An issue with the DL900 and DL950 Series Holter Recorder was recognized on January 1, 2020. This issue can prevent the recorder from operating normally. Braemar confirmed that beginning on January 1, 2020 if a AAA battery is inserted in the recorder and a user attempts to start it, the recorder will display Error: 602 and fail to function for a new patient study. There are no actions that a clinical user can take to clear this error. Even if this error code is displayed, existing ECG recordings stored on the device can still be downloaded via the standard Holter Application Software. Error 602 may be able to be cleared by technical or engineering staff at the customer site.

Jan 22, 2020 Diagnostic Equipment Nationwide View Details β†’

Non-sterile product was shipped to customers. Product package is labeled as both "STERILE" and "NON-STERILR" ; the product did not undergo sterilization process and is non-sterile. The usage of a needle that is not sterile could result in infection. Firm is not aware of any reports of patient harm.

Jan 15, 2020 Surgical Instruments Nationwide View Details β†’

Potential for an HVAD System user to mistakenly insert the Battery Charger AC Adapter into an HVAD Controller power port.

Jan 23, 2020 Other Medical Devices View Details β†’

A deterioration of performance was identified with influenza type B Enzyme Immunoassay Kits with findings of a positive bias with kit controls. Patient sample results may be affected with falsely elevated results incorrectly indicating a protective level of anti-Hib antibody in the assessment of immunodeficiency.

Aug 12, 2018 Diagnostic Equipment Nationwide View Details β†’

The form received three complaints related to the user obtaining a result that fell below the target range for CK-MB provided in the RAMP Cardiac Controls Instructions for Use. Via in-house monitoring, determined that liquid Cardiac Control has seen a decrease in CK-MB Levels of up to 57% signal loss within 11 months. Shelf-life of product is 24 months.

Mar 15, 2019 Other Medical Devices Nationwide View Details β†’

An internal component was assembled in the incorrect orientation, which may lead to the tulip head disengaging from the crew shank during or after screw insertion, or during the reduction step.

Jan 28, 2020 Other Medical Devices Nationwide View Details β†’

When using the Rotor-Gene Q with Software version 2.3.4 completing a LIMs export, the .csv file reports the calculated concentration result value as a logarithmic value, and could lead to a false negative result, which could lead to serious medical consequences such as the suspension or non-initiation of treatment

Jan 28, 2020 Other Medical Devices Nationwide View Details β†’

An issue with the DL900 and DL950 Series Holter Recorder was recognized on January 1, 2020. This issue can prevent the recorder from operating normally. Braemar confirmed that beginning on January 1, 2020 if a AAA battery is inserted in the recorder and a user attempts to start it, the recorder will display Error: 602 and fail to function for a new patient study. There are no actions that a clinical user can take to clear this error. Even if this error code is displayed, existing ECG recordings stored on the device can still be downloaded via the standard Holter Application Software. Error 602 may be able to be cleared by technical or engineering staff at the customer site.

Jan 22, 2020 Diagnostic Equipment Nationwide View Details β†’

Non-sterile product was shipped to customers. Product package is labeled as both "STERILE" and "NON-STERILR" ; the product did not undergo sterilization process and is non-sterile. The usage of a needle that is not sterile could result in infection. Firm is not aware of any reports of patient harm.

Jan 15, 2020 Other Medical Devices Nationwide View Details β†’

An internal component was assembled in the incorrect orientation, which may lead to the tulip head disengaging from the crew shank during or after screw insertion, or during the reduction step.

Jan 28, 2020 Other Medical Devices Nationwide View Details β†’

1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion alarms may result in unrecognized infusion completion 5)Custom concentration data entry errors results in concentrations being lower/higher than medication orders

Feb 4, 2020 Other Medical Devices Nationwide View Details β†’

A patch was released for a critical remote code execution vulnerability in remote desktop services. This vulnerability can be exploited remotely without authentication on certain operating systems. An attacker who successfully exploited this vulnerability could execute arbitrary code on the target system; then install programs; view, change, or delete data; or create new accounts with full user rights. If the vulnerability could be exploited locally, it would lock down the computer.

Aug 15, 2019 Other Medical Devices Nationwide View Details β†’

It has been determined that some units of catheters were not tightly connected to the thumb port. This may result in the catheter disconnecting from the thumb port which could reduce the suction performance and make it difficult to remove from the patient's trachea.

Aug 19, 2019 Diagnostic Equipment Nationwide View Details β†’