Nonconforming devices while manufactured correctly, do not comply to their design requirements. Potential harms associated with balloon detachment are endoscopic retrieval of an impacted or non-impacted object, bleeding caused by the retrieval device, or the incident can occur without injury and the object is left to pass naturally.
Nonconforming devices while manufactured correctly, do not comply to their design requirements. Potential harms associated with balloon detachment are endoscopic retrieval of an impacted or non-impacted object, bleeding caused by the retrieval device, or the incident can occur without injury and the object is left to pass naturally.
Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.
Defect in the outer packaging of the Cranial Access Kits can cause the packaging to split without any additional forces outside compromising the sterility of the device
Product was re-processed and re-sterilized by a third party using packaging configurations and a sterilization process which are not approved by the manufacture and lack validation.
Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.
Nonconforming devices while manufactured correctly, do not comply to their design requirements. Potential harms associated with balloon detachment are endoscopic retrieval of an impacted or non-impacted object, bleeding caused by the retrieval device, or the incident can occur without injury and the object is left to pass naturally.
Product was re-processed and re-sterilized by a third party using packaging configurations and a sterilization process which are not approved by the manufacture and lack validation.
Nonconforming devices while manufactured correctly, do not comply to their design requirements. Potential harms associated with balloon detachment are endoscopic retrieval of an impacted or non-impacted object, bleeding caused by the retrieval device, or the incident can occur without injury and the object is left to pass naturally.
The instruction manuals that were included with the Multi-Mode Stimulators distributed from 2018 to 2023 did not contain all of the appropriate warnings and contraindications.
Routine quality control testing of affected blood culture identification panel with affected control panel may need to be amended, due to control panel/instrument characteristic changes, that lead C. tropicalis to be amplified in some portion of test runs, but then it's reported as "Not Detected" because the assay amplicon Tm is outside the acceptable limit assigned for calling positive results.
Nonconforming devices while manufactured correctly, do not comply to their design requirements. Potential harms associated with balloon detachment are endoscopic retrieval of an impacted or non-impacted object, bleeding caused by the retrieval device, or the incident can occur without injury and the object is left to pass naturally.