software bug issues for SW-Version VA48A_SP0. The following safety issues were resolved: 1) Correction to improve visual warning and error indication son the gantry display. 2) Correction to improve acquisition data in order to optimize image quality. 3) Correction to improve robustness and general system behavior in some exception handling procedures. 4) Correction to improve auto post processin

Aug 14, 2015 Other Medical Devices Nationwide View Details →

Product labeling of these sterile guide wires (external carton label and patient record label) states a 10-year expiration date. Package testing supports a 5-year expiration date for these same guide wires.

Aug 27, 2015 Other Medical Devices Nationwide View Details →

Lightspeed 16 CT

GE Healthcare

Class I - Dangerous

Potential for expulsion of the Power Supply on Lightspeed CT scanner or Discovery ST PET CT or Discovery LS PET CT scanner.

Jul 17, 2015 Other Medical Devices Nationwide View Details →

PET Discovery LS

GE Healthcare

Class I - Dangerous

Potential for expulsion of the Power Supply on Lightspeed CT scanner or Discovery ST PET CT or Discovery LS PET CT scanner.

Jul 17, 2015 Other Medical Devices Nationwide View Details →

The Synthes Cranial Flap Tube Clamp and Crimping Device for Cranial Tube Clamp had been labeled MR Safe although they do not have the testing protocols currently required to designate them as MR Safe.

Aug 31, 2015 Other Medical Devices Nationwide View Details →

PET Discovery ST 16

GE Healthcare

Class I - Dangerous

Potential for expulsion of the Power Supply on Lightspeed CT scanner or Discovery ST PET CT or Discovery LS PET CT scanner.

Jul 17, 2015 Other Medical Devices Nationwide View Details →

The EPX-2500 Operation Manual and sales brochure incorrectly identify that the ED-530XT endoscope is meant to be used with the EPX-2500 Processor. The Operation Manual and sales brochure state that when the ED-530XT endoscope is used with the 2500 processor, a "super image" is displayed when it is actually a "standard image" that is displayed.

Jun 30, 2015 Other Medical Devices Nationwide View Details →

The EPX-2500 Operation Manual and sales brochure incorrectly identify that the ED-530XT endoscope is meant to be used with the EPX-2500 Processor. The Operation Manual and sales brochure state that when the ED-530XT endoscope is used with the 2500 processor, a "super image" is displayed when it is actually a "standard image" that is displayed.

Jun 30, 2015 Other Medical Devices Nationwide View Details →

The ASP6025 Tissue Processor instrument is incorrectly getting into a state of bottle empty during retort filling, after passing the fill lever sensor 2 and prior to reaching level sensor 3, thus not completing a 5 litre fill.

Jul 27, 2015 Other Medical Devices Nationwide View Details →

DLS pin breakage during planned implant removal, after uneventful and successful healing of the fracture. This notice to additional consignees identified after recall Z-0390-2014, related to complaints of breakage located at the bottom of the screws and under the locking screw head.

Aug 11, 2015 Other Medical Devices Nationwide View Details →

Cross-threading of the hemostasis valve may occur if it is tightened with the dilator in place, possibly preventing subsequent sealing of the valve when desired. This Field Correction reinforces the Directions for Use (DFU) and gives further guidance on correct use of the hemostasis valve to avoid cross-threading and to securely seal the valve, minimizing potential for undesired blood leakage.

Aug 25, 2015 Other Medical Devices Nationwide View Details →

Software Version SW 005.006_SH, resulted in the removal of purge enforcement from traditional New-Case, Power Up and Syringe Change use cases while the injector head is in the upright position.

Aug 4, 2015 Other Medical Devices Nationwide View Details →

The Leica ASP200S/ASP300S Tissue Processor have an incorrect internal wiring to the remote alarm plug, which can cause the remote alarm to not function correctly in the case of an instrument failure.

Aug 5, 2015 Other Medical Devices Nationwide View Details →