Evaluation handsets may not be able to communicate with the neurostimulator due to handsets having a faster processing speed than prior handsets. An inability to connect may cause the therapy to stop being delivered and a system error will be displayed. Patient may experience return of underlying disease symptoms (e.g. overactive bladder, urinary retention, or fecal incontinence) based on their response to stimulation during the evaluation period.
Medtronic Neuromodulation
There is a software issue that can permanently disable communication with an implantable neurostimulator (INS) when a specific intraoperative programming sequence is used.
Product complaints were received describing the A820 myPTM app taking longer than expected for patients to interact with their implantable pump.
Software issues were identified in application version 2.x.
Software issues were identified in application version 2.x.
Software issues were identified in application version 2.x.
Software issues were identified in application version 2.x.
The Wireless Rechargers in the kits may be unable to enter a recharging session upon first use.
Potential for error and inability to interrogate pump due to data corruption in the pump memory.
A limited number of Percept RC Implantable Neurostimulators (INS), Model B35300 will not be able to communicate with the A610 DBS Clinician Programmer Application upon use and an "Invalid Device" message will appear.
Model 977006 Vanta INS batterty life has depleted faster than HCPs or patients expected leading to outpatient reprogramming or earlier than anticipated surgical replacements.
A subset of the handsets within the kit may not be able to complete the pairing process with the communicator upon initial setup.
Patients implanted with a pocket adaptor (Model 64001 and/or 64002) are limited to HEAD ONLY MRI eligibility.
Design update to the Ascenda Intrathecal Catheter Models 8780, 8781, and 8784. The intent of the design update is to reduce the potential for tissue growth into the Ascenda catheter connector which may potentially lead to catheter occlusion.
inability to reprogram one device
Medtronic has discovered an instance in which the Intellis A710 Clinician Programmer Application (CP App) could no longer establish a session with the Implantable Neurostimulator (INS) which resulted in the inability for the clinician to adjust stimulation parameters. In the event of this issue, the patient may receive inadequate or excessive therapy and experience non-optimal pain relief.
Medtronic recently identified that if the SynchroMed II pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will not resume drug delivery after leaving the MRI magnetic field, which is inconsistent with the current labeling.
Medtronic recently identified that if the SynchroMed II pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will not resume drug delivery after leaving the MRI magnetic field, which is inconsistent with the current labeling.
Units distributed in Korea and Turkey that did not have the correct firmware installed to support the user interface in the local language.
There are a small number of units of the DBS Lead with StimLoc kit (3387S-40) due to a discrepancy in the Use By Date (UBD) printed on the outer packaged kit label versus the UBD printed on the individual product label.
Medtronic is requesting return of a small number of units of the SenSight Connector Plug, Model B31061, due to a discrepancy in the Use By Date (UBD) printed on the label.
The original version of the A71100 Restore Clinician Programmer Application has been identified to have a compatibility issue with some legacy clinician programmer software resulting in the programmer having an inability to establish communication with the implanted neurostimulators.
Lot of InterStim Self-Adhesive Ground Pad is being recalled due to the product being labeled with the incorrect expiration date.
When an update to the A710 Clinician Programmer software application for the Intellis Neurostimulation System was released, the corresponding version of the A901 Communication Manager software application was not available for download.
A cardioversion procedure may damage the electronics in the Vanta INS device, making the INS unresponsive and non-functional. As of April 19, 2023, Medtronic has received a total of two complaints concerning this issue from patients implanted with a Vanta INS, both of which have resulted in explant procedures.
Medtronic has received reports of SenSight Extension Tunneler kits that contain dual carriers with have been machined on only one side. As a result, the dual carrier cannot be used to pass two extensions simultaneously.
Software anomalies, CP App messages: "Too Many Device Found", "Unexpected Device Error Code 1502", and "System Update Needed, Service Code 303"
Software anomalies, CP App messages: "Too Many Device Found", "Unexpected Device Error Code 1502", and "System Update Needed, Service Code 303"
Software anomalies, CP App messages: "Too Many Device Found", "Unexpected Device Error Code 1502", and "System Update Needed, Service Code 303"
Software anomalies, CP App messages: "Too Many Device Found", "Unexpected Device Error Code 1502", and "System Update Needed, Service Code 303"
Software anomalies, CP App messages: "Too Many Device Found", "Unexpected Device Error Code 1502", and "System Update Needed, Service Code 303"
Potential communication issue in which the Vanta Clinician Programmer Application (CP App) A71200 v2.0.2455 may be unable to connect with a Vanta Implantable Neurostimulator (INS) Model 977006 when a patient is implanted with more than one neurostimulator.
Potential for Vanta Clinician Programmer Application (CP App) A71200 v2.0.2455 to display an Unexpected Device Error Code 1502 Message, and the user will be unable to perform programming of the Vanta Implantable Neurostimulator (INS) Model 977006.
The Implantable Neurostimulator Model may be susceptible to not communicating with the clinical programmer and/or patient programmer system.
The Implantable Neurostimulator (INS) cannot communicate with the clinical programmer and/or the patient programmer system.
The Implantable Neurostimulator (INS) cannot communicate with the clinical programmer and/or the patient programmer system.
There is a software anomaly with the InterStim X Clinician software application with results in the data entered in the Patient Information fields not e retained and a " Data Lost" notification being displayed.
During cardioversion events, the Model B35200 Percept PC Implantable Neurostimulator could become unresponsive and non-functional. Cardioversion is a procedure performed by a cardiologist to correct an abnormal heartbeat. Medtronic has determined that the cardioversion may damage the circuit electronics in the Percept PC stimulator, thereby, causing the Percept PC stimulator to become non-responsive.
The wireless rechargers (WR; WR9200 and WR9220; included in the Recharger Kits) become unresponsive if the charging steps, indicated in the Recharger Patient User Guide, are not followed.
The wireless rechargers (WR; WR9200 and WR9220; included in the Recharger Kits) become unresponsive if the charging steps, indicated in the Recharger Patient User Guide, are not followed.
The wireless rechargers (WR; WR9200 and WR9220; included in the Recharger Kits) become unresponsive if the charging steps, indicated in the Recharger Patient User Guide, are not followed.
A software anomaly may occur with the clinician programmer application.
Excess nylon flash on the clip.
Specific PINs of the Vectris SureScan MRI lead kits contain the incorrect lead electrode spacing information printed on the shelf box and the sterile pack labeling.
There is a potential for unintended heating that can occur with the Intellis Model 97755 Recharger.
There are two issues within this correction action: 1. Cycling Issue: When copying settings from a previously used Activa device to a new Percept PC Implantable Neurostimulator (INS) using the "replacement" function of the A610 clinician programmer application, the cycling feature is not copied properly. 2. Log Fill Issue: When the Percept PC INS System Event log is full, the clinician programmer A610 Software application version 2.0.4584, 2.0.4594, 2.0.4605, or 3.0.1057 crashes when it attempts a reading of the last records in the log.
A710 Intellis Clinician Application has a software issue that can result in the inability to program the Intellis implantable neurostimulation device.
A software anomaly in the A620 Patient Programmer application was identified that results in failure to connect with the Percept PC device.
There is a potential for the presence of a foreign particle inside the pump motor assembly which could interfere with the motor gear rotation and lead to a permanent motor stall.
There is a potential for an unexpected increase in stimulation during InterStim programming with the A10 Clinician Application (on Medtronic's smart programmer).
This voluntary recall is being conducted due to the curved tip introducer needle which is included in the kit, to have a potential manufacturing defect. This could result in difficulty advancing or withdrawing the Vectris lead through the curved tip introducer needle.
This voluntary recall is being conducted due to the curved tip introducer needle which is included in the kit, to have a potential manufacturing defect. This could result in difficulty advancing or withdrawing the Vectris lead through the curved tip introducer needle.
Medtronic implemented a final design change intended to address the motor stall due to corrosion and is recalling devices with the previous design that are not implanted.
There is a possibility that specific lots of catheters that have recently been distributed may be at risk for an increased potential for kinking at the proximal end where the catheter connects to the drug infusion pump.
Medtronic received reports that the DBS depth stop did not adequately secure to the lead, which can result in DBS lead placement beyond the intended target.
Medtronic received a complaint that there was an error code displayed on the programmer when the physician attempted to interrogate an implanted SynchroMed II pump. The error code prevented the physician from updating the pump; however the pump was providing therapy.
Medtronic is following up to a May 2013 communication regarding the Priming Bolus function for the SynchroMed Infusion System. Medtronic is updating the Model 8870 software application card (to version AAU01) and the SynchroMed pump labeling to address the priming bolus issue.
Medtronic has identified an increased number of complaints from customers involving reports of Rechargers (Medtronic Model 37751 Recharger, which is included in Models 37754 and 97754 Charging Systems,Spinal Cord Stimulation)that are in an unresponsive error state, where the Recharger is non-functional with a blank display screen and is beeping every 5 seconds. Medtronic has determined all Rechargers manufactured starting in November 2014 (indicated by serial numbers beginning with NKA4 or NKU4) are more susceptible to this error state.
Medtronic has voluntarily decided to retrieve unused Model 3387S-40 and 3389S-40 DBS leads from three manufacturing lot numbers. During the manufacturing process of a DBS lead component, Medtronic identified the potential for lead insulation damage.
Due to the potential for pocket adaptor conductor wire fractures involving DBS Pocket Adaptor Model 64001 and Model 64002, a safety notification was sent to healthcare professionals to reinforce device specific labeling for the handling of pocket adaptors and system integrity checking during implant procedures.
Loss of therapy during recharging of device. Reportedly, a charging session was terminated prior to obtaining recharge threshold voltage of 3.615, which triggered rapid battery depletion state. As a result, the device depleted to 1.925 volts in one to two days rather than the typical 30 days. Insufficient coupling between recharger and device during recharge session was found to be key factor.
Medtronic Neuromodulation is initiating a voluntary removal of a single Model 8781 Ascenda" Intrathecal Catheter kit. We have identified that this kit may have been produced with incorrect outer package labels. As a result, it may not contain all components necessary to complete the implant procedure.
Some Specify 5-6-5 and 2x8 surgical lead product labeling does not adequately distinguish between permanent surgical implant and trialing/screening use.
Medtronic is conducting a recall of a specific subset of Model 8637-20 and 8637-40 SynchroMed II implantable drug pumps because the audible alarm could cease to function.
Medtronic is providing Important information regarding extension handling during implant procedure after receiving reports of high impedances. This applies to extension Models 37085 and 37086 which can be used with the following implantable Neurostimulators: Activa¿ PC (Model 37601), Activa¿ RC (Model 37612), and the Activa¿ SC (Model 37603).
Current labeling indicates that cycling improves device longevity in every programming scenario. However, enabling cycling at certain parameter settings may decrease the device battery longevity. With cycling enabled, longevity estimate displayed on the Model 8840Programmer may be inaccurate.
Medtronic is recalling six Activa PC (model 37601) Implantable Neurostimulators due to the potential for a damaged electrical component during manufacturing.
Medtronic is removing specific lots of the Ascenda Intrathecal Catheters and Revision Kits, which are used with the implantable SynchroMed drug infusion pump. This recall is being conducted due to a single component of the catheter, the retainer ring, not meeting specification criteria. There is a possibility of unintentional disconnection of the catheter from the pump, or difficulty in disconne
Potential for misalignment of the Z-stage scale. Using one of these devices for a procedure could result in the microelectrode being inserted to an incorrect target depth.
Medtronic has determined that an Ascenda catheter kit may have had endotoxin in excess of USP limits.
Some refill kits in the affected lots may contain an extension tubing set that is occluded and healthcare professionals may be unable to aspirate or fill the SynchroMed pumps. There is no problem with the Lioresal Intrathecal (baclofen injection) contained in the affected drug refill kits. There have been no patient injuries beyond additional needle sticks associated with this issue.
This recall provides important new information regarding overinfusion associated with the Medtronic SynchroMed II Implantable Pump. Overinfusion can result in a life-threatening overdose and can also result in drug withdrawal due to premature emptying of the pump. Due to the low reported rate of occurrence of this issue and the inability to predict which pumps may be at risk, Medtronic is not re
Medtronic neurostimulators (used with pain and deep brain stimulation therapy) have a cycling feature that, when enabled, allows stimulation output to be toggled on and off based on the programmed settings. The use of this feature is determined by the physician and set using the clinician programmer. Current labeling indicates the use of cycling improves device longevity and recharge interval
Medtronic neurostimulators (used with pain and deep brain stimulation therapy) have a cycling feature that, when enabled, allows stimulation output to be toggled on and off based on the programmed settings. The use of this feature is determined by the physician and set using the clinician programmer. Current labeling indicates the use of cycling improves device longevity and recharge interval
Medtronic neurostimulators (used with pain and deep brain stimulation therapy) have a cycling feature that, when enabled, allows stimulation output to be toggled on and off based on the programmed settings. The use of this feature is determined by the physician and set using the clinician programmer. Current labeling indicates the use of cycling improves device longevity and recharge interval
Medtronic notified customers regarding Loss of Stimulation of specific Implantable Neuro-Stimulators under the following specific conditions: Switching between groups with Multiple programs or creating program groups. Therapy can be restored with the Physician Recharge mode (RPM) of the Implantable Neurostimulator Recharger (INSR).
Medtronic notified customers regarding Loss of Stimulation of specific Implantable Neuro-Stimulators under the following specific conditions: Switching between groups with Multiple programs or creating program groups. Therapy can be restored with the Physician Recharge mode (RPM) of the Implantable Neurostimulator Recharger (INSR).
Medtronic notified customers regarding over stimulation or stimulation in the wrong area related to select Medtronic neurostimulators. Upon a specific set of conditions, typically related to device recovery from an over discharge, there is a potential for over stimulation or stimulation directed to a lead electrode other than what was intended.
Medtronic notified customers regarding over stimulation or stimulation in the wrong area related to select Medtronic neurostimulators. Upon a specific set of conditions, typically related to device recovery from an over discharge, there is a potential for over stimulation or stimulation directed to a lead electrode other than what was intended.
The Sutureless Connector (SC) Intrathecal Catheter connector has been redesigned to reduce the potential for occlusion at the catheter to pump interface. Medtronic is removing the unused products from the market that were manufactured with the previous design, and recommend the previous design no longer be used due to greater potential for misalignment and subsequent occlusion.
The Sutureless Connector (SC) Intrathecal Catheter connector has been redesigned to reduce the potential for occlusion at the catheter to pump interface. Medtronic is removing the unused products from the market that were manufactured with the previous design, and recommend the previous design no longer be used due to greater potential for misalignment and subsequent occlusion.
Medtronic Neuromodulation is providing Healthcare Providers with safety information and patient management recommendations related to the potential for electrical shorting internal to they SynchroMed infusion pump. An electrical short circuit in a feedthrough may present as a motor stall or low battery reset/alarm and lead to a loss of or reduction in therapy which may result in the return of und
The Sutureless Connector (SC) Intrathecal Catheter connector has been redesigned to reduce the potential for occlusion at the catheter to pump interface. Medtronic is removing the unused products from the market that were manufactured with the previous design, and recommend the previous design no longer be used due to greater potential for misalignment and subsequent occlusion.
Medtronic is providing Healthcare Professionals with important safety information and patient management recommendations regarding the unintended delivery of drug during the priming bolus function for the SynchroMed implantable infusion pump. The unintended delivery of drug can contribute to patient overdose or underdose symptoms which may be clinically relevant.
Medtronic is providing Healthcare Professionals with important safety information and patient management recommendations regarding the unintended delivery of drug during the priming bolus function for the SynchroMed implantable infusion pump. The unintended delivery of drug can contribute to patient overdose or underdose symptoms which may be clinically relevant.
The Sutureless Connector (SC) Intrathecal Catheter connector has been redesigned to reduce the potential for occlusion at the catheter to pump interface. Medtronic is removing the unused products from the market that were manufactured with the previous design, and recommend the previous design no longer be used due to greater potential for misalignment and subsequent occlusion.
Medtronic Neuromodulation is providing Healthcare Providers with safety information and patient management recommendations related to the potential for electrical shorting internal to they SynchroMed infusion pump. An electrical short circuit in a feedthrough may present as a motor stall or low battery reset/alarm and lead to a loss of or reduction in therapy which may result in the return of und
There is a potential for lead damage due to the use of the lead cap provided in DBS and Dystonia kits. Medtronic has received reports of DBS leads being damaged at the connector end of the lead when the lead cap is used. The connector end of the lead is the end of the lead connected to the lead extension. Tightening or loosening of the setscrew may twist the setscrew connector block and may damag
There is a potential for lead damage due to the use of the lead cap provided in DBS and Dystonia kits. Medtronic has received reports of DBS leads being damaged at the connector end of the lead when the lead cap is used. The connector end of the lead is the end of the lead connected to the lead extension. Tightening or loosening of the setscrew may twist the setscrew connector block and may damag
Medtronic has found through device testing that if the optional footswitch is not connected to the Model 8930 RF generator during use, an electrostatic discharge (i.e. static electricity) to the footswitch connector port may toggle the RF output on or off. If this were to occur, it could result in unintentionally starting or stopping the lesion process, depending on the current status of the equi
Medtronic has identified 140 Restore DBS Extension Kits (model numbers 7482, 37085, and 37086), which may contain a damaged winged connector boot. The connector boot may not have been molded properly and may be cracked. Approximately 14% of the 140 affected extension kits may contain a damaged radiopaque winged connector boot.
Medtronic notified Healthcare Professionals of the impact of unapproved drugs on the performance of the SynchroMed infusion pump system. Use of unapproved drugs with SynchroMed pumps can result in an increased risk of permanent motor stall and cessation of drug infusion. UPDATE 2/1/2016: Medtronic received approval to incorporate enhancements to the SynchroMed II pump which decreases the potential for internal electrical shorting and motor corrosion, and is retrieving any unused pumps.
Medtronic notified Healthcare Professionals of the impact of unapproved drugs on the performance of the SynchroMed infusion pump system. Use of unapproved drugs with SynchroMed pumps can result in an increased risk of permanent motor stall and cessation of drug infusion. UPDATE 2/1/2016: Medtronic received approval to incorporate enhancements to the SynchroMed II pump which decreases the potential for internal electrical shorting and motor corrosion, and is retrieving any unused pumps.