Ultraview SL 91493 Integrated Module Housing. PCIS patient monitors is intended to sound various alarm tones that are generated by the specific PCIS patient parameter modules. Each patient parameter module determines (a) when an alarm condition is violated; (b) the alarm priority (i.e. high, medium or low); (c) alarm limits; and (d) when to generate and terminate alarms. PCIS patient monitors are also capable of displaying alarm conditions on other monitors that are on the network through the Alarm Watch feature. Spacelabs Medical PCIS patient monitors are designed to communicate with a variety of external devices such as display devices, network devices, serial devices, user input devices, audio system, local/remote recorders and Spacelabs Medical patient parameter modules. In all applications, Spacelabs Medical PCIS patient monitors provide access to that data on the Spacelabs Medical network either via hardwire or wireless transmission. Through this network, a patient's data may be displayed at the bedside monitor, central station or clinical information system.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Domestic - 1493-000275, 1493-000276, 1493-000277, 1493-002469, 1493-002616, 1493-003358, 1493-003359, 1493-003667, 1493-003668, 1493-003669, 1493-003670, 1493-003671, 1493-003672, 1493-003673, 1493-003674, 1493-003675, 1493-003676, 1493-003677, 1493-003678, 1493-003679, 1493-003680, 1493-003681, 1493-003682, 1493-003683, 1493-003684, 1493-003685, 1493-003686, 1493-003687, 1493-003688, 1493-003693, 1493-003694, 1493-003695, 1493-003696, 1493-003697, 1493-003879, 1493-003880, 1493-003881, 1493-003882, 1493-003883, 1493-003884, 1493-003885, 1493-003886, 1493-003887, 1493-003888, 1493-003889, 1493-003890, 1493-003891, 1493-003892, 1493-003893, 1493-003894, 1493-003895, 1493-003896, 1493-003897, 1493-003898, 1493-003899, 1493-003900, 1493-003901, 1493-003902, 1493-003903, 1493-003904, 1493-003905, 1493-003906, 1493-003907, 1493-003908, 1493-003909, 1493-003914, 1493-003915, 1493-003916, 1493-003917, 1493-003918, 1493-003919, 1493-003920, 1493-003921, 1493-003922, 1493-003923, 1493-003924, 1493-003925, 1493-003926, 1493-003927, 1493-003928, 1493-003929, 1493-003930, 1493-003931, 1493-003932, 1493-003933, 1493-003934, 1493-003935, 1493-003936, 1493-003937, 1493-003938, 1493-003939, 1493-003940, 1493-003941, 1493-003942, 1493-003943, 1493-003944, 1493-003945, 1493-003957, 1493-003958, 1493-003959, 1493-003960, 1493-003961, 1493-003962, 1493-003963, 1493-003964, 1493-003965, 1493-003985, 1493-004007, 1493-004011, 1493-004023, 1493-004026, 1493-004027, 1493-004028, 1493-004029, 1493-004030, 1493-004031, 1493-004032, 1493-004033, 1493-004034, 1493-004043, 1493-004044, 1493-004047, 1493-004048, 1493-004050, 1493-004051, 1493-004052, 1493-004053, 1493-004054, 1493-004055, 1493-004056, 1493-004058, 1493-004060, 1493-004064, 1493-004067, 1493-004068, 1493-004069, 1493-004070, 1493-004071, 1493-004075, 1493-004084, 1493-004085, 1493-004086, 1493-004106, 1493-004112, 1493-004113, 1493-004114, 1493-004115, 1493-004116, 1493-004117, 1493-004118, 1493-004119, 1493-004120, 1493-004121, 1493-004122, 1493-004123, 1493-004124, 1493-004125, 1493-004126, 1493-004127, 1493-004128, 1493-004130, 1493-004158, 1493-004159, 1493-004160, 1493-004161, 1493-004162, 1493-004163, 1493-004164, 1493-004172, 1493-004173, 1493-004174, 1493-004175, 1493-004176, 1493-004177, 1493-004178, 1493-004179, 1493-004180, 1493-004181, 1493-004182, 1493-004183, 1493-004184, 1493-004185, 1493-004205, 1493-004206, 1493-004207, 1493-004208, 1493-004209, 1493-004210, 1493-004211, 1493-004212, 1493-004214, 1493-004215, 1493-004216, 1493-004217, 1493-004218, 1493-004219, 1493-004220, 1493-004221, 1493-004222, 1493-004223, 1493-004224, 1493-004225, 1493-004226, 1493-004227, 1493-004228, 1493-004229, 1493-004230, 1493-004231, 1493-004232, 1493-004233, 1493-004234, 1493-004235, 1493-004236, 1493-004237, 1493-004238, 1493-004239, 1493-004240, 1493-004241, 1493-004242, 1493-004243, 1493-004256, 1493-004257, 1493-004258, 1493-004259, 1493-004260, 1493-004261, 1493-004262, 1493-004263, 1493-004264, 1493-004265, 1493-004266, 1493-004267, 1493-004268, 1493-004269, 1493-004270, 1493-004271, 1493-004272, 1493-004273, 1493-004274, 1493-004275, 1493-004276, 1493-004277, 1493-004278, 1493-004279, 1493-004280, 1493-004281, 1493-004282, 1493-004283, 1493-004284, 1493-004285, 1493-004286, 1493-004287, 1493-004288, 1493-004289, 1493-004290, 1493-004291, 1493-004292, 1493-004293, 1493-004294, 1493-004295, 1493-004296, 1493-004297, 1493-004298, 1493-004299, 1493-004301, 1493-004302, 1493-004306, 1493-004307, 1493-004308, 1493-004309, 1493-004310, 1493-004311, 1493-004312, 1493-004313, 1493-004314, 1493-004315, 1493-004316, 1493-004317, 1493-004318, 1493-004319, 1493-004320, 1493-004321, 1493-004322, 1493-004323, 1493-004324, 1493-004325, 1493-004326, 1493-004327, 1493-004328, 1493-004329, 1493-004330, 1493-004331, 1493-004332, 1493-004333, 1493-004334, 1493-004335, 1493-004336, 1493-004337, 1493-004338, 1493-004339, 1493-004340, 1493-004341, 1493-004342, 1493-004343, 1493-004344, 1493-004345, 1493-004346, 1493-004347, 1493-004348, 1493-004349, 1493-004350, 1493-004351, 1493-004352, 1493-004353, 1493-004354, 1493-004355, 1493-004356, 1493-004357, 1493-004358, 1493-004359, 1493-004360, 1493-004361, 1493-004362, 1493-004363, 1493-004364, 1493-004365, 1493-004366, 1493-004367, 1493-004368, 1493-004369, 1493-004370, 1493-004371, 1493-004372, 1493-004373, 1493-004374, 1493-004375, 1493-004376, 1493-004377, 1493-004378, 1493-004379, 1493-004380, 1493-004381, 1493-004382, 1493-004383, 1493-004384, 1493-004385, 1493-004405, 1493-004406, 1493-004407, and 1493-004408. International - 1493-002941, 1493-003666, 1493-003689, 1493-003690, 1493-003691, 1493-003692, 1493-003910, 1493-003911, 1493-003912, 1493-003913, 1493-003946, 1493-003947, 1493-003948, 1493-003949, 1493-003950, 1493-003951, 1493-003952, 1493-003953, 1493-003954, 1493-003955, 1493-003956, 1493-003966, 1493-003967, 1493-003968, 1493-003969, 1493-003970, 1493-003971, 1493-003972, 1493-003973, 1493-003974, 1493-003975, 1493-003976, 1493-003977, 1493-003978, 1493-003979, 1493-003980, 1493-003981, 1493-003982, 1493-003983, 1493-003984, 1493-003986, 1493-003987, 1493-003988, 1493-003989, 1493-003990, 1493-003991, 1493-003992, 1493-003993, 1493-003994, 1493-003995, 1493-003996, 1493-003997, 1493-003998, 1493-003999, 1493-004000, 1493-004001, 1493-004002, 1493-004003, 1493-004004, 1493-004005, 1493-004008, 1493-004009, 1493-004010, 1493-004012, 1493-004013, 1493-004014, 1493-004015, 1493-004016, 1493-004017, 1493-004018, 1493-004019, 1493-004020, 1493-004021, 1493-004022, 1493-004024, 1493-004025, 1493-004035, 1493-004036, 1493-004037, 1493-004038, 1493-004039, 1493-004040, 1493-004041, 1493-004042, 1493-004045, 1493-004046, 1493-004065, 1493-004066, 1493-004072, 1493-004073, 1493-004074, 1493-004076, 1493-004078, 1493-004079, 1493-004080, 1493-004081, 1493-004083, 1493-004087, 1493-004088, 1493-004089, 1493-004090, 1493-004091, 1493-004092, 1493-004093, 1493-004094, 1493-004095, 1493-004096, 1493-004097, 1493-004098, 1493-004099, 1493-004100, 1493-004101, 1493-004102, 1493-004103, 1493-004104, 1493-004105, 1493-004107, 1493-004108, 1493-004109, 1493-004110, 1493-004111, 1493-004129, 1493-004131, 1493-004132, 1493-004133, 1493-004134, 1493-004135, 1493-004136, 1493-004137, 1493-004138, 1493-004139, 1493-004140, 1493-004141, 1493-004142, 1493-004143, 1493-004144, 1493-004145, 1493-004146, 1493-004147, 1493-004148, 1493-004149, 1493-004150, 1493-004151, 1493-004152, 1493-004153, 1493-004154, 1493-004155, 1493-004165, 1493-004166, 1493-004167, 1493-004168, 1493-004169, 1493-004170, 1493-004171, 1493-004186, 1493-004187, 1493-004188, 1493-004189, 1493-004190, 1493-004191, 1493-004192, 1493-004193, 1493-004194, 1493-004195, 1493-004196, 1493-004197, 1493-004198, 1493-004199, 1493-004200, 1493-004201, 1493-004202, 1493-004203, 1493-004204, 1493-004300, 1493-004304, 1493-004305, and 670-1350-00.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Spacelabs Healthcare, Llc
Reason for Recall:
There is a potential failure of the Integrated Module Housing causing the patient module parameters to drop off the monitor display or the patient module will fail to sign on when inserted into the housing.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Ultraview SL 91493 Integrated Module Housing. PCIS patient monitors is intended to sound various alarm tones that are generated by the specific PCIS patient parameter modules. Each patient parameter module determines (a) when an alarm condition is violated; (b) the alarm priority (i.e. high, medium or low); (c) alarm limits; and (d) when to generate and terminate alarms. PCIS patient monitors are also capable of displaying alarm conditions on other monitors that are on the network through the Alarm Watch feature. Spacelabs Medical PCIS patient monitors are designed to communicate with a variety of external devices such as display devices, network devices, serial devices, user input devices, audio system, local/remote recorders and Spacelabs Medical patient parameter modules. In all applications, Spacelabs Medical PCIS patient monitors provide access to that data on the Spacelabs Medical network either via hardwire or wireless transmission. Through this network, a patient's data may be displayed at the bedside monitor, central station or clinical information system.

Product Codes/Lot Numbers:

Domestic - 1493-000275, 1493-000276, 1493-000277, 1493-002469, 1493-002616, 1493-003358, 1493-003359, 1493-003667, 1493-003668, 1493-003669, 1493-003670, 1493-003671, 1493-003672, 1493-003673, 1493-003674, 1493-003675, 1493-003676, 1493-003677, 1493-003678, 1493-003679, 1493-003680, 1493-003681, 1493-003682, 1493-003683, 1493-003684, 1493-003685, 1493-003686, 1493-003687, 1493-003688, 1493-003693, 1493-003694, 1493-003695, 1493-003696, 1493-003697, 1493-003879, 1493-003880, 1493-003881, 1493-003882, 1493-003883, 1493-003884, 1493-003885, 1493-003886, 1493-003887, 1493-003888, 1493-003889, 1493-003890, 1493-003891, 1493-003892, 1493-003893, 1493-003894, 1493-003895, 1493-003896, 1493-003897, 1493-003898, 1493-003899, 1493-003900, 1493-003901, 1493-003902, 1493-003903, 1493-003904, 1493-003905, 1493-003906, 1493-003907, 1493-003908, 1493-003909, 1493-003914, 1493-003915, 1493-003916, 1493-003917, 1493-003918, 1493-003919, 1493-003920, 1493-003921, 1493-003922, 1493-003923, 1493-003924, 1493-003925, 1493-003926, 1493-003927, 1493-003928, 1493-003929, 1493-003930, 1493-003931, 1493-003932, 1493-003933, 1493-003934, 1493-003935, 1493-003936, 1493-003937, 1493-003938, 1493-003939, 1493-003940, 1493-003941, 1493-003942, 1493-003943, 1493-003944, 1493-003945, 1493-003957, 1493-003958, 1493-003959, 1493-003960, 1493-003961, 1493-003962, 1493-003963, 1493-003964, 1493-003965, 1493-003985, 1493-004007, 1493-004011, 1493-004023, 1493-004026, 1493-004027, 1493-004028, 1493-004029, 1493-004030, 1493-004031, 1493-004032, 1493-004033, 1493-004034, 1493-004043, 1493-004044, 1493-004047, 1493-004048, 1493-004050, 1493-004051, 1493-004052, 1493-004053, 1493-004054, 1493-004055, 1493-004056, 1493-004058, 1493-004060, 1493-004064, 1493-004067, 1493-004068, 1493-004069, 1493-004070, 1493-004071, 1493-004075, 1493-004084, 1493-004085, 1493-004086, 1493-004106, 1493-004112, 1493-004113, 1493-004114, 1493-004115, 1493-004116, 1493-004117, 1493-004118, 1493-004119, 1493-004120, 1493-004121, 1493-004122, 1493-004123, 1493-004124, 1493-004125, 1493-004126, 1493-004127, 1493-004128, 1493-004130, 1493-004158, 1493-004159, 1493-004160, 1493-004161, 1493-004162, 1493-004163, 1493-004164, 1493-004172, 1493-004173, 1493-004174, 1493-004175, 1493-004176, 1493-004177, 1493-004178, 1493-004179, 1493-004180, 1493-004181, 1493-004182, 1493-004183, 1493-004184, 1493-004185, 1493-004205, 1493-004206, 1493-004207, 1493-004208, 1493-004209, 1493-004210, 1493-004211, 1493-004212, 1493-004214, 1493-004215, 1493-004216, 1493-004217, 1493-004218, 1493-004219, 1493-004220, 1493-004221, 1493-004222, 1493-004223, 1493-004224, 1493-004225, 1493-004226, 1493-004227, 1493-004228, 1493-004229, 1493-004230, 1493-004231, 1493-004232, 1493-004233, 1493-004234, 1493-004235, 1493-004236, 1493-004237, 1493-004238, 1493-004239, 1493-004240, 1493-004241, 1493-004242, 1493-004243, 1493-004256, 1493-004257, 1493-004258, 1493-004259, 1493-004260, 1493-004261, 1493-004262, 1493-004263, 1493-004264, 1493-004265, 1493-004266, 1493-004267, 1493-004268, 1493-004269, 1493-004270, 1493-004271, 1493-004272, 1493-004273, 1493-004274, 1493-004275, 1493-004276, 1493-004277, 1493-004278, 1493-004279, 1493-004280, 1493-004281, 1493-004282, 1493-004283, 1493-004284, 1493-004285, 1493-004286, 1493-004287, 1493-004288, 1493-004289, 1493-004290, 1493-004291, 1493-004292, 1493-004293, 1493-004294, 1493-004295, 1493-004296, 1493-004297, 1493-004298, 1493-004299, 1493-004301, 1493-004302, 1493-004306, 1493-004307, 1493-004308, 1493-004309, 1493-004310, 1493-004311, 1493-004312, 1493-004313, 1493-004314, 1493-004315, 1493-004316, 1493-004317, 1493-004318, 1493-004319, 1493-004320, 1493-004321, 1493-004322, 1493-004323, 1493-004324, 1493-004325, 1493-004326, 1493-004327, 1493-004328, 1493-004329, 1493-004330, 1493-004331, 1493-004332, 1493-004333, 1493-004334, 1493-004335, 1493-004336, 1493-004337, 1493-004338, 1493-004339, 1493-004340, 1493-004341, 1493-004342, 1493-004343, 1493-004344, 1493-004345, 1493-004346, 1493-004347, 1493-004348, 1493-004349, 1493-004350, 1493-004351, 1493-004352, 1493-004353, 1493-004354, 1493-004355, 1493-004356, 1493-004357, 1493-004358, 1493-004359, 1493-004360, 1493-004361, 1493-004362, 1493-004363, 1493-004364, 1493-004365, 1493-004366, 1493-004367, 1493-004368, 1493-004369, 1493-004370, 1493-004371, 1493-004372, 1493-004373, 1493-004374, 1493-004375, 1493-004376, 1493-004377, 1493-004378, 1493-004379, 1493-004380, 1493-004381, 1493-004382, 1493-004383, 1493-004384, 1493-004385, 1493-004405, 1493-004406, 1493-004407, and 1493-004408. International - 1493-002941, 1493-003666, 1493-003689, 1493-003690, 1493-003691, 1493-003692, 1493-003910, 1493-003911, 1493-003912, 1493-003913, 1493-003946, 1493-003947, 1493-003948, 1493-003949, 1493-003950, 1493-003951, 1493-003952, 1493-003953, 1493-003954, 1493-003955, 1493-003956, 1493-003966, 1493-003967, 1493-003968, 1493-003969, 1493-003970, 1493-003971, 1493-003972, 1493-003973, 1493-003974, 1493-003975, 1493-003976, 1493-003977, 1493-003978, 1493-003979, 1493-003980, 1493-003981, 1493-003982, 1493-003983, 1493-003984, 1493-003986, 1493-003987, 1493-003988, 1493-003989, 1493-003990, 1493-003991, 1493-003992, 1493-003993, 1493-003994, 1493-003995, 1493-003996, 1493-003997, 1493-003998, 1493-003999, 1493-004000, 1493-004001, 1493-004002, 1493-004003, 1493-004004, 1493-004005, 1493-004008, 1493-004009, 1493-004010, 1493-004012, 1493-004013, 1493-004014, 1493-004015, 1493-004016, 1493-004017, 1493-004018, 1493-004019, 1493-004020, 1493-004021, 1493-004022, 1493-004024, 1493-004025, 1493-004035, 1493-004036, 1493-004037, 1493-004038, 1493-004039, 1493-004040, 1493-004041, 1493-004042, 1493-004045, 1493-004046, 1493-004065, 1493-004066, 1493-004072, 1493-004073, 1493-004074, 1493-004076, 1493-004078, 1493-004079, 1493-004080, 1493-004081, 1493-004083, 1493-004087, 1493-004088, 1493-004089, 1493-004090, 1493-004091, 1493-004092, 1493-004093, 1493-004094, 1493-004095, 1493-004096, 1493-004097, 1493-004098, 1493-004099, 1493-004100, 1493-004101, 1493-004102, 1493-004103, 1493-004104, 1493-004105, 1493-004107, 1493-004108, 1493-004109, 1493-004110, 1493-004111, 1493-004129, 1493-004131, 1493-004132, 1493-004133, 1493-004134, 1493-004135, 1493-004136, 1493-004137, 1493-004138, 1493-004139, 1493-004140, 1493-004141, 1493-004142, 1493-004143, 1493-004144, 1493-004145, 1493-004146, 1493-004147, 1493-004148, 1493-004149, 1493-004150, 1493-004151, 1493-004152, 1493-004153, 1493-004154, 1493-004155, 1493-004165, 1493-004166, 1493-004167, 1493-004168, 1493-004169, 1493-004170, 1493-004171, 1493-004186, 1493-004187, 1493-004188, 1493-004189, 1493-004190, 1493-004191, 1493-004192, 1493-004193, 1493-004194, 1493-004195, 1493-004196, 1493-004197, 1493-004198, 1493-004199, 1493-004200, 1493-004201, 1493-004202, 1493-004203, 1493-004204, 1493-004300, 1493-004304, 1493-004305, and 670-1350-00.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2382-2012

Related Recalls

Due to a manufacturing issue (i.e. malfunctioning of electrical Leakage tester).

Mar 26, 2026 Diagnostic Equipment Nationwide View Details →

Due to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific navigation conditions, test data intended for one patient may be saved under another patient's record

Sep 16, 2025 Other Medical Devices Nationwide View Details →

Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.

Sep 5, 2025 Other Medical Devices Nationwide View Details →