🏨

Other Medical Devices

🏥 Medical Devices • 15,809 recalls

GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.

Nov 5, 2024 Other Medical Devices View Details →

GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.

Nov 5, 2024 Other Medical Devices View Details →

GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.

Nov 5, 2024 Other Medical Devices View Details →

MEDLINE convenience kits labeled as: 1) BASIC NEURO PACK, REF CDS780202P; 2) BASIC NEURO PACK, REF CDS780202Q; 3) BASIC NEURO PACK, REF CDS780202R; 4) ACDF CDS, REF CDS941061M; 5) NEURO KIT, REF CDS981327Q; 6) ZALE V-P SHUNT CDS, REF CDS983186J; 7) NEURO PACK-LF, REF DYNJ0038440C; 8) NEURO PACK-LF, REF DYNJ0406347M; 9) NEURO-LF, REF DYNJ0484186M; 10) NEURO-LF, REF DYNJ0484186N; 11) NEURO VP SHUNT PACK-LF, REF DYNJ0843121K; 12) NEURO VP SHUNT/VAGAL NERVE-RF, REF DYNJ21925S; 13) NEURO-ORTHO EXTRAS, REF DYNJ44882L; 14) NEURO PACK, REF DYNJ57785B; 15) NEURO PACK, REF DYNJ60825B; 16) CHRISTUS MC NEURO PACK, REF DYNJ61157C; 17) CHRISTUS CHILDRENS NEURO PK, REF DYNJ61168C; 18) CHRISTUS CHILDRENS VP SHUNT PK, REF DYNJ61174C; 19) AFCH SHUNT, REF DYNJ63317C; 20) NEURO SPINE PK, REF DYNJ63319D; 21) NEURO CAROTID ENDARTERECTO, REF DYNJ63321D; 22) DBS PACK, REF DYNJ63322A; 23) NEURO LAM, REF DYNJ64188B; 24) PACK,V-P SHUNT, REF DYNJ65472B; 25) NEURO PK, REF DYNJ66735A; 26) DR. FONTES ADD-ON PK-LF, REF DYNJ66766B; 27) NEURO PACK, REF DYNJ68558A; 28) NEURO PACK, REF DYNJ69019C; 29) VP SHUNT, REF DYNJ81866A; 30) VP SHUNT, REF DYNJ81866B; 31) VP SHUNT, REF DYNJ81866C; 32) NEURO PACK, REF DYNJ84058A; 33) NEURO, REF DYNJ900420O; 34) CRANIO / NEURO / SPINE, REF DYNJ901072G; 35) CRANIO / NEURO / SPINE, REF DYNJ901072I; 36) LAMINECTOMY, REF DYNJ903870F; 37) MINOR NEURO, REF DYNJ904859C; 38) MINOR NEURO, REF DYNJ904859D; 39) MINOR NEURO, REF DYNJ904859F; 40) NEURO, REF DYNJ904861B; 41) NEURO, REF DYNJ904861D; 42) NEURO, REF DYNJ904861F; 43) NEURO, REF DYNJ904861G; 44) KIT V P SHUNT, REF DYNJ908835A; 45) GENERAL NEURO, REF DYNJ909744

MEDLINE INDUSTRIES, LP - Northfield

Class I - Dangerous

Medline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker.

Oct 9, 2024 Other Medical Devices Nationwide View Details →

GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.

Nov 5, 2024 Other Medical Devices View Details →
Class I - Dangerous

Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit.

Aug 8, 2024 Other Medical Devices View Details →

Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit.

Aug 8, 2024 Other Medical Devices View Details →

Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit.

Aug 8, 2024 Other Medical Devices View Details →

GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.

Nov 5, 2024 Other Medical Devices View Details →

GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.

Nov 5, 2024 Other Medical Devices View Details →
Class I - Dangerous

Potential that the IMT Diluent volume remaining (% remaining) does not decrease as expected on the Atellica CI Analyzer, potentially leading to the IMT Diluent being empty while still displaying that volume is remaining. In this case Sodium (Na), Potassium (K) and Chloride (Cl) test results may be falsely elevated. Quality Control materials demonstrate the same behavior.

Nov 4, 2024 Other Medical Devices Nationwide View Details →

BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.

Oct 31, 2024 Other Medical Devices Nationwide View Details →

MEDLINE convenience kits labeled as: 1) GYN LAPAROSCOPY-LF, REF CDS760059I; 2) GYN LAPAROSCOPY CDS #34-RF, REF CDS920099S; 3) GYN MAJOR, REF CDS980998L; 4) ROBOTIC PROCEDURE, REF CDS982128P; 5) ROBOTIC, REF CDS984262N; 6) GYN LAPAROSCOPY, REF CDS984900M; 7) GYN LAPAROSCOPY, REF CDS984900N; 8) DAVINCI PACK-LF, REF DYNJ0101304G; 9) PROSTATE ROBOTIC PACK, REF DYNJ0347372J; 10) URO GYN PACK-LF, REF DYNJ17454N; 11) DB DAVINCI ROBOT PACK-LF, REF DYNJ33592I; 12) ROBOTIC UROLOGY-RF, REF DYNJ38844N; 13) TLH PACK, REF DYNJ42792D; 14) SVMMC GEN ROBOTIC PACK, REF DYNJ43770L; 15) SVMMC ROBOTIC KIDNEY PACK, REF DYNJ43778L; 16) SVMMC GYN ROBOTIC PACK, REF DYNJ43782K; 17) SVMMC ROBOTIC PROSTATE PACK, REF DYNJ43862J; 18) DA VINCI HYSTERECTOMY, REF DYNJ44863O; 19) DA VINCI PROSTATE/COLPOPEXY, REF DYNJ44864S; 20) LAP-GYN PACK, REF DYNJ44876L; 21) LAP-GYN PACK, REF DYNJ44876M; 22) ROBOT PACK, REF DYNJ45184F; 23) GYN ROBOTIC PACK, REF DYNJ47766Q; 24) ST. ANNE'S MULTI PORT PACK, REF DYNJ48638G; 25) RBTC COLON/KIDNEY PK RFID-LF, REF DYNJ52397M; 26) GYN LAP PACK, REF DYNJ52524C; 27) ROBOTIC PACK, REF DYNJ53650D; 28) LAPAROTOMY PACK, REF DYNJ59079D; 29) ROBOTIC GYN PACK, REF DYNJ62107B; 30) GYN UROL BASIC LAPAROSCOPY, REF DYNJ63326D; 31) FAIRFAX DAVINCI TRAY, REF DYNJ63606C; 32) GYN LAP PACK, REF DYNJ64234B; 33) ROBOT GYN SHARED, REF DYNJ64793A; 34) ROBOT GYN SHARED, REF DYNJ64793B; 35) ROBOT GYN SHARED, REF DYNJ64793C; 36) GYN DAVINCI PACK, REF DYNJ65373A; 37) ROBOT PACK, REF DYNJ67190C; 38) TUBAL LIGATION PACK, REF DYNJ68563A; 39) TUBAL LIGATION PACK, REF DYNJ68563B; 40) DAVINCI GYN PACK, REF DYNJ69169A; 41) GYN LAPAROSCOPY PACK, REF DYNJ69490C; 42) ROBOTIC PACK, REF DYNJ84776A; 43) ROBOTIC LAVH, REF DYNJ901281G; 44) ROBOTIC LAP CHOLE, REF DYNJ901282J; 45) GYN LAPAROTOMY, REF DYNJ902498J; 46) WMC ROBOTIC BASE-LF, REF DYNJ902583M; 47) HHI ROBOTIC CDS, REF DYNJ902953G; 48) HHI ROBOTIC CDS, REF DYNJ902953I; 49) HHI ROBOTIC CDS, REF DYNJ902953J; 50) HHI ROBOTIC CDS, REF DYNJ902953K; 51) ROBOTIC GYN & GU, REF DYNJ903459F; 52) SM DAVINCI, REF DYNJ903539G; 53) WEST GYN LAPAROSCOPY, REF DYNJ904490I; 54) ROBOTIC GYN, REF DYNJ904792C; 55) ROBOTIC GENERAL, REF DYNJ904793B; 56) ROBOT, REF DYNJ906112B; 57) ROBOTIC URO LH, REF DYNJ906445A; 58) ROBOTIC URO LH, REF DYNJ906445B; 59) ROBOTIC URO LH, REF DYNJ906445C; 60) WMC THORACIC ROBOTIC, REF DYNJ907153C; 61) DAVINCI, REF DYNJ907392A; 62) DAVINCI, REF DYNJ907392B; 63) DAVINCI ROBOTICS-HERNIA, REF DYNJ907396B; 64) DAVINCI ROBOTICS-HERNIA, REF DYNJ907396C; 65) GYN LAPAROSCOPY - NTSC, REF DYNJ907407A; 66) GYN LAPAROSCOPY - NTSC, REF DYNJ907407B; 67) ROBOTIC-LF, REF DYNJ907910; 68) ROBOTIC-LF, REF DYNJ907910A; 69) GU DAVINCI KIT, REF DYNJ908783; 70) GU DAVINCI KIT, REF DYNJ908783A; 71) GU DAVINCI PACK - LF, REF DYNJ909398A; 72) GYN LAPAROSCOPY, REF DYNJ910084; 73) MHPB GYN ROBOTIC PACK, REF DYNJ910125;

MEDLINE INDUSTRIES, LP - Northfield

Class I - Dangerous

Medline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker.

Oct 9, 2024 Other Medical Devices Nationwide View Details →