BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, used with Maine Molecular Quality Controls, Inc (MMQCI) FilmArray BCID2 Control Panel M416, which is referenced in Quality Control Materials for use with BIOFIRE FILMARRAY and BIOFIRE SPOTFIRE Panels - FLM1-PRT-0251
Class I - DangerousWhat Should You Do?
- Check if you have this product: REF/UDI-DI: RFIT-ASY-0147/00815381020338. FLM1-PRT-0251-16 and prior.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- BioFire Diagnostics, LLC
- Reason for Recall:
- Routine quality control testing of affected blood culture identification panel with affected control panel may need to be amended, due to control panel/instrument characteristic changes, that lead C. tropicalis to be amplified in some portion of test runs, but then it's reported as "Not Detected" because the assay amplicon Tm is outside the acceptable limit assigned for calling positive results.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, used with Maine Molecular Quality Controls, Inc (MMQCI) FilmArray BCID2 Control Panel M416, which is referenced in Quality Control Materials for use with BIOFIRE FILMARRAY and BIOFIRE SPOTFIRE Panels - FLM1-PRT-0251
Product Codes/Lot Numbers:
REF/UDI-DI: RFIT-ASY-0147/00815381020338. FLM1-PRT-0251-16 and prior.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0451-2024
Related Recalls
BioFire Joint Infection (JI) Panel REF: RFIT-ASY-0138, 30 test Kit.
BioFire Diagnostics
Contamination to in-vitro diagnostic test may result in false positives.
BioFire Respiratory Panel 2.1, IVD, REF: 423742 (30 test kit)
BioFire Diagnostics
Due to unintended movement of pouch chemistry, nucleic acid test may result in elevated risk of false negative results and risk of control failures.
BioFire Respiratory Panel 2.1 plus Kit, REF: 423740, For Filmarray systems
BioFire Diagnostics
Due to manufacturing error, respiratory panel may result in elevated control failures and/or false negative results.