Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external drainage and monitoring system (EDMS).
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately.
Due to complaints received for the marker band detaching from the distal tip of the catheter and remaining in the patient as laser catheter was retracted.
MICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, # MXU-RV71
Micro-X
Mobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator.
Potential safety issues in Allura Xper R7.6- R8.1 systems where the Lifetime Extension (LTE) kit is not installed in a timely manner. Root cause is under investigation, but the recall reason is listed as for systems over 10 years of life with LTE kits are experiencing propeller motor clamping bolt and x-ray tube locking bolt breaks. If this happens, it may cause erratic C-arm movements and potential collision with patients or bystanders, or detection of a false collisions and subsequent blocked movement of the c-arm. Movement may also cause image noise and black area on the x-ray image, and possible need to expose the patient to additional unnecessary radiation during re-examination.
IFU was not provided with the Sizer Set sterilization tray Revised Sterilization Instructions for Sizer Set
Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
During internal testing it was found that if the system is Ready to expose and the hand switch is then dropped into the stowage unit such that it lands on the prep/expose switch, the prep switch can be activated for a very short time such that the orange ring of death is displayed in the tube cooling area of the screen. The system then becomes unresponsive to kV/mAs changes, prep/expose presses. The only way to proceed is to suspend the patient, Exit Nexus DR and relogin and continue the exam.
BD Pyxis CII Safe ES, REF: 1116-00
CareFusion 303
Due to a software issue, automated dispensing cabinet devices may open the wrong cubie pocket or position, which can lead to: 1) Unintended and incorrect medication removal, or 2) Potential for a different medication (medication pulled in error) to be refilled/loaded, which may lead to may lead to low inventory or a stockout.
Cardinal Health SMS Elastic Cuff Thumbhook Level 3 Isolation Gown, Full-Back, Product Codes AT6588-2XL, AT6588-BD, AT6588-XL, AT6688-2XL, AT6688-BD, and AT6688-XL
Cardinal Health 200
Isolation Gowns were manufactured with raw materials that may fail the hydrostatic pressure specification and therefore would not comply with AAMI (The Association for the Advancement of Medical Instrumentation) Level 3 protection requirements.
BD Pyxis MedStation ES 7 Drawer Auxiliary Tower, REF: 324
CareFusion 303
Due to a software issue, automated dispensing cabinet devices may open the wrong cubie pocket or position, which can lead to: 1) Unintended and incorrect medication removal, or 2) Potential for a different medication (medication pulled in error) to be refilled/loaded, which may lead to may lead to low inventory or a stockout.
Convertor Isolation Gown, AAMI Level 3 protection, Product Codes AT6100, AT6100-XL, AT6200, and AT6200-XL
Cardinal Health 200
Isolation Gowns were manufactured with raw materials that may fail the hydrostatic pressure specification and therefore would not comply with AAMI (The Association for the Advancement of Medical Instrumentation) Level 3 protection requirements.
Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive force may result in tissue injury/pneumothorax; instruments with long rigid distal sections may exhibit insertion resistance when bronchoscope bending section significantly articulated.
Due to finished goods incorrectly packaged and labeled.
BD Pyxis MedStation ES, REF: 323
CareFusion 303
Due to a software issue, automated dispensing cabinet devices may open the wrong cubie pocket or position, which can lead to: 1) Unintended and incorrect medication removal, or 2) Potential for a different medication (medication pulled in error) to be refilled/loaded, which may lead to may lead to low inventory or a stockout.
Customer complaints were received alleging leaking devices and insufficient buffer volume. Diasorin's internal investigation determined that 0.14% of the LIAISON Q.S.E.T. Device Plus in the lots identified have loose clear caps, which may allow buffer to leak from the tube. A stool sample prepared with less than the correct amount of buffer may result an incorrectly high patient result.
Hybrid Recon software on Hermes Medical Imaging Suite version 6.1 Hybrid Recon software versions affected are 4.0.0x and 5.0.0
Hermes Medical Solutions AB
Due a potential software/configuration issue that may result is incorrect alignment during reconstructing a SPECT/CT study.
Units were not intended for commercial distribution and inadvertently distributed.
Microscissor Tip (Disposable); Microline Catalog #: 3122.
Microline Surgical
A lot of ReNew Microscissor Tips exhibited a 2% rate of failure in terms of the heat shrink breaking off.
The reason for the recall is that EDTA and lithium heparin plasma samples that are stored in the DxI 9000 Access Immunoassay Analyzer sample wheel for a prolonged period before performing the intact PTH (parathyroid hormone) test may produce falsely decreased intact parathyroid hormone results. This could lead to a missed diagnosis, a delay in treatment, or inappropriate clinical management, depending upon the patient s clinical condition.