STERIS has learned from Customer feedback and field service experience that damage from external impact to the table hydraulic column can cause the stainless steel shroud sections covering the table hydraulic column to become misaligned; if the damage is severe, the up/down movement of the table may be impeded. In addition, the table hand control may indicate that the table floor locks are unlock
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
This letter is to notify you that several of our sterile screw products are being voluntarily recalled. Product sterility may be compromised due to a mechanical failure of the protective packaging tube. Tube end plugs may dislodge creating the potential for the screw to come out of the protective tube and then the sharp end of the screw pierce the outer sterile barrier pouch.
BR980-9600 Hysteroscopy Sterile Procedure Kit containing 2 oz. bottle PVP Prep Solution Product Usage: Usage: The PVP contained in the kit is used in skin prep
Custom Medical Specialties
The firm saw on the FDA web site a recall by the manufacturer of the Povidone Iodine Prep Solution due to no microbial testing and product did not meet proper finished goods specifications.
Shipped with incorrect software
Incorrect version of the CD was included with BD FACS 7 Color Setup Beads.
CMS-4501-R1 Custom HSG Tray containing 2 oz. PVP Prep Bottle Product Usage: Usage: The PVP contained in the kit is used in skin prep
Custom Medical Specialties
The firm saw on the FDA web site a recall by the manufacturer of the Povidone Iodine Prep Solution due to no microbial testing in place at the manufacturer and product did not meet proper finished goods specifications.
There is a potential failure of the Integrated Module Housing causing the patient module parameters to drop off the monitor display or the patient module will fail to sign on when inserted into the housing.
Model 324JL On-Board Suction System Intended use: On-board suction system for Ambulance manufacturers.
Impact Instrumentation
A shipment mix-up took place.
Asahi KASEI, REXEED(TM)- 18 SX, HEMODIALYZER dialysis filter
Asahi Kasei Medical Co., Ltd., Okatomi Plant
Asahi became aware of a negative trend for rate of the adverse events that may be associated with a manufacturing change to certain lot numbers of the products, though there is no sufficient information to date to conclusively link these adverse events to the products.
SPECTROLYSE PAI-1 activity assay producing lower than expected absorbance values with "0" standard. This will reduce the slope of the standard curve requiring the user to make more frequent sample dilutions.
ADVIA Centaur Prolactin Assay and FSH, Calibrator B fails the analytical sensitivity stated in the IFU for lots below 40
CMS-8162-R1 Fox Chase Specials Pack containing 2 oz. bottle PVP Prep Solution Product Usage: Usage: The PVP contained in the kit is used in skin prep
Custom Medical Specialties
The firm saw on the FDA web site a recall by the manufacturer of the Povidone Iodine Prep Solution due to no microbial testing and product did not meet proper finished goods specifications.
The placement guidewire diameter is too large and does not pass-through the introducer needle during the placement procedure of the PEG Feeding Tube.
Asahi KASEI, REXEED(TM)- 25 SX, HEMODIALYZER dialysis filter
Asahi Kasei Medical Co., Ltd., Okatomi Plant
Asahi became aware of a negative trend for rate of the adverse events that may be associated with a manufacturing change to certain lot numbers of the products, though there is no sufficient information to date to conclusively link these adverse events to the products.
Vascular Solutions, Inc., became aware of a potential problem involving pouches missing a seal from Guardian II (Model 8210) and Guardian II NC (Model FH102) Hemostasis Valve which poses a risk of product contamination.
Investigation found that the current design of the product allows the press-fit of the knob/shaft interface to be overcome by the impaction force of a mallet, causing the knob to potentially unfasten and slide down the shaft.
There have been a small number of instances in hospitals where the joint between the drive gear and the gantry base drive wheel has become loose, leading to fatigue failure of some of the securing bolts.
CMS-4975 HSG Tray containing 4 oz. bottle PVP Prep Solution Product Usage: Usage: The PVP contained in the kit is used in skin prep
Custom Medical Specialties
The firm saw on the FDA web site a recall by the manufacturer of the Povidone Iodine Prep Solution due to no microbial testing and product did not meet proper finished goods specifications.
STERIS has learned that the Verify Bowie Dick Test Cards are not performing to product specifications. Specifically, test cards have resulted in false fail results when sterilizer performance is within acceptable ranges.
CMS-8325-R2 Abington Radiology Drainage Pack and CMS-8325-R3 Abington Radiology Drainage Pack containing 4 oz. bottle PVP Prep Solution Product Usage: Usage: The PVP contained in the kit is used in skin prep
Custom Medical Specialties
The firm saw on the FDA web site a recall by the manufacturer of the Povidone Iodine Prep Solution due to no microbial testing and product did not meet proper finished goods specifications.