The TCS Straight Hexlobular #6 Drivers (5210-1004) tip sheared off during use in procedures that involved patients with dense bones. As a result of this malfunction, Titan Spine, LLC is removing all of the TCS Straight Hex lobular #6 Drivers (5210-1004) Lot# G150101 from the field.

Feb 20, 2015 Surgical Instruments View Details →

In May 2013, the original recall was initiated due to complaints the Hand Switch for the Electric Pen Drive was not labeled correctly, and the on and lock labels were interchanged.

Dec 19, 2014 Surgical Instruments Nationwide View Details →

Correction due to a fault that causes distal SUJ (DSUJ) failures that indicates a mismatch between the positions reported by the redundant position sensors on one of the patient cart arm joints.

Mar 3, 2015 Surgical Instruments Nationwide View Details →

Correction due to a fault that causes distal SUJ (DSUJ) failures that indicates a mismatch between the positions reported by the redundant position sensors on one of the patient cart arm joints.

Mar 3, 2015 Surgical Instruments Nationwide View Details →

Correction due to a fault that causes distal SUJ (DSUJ) failures that indicates a mismatch between the positions reported by the redundant position sensors on one of the patient cart arm joints.

Mar 3, 2015 Surgical Instruments Nationwide View Details →

The recalling firm was notified by the contract manufacturer that they had reversed the L2, L3, L4, or L5 components on the P/N09010 RoHS boards of the Coblator II Controller (Model RF8000E).

Feb 20, 2015 Surgical Instruments Nationwide View Details →

BD Posiflush Saline Flush Syringe 10 ml inlcuded in the following device kits: Bard Poly Midline 3F Full Tray (Single-Lumen) (with Microintroducer), Bard Poly Midline 3F Maximal Barrier Tray (Single-Lumen), Bard Poly Midline 4F Maximal Barrier Tray (Single-Lumen) (with Microintroducer), Bard Poly Midline BARD MYPICC KIT (4F) (Single Lumen), Groshong NXT Catheter 5F Maximal Barrier Tray (Dual-Lumen) (with MicroIntroducer) , Groshong NXT BARD MYPICC KIT (5F) (Dual-Lumen) (45cm), PowerPICC Catheter 5F Full Tray (Single-Lumen), PowerPICC Catheter 5F Maximal Barrier Tray (Single-Lumen), Poly Per-Q-Cath Catheter 3F Single-Lumen Full Tray (with Microintroducer), Poly Per-Q-Cath Catheter 4F Single-Lumen Full Tray (with Mircrointroducer), Poly Per-Q-Cath 5F Dual-Lumen Full Tray (with Microintroducer), Poly Per-Q-Cath Catheter 3F Full Tray (Single-Lumen) (with Microintroducer) (60cm), Poly Per-Q-Cath Catheter 4F Full Tray (Single-Lumen) (with Microintroducer) (60cm), Poly Per-Q-Cath Catheter 5F Full Tray (Dual-Lumen) (with Microintroducer) (60cm), Groshong NXT ClearVue 4F Catheter Full Tray (Single-Lumen) (60cm), Groshong NXT ClearVue 4F Catheter Full Tray (Single-Lumen) (60cm) (with Microintroducer), Groshong NXT ClearVue Catheter 4F Maximal Barrier Tray (Single-Lumen), PowerPICC 5F Dual-Lumen Polyurethane Catheter with MicroEZ Microintroducer Nurse Full Tray (with Lidocaine), PowerPICC Catheter 5F Maximal Barrier Tray with Microintroducer (Dual-Lumen), PowerPICC FT Catheter 5F Maximal Barrier Tray Tray (Dual-Lumen), PowerPICC Bard MYPICC Kit 5F (Dual-Lumen)(with Nitinol Guidewire)(70cm), PowerPICC Catheter 5F Maximal Barrier Tray (Dual-Lumen) (55cm) (with Microintroducer), PowerPICC Catheter 5F Full Tray (Dual-Lumen) (70cm Nitinol Guidewire) (with Microintroducer), PowerPICC Bard MYPICC Kit 5F (Dual-Lumen) RN, PowerPICC BARD MYPICC KIT (5F) (Dual-Lumen), PowerPICC Catheter 6F Full Tray (Triple-Lumen)(with Microintroducer) (with Lidocaine), PowerPICC HF Catheter 5F Full Tray (Triple-Lumen), PowerPICC HF Catheter 5F Maximal Barrier Tray (Triple-Lumen), PowerPICC Catheter 6F (Triple-Lumen) Maximal Barrier Tray, PowerPICC SOLO HF Catheter with Sherlock Tip Location System (TLS) Stylet 5F Full Tray (Triple-Lumen), PowerPICC HF Catheter with Sherlock Tip Location System (TLS) Stylet 5F Maximal Barrier Tray (Triple-Lumen), PowerPICC HF Bard MYPICC Kit 5F (Triple-Lumen), PowerPICC HF BARD MYPICC KIT (5F) (Triple-Lumen), PowerPICC HF Catheter Maximal Barrier Tray (5F) (Triple-Lumen) (55cm), Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet 4F Full Tray (Single-Lumen), Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet 4F Full Tray (Single-Lumen) (with Microintroducer), Groshong NXT Catheter with Sherlock Tip Location System (TLS) Stylet 5F Full Tray (Dual-Lumen) (45cm), Groshong NXT Catheter with Sherlock Tip Location System (TLS) Stylet 5F Full Tray (Dual-Lumen) (45cm) (with Microintroducer), Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet 4F Maximal Barrier Tray (Single-Lumen), Groshong NXT Catheter with Sherlock Tip Location System (TLS) Stylet 5F Maximal Barrier Tray (Dual-Lumen) (45cm), Groshong NXT Sherlock II BARD MYPICC KIT (4F) (Single-Lumen) (TLS), Poly Per-Q-Cath Catheter with Sherlock Tip Location System (TLS) Stylet 3F Full Tray (with Microintroducer), Poly Per-Q-Cath Catheter with Sherlock Tip Location System (TLS) Stylet 4F Full Tray (with Microintroducer), Poly Per-Q-Cath Catheter with Sherlock Tip Location System (TLS) Stylet 4F Full Tray, Poly Per-Q-Cath Catheter with Sherlock Tip Location System (TLS) Stylet 5F Full Tray, Poly Per-Q-Cath Sherlock II Bard MYPICC KIT (4F) (Sinlge-Lumen) (TLS), Poly Per-Q-Cath Sherlock II BARD MYPICC KIT (4F) (Single-Lumen) (TLS), PowerPICC Catheter with Sherlock Tip Location System (TLS) Stylet 5F Full Tray (Single-Lumen) (with Microintroducer), PowerPICC 5F Dual-Lumen Polyurethane Catheter with Sherlock Tip L

Bard Access Systems

Class I - Dangerous

Some unit packages of BD PosiFlush SF Saline Flush Syringe 10 mL may exhibit open seals, which could impact product sterility. Bard includes these pouched syringes, attached to some configurations of PICC kits, Port Access kits (PAK), PowerGlide kits, Midline kits, and Convenience kits.

Aug 8, 2014 Surgical Instruments Nationwide View Details →

A configuration update needs to be done on TactiSys to appropriately recognize all TactiCath catheters.

Nov 25, 2014 Surgical Instruments Nationwide View Details →

BD Posiflush Saline Flush Syringe 10 ml inlcuded in the following device kits: SafeStep Port Access Kit 22G x 0.75in , SafeStep Port Access Kit 22G x 0.75in with Y-Site, SafeStep Port Access Kit 22G x 1 in , SafeStep Port Access Kit 20G x 0.75in , SafeStep Port Access Kit 20G x 0.75in with Y-Site, SafeStep Port Access Kit 20G x 1 in , SafeStep Port Access Kit 20G x 1 in with Y-Site, SafeStep Port Access Kit 19G x 0.75in , SafeStep Port Access Kit 19G x 0.75in with Y-Site, SafeStep Port Access Kit 19G x 1 in , SafeStep Port Access Kit 19G x 1 in with Y-Site, SafeStep Port Access Kit 20G x 1.5in , SafeStep Port Access Kit 20G x 1.5in with Y-Site, MiniLoc Port Access Kit 20G x 1 in, MiniLoc Port Access Kit 20G x 0.75in, MiniLoc Port Access Kit 22G x 1 in, MiniLoc Port Access Kit 22G x 0.75in, MiniLoc Port Access Kit 19G x 1 in with Y-Site, MiniLoc Port Access Kit 20G x 1 in with Y-Site, MiniLoc Port Access Kit 20G x 0.75in with Y-Site, MiniLoc Port Access Kit 22G x 1.5in with Y-Site, MiniLoc Port Access Kit 22G x 0.75in with Y-Site, PowerLoc Max Port Access Kit 19G x 1.0 in with Y-Site, PowerLoc Max Port Access Kit 19G x 0.75in with Y-Site, PowerLoc Max Port Access Kit 20G x 0.75in with Y-Site, PowerLoc Max Port Access Kit 19G x 1.0 in , PowerLoc Max Port Access Kit 20G x 1.0 in , PowerLoc Max Port Access Kit 20G x 0.75in , PowerLoc Max Port Access Kit 22G x 1.0 in , PowerLoc Max Port Access Kit 22G x 0.75in The syringe is used to pre-flush the catheter prior to placement and may also be used to verify patency post-placement.

Bard Access Systems

Class I - Dangerous

Some unit packages of BD PosiFlush SF Saline Flush Syringe 10 mL may exhibit open seals, which could impact product sterility. Bard includes these pouched syringes, attached to some configurations of PICC kits, Port Access kits (PAK), PowerGlide kits, Midline kits, and Convenience kits.

Aug 8, 2014 Surgical Instruments Nationwide View Details →

Some unit packages of BD PosiFlush SF Saline Flush Syringe 10 mL may exhibit open seals, which could impact product sterility. Bard includes these pouched syringes, attached to some configurations of PICC kits, Port Access kits (PAK), PowerGlide kits, Midline kits, and Convenience kits.

Aug 8, 2014 Surgical Instruments Nationwide View Details →

Some unit packages of BD PosiFlush SF Saline Flush Syringe 10 mL may exhibit open seals, which could impact product sterility. Bard includes these pouched syringes, attached to some configurations of PICC kits, Port Access kits (PAK), PowerGlide kits, Midline kits, and Convenience kits.

Aug 8, 2014 Surgical Instruments Nationwide View Details →

Some unit packages of BD PosiFlush SF Saline Flush Syringe 10 mL may exhibit open seals, which could impact product sterility. Bard includes these pouched syringes, attached to some configurations of PICC kits, Port Access kits (PAK), PowerGlide kits, Midline kits, and Convenience kits.

Aug 8, 2014 Surgical Instruments Nationwide View Details →

The firm is recalling sterile enteral feeding syringes due to determining that the sterility of the affected products may be compromised as a result of a potential breach in packaging.

Feb 13, 2015 Surgical Instruments Nationwide View Details →

There is the possibility that the Vacuum Source Tubing set for the Endo Wrist Stabilizer may have the outer pouch compromised during shipping, resulting in small pinholes which compromise the sterility of the pouches outer surface.

Feb 16, 2015 Surgical Instruments View Details →

Biosense Webster is recalling the LASSO NAV Duo Loop eco Catheters because of reports of difficulty retracting and damage to the spine cover.

Feb 25, 2015 Surgical Instruments Nationwide View Details →

Outer package product label incorrectly identifies the product as FiberOptix Flex IAB 7.5Fr rather than the correct FiberOptix IAB 8Fr

Feb 5, 2015 Surgical Instruments Nationwide View Details →