The product was mislabeled with the incorrect label.
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
ISH Protease 3, Catalog Number 05273331001, model 780-4149 Immunohistochemistry (IHC) for in vitro diagnostic use.
Ventana Medical Systems
Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
NEXES VEN IVIEW DAB DET KT JPN-US EXPORT, Catalog Number 05266084001 Model 760-041 Immunohistochemistry (IHC) for in vitro diagnostic use.
Ventana Medical Systems
Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
The Gentle Thread PLGA Round Head, 7x20mm and the Gentle Thread PLGA Full Thread 10x30 have been commingled.
The temperature sensor catheters may be defective in that they will show a lower body temperature measurement than the actual temperature of the patient.
The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
CONFIRM anti-Progesterone Receptor (PR) (1E2) Rabbit Monoclonal Primary Antibody, Catalog Number 05278392001, model 790-4296 Immunohistochemistry (IHC) for in vitro diagnostic use.
Ventana Medical Systems
Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
The temperature sensor catheters may be defective in that they will show a lower body temperature measurement than the actual temperature of the patient.
PATHFAST NTproBNP Catalog Number: PF1061-KUS PATHFASTΒΏ NTproBNP is a product for in-vitro diagnostic use with the in vitro diagnostic system PATHFASTΒΏ for the quantitative measurement of N-terminal-pro B-type Natriuretic Peptide (NT-proBNP) in heparinized or EDTA whole blood and plasma.
Polymedco, Inc.,/Polymedco Cancer Diagnostic Products
A pinhole caused by the laser printer in the aluminum sheet over the ALP reagent wells caused leakage through the pinhole. If the cartridge with leakage is used, higher measurement value as low as +12% and no higher than +20% bias may be obtained.
The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
DORO LUCENT Transitional Member, REF 1101.026 Component of headrest system used in open and percutaneous craniotomies and spinal surgeries
Pro-Med Instruments Gmbh
The combination of two different design revisions of DORO LUCENT Base Unit, Locking Transitional Member, and Transitional Member within one set up might lead to a malfunction and failure of the headrest system.
sterEOS workstation
EOS Imaging
3D projections may be incorrect when the pair of images used is a secondary capture generated from a series containing more than two images, where one of the images has been manually flipped. Projection errors may lead to errors in calculation of 3D clinical parameters when adjusting the position, size and shape of the 3D objects used. Incorrect clinical parameters can be a contributing factor in incorrect clinical decisions, or sub-optimal surgical interventions.
The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
Potential for bucky (the part that holds the grid and is moveable to position the patient) moving unexpectedly.