Rotating Tibial Platform ATTUNE INTUITION Impactor. The ATTUNE INTUITION Impactors are re-useable instruments utilized in knee replacement procedures. The three affected impactors are used at various times within the surgical flow, each impactor is used to impact and drive home its relevant mating part: Fixed Tibial bearing, (254401003) Rotating Tibial Platform, (254401004) and Femoral impactor, (254401006).
Class I - DangerousWhat Should You Do?
- Check if you have this product: CATALOG NO. : ATTUNE INTUITION Impactors Rotating Tibial Platform, (254401004) Barcode GTIN Number: Rotating Tibial Platform, ......... 10603295130208 ALL LOTS
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- DePuy Orthopaedics, Inc.
- Reason for Recall:
- ATTUNE INTUITION Impactors (cat. nos. 254401003, 254401004, and 254401006) have broken and produced small pieces. Should a fracture occur and not be observed during surgery, there is the potential for these small fractured pieces of the instrument to be left in the patient.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Rotating Tibial Platform ATTUNE INTUITION Impactor. The ATTUNE INTUITION Impactors are re-useable instruments utilized in knee replacement procedures. The three affected impactors are used at various times within the surgical flow, each impactor is used to impact and drive home its relevant mating part: Fixed Tibial bearing, (254401003) Rotating Tibial Platform, (254401004) and Femoral impactor, (254401006).
Product Codes/Lot Numbers:
CATALOG NO. : ATTUNE INTUITION Impactors Rotating Tibial Platform, (254401004) Barcode GTIN Number: Rotating Tibial Platform, ......... 10603295130208 ALL LOTS
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0904-2015
Related Recalls
ATTUNE AFFIXIUM Cementless Fixed Bearing Knee with 3DP Technology, Size 6. Component of the ATTUNE Total Knee Replacement System.
DePuy Orthopaedics
Product incorrectly labelled.
Attune Revision Limb Preservation System (LPS) Tibial Insert XXSM 18MM, Part Number 151760118
DePuy Orthopaedics
The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to the implant material properties.
Attune Revision Limb Preservation System (LPS) Tibial Insert SM 12MM, Part Number 151760312
DePuy Orthopaedics
The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to the implant material properties.