BD Pyxis Mini Main REF: 349 Medication cabinet
CareFusion 303
Fingerprint scanner may overheat to a temperature to cause 1st degree burn.
π₯ Medical Devices β’ 15,809 recalls
CareFusion 303
Fingerprint scanner may overheat to a temperature to cause 1st degree burn.
It has been identified that the impacted docking station (SN 2521008873) was not product released from manufacturing and may not have completed required testing prior to being shipped to your facility. There is a potential risk of electrical shock due to the device s safety testing being unverifiable.
KAYSERBETTEN GMBH & CO. KG
If the adjustment functions of the hand control, as instructed in the Manual, are not locked with the provided key when the patient (child) is unattended, there is an increased risk that the patient (child) or third parties (children) will operate the hand control and trap themselves or others under the bed frame or between the bed frame and the floor. This can lead to serious injuries or even death of the patient (child) or third parties (children).
Daavlin Distributing Company
Some of the Calibration/Output Certificates had the outputs reversed.
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
B. Braun was notified by our supplier, Ferndale Laboratories, Inc., the manufacturer of MASTISOLΒΏ Liquid Adhesive, that they have issued a voluntary recall due to potential cracking of the butyrate tubing (outside tube), which renders the product unusable.
CareFusion 303
Fingerprint scanner may overheat to a temperature to cause 1st degree burn.
B. Braun was notified by our supplier, Ferndale Laboratories, Inc., the manufacturer of MASTISOLΒΏ Liquid Adhesive, that they have issued a voluntary recall due to potential cracking of the butyrate tubing (outside tube), which renders the product unusable.
Due to customer complaint regarding incorrect display box labeling.
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
CareFusion 303
Fingerprint scanner may overheat to a temperature to cause 1st degree burn.
B. Braun was notified by our supplier, Ferndale Laboratories, Inc., the manufacturer of MASTISOLΒΏ Liquid Adhesive, that they have issued a voluntary recall due to potential cracking of the butyrate tubing (outside tube), which renders the product unusable.
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Siemens Medical Solutions USA
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Grace Medical
Due to incorrect functional length on device labeling.
Oculus Technologies of Mexico, S.A. de C.V.
The recall is being conducted due to corrosion of the internal metallic spring within the sprayer component. This condition is caused by material incompatibility between the product solution and the stainless steel spring and may result in visible discoloration and the presence of trace levels of metallic elements (including iron, chromium, and nickel) in the solution during use.
Siemens Medical Solutions USA
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console.
Medline Industries
Specific Medline Kits contain Cardinal Health Monoject 1mL Luer Lock Syringes, Tuberculin Print, Sterile. Cardinal Health has recalled these syringes because the outer carton and blister pack are labeled as a 1 mL Luer Lock Tuberculin Syringe (Product Code 1180100777), but the syringes inside are U-100 insulin syringes.