13586817 13586856 13586874 13586876 13586879 13586953 13586974 13587001 13587095 13587096 13588738 13589017 13589295 13589445 13589479 3701183 4003048 12462783 13568983 13569122 13569136 13569148 13543131 13543445 13543487 13543855 13543861 13543862 13543908 13543922 13543927 13543931 13543934 13543935 13543961 13543982 13543983 13543984 13543998 13543999 13544008 13544009 13544010 13544012 13544013 13544014 13544016 13544017 13544044 13544055 13544070 13544072 13544074 13544083 13544093 13544097 13544099 13544100 13544101 13544105 13544108 13544111 13544114 13544121 13544122 13544123 13544131 13544132 13544146 13544152 13544155 13544156 13544158 13544164 13544180 13544207 13544214 13544270 13544273 13544277 13544278 13544281 13544284 13544300 13544302 13544333 13544401 13544434 13544447 13544456 13545020 13545021 13545039 13545042 13545044 13545049 13545052 13545054 13545057 13545063 13545064 13545065 13545066 13545069 13545074 13545075 13545076 13545081 13545083 13545084 13545086 13545087 13545088 13545089 13545095 13545099 13545100 13545102 13545103 13545107 13545109 13545117 13545119 13545124 13545127 13545130 13545134 13545135 13545136 13545139 13545140 13545141 13545142 13545154 13545159 13545160 13545161 13545162 13545163 13545164 13545166 13545167 13545168 13545169 13545171 13545172 13545173 13545175 13545176 13545177 13545179 13545180 13545181 13545182 13545183 13545185 13545188 13545189 13545190 13545193 13545194 13545195 13545196 13545200 13545201 13545203 13545205 13545208 13545209 13545210 13545212 13545213 13545214 13545216 13545218 13545220 13545221 13545223 13545226 13545228 13545229 13545238 13545240 13545241 13545249 13545251 13545254 13545259 13545261 13545263 13545266 13545267 13545463 13545464 13545471 13545472 13545473 13545474 13545475 13545489 13545662 13545663 13545715 13545717 13545748 13545810 13545884 13545886 13545889 13545919 13545940 13545945 13545946 13545953 13545960 13545970 13545991 13546030 13546050 13546053 13546060 13546063 13546064 13546067 13546068 13546071 13546073 13546079 13546080 13546092 13546094 13546095 13546096 13546111 13546112 13546113 13546114 13546121 13546122 13546124 13546127 13546128 13546131 13546132 13546137 13546139 13546141 13546155 13546159 13546160 13546166 13546169 13546176 13546184 13546190 13546191 13546193 13546199 13546205 13546206 13546209 13546214 13546218 13546219 13546223 13546225 13546228 13546229 13546230 13546231 13546233 13546263 13546280 13546285 13546287 13546295 13546296 13546300 13546302 13546313 13546316 13546317 13546322 13546323 13546329 13546330 13546334 13546336 13546340 13546342 13546347 13546349 13546351 13546354 13546358 13546375 13546382 13546385 13546388 13546390 13546393 13546402 13546407 13546410 13546411 13546413 13546419 13546420 13546421 13546423 13546424 13546425 13546427 13546430 13546434 13546435 13546437 13546438 13546440 13546446 13546447 13546449 13546450 13546458 13546461 13546465 13546468 13546469 13546477 13546502 13546505 13546506 13546510 13546513 13546519 13546520 13546528 13546530 13546532 13546542 13546547 13546548 13546561 13546659 13546661 13546827 13546857 13546869 13546870 13546873 13546881 13546882 13546883 13546898 13546899 13546900 13546920 13546922 13546935 13547013 13547020 13547037 13547050 13547051 13547064 13547082 13547083 13547085 13547088 13547148 13547175 13547197 13547198 13547206 13547211 13547217 13547245 13547251 13547287 12830454 12858882 4124663 9870359 12689982 12443023 13552780 13552783 13553072 13553240 13553263 13553283 13553302 13553324 13553414 13554003 13554004 13554005 13554020 13554023 13554025 13554026 13554029 13554030 13554031 13554033 13554036 13554051 13554052 13554058 13554062 13554071 13554075 13554087 13554088 13554097 13554098 13554108 13554114 13554118 13554126 13554129 13554130 13554132 13554133 13554137 13554138 13554141 13554143 13554146 13554148 13554151 13554154 13554180 13554195 12562674 13556050 13556222 13556232 13556247 13556251 13556276 13556310 13556369 13556467 13556479 13556484 13556501 13556527 13556535 13556586 13556610 13556611 13556656 13556661 13556666 13556670 13556675 13556682 13556686 13556692 13616640 13617410 13617746 12563814 13551897 13589063 13589083 13589123 13589138 13589219 12356973 13608884 13609053 13609190 13609556 13609673 13616712 13616715 13616949 13616983 12691937 13581261 13582463 12467280 13512991 13567722 13620349 13626407 9974917 9938571 13548354 13548435 13550777 13554079 13555191 13555252 13555672 13555770 13555848 13555911 13556007 13556015 13556022 13556094 13557686 13557729 13557869 13557924 13558494 13558798 13558834 13559206 13559378 13559380 13559512 13559534 13559571 13560423 13560734 13561283 13561300 13561850 13561856 13561861 13561918 13561997 13562245 13562343 13562373 13562441 13562574 13562872 13562977 13563030 13563125 13563188 13565005 13565314 13565376 13565415 13565420 13565457 13565474 13565492 13565502 13565690 13565698 13566071 13566652 13566653 13566682 13566737 13566806 13566809 13567178 13567302 13567382 13567503 13567572 13567581 13567705 13569222 13569237 13569338 13569630 13569805 13569813 13569952 13569956 13570068 13570235 13570252 13570417 13570454 13570517 13570523 13570552 13570599 13570634 13570636 13570660 13570665 13570710 13570766 13570849 13570892 13570894 13570976 13571347 13572032 13572285 13572647 13572704 13572722 13573032 13573335 13573341 13573619 13573627 13573692 13574136 13574143 13574221 13574339 13574851 13575330 13575381 13575391 13575705 13575780 13575924 13576266 13576550 13576933 13577083 13577197 13577304 13577305 13577352 13577387 13577394 13577433 13579994 13580052 13584049 9864812 9865832 9866037 9866112 9866301 9866331 9866396 9866545 9866882 9867044 9867172 9867485 9867663 9867865 2000674 13110969 13374022 13374456 13375213 13375219 13375282 13376221 12463176 12463789 13565149 13568880 13568899 13569158 13569305 13576964 13581486 13585305 13585609 13585646 13585666 13585689 13585708 13585725 13585744 13585767 13585889 13585946 13620965 13620967 13620973 12580577 13015820 13185969 13186052 4120867 4121779 4122901 13547289 13574079 13192802 13194564 12777319 13537247 13537248 13538827 13541386 13541825 13543196 13544356 13545227 13545236 13545237 13545239 13545242 13545243 13545245 13545246 13545247 13545248 13545255 13545256 13545258 13545264 13545265 13545456 13545459 13545460 13545465 13545478 13545479 13545486 13545487 13545488 13545491 13545500 13545501 13545503 13545504 13545505 13545506 13545520 13545521 13545523 13545528 13545529 13545530 13545531 13545532 13545533 13545540 1354554
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
CareFusion 303, Inc.
Reason for Recall:
The recalling firm has received reports of increased or decreased flows that have occurred in certain pumps.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Alaris Pump Module model 8100 manufactured between November 2011 and March 2012; Alaris Pump Module serviced with LVP Mechanism Sub Assembly (P/N) 10942012 between November 2011 and March 2012; Alaris Pump module Bezel Kit Assembly (P/N) 10964559) shipped between November 2011 and March 2012. The Alaris Pump module is a large volume infusion pump offered under the Alaris System. The Alaris Pump module will deliver medication and fluids using the IV administration sets for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous or epidural.

Product Codes/Lot Numbers:

All Alaris Pump module model 8100 manufactured and serviced with the bezel assembly manufactured between October 2011 and November 2011; Serial Numbers - Serial 13537720 13537735 13538005 13538811 13540340 13541935 13543108 13543503 13543666 13544201 13544350 13545092 13545892 13546315 13546638 13548091 13550608 13550614 13550629 13550670 13552666 13552667 13552781 13552782 13553628 13553661 13553714 13553730 13553739 13554035 12851369 12851850 9879510 9879561 9880323 9880838 9881158 12564799 13157405 13563286 13566911 13566932 13566933 13567118 13585121 13585213 13585493 13586248 13586314 13586329 13586339 13586340 13586408 13586578 13586723 13586731 13586733 13586738 13586746 13586758 13586763 13586764 13586768 13586775 13586782 13586784 13586793 13586795 13586797 13586798 13586819 13586820 13586835 13586836 13586839 13586855 13586860 13586877 13586878 13586898 13586952 13586957 13586960 13586967 13586976 13587076 13587077 13587118 13587220 13587231 13587232 13587233 13587262 13587263 13587266 13587289 13587290 13587291 13587297 13587312 13587315 13587345 13587398 13196293 13203541 13207333 13213463 13214795 13217554 13219137 13219971 13375015 13375617 13375737 12471635 9857470 9858862 9858865 9859690 9860106 9860746 13184877 13186091 13582527 13582709 13599442 12460530 12474814 12771113 13619912 13620162 13585671 13566067 13616956 13545222 13545224 13545250 13545260 13545262 13550278 13550317 13550318 13550319 13550325 13550337 13550365 13550369 13550370 13550379 13553239 13549816 13549832 13549886 13549901 13549903 13550267 13550271 13550339 13550340 13550345 13550346 13550355 13550356 13550357 13550366 13553126 13553134 13553403 13553415 13568914 13568950 13569155 13569179 13569248 13569694 13569829 13569830 13569838 13569839 13569854 13569876 13569947 13569948 13569953 13569955 13569957 13570013 13570014 13570015 13570017 13570022 13570023 13570052 13570063 13570076 13570077 13570127 13570128 13570137 13570138 13570168 13570184 13570198 13570222 13570236 13570249 13570255 13570256 13570263 12518795 12520618 12520990 13139496 13152562 13153423 13550976 13605208 13605881 13605956 13620375 13620540 13620587 13620902 13620923 13552790 13556036 13556048 13556052 13556056 13558162 13558197 13558234 13142008 13142059 13142277 13142311 13550252 13552900 13552901 13554258 13554809 13554826 13554865 13554878 13554887 13555000 13555023 13036704 4130472 13551713 13551714 13551715 13551750 13552281 13122441 13606144 13606620 13606680 13606687 12652922 13551874 13554663 13555012 13555162 13555172 13555189 13555277 13555711 13555714 13555740 13555743 13562785 13562126 13562559 13566817 13566828 13566873 13618695 13618974 13620108 13620109 13623007 13626334 4049223 12389924 12390048 12390059 12390063 12390110 12421349 13410265 13481070 13415581 13415684 13546628 13546711 13550790 13550951 9895104 12548328 12366930 12367649 12367659 12367775 12367815 12368485 13550687 13550967 13537121 13537307 13537615 13537708 13537929 13538004 13538163 13538907 13539151 13539520 13539841 13539844 13540113 13540190 13540375 13540560 13540586 13541136 13541300 13541571 13541863 13541874 13542038 13542061 13542222 13542379 13542436 13542446 13542474 13542525 13542684 13542888 13542903 13542957 13542989 13543284 13543448 13543508 13543610 13543667 13543668 13543746 13543917 13544255 13544355 13545718 13546268 13546294 13546518 13546684 13546696 13546821 13547115 13547166 13547222 13547288 13547340 13547345 13547378 13547382 13547407 13547413 13547427 13547530 13547563 13548096 13548131 13548135 13548173 13548309 13548311 13548425 13548455 13549072 13549099 13549100 13549225 13549229 13549244 13549258 13549279 13549285 13549314 13549315 13549319 13549324 13549328 13549939 13549974 13549992 13549996 13550004 13550008 13550033 13550039 13550048 13550054 13550058 13550068 13550076 13550078 13550081 13550087 13550102 13550107 13550113 13550127 13550131 13550134 13550136 13550138 13550141 13550146 13550147 13550149 13550150 13550151 13550152 13550153 13550164 13550165 13550168 13550171 13550175 13550178 13550180 13550212 13550224 13550230 13550232 13550235 13550635 13550654 13550673 13550678 13550684 13550697 13550706 13550707 13550725 13550728 13550729 13550732 13550742 13550750 13550751 13550753 13550774 13550776 13550787 4071770 4072996 13582245 13582462 13582640 13582680 4107349 12340716 13571569 13571727 13571729 13571749 13571770 13571812 13571813 13571820 13571828 13571848 13571854 13571856 13571878 13571894 13571917 13571923 13571930 13571951 13572147 13572149 13572183 13572202 13572205 13572211 13572212 13572214 13572226 13572229 13572241 13572249 13556663 13556669 13556672 13556673 13556683 13574132 13576370 13587807 13612188 13616914 13617837 13618602 13619396 13619872 13619896 13619956 13620006 13620357 13620552 13621706 13621787 13621968 13622945 13623052 13623339 13623429 13623619 13623843 13623879 13623889 13624255 13624679 13624798 13624859 13625544 13626012 13626042 13626134 13626210 13626383 13626599 13626685 13627074 13627153 13627289 13627372 13627503 13585554 13587666 13598417 13601558 13601987 13610893 13613150 13585095 13590080 13591941 13592282 13593366 13595657 13595818 13596981 13597308 13598495 13599219 13599275 13599409 13599449 13600261 13600345 13600374 13600432 13600444 13600445 13600471 13600479 13600503 13600524 13600605 13600619 13600622 13600623 13600672 13600683 13600766 13600769 13600777 13600897 13600959 13601035 13601054 13601079 13601081 13601098 13601112 13601145 13601165 13604013 13604088 13604104 13604105 13604109 13604112 13604139 4034918 4035880 12745455 12781937 13536784 13562891 13584567 13585555 13588652 13593516 13594815 13596828 13598383 13598766 13599103 13599391 13599506 13599901 13600003 13600618 13600774 13600783 13600834 13600876 13600900 13600910 13601018 13601050 13601237 13601919 13602609 13602893 13602955 13603251 13603340 13603367 13603377 13603460 13603963 13604083 13604921 13604963 13605204 13605205 13605498 13605691 13605754 13605767 13605792 13605821 13606065 13606337 13606377 13606447 13606583 13606701 13606810 13606866 13606911 13606936 13606945 13606951 13606989 13607027 13607087 13607115 13607212 13607629 13607653 13607681 13607725 13607851 13607896 13608010 13608102 13608160 13608168 13608240 13608274 13608302 13608314 13608346 13608413 13608425 13609048 13609052 13609069 13609244 13610420 13610440 13610465 13610581 13610653 13610667 13610710 13610713 13610732 13610735 13610795 13610823 13610872 13610892 13610901 13610902 13610904 13610907 13611369 13611502 13611720 13611817 13612071 13612848 13612879 13612913 13612918 13612955 13612959 13612970 13612980 13612982 13612987 13612992 13612997 13613000 13613008 13613009 13613010 13613012 13613022 13613034 13613037 13613038 13613039 13613040 13613048 13613051 13613053 13613057 13613058 13613071 13613073 13613083 13613096 13613167 13613194 13613252 13613258 13613265 13613280 13613291 13613306 13613318 13613331 13613341 13613352 13613359 13613364 13613415 13613433 13613674 13613718 13613719 13613720 13613729 13613753 13613780 13613793 13613796 13613811 13613815 13613816 13613831 13613841 13613846 13613851 13613858 13613878 13613881 13613890 13613897 13613898 13613905 13613907 13613909 13613925 13613926 13613929 13613933 13613934 13613943 13613944 13613951 13613954 13613956 13613959 13613962 13613977 13613981 13613984 13613985 13613987 13613989 13613990 13613996 13613999 13614002 13614005 13614007 13614009 13614024 13614029 13614031 13614034 13614035 13614039 13614042 13614044 13614053 13614061 13614063 13614066 13614067 13614069 13614080 13614083 13614084 13614085 13614088 13614091 13614093 13614095 13614108 13614110 13614111 13614114 13614115 13614116 13614117 13614118 13614119 13614120 13614125 13614126 13614135 13614138 13614141 13614147 13614149 13614152 13614153 13614154 13614156 13614157 13614165 13614166 13614172 13614173 13614175 13614178 13614194 13614200 13614204 13614207 13614228 13614248 13614249 13614255 13614268 13614274 13614277 13614300 13614301 13614315 13614319 13614324 13614328 13614336 13614349 13614351 13614352 13614355 13614358 13614361 13614362 13614371 13614374 13614396 13614405 13614409 13614414 13614443 13614445 13614446 13614453 13614454 13614455 13614457 13614458 13614464 13614465 13614466 13614467 13614468 13614472 13614473 13614479 13614481 13614487 13614489 13614490 13614491 13614492 13614503 13614504 13614505 13614507 13614514 13614521 13614522 13614531 13614535 13614549 13614557 13614560 13614561 13614563 13614565 13614582 13614584 13614586 13614587 13614588 13614591 13614595 13614601 13614608 13614642 13614650 13614682 13614683 13614684 13614691 13614692 13614695 13614696 13614698 13614702 13614709 13614714 13614715 13614716 13614717 13614736 13614777 13614793 13614795 13614807 13614830 13614831 13614832 13614907 13614928 13614975 13614979 13614980 13614981 13614982 13614987 13614988 13614993 13614994 13615000 13615010 13615017 13615018 13615019 13615021 13615024 13615032 13615033 13615035 13615038 13615040 13615041 13615046 13615057 13615059 13615060 13615062 13615063 13615071 13615073 13615075 13615076 13615077 13615079 13615084 13615086 13615091 13615092 13615107 13615108 13615126 13615128 13615130 13615138 13615139 13615149 13615155 13615165 13615166 13615168 13615169 13615176 13615181 13615183 13615184 13615187 13615190 13615196 13615197 13615202 13615205 13615206 13615207 13615208 13615209 13615210 13615211 13615228 13615234 13615236 13615237 13615255 13615258 13615261 13615262 13615264 13615265 13615268 13615269 13615270 13615276 13615280 13615285 13615288 13615291 13615292 13615298 13615300 13615301 13615306 13615308 13615310 13615318 13615320 13615324 13615326 13615328 13615335 13615337 13615338 13615339 13615341 13615343 13615344 13615345 13615346 13615354 13615355 13615364 13615368 13615372 13615374 13615375 13615379 13615380 13615381 13615392 13615394 13615400 13615407 13615416 13615417 13615418 13615421 13615422 13615423 13615426 13615427 13615428 13615429 13615434 13615437 13615438 13615439 13615440 13615441 13615442 13615443 13615446 13615447 13615449 13615451 13615454 13615460 13615462 13615464 13615481 13615490 13615493 13615498 13615499 13615500 13615517 13615527 13615536 13615537 13615544 13615545 13615546 13615548 13615549 13615552 13615554 13615555 13615556 13615558 13615559 13615560 13615561 13615562 13615564 13615569 13615570 13615573 13615576 13615581 13615584 13615587 13615588 13615590 13615595 13615596 13615602 13615603 13615604 13615621 13615622 13615623 13615624 13615626 13615627 13615632 13615634 13615635 13615638 13615639 13615640 13615641 13615644 13615648 13615651 13615655 13615656 13615666 13615677 13615679 13615681 13615683 13615689 13615690 13615691 13615698 13615699 13615700 13615701 13615706 13615708 13615711 13615712 13615718 13615720 13615721 13615723 13615724 13615725 13615726 13615727 13615734 13615741 13615742 13615744 13615749 13615750 13615751 13615752 13615754 13615755 13615757 13615758 13615759 13615762 13615763 13615764 13615766 13615767 13615770 13615775 13615776 13615781 13615782 13615783 13615786 13615788 13615789 13615790 13615791 13615795 13615796 13615797 13615799 13615800 13615801 13615803 13615805 13615806 13615807 13615808 13615810 13615811 13615812 13615815 13615818 13615832 13615833 13615834 13615835 13615848 13615853 13615856 13615866 13615878 13615879 13615880 13615881 13615891 13615893 13615901 13615906 13615907 13615916 13615925 13615969 13615970 13615977 13615981 13615988 13615989 13615993 13615994 13616007 13616009 13616014 13616022 13616024 13616026 13616027 13616030 13616034 13616037 13616041 13616047 13616048 13616125 13616132 13616133 13616134 13616136 13616137 13616140 13616142 13616143 13616144 13616146 13616148 13616153 13616157 13616160 13616162 13616163 13616170 13616171 13616172 13616176 13616178 13616180 13616181 13616183 13616186 13616187 13616190 13616196 13616201 13616203 13616205 13616214 13616220 13616225 13616228 13616229 13616231 13616238 13616239 13616240 13616242 13616247 13616265 13616271 13616273 13616274 13616275 13616276 13616277 13616281 13616282 13616289 13616298 13616300 13616301 13616304 13616305 13616306 13616308 13616310 13616311 13616313 13616316 13616317 13616319 13616320 13616321 13616322 13616323 13616324 13616325 13616326 13616328 13616331 13616332 13616333 13616334 13616335 13616336 13616337 13616339 13616340 13616341 13616342 13616343 13616344 13616345 13616346 13616348 13616350 13616351 13616352 13616353 13616355 13616356 13616357 13616358 13616359 13616360 13616361 13616362 13616363 13616364 13616365 13616367 13616368 13616370 13616371 13616375 13616377 13616379 13616381 13616384 13616387 13616388 13616389 13616390 13616391 13616393 13616394 13616395 13616397 13616399 13616400 13616401 13616404 13616407 13616409 13616410 13616411 13616413 13616414 13616416 13616417 13616419 13616420 13616421 13616423 13616424 13616427 13616428 13616430 13616431 13616434 13616435 13616436 13616437 13616438 13616439 13616440 13616441 13616442 13616443 13616444 13616445 13616446 13616447 13616450 13616451 13616452 13616453 13616454 13616455 13616458 13616463 13616464 13616466 13616468 13616469 13616471 13616472 13616473 13616478 13616480 13616481 13616483 13616484 13616485 13616486 13616487 13616488 13616489 13616490 13616491 13616494 13616495 13616496 13616500 13616502 13616503 13616505 13616506 13616507 13616509 13616510 13616512 13616513 13616514 13616515 13616590 13616592 13616593 13616598 13616600 13616602 13616603 13616605 13616606 13616607 13616609 13616610 13616612 13616613 13616620 13616622 13616623 13616625 13616630 13616631 13616633 13616634 13616635 13616636 13616638 13616641 13616643 13616645 13616650 13616653 13616654 13616655 13616658 13616670 13616671 13616708 13616710 13616713 13616716 13616719 13616728 13616735 13616740 13616750 13616755 13616756 13616768 13616776 13616779 13616793 13616801 13616839 13616840 13616841 13616846 13616848 13616849 13616858 13616862 13616863 13616866 13616872 13616881 13616889 13616891 13616899 13616909 13616920 13616927 13616929 13616934 13616943 13616946 13616954 13616958 13616968 13616973 13616974 13616975 13616980 13616989 13616995 13617005 13617008 13617038 13617039 13617050 13617059 13617068 13617080 13617098 13617121 13617152 13617158 13617203 13617224 13617230 13617235 13617253 13617254 13617267 13617284 13617297 13617303 13617307 13617308 13617315 13617336 13617340 13617344 13617346 13617350 13617352 13617353 13617357 13617368 13617370 13617372 13617373 13617374 13617377 13617378 13617380 13617391 13617394 13617399 13617403 13617404 13617405 13617406 13617407 13617408 13617409 13617413 13617415 13617425 13617428 13617430 13617435 13617436 13617438 13617439 13617440 13617441 13617448 13617454 13617456 13617457 13617459 13617461 13617468 13617469 13617473 13617476 13617483 13617492 13617493 13617495 13617496 13617497 13617502 13617506 13617507 13617508 13617509 13617510 13617511 13617518 13617519 13617526 13617529 13617530 13617533 13617534 13617535 13617538 13617539 13617540 13617546 13617556 13617565 13617566 13617581 13617586 13617587 13617588 13617589 13617590 13617591 13617593 13617594 13617596 13617599 13617601 13617604 13617607 13617613 13617614 13617621 13617622 13617623 13617624 13617626 13617627 13617634 13617635 13617636 13617637 13617638 13617639 13617641 13617642 13617643 13617644 13617646 13617654 13617668 13617674 13617686 13617687 13617689 13617690 13617691 13617694 13617695 13617696 13617698 13617699 13617709 13617710 13617714 13617718 13617724 13617725 13617740 13617742 13617745 13617747 13617749 13617750 13617762 13617763 13617764 13617767 13617768 13617770 13617777 13617784 13617789 13617790 13617791 13617792 13617793 13617794 13617795 13617796 13617799 13617800 13617801 13617804 13617807 13617810 13617811 13617812 13617813 13617814 13617818 13617819 13617820 13617821 13617822 13617823 13617824 13617826 13617829 13617830 13617831 13617833 13617834 13617836 13617838 13617840 13617841 13617842 13617844 13617845 13617847 13617848 13617849 13617851 13617853 13617854 13617856 13617857 13617858 13617859 13617860 13617861 13617863 13617865 13617867 13617868 13617869 13617871 13617872 13617874 13617875 13617881 13617882 13617883 13617884 13617885 13617890 13617894 13617895 13617897 13617898 13617904 13617912 13617913 13617914 13617916 13617923 13617925 13617927 13617930 13617931 13617934 13617935 13617937 13617942 13617946 13617953 13617963 13617973 13617979 13618051 13618052 13618055 13618067 13618069 13618071 13618072 13618074 13618075 13618077 13618083 13618084 13618086 13618099 13618105 13618111 13618116 13618344 13618351 13618353 13618354 13618367 13618384 13618389 13618390 13618392 13618393 13618396 13618399 13618401 13618426 13618427 13618441 13618442 13618446 13618447 13618449 13618450 13618451 13618459 13618474 13618496 13618504 13618507 13618511 13618513 13618514 13618516 13618518 13618520 13618522 13618523 13618524 13618530 13618532 13618533 13618534 13618538 13618549 13618551 13618561 13618562 13618570 13618575 13618578 13618579 13618583 13618586 13618588 13618590 13618591 13618604 13618612 13618619 13618622 13618630 13618636 13618637 13618644 13618682 13618697 13618699 13618737 13618747 13618750 13618751 13618752 13618754 13618755 13618756 13618757 13618758 13618761 13618763 13618782 13618785 13618787 13618788 13618790 13618793 13618806 13618809 13618810 13618811 13618812 13618821 13618824 13618825 13618826 13618827 13618830 13618838 13618840 13618853 13618854 13618856 13618857 13618858 13618872 13618874 13618878 13618880 13618885 13618886 13618910 13618917 13618921 13618932 13618944 13618950 13619032 13619774 13619871 13619945 13619959 13619963 13620091 13620101 13620131 13620703 13621052 13621116 13621451 13621676 13623848 13626838 13627267 13627477 13627515 13627540 13627553 13627563 13627565 13627573 13627574 13627575 13627579 13627580 13627581 13388491 13612933 13556320 13560644 13585491 13585619 13588572 13589270 13589294 13589467 13600017 13600075 13600103 13600110 13600116 13600119 13600133 13600135 13600161 13600218 13600239 13600365 13600521 13600540 13600548 13600829 13600887 13601006 13601043 13601049 13601058 13601085 13601086 13601087 13601117 13601120 13601121 13601137 13601140 13601175 12943397 12943427 12943590 12943594 12943684 12943724 3687431 3687533 3689010 3689852 3691237 13588966 13595492 13599714 13599911 13603990 13604073 13604120 13604125 13604132 13604176 13604177 13604371 13604465 13604475 13604477 13604478 13604491 13604501 13604539 13604550 13604552 13604615 13604629 13604631 13604633 13604634 13604635 13604636 13604638 13604639 13604646 13604658 13604659 13604669 13604682 13604683 13604696 13604699 13604706 13604792 13604824 13604858 13604871 13604897 13604898 13604903 13604919 13604922 13604959 13604962 13604975 13605010 13605030 13605603 13411475 13411465 13411659 13186123 13558319 12920284 12643797 12643800 13605042 13605415 13605517 13605546 13605575 13605919 13616652 13617371 13617732 13622175 9905416 9905925 9906038 9906070 9912145 9914030 9914180 9914202 9914703 9914948 9914960 9915067 9915076 9915589 9915818 9915937 9915958 9916063 9916098 9916179 9916529 9916547 9916722 9916737 9916970 9917140 9917414 9917451 9917485 9917631 12341830 12344336 13558214 13559397 13562422 13562473 13565038 13565416 13565865 13566039 13566143 13568862 13568879 13568882 13568883 13568885 13568907 13568921 13568937 13568938 13568940 13568941 13568943 13568944 13568948 13568949 13568951 13568953 13568955 13568977 13568993 13568995 13568996 13568998 13568999 13569013 13569022 13569091 13569108 13569113 13569118 13569134 13569215 13569221 13569233 13569234 13569241 13569247 13569249 13569256 13569257 13569258 13569277 13569296 13569301 13569327 13569436 13569445 13569446 13569459 13569475 13569632 13569644 13569646 13569650 13569659 13569662 13569663 13569664 13569666 13569669 13569682 13569683 13569692 13569693 13569699 13569706 13569714 13569721 13569727 13569732 13569740 13569779 13569782 13569783 13569784 13569797 13569804 13569807 13569814 13569817 13569831 13569833 13569849 13569870 13569871 13569877 13569878 13569879 13569895 13569897 13569901 13569934 13569939 13569940 13569942 13569943 13569944 13569946 13569949 13569950 13569954 13569992 13570011 13570012 13570016 13570034 13570040 13570042 13570046 13570047 13570048 13570053 13570054 13570064 13570065 13570066 13570067 13570073 13570125 13570130 13570134 13570149 13570189 13570190 13570192 13570199 13570208 13570231 13570288 13570290 13570326 13570327 13570329 13570330 13570343 13570345 13570346 13570347 13570348 13570349 13570350 13570373 13570378 13570383 13570384 13570386 13570389 13570396 13570397 13570398 13570399 13570400 13570401 13570404 13570405 13570409 13570410 13570415 13570419 13570422 13570427 13570436 13570442 13570444 13570447 13570451 13570453 13570456 13570463 13570465 13570472 13570479 13570483 13570490 13570491 13570492 13570495 13570496 13570507 13570508 13570511 13570512 13570514 13570515 13570516 13570519 13570520 13570521 13570522 13570525 13570526 13570529 13570530 13570531 13570533 13570534 13570535 13570536 13570537 13570538 13570539 13570540 13570542 13570543 13570544 13570546 13570548 13570561 13570562 13570563 13570564 13570567 13570568 13570569 13570570 13570571 13570572 13570573 13570574 13570575 13570577 13570581 13570583 13570590 13570595 13570603 13570604 13570606 13570801 13570811 13570823 13570847 13570859 13570866 13570947 13570960 13570965 13570966 13570979 13571030 13571032 13551107 12389182 12389259 12389723 13585503 12583308 9896323 12344631 12951526 12792273 12816712 12827932 2000687 12594225 9870369 9870862 9871285 9871311 4055327 4055340 4055686 4055696 4055728 4055758 4055864 4055938 4055997 4060066 4060332 4060373 4063251 4063254 4063264 4063301 4063327 4063328 4063368 4063372 4063528 4063580 4063585 4063589 4063603 4063829 4064628 4064712 4064797 4064802 4064818 4064861 4064875 4065047 4065075 4065178 4065215 4065318 4065361 4065374 4065561 4065564 4065566 4065585 4065655 4065690 4068327 4068409 4068456 4068463 4068473 4068773 4068885 4068906 4068909 4068915 4069042 4069113 4069163 4069175 4069200 4069205 4069215 4069226 4069229 4069249 4069296 4069310 4069312 4069339 4069359 4069485 4069515 4069619 4069648 4069649 4069654 4069658 4069697 4069712 4069760 4069763 4069766 4069771 4069774 4069780 4069782 4069793 4069798 4069821 4069832 4069970 4070038 4090905 4090909 12390952 12392611 12392763 12426401 12430945 12432027 12432057 12432310 12432375 12432443 12432536 12433282 12433635 9360734 13585621 13586208 13586229 13586235 13586243 13586298 13586315 13586325 13586328 13586342 13586357 13586396 13586409 13586416 13586600 13586729 13586737 13586756 13586760 13586762 13586766 13586770 13586774 13586776 13586781 13586794 13586796 13586813 13586815 13586816 13586817 13586856 13586874 13586876 13586879 13586953 13586974 13587001 13587095 13587096 13588738 13589017 13589295 13589445 13589479 3701183 4003048 12462783 13568983 13569122 13569136 13569148 13543131 13543445 13543487 13543855 13543861 13543862 13543908 13543922 13543927 13543931 13543934 13543935 13543961 13543982 13543983 13543984 13543998 13543999 13544008 13544009 13544010 13544012 13544013 13544014 13544016 13544017 13544044 13544055 13544070 13544072 13544074 13544083 13544093 13544097 13544099 13544100 13544101 13544105 13544108 13544111 13544114 13544121 13544122 13544123 13544131 13544132 13544146 13544152 13544155 13544156 13544158 13544164 13544180 13544207 13544214 13544270 13544273 13544277 13544278 13544281 13544284 13544300 13544302 13544333 13544401 13544434 13544447 13544456 13545020 13545021 13545039 13545042 13545044 13545049 13545052 13545054 13545057 13545063 13545064 13545065 13545066 13545069 13545074 13545075 13545076 13545081 13545083 13545084 13545086 13545087 13545088 13545089 13545095 13545099 13545100 13545102 13545103 13545107 13545109 13545117 13545119 13545124 13545127 13545130 13545134 13545135 13545136 13545139 13545140 13545141 13545142 13545154 13545159 13545160 13545161 13545162 13545163 13545164 13545166 13545167 13545168 13545169 13545171 13545172 13545173 13545175 13545176 13545177 13545179 13545180 13545181 13545182 13545183 13545185 13545188 13545189 13545190 13545193 13545194 13545195 13545196 13545200 13545201 13545203 13545205 13545208 13545209 13545210 13545212 13545213 13545214 13545216 13545218 13545220 13545221 13545223 13545226 13545228 13545229 13545238 13545240 13545241 13545249 13545251 13545254 13545259 13545261 13545263 13545266 13545267 13545463 13545464 13545471 13545472 13545473 13545474 13545475 13545489 13545662 13545663 13545715 13545717 13545748 13545810 13545884 13545886 13545889 13545919 13545940 13545945 13545946 13545953 13545960 13545970 13545991 13546030 13546050 13546053 13546060 13546063 13546064 13546067 13546068 13546071 13546073 13546079 13546080 13546092 13546094 13546095 13546096 13546111 13546112 13546113 13546114 13546121 13546122 13546124 13546127 13546128 13546131 13546132 13546137 13546139 13546141 13546155 13546159 13546160 13546166 13546169 13546176 13546184 13546190 13546191 13546193 13546199 13546205 13546206 13546209 13546214 13546218 13546219 13546223 13546225 13546228 13546229 13546230 13546231 13546233 13546263 13546280 13546285 13546287 13546295 13546296 13546300 13546302 13546313 13546316 13546317 13546322 13546323 13546329 13546330 13546334 13546336 13546340 13546342 13546347 13546349 13546351 13546354 13546358 13546375 13546382 13546385 13546388 13546390 13546393 13546402 13546407 13546410 13546411 13546413 13546419 13546420 13546421 13546423 13546424 13546425 13546427 13546430 13546434 13546435 13546437 13546438 13546440 13546446 13546447 13546449 13546450 13546458 13546461 13546465 13546468 13546469 13546477 13546502 13546505 13546506 13546510 13546513 13546519 13546520 13546528 13546530 13546532 13546542 13546547 13546548 13546561 13546659 13546661 13546827 13546857 13546869 13546870 13546873 13546881 13546882 13546883 13546898 13546899 13546900 13546920 13546922 13546935 13547013 13547020 13547037 13547050 13547051 13547064 13547082 13547083 13547085 13547088 13547148 13547175 13547197 13547198 13547206 13547211 13547217 13547245 13547251 13547287 12830454 12858882 4124663 9870359 12689982 12443023 13552780 13552783 13553072 13553240 13553263 13553283 13553302 13553324 13553414 13554003 13554004 13554005 13554020 13554023 13554025 13554026 13554029 13554030 13554031 13554033 13554036 13554051 13554052 13554058 13554062 13554071 13554075 13554087 13554088 13554097 13554098 13554108 13554114 13554118 13554126 13554129 13554130 13554132 13554133 13554137 13554138 13554141 13554143 13554146 13554148 13554151 13554154 13554180 13554195 12562674 13556050 13556222 13556232 13556247 13556251 13556276 13556310 13556369 13556467 13556479 13556484 13556501 13556527 13556535 13556586 13556610 13556611 13556656 13556661 13556666 13556670 13556675 13556682 13556686 13556692 13616640 13617410 13617746 12563814 13551897 13589063 13589083 13589123 13589138 13589219 12356973 13608884 13609053 13609190 13609556 13609673 13616712 13616715 13616949 13616983 12691937 13581261 13582463 12467280 13512991 13567722 13620349 13626407 9974917 9938571 13548354 13548435 13550777 13554079 13555191 13555252 13555672 13555770 13555848 13555911 13556007 13556015 13556022 13556094 13557686 13557729 13557869 13557924 13558494 13558798 13558834 13559206 13559378 13559380 13559512 13559534 13559571 13560423 13560734 13561283 13561300 13561850 13561856 13561861 13561918 13561997 13562245 13562343 13562373 13562441 13562574 13562872 13562977 13563030 13563125 13563188 13565005 13565314 13565376 13565415 13565420 13565457 13565474 13565492 13565502 13565690 13565698 13566071 13566652 13566653 13566682 13566737 13566806 13566809 13567178 13567302 13567382 13567503 13567572 13567581 13567705 13569222 13569237 13569338 13569630 13569805 13569813 13569952 13569956 13570068 13570235 13570252 13570417 13570454 13570517 13570523 13570552 13570599 13570634 13570636 13570660 13570665 13570710 13570766 13570849 13570892 13570894 13570976 13571347 13572032 13572285 13572647 13572704 13572722 13573032 13573335 13573341 13573619 13573627 13573692 13574136 13574143 13574221 13574339 13574851 13575330 13575381 13575391 13575705 13575780 13575924 13576266 13576550 13576933 13577083 13577197 13577304 13577305 13577352 13577387 13577394 13577433 13579994 13580052 13584049 9864812 9865832 9866037 9866112 9866301 9866331 9866396 9866545 9866882 9867044 9867172 9867485 9867663 9867865 2000674 13110969 13374022 13374456 13375213 13375219 13375282 13376221 12463176 12463789 13565149 13568880 13568899 13569158 13569305 13576964 13581486 13585305 13585609 13585646 13585666 13585689 13585708 13585725 13585744 13585767 13585889 13585946 13620965 13620967 13620973 12580577 13015820 13185969 13186052 4120867 4121779 4122901 13547289 13574079 13192802 13194564 12777319 13537247 13537248 13538827 13541386 13541825 13543196 13544356 13545227 13545236 13545237 13545239 13545242 13545243 13545245 13545246 13545247 13545248 13545255 13545256 13545258 13545264 13545265 13545456 13545459 13545460 13545465 13545478 13545479 13545486 13545487 13545488 13545491 13545500 13545501 13545503 13545504 13545505 13545506 13545520 13545521 13545523 13545528 13545529 13545530 13545531 13545532 13545533 13545540 1354554

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0026-2018

Related Recalls

If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated.