Class I - Dangerous

Devices with affected software may experience two unintended motion issues that may lead to contact between the Gantry or table with the operator or patient, along with additional software issues that affect CT performance.

May 29, 2025 Other Medical Devices View Details →

Due to an error in assembly of the thermistor within some units, the affected probes will experience inaccuracy in measured shunt sensor temperature. Additionally, following changes in patient temperature without an in vivo recalibration, affected probes may experience inaccuracy in other measured parameters, including pCO2, pO2, pH, Potassium and calculated parameters, including BE, Bicarbonate, Arterial Oxygen Saturation, Delivered Oxygen, Area Under of the Curve of Delivered Oxygen, Oxygen Consumption, and Oxygen Extraction Ratio, which utilize temperature as an input. Inaccurate data may lead to inappropriate user response.

May 28, 2025 Other Medical Devices Nationwide View Details →

The impacted tape was manufactured with incorrect information on the tape. Incorrect values are printed in the Red zone, Orange zone, and Grey zone. Using the incorrect values in the Red zone could lead to shocking a patient with an excessive dose of joules, causing significant harm including burns, heart damage, and cardiac arrest. Incorrect values in the Orange zone could lead to overdosing the patient with sodium bicarbonate and cause metabolic alkalosis, electrolyte imbalances, tissue damage, and worsen respiratory status. Incorrect information in the Grey may lead to underdosing the patient with sodium bicarbonate and may cause reduced myocardial contractility, decreased response to vasopressors, and increased risk of dysrhythmia.

May 15, 2025 Other Medical Devices Nationwide View Details →
Class I - Dangerous

Analyzer has calibration issues where curves switched from passed to failed due to system errors during aspiration. This anomaly causes erroneous results and delays if unnoticed. RLU values may be misinterpreted as valid measurements by LIS.

May 7, 2025 Other Medical Devices Nationwide View Details →

Cathode ray tube display system encoders and decoders have same MAC address; new system setup with two or more affected units can lead to systems not functioning reliably:audio, video and/or control signals might be interrupted or fall away, displays might switch to failover and back multiple times, devices/sources/sinks might not be discovered/available, might lead to delayed/prolonged procedures

May 9, 2025 Other Medical Devices Nationwide View Details →

AcoSound. Model Number: LW12-BTE-M

HANGZHOU ACOSOUND TECHNOLOGY CO.

Class I - Dangerous

Three issues 1. Label error: The manufacturer information in the labeling is not correct, the manufacturer's name is mislabeled as "Developed by AcoSound," Correct should be "Manufactured by Hangzhou AcoSound." And the labeling has no statement of the place of business for the manufacturer. 2. Customer information: The customer "Blaid" information is not correctly marked; Blaid is not the manufacturer. 3. UDI Information: The UDI information is incomplete and without DI information and correct Pl information, not complied with UDI label requirements.

Apr 25, 2025 Other Medical Devices Nationwide View Details →

Ami HTX.

Spectral Instruments

Class I - Dangerous

The X-ray safety and manufacturing date labels affixed to the Lago X/ AMI HTX instrument do not correspond exactly to the labeling stipulated by the FDA / CDRH.

May 13, 2025 Other Medical Devices View Details →