Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in affected systems.
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
Lot #0001047501 of the 9025RHTR ABG sampling kits have been manufactured containing the incorrect sub-assembly which includes a larger needle (23 Gage, 1" needle), larger syringe (3mL), and larger amount of heparin (113IU), than what is intended for the kit.
The diameter of the Fogarty Catheter balloon, measures 6mm instead of 10mm.
Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in affected systems.
It was reported that the models used to manufacture S-200956/958 had the wrong dimension used to create the spherical radius cutting surface. Device design allows a larger spherical radius cutting surface than intended.
FDA determined that a 510(k) is needed for the Vault Mapping software (also known as Lens Fitting).
The kit box label incorrectly identifies PCP (instead of TCA) as one of the 10 drug analytes included in the kit. The pouch label, product insert, and the device itself are labeled correctly.
Potential failure in the balloon bond in the inner catheter stem, leading to the separation of the balloon from the inner stem, leaving it in the rectal cavity.
FDA determined that a 510(k) is needed for the Vault Mapping software (also known as Lens Fitting).
Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in affected systems.
Rigid Arm Supports (PN: 106015 and 106047), which were used as part of the Posiboard-2 and Posirest, and were distributed as PN: 106015, 106047, 106521, 106522, 109030, and 109040 between Ocotober 2009 and July 2012 may have a height difference of approximately 17mm.
Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in affected systems.
Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in affected systems.
The issue was identified as a firmware problem with the generator.
Software defect
Rigid Arm Supports (PN: 106015 and 106047), which were used as part of the Posiboard-2 and Posirest, and were distributed as PN: 106015, 106047, 106521, 106522, 109030, and 109040 between Ocotober 2009 and July 2012 may have a height difference of approximately 17mm.
The firm has identified that a fitting in some Slidemaker Stainer units may not completely seal after installation.
Tissu-Trans Filtron 500, 3-TT-FILTRON 500. Product Usage: Intended for fat transfer and liposuction.
Shippert Medical Technologies
Seal integrity of the Tissu-Trans product sterile packages cannot be assured.
Tissu-Trans Filtron 250, 3-TT-FILTRON 250. Product Usage: Intended for fat transfer and liposuction.
Shippert Medical Technologies
Seal integrity of the Tissu-Trans product sterile packages cannot be assured.
The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.