IsoFlex LAL Support Surface, Model Number 2860 Product Usage: : The IsoFlex LAL support surface assists in the prevention and treatment of all pressure injuries (including stages 1,2,3,4, unstageable injury, and deep tissue injury) and is recommended to be implemented in combination with clinical evaluation of risk factors and skin assessments made by a health care professional. This support surface is for use with human patients with existing or at risk of incurring pressure injuries. The patient must not exceed the safe working load as specified by the support surface. A healthcare professional should determine the use of this support surface by patients outside of the therapeutic weight range. The minimum patient age requirement for this support surface is two years. IsoFlex LAL shall be used with a support surface cover at all times. The support surface cover can interact with all external skin. This support surface is for use by patients in an acute care setting. This may include critical care, step down, progressive care, med/surg, subacute care, and post anesthesia care unit (PACU), or other locations as prescribed by a physician. Operators of this support surface include healthcare professionals (such as nurses, nurse aids, or doctors). IsoFlex LAL support surface is not intended to: be used in a home health environment setting be used as a sterile product include a measuring function be used with a patient less than two years of age Stryker promotes the clinical assessment of each patient and appropriate usage by the operator.

Stryker Medical Division of Stryker

Class I - Dangerous

Certain 2860 IsoFlex LAL support surfaces were assembled with a non-conforming fire barrier.

Aug 13, 2018 Infusion Pumps Nationwide View Details β†’

The slides did not meet the current claims for Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantification (LoQ) for serum and urine samples and therefore affecting the low end of the measuring range for this assay. The current low end measuring range is 0.05mg/dL for serum and 1.2mg/dL for urine. Based on the CREA testing results for the LoQ, the revised low end CREA measuring range is 0.15mg/dL for serum and 3.2mg/dL for urine.

Jul 11, 2018 Diagnostic Equipment Nationwide View Details β†’
Class I - Dangerous

The Tube Current Modulation feature (Modulated Scans) is not working in AIRO systems with software version 2.0.0.0, and operators would not be able to detect this fault until after a scan is completed.

Aug 1, 2018 Other Medical Devices View Details β†’

The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.

Jun 21, 2018 Diagnostic Equipment Nationwide View Details β†’

The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.

Jun 21, 2018 Diagnostic Equipment Nationwide View Details β†’

The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.

Jun 21, 2018 Diagnostic Equipment Nationwide View Details β†’

Communication-connection error.

Dec 27, 2017 Implants & Prosthetics Nationwide View Details β†’

The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.

Jun 21, 2018 Diagnostic Equipment Nationwide View Details β†’

In the hemodialysis machine during the cleaning/disinfection program, when there is a power failure or interruption, the dialysis program button should be greyed out when the machine is powered up and a mandatory rinse should be required. However, if a power failure or interruption occurs during the cleaning/disinfection program with the auto-start feature enabled, the machine enters the dialysis program even though the dialysis program button is disabled and the mandatory rinse has not been performed. As a result, disinfecting chemical may not be completely removed from the machine before starting a treatment and disinfectant could exist in the dialysate delivered to the patient.

Jul 13, 2018 Other Medical Devices Nationwide View Details β†’

The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.

Jun 21, 2018 Diagnostic Equipment Nationwide View Details β†’

In the hemodialysis machine during the cleaning/disinfection program, when there is a power failure or interruption, the dialysis program button should be greyed out when the machine is powered up and a mandatory rinse should be required. However, if a power failure or interruption occurs during the cleaning/disinfection program with the auto-start feature enabled, the machine enters the dialysis program even though the dialysis program button is disabled and the mandatory rinse has not been performed. As a result, disinfecting chemical may not be completely removed from the machine before starting a treatment and disinfectant could exist in the dialysate delivered to the patient.

Jul 13, 2018 Other Medical Devices Nationwide View Details β†’

The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.

Jun 21, 2018 Diagnostic Equipment Nationwide View Details β†’

A pinhole caused by the laser printer in the aluminum sheet over the ALP reagent wells caused leakage through the pinhole. If the cartridge with leakage is used, higher measurement value as low as +12% and no higher than +20% bias may be obtained.

Apr 30, 2018 Diagnostic Equipment View Details β†’

The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.

Jun 21, 2018 Diagnostic Equipment Nationwide View Details β†’