One unit from a demo lot of POWERFLEX PRO PTA Dilatation Catheter was left at a hospital and was not sterilized.

Dec 22, 2014 Implants & Prosthetics Nationwide View Details β†’

Complaints have indicated that the ATTUNE INTUITION Impaction Handle lever has fractured during impaction. If the lever fractures, there is the potential for fractured pieces of the device, or the inner spring, if released, to be left in the patient if not observed during surgery.

Nov 10, 2014 Implants & Prosthetics Nationwide View Details β†’

Difficulty locking EXp tibial inserts into tibial trays, caused by a manufacturing condition where a lead in radius on the anterior locking tab is not present on certain insert lots. The absence of this lead in may prevent or make assembly of the insert and tray difficult.

Dec 1, 2014 Implants & Prosthetics View Details β†’

ATTUNE INTUITION Impactors (cat. nos. 254401003, 254401004, and 254401006) have broken and produced small pieces. Should a fracture occur and not be observed during surgery, there is the potential for these small fractured pieces of the instrument to be left in the patient.

Nov 10, 2014 Implants & Prosthetics Nationwide View Details β†’

ATTUNE INTUITION Impactors (cat. nos. 254401003, 254401004, and 254401006) have broken and produced small pieces. Should a fracture occur and not be observed during surgery, there is the potential for these small fractured pieces of the instrument to be left in the patient.

Nov 10, 2014 Implants & Prosthetics Nationwide View Details β†’

Inoperative screw head seat splay and/or fracture as a result of positioning of the rod which is a part of the Cypher MIS Screw System.

Oct 8, 2014 Implants & Prosthetics View Details β†’

ATTUNE INTUITION Impactors (cat. nos. 254401003, 254401004, and 254401006) have broken and produced small pieces. Should a fracture occur and not be observed during surgery, there is the potential for these small fractured pieces of the instrument to be left in the patient.

Nov 10, 2014 Implants & Prosthetics Nationwide View Details β†’

Samples of product from Supplier lot 14D0809 and 14D0808 showed little adhesive transfer between the Tyvek lid and cavity when opened for internal bone cement cure and pairing tests. Visually these samples do not meet the requirements of ZWI 43.109 and corresponding visual aid 55-0000-310-01.

Dec 8, 2014 Implants & Prosthetics View Details β†’

Two production lots of the Protege Rx Tapered Carotid Stent Systems (SECX-10-7-40-135, Lot # 9922452 & SEPX-8-6-40-135, Lot # 9922795) because of a product labeling error.

Dec 4, 2014 Implants & Prosthetics View Details β†’

The Set Up Joint subassembly of the Sensei X Robotic Catheter System may become loose relative to the operating table with the potential for the device to fall causing, uncontrolled catheter movement. cardiac tampenade, delayed procedure, or crush injury to the patient and/or user.

Dec 4, 2014 Implants & Prosthetics Nationwide View Details β†’

Two production lots of the Protege Rx Tapered Carotid Stent Systems (SECX-10-7-40-135, Lot # 9922452 & SEPX-8-6-40-135, Lot # 9922795) because of a product labeling error.

Dec 4, 2014 Implants & Prosthetics View Details β†’

The affected lot was supplied by contract manufacturer and are potentially subject to disassociation of the magnet cover. Zimmer received one complaint indicating the magnet cover of a Persona Cemented Tibial Broach Inserter/Extractor Handle lot 62085420 was missing after broaching the tibia. Postoperative radiographs were inspected and the magnet cover was not identified. The investigation indi

Dec 4, 2014 Implants & Prosthetics View Details β†’

Zimmer received a single complaint wherein a Trilogy® Fiber Metal Multi-Hole Acetabular Shell would not attach to the inserter/impactor instrumentation. Subsequent investigation found the polar-hole threads to be present but non-conforming to specifications, in that the minor diameter was undersized. Investigation also indicates that the issue is isolated and affects only lot 62714305, which

Dec 4, 2014 Implants & Prosthetics Nationwide View Details β†’

Product not approved for use in the US

Aug 11, 2014 Implants & Prosthetics Nationwide View Details β†’

The recalled product was distributed with an incorrect serial number printed on the device's label.

Nov 20, 2014 Implants & Prosthetics View Details β†’