Zimmer Guide Wires for orthopedic procedures.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    00-2228-024-00 2.4MM Diameter Bullet Tip Guide Wire; 00-2255-025-00 Humeral Smooth Guide Wire 2.4mm 70cm Length; 00-2255-026-00 Humeral Bullet Tip Guide Wire 2.4mm Diameter 70cm Length; 47-2237-033-00 Smooth Guide Wire 2.4mm Diameter 100cm Length; 47-2237-037-00 3.0MM Diameter Smooth Guide Wire 100cm Length; 47-2237-038-00 Bullet Tip Guide Wire 3.0mm Diameter 100cm Length; 47-2255-008-00 Ball Tip Guide Wire 2.4mm Diameter 70cm Length; 47-2255-008-01 Ball Tip Guide Wire 3.0mm Diameter 100cm Length All non-expired, distributed lots with a 10 year expiration date on the label.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Zimmer, Inc.
Reason for Recall:
Product labeling of these sterile guide wires (external carton label and patient record label) states a 10-year expiration date. Package testing supports a 5-year expiration date for these same guide wires.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Zimmer Guide Wires for orthopedic procedures.

Product Codes/Lot Numbers:

00-2228-024-00 2.4MM Diameter Bullet Tip Guide Wire; 00-2255-025-00 Humeral Smooth Guide Wire 2.4mm 70cm Length; 00-2255-026-00 Humeral Bullet Tip Guide Wire 2.4mm Diameter 70cm Length; 47-2237-033-00 Smooth Guide Wire 2.4mm Diameter 100cm Length; 47-2237-037-00 3.0MM Diameter Smooth Guide Wire 100cm Length; 47-2237-038-00 Bullet Tip Guide Wire 3.0mm Diameter 100cm Length; 47-2255-008-00 Ball Tip Guide Wire 2.4mm Diameter 70cm Length; 47-2255-008-01 Ball Tip Guide Wire 3.0mm Diameter 100cm Length All non-expired, distributed lots with a 10 year expiration date on the label.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2820-2015

Related Recalls

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.

Dec 2, 2025 Other Medical Devices Nationwide View Details →

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.

Dec 2, 2025 Other Medical Devices Nationwide View Details →

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.

Dec 2, 2025 Other Medical Devices Nationwide View Details →