Spectral CT. Computed tomography X-ray system.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Part No. 728333; UDI: (01)00884838101111(21)10014, (01)00884838101111(21)10015, (01)00884838101111(21)10016, (01)00884838101111(21)10017, (01)00884838101111(21)10018, (01)00884838101111(21)10019, (01)00884838101111(21)10020, (01)00884838101111(21)10021, (01)00884838101111(21)10022, (01)00884838101111(21)10023, (01)00884838101111(21)10024, (01)00884838101111(21)10025, (01)00884838101111(21)10027, (01)00884838101111(21)10031, (01)00884838101111(21)10032, (01)00884838101111(21)10033, (01)00884838101111(21)10034, (01)00884838101111(21)10035, (01)00884838101111(21)10036, (01)00884838101111(21)10038, (01)00884838101111(21)10039, (01)00884838101111(21)10040, (01)00884838101111(21)10041, (01)00884838101111(21)10042, (01)00884838101111(21)10043, (01)00884838101111(21)10044, (01)00884838101111(21)10045, (01)00884838101111(21)10046, (01)00884838101111(21)10047, (01)00884838101111(21)10048, (01)00884838101111(21)10049, (01)00884838101111(21)10050, (01)00884838101111(21)10052, (01)00884838101111(21)10053, (01)00884838101111(21)10055, (01)00884838101111(21)10056, (01)00884838101111(21)10057, (01)00884838101111(21)10058, (01)00884838101111(21)10059, (01)00884838101111(21)10060, (01)00884838101111(21)10061, (01)00884838101111(21)10062, (01)00884838101111(21)10063, (01)00884838101111(21)10065, (01)00884838101111(21)10066, (01)00884838101111(21)10067, (01)00884838101111(21)10068, (01)00884838101111(21)10069, (01)00884838101111(21)10070, (01)00884838101111(21)10071, (01)00884838101111(21)10072, (01)00884838101111(21)10074, (01)00884838101111(21)10075, (01)00884838101111(21)10076, (01)00884838101111(21)10077, (01)00884838101111(21)10078, (01)00884838101111(21)10079, (01)00884838101111(21)10080, (01)00884838101111(21)10081, (01)00884838101111(21)10082, (01)00884838101111(21)10084, (01)00884838101111(21)10085, (01)00884838101111(21)10086, (01)00884838101111(21)10087, (01)00884838101111(21)10088, (01)00884838101111(21)10089, (01)00884838101111(21)10090, (01)00884838101111(21)10091, (01)00884838101111(21)10092, (01)00884838101111(21)10093, (01)00884838101111(21)10094, (01)00884838101111(21)10095, (01)00884838101111(21)10096, (01)00884838101111(21)10097, (01)00884838101111(21)10098, (01)00884838101111(21)10099, (01)00884838101111(21)10100, (01)00884838101111(21)10101, (01)00884838101111(21)10102, (01)00884838101111(21)10103, (01)00884838101111(21)10104, (01)00884838101111(21)10105, (01)00884838101111(21)10106, (01)00884838101111(21)10107, (01)00884838101111(21)10108, (01)00884838101111(21)10109, (01)00884838101111(21)10110, (01)00884838101111(21)10111, (01)00884838101111(21)10112, (01)00884838101111(21)10113, (01)00884838101111(21)10114, (01)00884838101111(21)10115, (01)00884838101111(21)10116, (01)00884838101111(21)10117, (01)00884838101111(21)10118, (01)00884838101111(21)10119, (01)00884838101111(21)10120, (01)00884838101111(21)10121, (01)00884838101111(21)10122, (01)00884838101111(21)10123, (01)00884838101111(21)10124, (01)00884838101111(21)10125, (01)00884838101111(21)10126, (01)00884838101111(21)10127, (01)00884838101111(21)10128, (01)00884838101111(21)10129, (01)00884838101111(21)10130, (01)00884838101111(21)10131, (01)00884838101111(21)10132, (01)00884838101111(21)10133, (01)00884838101111(21)10134, (01)00884838101111(21)10135, (01)00884838101111(21)10136, (01)00884838101111(21)10137, (01)00884838101111(21)10138, (01)00884838101111(21)10139, (01)00884838101111(21)10140, (01)00884838101111(21)10141, (01)00884838101111(21)10142, (01)00884838101111(21)10143, (01)00884838101111(21)10144, (01)00884838101111(21)10145, (01)00884838101111(21)10146, (01)00884838101111(21)10147, (01)00884838101111(21)10148, (01)00884838101111(21)10149, (01)00884838101111(21)10150, (01)00884838101111(21)10151, (01)00884838101111(21)10152, (01)00884838101111(21)10153, (01)00884838101111(21)10154, (01)00884838101111(21)10155, (01)00884838101111(21)10156, (01)00884838101111(21)10157, (01)00884838101111(21)10158, (01)00884838101111(21)10159, (01)00884838101111(21)10160, (01)00884838101111(21)10161, (01)00884838101111(21)10162, (01)00884838101111(21)10163, (01)00884838101111(21)10164, (01)00884838101111(21)10165, (01)00884838101111(21)10166, (01)00884838101111(21)10167, (01)00884838101111(21)10168, (01)00884838101111(21)10169, (01)00884838101111(21)10170, (01)00884838101111(21)10171, (01)00884838101111(21)10172, (01)00884838101111(21)10173, (01)00884838101111(21)10174, (01)00884838101111(21)10175, (01)00884838101111(21)10176, (01)00884838101111(21)10177, (01)00884838101111(21)10178, (01)00884838101111(21)10179, (01)00884838101111(21)10180, (01)00884838101111(21)10181, (01)00884838101111(21)10182, (01)00884838101111(21)10183, (01)00884838101111(21)10184, (01)00884838101111(21)10185, (01)00884838101111(21)10186, (01)00884838101111(21)10187, (01)00884838101111(21)10188, (01)00884838101111(21)10189, (01)00884838101111(21)10190, (01)00884838101111(21)10191, (01)00884838101111(21)10192, (01)00884838101111(21)10193, (01)00884838101111(21)10194, (01)00884838101111(21)10195, (01)00884838101111(21)10196, (01)00884838101111(21)10197, (01)00884838101111(21)10198, (01)00884838101111(21)10199, (01)00884838101111(21)10200, (01)00884838101111(21)10201, (01)00884838101111(21)10202, (01)00884838101111(21)10203, (01)00884838101111(21)10204, (01)00884838101111(21)10205, (01)00884838101111(21)10206, (01)00884838101111(21)10207, (01)00884838101111(21)10208, (01)00884838101111(21)10209, (01)00884838101111(21)10210, (01)00884838101111(21)10211, (01)00884838101111(21)10212, (01)00884838101111(21)10213, (01)00884838101111(21)10214, (01)00884838101111(21)10215, (01)00884838101111(21)10217, (01)00884838101111(21)10218, (01)00884838101111(21)10219, (01)00884838101111(21)10220, (01)00884838101111(21)10221, (01)00884838101111(21)10222, (01)00884838101111(21)10223, (01)00884838101111(21)10224, (01)00884838101111(21)10225, (01)00884838101111(21)10226, (01)00884838101111(21)10227, (01)00884838101111(21)10228, (01)00884838101111(21)10229, (01)00884838101111(21)10230, (01)00884838101111(21)10231, (01)00884838101111(21)10232, (01)00884838101111(21)10233, (01)00884838101111(21)10234, (01)00884838101111(21)10235, (01)00884838101111(21)10236, (01)00884838101111(21)10237, (01)00884838101111(21)10238, (01)00884838101111(21)10239, (01)00884838101111(21)10240, (01)00884838101111(21)10241, (01)00884838101111(21)10242, (01)00884838101111(21)10243, (01)00884838101111(21)10244, (01)00884838101111(21)10245, (01)00884838101111(21)10246, (01)00884838101111(21)10247, (01)00884838101111(21)10248, (01)00884838101111(21)10249, (01)00884838101111(21)10250, (01)00884838101111(21)10251, (01)00884838101111(21)10252, (01)00884838101111(21)10253, (01)00884838101111(21)10254, (01)00884838101111(21)10255, (01)00884838101111(21)10256, (01)00884838101111(21)10257, (01)00884838101111(21)10258, (01)00884838101111(21)10259, (01)00884838101111(21)10260, (01)00884838101111(21)10261, (01)00884838101111(21)10262, (01)00884838101111(21)10263, (01)00884838101111(21)10264, (01)00884838101111(21)10265, (01)00884838101111(21)10266, (01)00884838101111(21)10267, (01)00884838101111(21)10268, (01)00884838101111(21)10269, (01)00884838101111(21)10270, (01)00884838101111(21)10271, (01)00884838101111(21)10272, (01)00884838101111(21)10273, (01)00884838101111(21)10274, (01)00884838101111(21)10275, (01)00884838101111(21)10276, (01)00884838101111(21)10277, (01)00884838101111(21)10278, (01)00884838101111(21)10279, (01)00884838101111(21)10280, (01)00884838101111(21)10281, (01)00884838101111(21)10284, (01)00884838101111(21)10285, (01)00884838101111(21)10286, (01)00884838101111(21)10287, (01)00884838101111(21)10288, (01)00884838101111(21)10289, (01)00884838101111(21)10290, (01)00884838101111(21)10291, (01)00884838101111(21)10292, (01)00884838101111(21)10293, (01)00884838101111(21)10294, (01)00884838101111(21)10298, (01)00884838101111(21)10299, (01)00884838101111(21)10300, (01)00884838101111(21)10301, (01)00884838101111(21)12000, (01)00884838101111(21)12001, (01)00884838101111(21)12002, (01)00884838101111(21)12003, (01)00884838101111(21)12004, (01)00884838101111(21)12005, (01)00884838101111(21)12006, (01)00884838101111(21)12007, (01)00884838101111(21)12008, (01)00884838101111(21)12009, (01)00884838101111(21)396001, (01)00884838101111(21)396003, (01)00884838101111(21)396004, (01)00884838101111(21)396005, (01)00884838101111(21)396006, (01)00884838101111(21)396007, (01)00884838101111(21)396008, (01)00884838101111(21)396009, (01)00884838101111(21)396010, (01)00884838101111(21)396011; Serial No. 10013, 10014, 10015, 10016, 10017, 10018, 10019, 10020, 10021, 10022, 10023, 10024, 10025, 10027, 10031, 10032, 10033, 10034, 10035, 10036, 10038, 10039, 10040, 10041, 10042, 10043, 10044, 10045, 10046, 10047, 10048, 10049, 10050, 10052, 10053, 10055, 10056, 10057, 10058, 10059, 10060, 10061, 10062, 10063, 10065, 10066, 10067, 10068, 10069, 10070, 10071, 10072, 10074, 10075, 10076, 10077, 10078, 10079, 10080, 10081, 10082, 10084, 10085, 10086, 10087, 10088, 10089, 10090, 10091, 10092, 10093, 10094, 10095, 10096, 10097, 10098, 10099, 10100, 10101, 10102, 10103, 10104, 10105, 10106, 10107, 10108, 10109, 10110, 10111, 10112, 10113, 10114, 10115, 10116, 10117, 10118, 10119, 10120, 10121, 10122, 10123, 10124, 10125, 10126, 10127, 10128, 10129, 10130, 10131, 10132, 10133, 10134, 10135, 10136, 10137, 10138, 10139, 10140, 10141, 10142, 10143, 10144, 10145, 10146, 10147, 10148, 10149, 10150, 10151, 10152, 10153, 10154, 10155, 10156, 10157, 10158, 10159, 10160, 10161, 10162, 10163, 10164, 10165, 10166, 10167, 10168, 10169, 10170, 10171, 10172, 10173, 10174, 10175, 10176, 10177, 10178, 10179, 10180, 10181, 10182, 10183, 10184, 10185, 10186, 10187, 10188, 10189, 10190, 10191, 10192, 10193, 10194, 10195, 10196, 10197, 10198, 10199, 10200, 10201, 10202, 10203, 10204, 10205, 10206, 10207, 10208, 10209, 10210, 10211, 10212, 10213, 10214, 10215, 10217, 10218, 10219, 10220, 10221, 10222, 10223, 10224, 10225, 10226, 10227, 10228, 10229, 10230, 10231, 10232, 10233, 10234, 10235, 10236, 10237, 10238, 10239, 10240, 10241, 10242, 10243, 10244, 10245, 10246, 10247, 10248, 10249, 10250, 10251, 10252, 10253, 10254, 10255, 10256, 10257, 10258, 10259, 10260, 10261, 10262, 10263, 10264, 10265, 10266, 10267, 10268, 10269, 10270, 10271, 10272, 10273, 10274, 10275, 10276, 10277, 10278, 10279, 10280, 10281, 10284, 10285, 10286, 10287, 10288, 10289, 10290, 10291, 10292, 10293, 10294, 10298, 10299, 10300, 10301, 12000, 12001, 12002, 12003, 12004, 12005, 12006, 12007, 12008, 12009, 396001, 396003, 396004, 396005, 396006, 396007, 396008, 396009, 396010, 396011.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips North America Llc
- Reason for Recall:
- Devices with affected software may experience two unintended motion issues that may lead to contact between the Gantry or table with the operator or patient, along with additional software issues that affect CT performance.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Spectral CT. Computed tomography X-ray system.
Product Codes/Lot Numbers:
Part No. 728333; UDI: (01)00884838101111(21)10014, (01)00884838101111(21)10015, (01)00884838101111(21)10016, (01)00884838101111(21)10017, (01)00884838101111(21)10018, (01)00884838101111(21)10019, (01)00884838101111(21)10020, (01)00884838101111(21)10021, (01)00884838101111(21)10022, (01)00884838101111(21)10023, (01)00884838101111(21)10024, (01)00884838101111(21)10025, (01)00884838101111(21)10027, (01)00884838101111(21)10031, (01)00884838101111(21)10032, (01)00884838101111(21)10033, (01)00884838101111(21)10034, (01)00884838101111(21)10035, (01)00884838101111(21)10036, (01)00884838101111(21)10038, (01)00884838101111(21)10039, (01)00884838101111(21)10040, (01)00884838101111(21)10041, (01)00884838101111(21)10042, (01)00884838101111(21)10043, (01)00884838101111(21)10044, (01)00884838101111(21)10045, (01)00884838101111(21)10046, (01)00884838101111(21)10047, (01)00884838101111(21)10048, (01)00884838101111(21)10049, (01)00884838101111(21)10050, (01)00884838101111(21)10052, (01)00884838101111(21)10053, (01)00884838101111(21)10055, (01)00884838101111(21)10056, (01)00884838101111(21)10057, (01)00884838101111(21)10058, (01)00884838101111(21)10059, (01)00884838101111(21)10060, (01)00884838101111(21)10061, (01)00884838101111(21)10062, (01)00884838101111(21)10063, (01)00884838101111(21)10065, (01)00884838101111(21)10066, (01)00884838101111(21)10067, (01)00884838101111(21)10068, (01)00884838101111(21)10069, (01)00884838101111(21)10070, (01)00884838101111(21)10071, (01)00884838101111(21)10072, (01)00884838101111(21)10074, (01)00884838101111(21)10075, (01)00884838101111(21)10076, (01)00884838101111(21)10077, (01)00884838101111(21)10078, (01)00884838101111(21)10079, (01)00884838101111(21)10080, (01)00884838101111(21)10081, (01)00884838101111(21)10082, (01)00884838101111(21)10084, (01)00884838101111(21)10085, (01)00884838101111(21)10086, (01)00884838101111(21)10087, (01)00884838101111(21)10088, (01)00884838101111(21)10089, (01)00884838101111(21)10090, (01)00884838101111(21)10091, (01)00884838101111(21)10092, (01)00884838101111(21)10093, (01)00884838101111(21)10094, (01)00884838101111(21)10095, (01)00884838101111(21)10096, (01)00884838101111(21)10097, (01)00884838101111(21)10098, (01)00884838101111(21)10099, (01)00884838101111(21)10100, (01)00884838101111(21)10101, (01)00884838101111(21)10102, (01)00884838101111(21)10103, (01)00884838101111(21)10104, (01)00884838101111(21)10105, (01)00884838101111(21)10106, (01)00884838101111(21)10107, (01)00884838101111(21)10108, (01)00884838101111(21)10109, (01)00884838101111(21)10110, (01)00884838101111(21)10111, (01)00884838101111(21)10112, (01)00884838101111(21)10113, (01)00884838101111(21)10114, (01)00884838101111(21)10115, (01)00884838101111(21)10116, (01)00884838101111(21)10117, (01)00884838101111(21)10118, (01)00884838101111(21)10119, (01)00884838101111(21)10120, (01)00884838101111(21)10121, (01)00884838101111(21)10122, (01)00884838101111(21)10123, (01)00884838101111(21)10124, (01)00884838101111(21)10125, (01)00884838101111(21)10126, (01)00884838101111(21)10127, (01)00884838101111(21)10128, (01)00884838101111(21)10129, (01)00884838101111(21)10130, (01)00884838101111(21)10131, (01)00884838101111(21)10132, (01)00884838101111(21)10133, (01)00884838101111(21)10134, (01)00884838101111(21)10135, (01)00884838101111(21)10136, (01)00884838101111(21)10137, (01)00884838101111(21)10138, (01)00884838101111(21)10139, (01)00884838101111(21)10140, (01)00884838101111(21)10141, (01)00884838101111(21)10142, (01)00884838101111(21)10143, (01)00884838101111(21)10144, (01)00884838101111(21)10145, (01)00884838101111(21)10146, (01)00884838101111(21)10147, (01)00884838101111(21)10148, (01)00884838101111(21)10149, (01)00884838101111(21)10150, (01)00884838101111(21)10151, (01)00884838101111(21)10152, (01)00884838101111(21)10153, (01)00884838101111(21)10154, (01)00884838101111(21)10155, (01)00884838101111(21)10156, (01)00884838101111(21)10157, (01)00884838101111(21)10158, (01)00884838101111(21)10159, (01)00884838101111(21)10160, (01)00884838101111(21)10161, (01)00884838101111(21)10162, (01)00884838101111(21)10163, (01)00884838101111(21)10164, (01)00884838101111(21)10165, (01)00884838101111(21)10166, (01)00884838101111(21)10167, (01)00884838101111(21)10168, (01)00884838101111(21)10169, (01)00884838101111(21)10170, (01)00884838101111(21)10171, (01)00884838101111(21)10172, (01)00884838101111(21)10173, (01)00884838101111(21)10174, (01)00884838101111(21)10175, (01)00884838101111(21)10176, (01)00884838101111(21)10177, (01)00884838101111(21)10178, (01)00884838101111(21)10179, (01)00884838101111(21)10180, (01)00884838101111(21)10181, (01)00884838101111(21)10182, (01)00884838101111(21)10183, (01)00884838101111(21)10184, (01)00884838101111(21)10185, (01)00884838101111(21)10186, (01)00884838101111(21)10187, (01)00884838101111(21)10188, (01)00884838101111(21)10189, (01)00884838101111(21)10190, (01)00884838101111(21)10191, (01)00884838101111(21)10192, (01)00884838101111(21)10193, (01)00884838101111(21)10194, (01)00884838101111(21)10195, (01)00884838101111(21)10196, (01)00884838101111(21)10197, (01)00884838101111(21)10198, (01)00884838101111(21)10199, (01)00884838101111(21)10200, (01)00884838101111(21)10201, (01)00884838101111(21)10202, (01)00884838101111(21)10203, (01)00884838101111(21)10204, (01)00884838101111(21)10205, (01)00884838101111(21)10206, (01)00884838101111(21)10207, (01)00884838101111(21)10208, (01)00884838101111(21)10209, (01)00884838101111(21)10210, (01)00884838101111(21)10211, (01)00884838101111(21)10212, (01)00884838101111(21)10213, (01)00884838101111(21)10214, (01)00884838101111(21)10215, (01)00884838101111(21)10217, (01)00884838101111(21)10218, (01)00884838101111(21)10219, (01)00884838101111(21)10220, (01)00884838101111(21)10221, (01)00884838101111(21)10222, (01)00884838101111(21)10223, (01)00884838101111(21)10224, (01)00884838101111(21)10225, (01)00884838101111(21)10226, (01)00884838101111(21)10227, (01)00884838101111(21)10228, (01)00884838101111(21)10229, (01)00884838101111(21)10230, (01)00884838101111(21)10231, (01)00884838101111(21)10232, (01)00884838101111(21)10233, (01)00884838101111(21)10234, (01)00884838101111(21)10235, (01)00884838101111(21)10236, (01)00884838101111(21)10237, (01)00884838101111(21)10238, (01)00884838101111(21)10239, (01)00884838101111(21)10240, (01)00884838101111(21)10241, (01)00884838101111(21)10242, (01)00884838101111(21)10243, (01)00884838101111(21)10244, (01)00884838101111(21)10245, (01)00884838101111(21)10246, (01)00884838101111(21)10247, (01)00884838101111(21)10248, (01)00884838101111(21)10249, (01)00884838101111(21)10250, (01)00884838101111(21)10251, (01)00884838101111(21)10252, (01)00884838101111(21)10253, (01)00884838101111(21)10254, (01)00884838101111(21)10255, (01)00884838101111(21)10256, (01)00884838101111(21)10257, (01)00884838101111(21)10258, (01)00884838101111(21)10259, (01)00884838101111(21)10260, (01)00884838101111(21)10261, (01)00884838101111(21)10262, (01)00884838101111(21)10263, (01)00884838101111(21)10264, (01)00884838101111(21)10265, (01)00884838101111(21)10266, (01)00884838101111(21)10267, (01)00884838101111(21)10268, (01)00884838101111(21)10269, (01)00884838101111(21)10270, (01)00884838101111(21)10271, (01)00884838101111(21)10272, (01)00884838101111(21)10273, (01)00884838101111(21)10274, (01)00884838101111(21)10275, (01)00884838101111(21)10276, (01)00884838101111(21)10277, (01)00884838101111(21)10278, (01)00884838101111(21)10279, (01)00884838101111(21)10280, (01)00884838101111(21)10281, (01)00884838101111(21)10284, (01)00884838101111(21)10285, (01)00884838101111(21)10286, (01)00884838101111(21)10287, (01)00884838101111(21)10288, (01)00884838101111(21)10289, (01)00884838101111(21)10290, (01)00884838101111(21)10291, (01)00884838101111(21)10292, (01)00884838101111(21)10293, (01)00884838101111(21)10294, (01)00884838101111(21)10298, (01)00884838101111(21)10299, (01)00884838101111(21)10300, (01)00884838101111(21)10301, (01)00884838101111(21)12000, (01)00884838101111(21)12001, (01)00884838101111(21)12002, (01)00884838101111(21)12003, (01)00884838101111(21)12004, (01)00884838101111(21)12005, (01)00884838101111(21)12006, (01)00884838101111(21)12007, (01)00884838101111(21)12008, (01)00884838101111(21)12009, (01)00884838101111(21)396001, (01)00884838101111(21)396003, (01)00884838101111(21)396004, (01)00884838101111(21)396005, (01)00884838101111(21)396006, (01)00884838101111(21)396007, (01)00884838101111(21)396008, (01)00884838101111(21)396009, (01)00884838101111(21)396010, (01)00884838101111(21)396011; Serial No. 10013, 10014, 10015, 10016, 10017, 10018, 10019, 10020, 10021, 10022, 10023, 10024, 10025, 10027, 10031, 10032, 10033, 10034, 10035, 10036, 10038, 10039, 10040, 10041, 10042, 10043, 10044, 10045, 10046, 10047, 10048, 10049, 10050, 10052, 10053, 10055, 10056, 10057, 10058, 10059, 10060, 10061, 10062, 10063, 10065, 10066, 10067, 10068, 10069, 10070, 10071, 10072, 10074, 10075, 10076, 10077, 10078, 10079, 10080, 10081, 10082, 10084, 10085, 10086, 10087, 10088, 10089, 10090, 10091, 10092, 10093, 10094, 10095, 10096, 10097, 10098, 10099, 10100, 10101, 10102, 10103, 10104, 10105, 10106, 10107, 10108, 10109, 10110, 10111, 10112, 10113, 10114, 10115, 10116, 10117, 10118, 10119, 10120, 10121, 10122, 10123, 10124, 10125, 10126, 10127, 10128, 10129, 10130, 10131, 10132, 10133, 10134, 10135, 10136, 10137, 10138, 10139, 10140, 10141, 10142, 10143, 10144, 10145, 10146, 10147, 10148, 10149, 10150, 10151, 10152, 10153, 10154, 10155, 10156, 10157, 10158, 10159, 10160, 10161, 10162, 10163, 10164, 10165, 10166, 10167, 10168, 10169, 10170, 10171, 10172, 10173, 10174, 10175, 10176, 10177, 10178, 10179, 10180, 10181, 10182, 10183, 10184, 10185, 10186, 10187, 10188, 10189, 10190, 10191, 10192, 10193, 10194, 10195, 10196, 10197, 10198, 10199, 10200, 10201, 10202, 10203, 10204, 10205, 10206, 10207, 10208, 10209, 10210, 10211, 10212, 10213, 10214, 10215, 10217, 10218, 10219, 10220, 10221, 10222, 10223, 10224, 10225, 10226, 10227, 10228, 10229, 10230, 10231, 10232, 10233, 10234, 10235, 10236, 10237, 10238, 10239, 10240, 10241, 10242, 10243, 10244, 10245, 10246, 10247, 10248, 10249, 10250, 10251, 10252, 10253, 10254, 10255, 10256, 10257, 10258, 10259, 10260, 10261, 10262, 10263, 10264, 10265, 10266, 10267, 10268, 10269, 10270, 10271, 10272, 10273, 10274, 10275, 10276, 10277, 10278, 10279, 10280, 10281, 10284, 10285, 10286, 10287, 10288, 10289, 10290, 10291, 10292, 10293, 10294, 10298, 10299, 10300, 10301, 12000, 12001, 12002, 12003, 12004, 12005, 12006, 12007, 12008, 12009, 396001, 396003, 396004, 396005, 396006, 396007, 396008, 396009, 396010, 396011.
Distribution:
Distributed in: AZ, CA, CO, FL, GA, HI, IN, KY, MA, MD, MN, NY, OH, OR, PA, TX, VT, WA, WV
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2034-2025
Related Recalls
Philips SmartPath to dStream for 3.0T with MR Elastography (MRE). 1. Model Number (REF): 782145.
Philips North America
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Philips SmartPath to Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 782118. 2. Model Number (REF): 782144. 3. Model Number (REF): 782163.
Philips North America
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.