da Vinci Xi Surgical System, IS4000; Assy, Distal SUJ, Inner, IS4000 Material number: 380662-20 and 380662-21. The Intuitive Surgical Endoscope Instrument Control System (da Vinci Surgical System, Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endocopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endscopic retractors, electrocautery and accessories for edoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximations, ligation, electrocautery, suturing and delivery and placement of microware and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic labaroscopic surgical procedures, general thorascopic surgical procedures and thorascopically-assisted cardiotomy procedures.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Material number: 3680662-20: Serial numbers: 1311140273, 1311140370, 1311140385, 1311140148, 1311140099, 1311140152, 1311140081, 1311140473, 1311140561, 1311140409. Material number: 380662-21: Serial numbers: 1311140225, 1311140437, 1311140476, 1311140336, 1311140472, 1311140438, 1311140353, 1311140195, 1311140545, 1311140098 1311140189, 1311140465, 1311140394, 1311140366, 1311140008, 1311140115 1311140014, 1311140371, 1311140238, 1311140306, 1311140332, 1311140404 1311140054, 1311140188, 1311140324, 1311140453, 1311140408, 1311140237 1311140220, 1311140372, 1311140325, 1311140088, 1311140535, 1311140355.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Intuitive Surgical, Inc.
- Reason for Recall:
- Correction due to a fault that causes distal SUJ (DSUJ) failures that indicates a mismatch between the positions reported by the redundant position sensors on one of the patient cart arm joints.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
da Vinci Xi Surgical System, IS4000; Assy, Distal SUJ, Inner, IS4000 Material number: 380662-20 and 380662-21. The Intuitive Surgical Endoscope Instrument Control System (da Vinci Surgical System, Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endocopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endscopic retractors, electrocautery and accessories for edoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximations, ligation, electrocautery, suturing and delivery and placement of microware and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic labaroscopic surgical procedures, general thorascopic surgical procedures and thorascopically-assisted cardiotomy procedures.
Product Codes/Lot Numbers:
Material number: 3680662-20: Serial numbers: 1311140273, 1311140370, 1311140385, 1311140148, 1311140099, 1311140152, 1311140081, 1311140473, 1311140561, 1311140409. Material number: 380662-21: Serial numbers: 1311140225, 1311140437, 1311140476, 1311140336, 1311140472, 1311140438, 1311140353, 1311140195, 1311140545, 1311140098 1311140189, 1311140465, 1311140394, 1311140366, 1311140008, 1311140115 1311140014, 1311140371, 1311140238, 1311140306, 1311140332, 1311140404 1311140054, 1311140188, 1311140324, 1311140453, 1311140408, 1311140237 1311140220, 1311140372, 1311140325, 1311140088, 1311140535, 1311140355.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1302-2015
Related Recalls
Due to an increase in complaints concerning foot tray pedal spring failing resulting in the pedal remaining pressed
8MM,HORIZON SMALL CLIP APPLIER,IS4000¿ REF 470401
Intuitive Surgical
Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.
8MM,MEGA SUTURECUT ND,IS4000 REF 471309
Intuitive Surgical
Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.