A610 Clinician Programmer (CP) Software Application (app) with the following versions: Version 2.0.x (2.0.4584, 2.0.4594, 2.0.4605, 2.0.4630, and 2.0.4648), Version 3.0.x (3.0.1057, 3.0.1062, 3.0.1081, and 3.0.1098), Version 4.0.x (4.0.1052). Includes the below product descriptions: 1. ADAPTOR 64001 1X4 MVMT POCKET ADAPTOR, 2. ADAPTOR 64002 2X4 MVMT POCKET ADAPTOR, 3. INS 37602 Activa SC Neurostimulator, 4. INS 37603 Activa SCx Neurostimulator, 5. INS B35200 DBS PERCEPT PC, 6. INS B35300 PERCEPT RC V1
Class I - DangerousWhat Should You Do?
- Check if you have this product: Versions: Version 2.0.x (2.0.4584, 2.0.4594, 2.0.4605, 2.0.4630, and 2.0.4648), Version 3.0.x (3.0.1057, 3.0.1062, 3.0.1081, and 3.0.1098), Version 4.0.x (4.0.1052). 1. Model Number 64001, GTIN (Serial Numbers): 00613994544940 (N733965, N733969, N753252, N775354, N776291, N778120, N782773); 00613994604477 (0212774778, 0212775744, 0212775759, 0212776600, 0212793056, 0212804629, 0212805718, 0212864338, 0212874932, 0212890560, 0212890575, 0212895213, 0212895214, 0212895215, 0212907133, 0212932970, 0212932973, 0212932974, 0212935897, 0213003936, 0213011214, 0213037083, 0213103554, 0213103836, 0213103947, 0213119751, 0213126252, 0213126253, 0213126254, 0213126257, 0213126258, 0213126259, 0213145327, 0213238728, 0213238729, 0213238730, 0213238731, 0213238732, 0213238733, 0213296336, 0213296338, 0213296636, 0213336983, 0213336984, 0213379975, 0213391332, 0213403188, 0213410729, 0213447187, 0213453300, 0213453301, 0213462755, 0213462757, 0213494778, 0213503190, 0213512318, 0213512796, 0213513281, 0213526864, 0213526865, 0213526866, 0213526867, 0213526868, 0213526869, 0213526870, 0213526871, 0213526872, 0213557150, 0213557151, 0213557152, 0213566074, 0213575790, 0213589299, 0213612423, 0213612424, 0213612425, 0213627695, 0213632311, 0213632313, 0213643879, 0213643880, 0213649861, 0213649862, 0213649863, 0213649865, 0213649867, 0213649868, 0213649869, 0213649870, 0213649872, 0213649873, 0213649874, 0213649875, 0213649876, 0213662356, 0213917402, 0213938623, 0213938624, 0213938625, 0213938626, 0213938638, 0214153105, 0214153106, 0214220352, 0214220353, 0214220354, 0214220355, 0214220357, 0214221962, 0214221963, 0214447748, 0214447749, 0214490072, 0214494047, 0214494053, 0214494054, 0214494055, 0214494056, 0214503001, 0214528433, 0214537122, 0214537123, 0214539376, 0214574012, 0214574462, 0214617689, 0214617821, 0214694530, 0214694659, 0214734073, 0214734078, 0214734591, 0214735109, 0214735110, 0214827966, 0214892177, 0214892179, 0214899543, 0215002059, 0215039974, 0215068207, 0215069053, 0215157787, 0215170344, 0215172048, 0215189210, 0215191813, 0215204255, 0215546041, 0215546042, 0215546043, 0215546044, 0215624458, 0215631140, 0215732518, 0215744823, 0215774217, 0215781838, 0215781840, 0215782467, 0215782468, 0215782722, 0215796016, 0215796020, 0215803678, 0215803679, 0215857526, 0215857530, 0215898037, 0215899117, 0215899118, 0216018172, 0216171381, 0216171382, 0216207616, 0216346119, 0216377773, 0216403976, 0216419016, 0216499028, 0216499029, 0216499030, 0216499031, 0216658942, 0216658943, 0216664566, 0216759284, 0216774939, 0216774940, 0216810077, 0216810078, 0216810079, 0216810080, 0216810081, 0216919717, 0216919718, 0216970034, 0216977281, 0217000199, 0217046494, 0217057275, 0217067679, 0217067680, 0217067722, 0217068512, 0217068513, 0217068514, 0217170714, 0217170715, 0217170716, 0217170717, 0217170718, 0217170719, 0217170720, 0217179190, 0217179191, 0217179192, 0217179193, 0217179195, 0217179196, 0217179197, 0217179198, 0217179199, 0217306044, 0217306045, 0217309986, 0217309987, 0217309988, 0217309989, 0217309990, 0217309991, 0217309992, 0217333857, 0217333858, 0217333859, 0217333860, 0217333861, 0217333862, 0217333863, 0217440249, 0217596169, 0217596170, 0217619906, 0217664192, 0217671944, 0217671945, 0217750243, 0217927627, 0218001371, 0218173517, 0218318172, 0218318175, 0218318178); 00613994604484 (0213209850, 0213288462, 0213293456, 0213293457, 0213293458, 0213352261, 0213419249, 0213419340, 0213473603, 0213495294, 0213495295, 0213495296, 0213495297, 0213495298, 0213495299, 0213495340, 0213495341, 0213495342, 0213559327, 0213716708, 0213766926, 0213890732, 0213890733, 0214094542, 0214311406, 0214311407, 0214311408, 0214311414, 0214311415, 0214311416, 0214311419, 0214311474, 0214311475, 0214311476, 0214311477, 0214319850, 0214319853, 0214319967, 0214319968, 0214331598, 0214331599, 0216688332, 0216812514, 0217671913, 0217826925, 0217829246, 0217829247, 0217829248); 00613994604491 (0213925833, 0214282967, 0214282968, 0214301619, 0214403827, 0214404408, 0214404409, 0214448984, 0214448987, 0214451613, 0215280226); 00643169090866 (N693604, N695628, N697903, N697905, N707785, N707814, N707818, N709873, N711340, N712986, N716557, N724809, N724820, N724826, N724838, N726518, N728574, N730697, N742633, N742635, N744671, N744678, N746637, N746640, N748178, N748179, N751469, N751477, N753246, N753250, N753252, N755339, N755413, N755415, N758481, N769619, N769621, N771699, N779249, N781452, N782773, N786955, N786955N01, N796290, N796384, N808556, N810477, N810477N01, N818532, VA1UQ8W, VA1VSZ8, VA1VXY6, VA1W50Y, VA1XYNJ, VA1ZYXF, VA20GDV, VA214QW, VA21566, VA22BQP, VA239A9, VA239AC, VA25V40, VA283DR, VA28AQV, VA28SB9, VA29QGY, VA2AQ0X, VA2AQ9F, VA2BZYK, VA2E6W7, VA2E6W9, VA2ET07, VA2FMEF, VA2UEHE); 00643169936584 (VA22LW4, VA28SB9, VA29B21, VA2BK04, VA2BZYJ, VA2BZYK, VA2FMEF, VA2S9QU); 00763000111847 (0218354155, 0218354156, 0218354157, 0218354158, 0218354159, 0218354160, 0218354161, 0218366043, 0218366044, 0218366045, 0218370795, 0218370796, 0218370797, 0218425698, 0218425699, 0218426758, 0218605857, 0218616490, 0218669761, 0218717411, 0218720328, 0218720329, 0218774777, 0218780695, 0218780696, 0218786120, 0218803244, 0218803245, 0218821512, 0218904841, 0218952397, 0218952399, 0218952401, 0218960557, 0218960559, 0218960561, 0218960563, 0219134852, 0219143990, 0219160408, 0219185911, 0219201429, 0219201440, 0219206196, 0219214603, 0219245968, 0219305171, 0219305172, 0219305173, 0219305174, 0219337011, 0219349459, 0219375530, 0219375531, 0219377400, 0219390335, 0219398981, 0219422636, 0219425617, 0219433666, 0219497903, 0219518491, 0219587572, 0219587573, 0219612248, 0219612249, 0219620526, 0219621110, 0219715005, 0219717420, 0219718438, 0219718440, 0219765247, 0219765251, 0219868602, 0219890615, 0220071933, 0220075959, 0220100014, 0220256276, 0220260478, 0220263746, 0220276913, 0220297268, 0220297270, 0220316650, 0220368431, 0220368432, 0220375984, 0220428236, 0220430237, 0220584282, 0220590246, 0220590247, 0220590248, 0220590249, 0220590250, 0220590251, 0220594791, 0220594792, 0220614692, 0220648615, 0220668846, 0221036743, 0221053170, 0221184563, 0221186755, 0221227258, 0221264033, 0221264034, 0221264960, 0221272023, 0221281956, 0221281957, 0221323103, 0221430617, 0221444627, 0221491002, 0221514322, 0221514323, 0221514324, 0221514325, 0221546530, 0221566668, 0221577081, 0221594832, 0221622517, 0221649969, 0221690470, 0221741062, 0221741063, 0221755012, 0221765184, 0221802766, 0221803888, 0221803889, 0221803890, 0221827095, 0221839656, 0221848787, 0221849670, 0221849671, 0221849672, 0221850724, 0221881224, 0221894232, 0221894233, 0221972044, 0221972045, 0221972506, 0221979343, 0222006670, 0222006671, 0222025355, 0222038363, 0222054122, 0222139672, 0222139673, 0222139674, 0222139675, 0222139676, 0222139677, 0222139678, 0222139679, 0222154978, 0222154979, 0222154980, 0222154981, 0222395543, 0222395544, 0222395545, 0222396586, 0222396587, 0222396588, 0222396589, 0222396590, 0222396591, 0222396592, 0222396593, 0222396594, 0222396595, 0222396596, 0222396597, 0222396598, 0222396599, 0222396600, 0222396601, 0222412689, 0222412690, 0222418531, 0222418532, 0222418588, 0222418589, 0222418591, 0222426616, 0222427334, 0222427335, 0222427336, 0222429509, 0222580214, 0222621260, 0222621261, 0222621262, 0222621263, 0222622281, 0222629620, 0222630230, 0222630231, 0222630232, 0222646465, 0222652428, 0222653147, 0222653148, 0222658928, 0222660622, 0222660853, 0222677422, 0222684379, 0222687577, 0222687578, 0222689822, 0222689823, 0222741781, 0222741782, 0222741783, 0222742206, 0222818437, 0222818438, 0222818439, 0222818440, 0222832453, 0222845566, 0222845585, 0222845627, 0223009455, 0223009456, 0223009457, 0223009458, 0223019858, 0223023195, 0223023200, 0223023202, 0223037118, 0223037120, 0223083064, 0223083065, 0223090114, 0223091943, 0223091944, 0223091945, 0223095856, 0223095857, 0223099274, 0223121417, 0223130455, 0223257772, 0223259688, 0223259689, 0223259690, 0223259691, 0223259692, 0223259693, 0223259694, 0223259695, 0223259696, 0223259697, 0223259698, 0223259699, 0223259700, 0223259701, 0223315885, 0223316126, 0223316130, 0223316131, 0223316132, 0223316133, 0223316134, 0223316135, 0223316136, 0223316137, 0223316140, 0223316141, 0223316142, 0223316143, 0223316144, 0223316145, 0223316166, 0223316167, 0223316168, 0223316169, 0223316170, 0223316171, 0223316172, 0223354587, 0223450726, 0223450727, 0223450728, 0223450729, 0223450730, 0223450731, 0223450732, 0223450733, 0223450734, 0223450735, 0223450736, 0223450737, 0223450738, 0223450739, 0223450740, 0223450745, 0223450746, 0223450747, 0223450748, 0223450749, 0223456972, 0223456974, 0223456975, 0223456976, 0223562242, 0223562243, 0223562244, 0223562245, 0223562246, 0223562247, 0223562248, 0223562249, 0223562250, 0223562251, 0223562252, 0223562253, 0223562254, 0223562255, 0223562256, 0223562257, 0223562258, 0223562259, 0223562260, 0223562261, 0223852618, 0223852619, 0223852620, 0223852621, 0223852622, 0223852623, 0223852624, 0223859033, 0223861157, 0223861158, 0223861160, 0223861778, 0223861779, 0223861800, 0223861814, 0223861879, 0223861880, 0223861882, 0223861939, 0223861940, 0223861949, 0223861950, 0223862582, 0223862601, 0223862609, 0223862666, 0224025144, 0224034627, 0224034666, 0224034667, 0224082804, 0224082805, 0224085442, 0224085443, 0224085444, 0224085445, 0224085586, 0224085626, 0224085627, 0224085628, 0224085629, 0224085630, 0224107431, 0224107432, 0224107434, 0224107435, 0224107436, 0224107437, 0224107439, 0224107440, 0224107441, 0224107442, 0224107443, 0224107444, 0224107445, 0224107806, 0224107808, 0224134495, 0224134496, 0224134497, 0224159135, 0224159136, 0224159137, 0224159138, 0224159139, 0224159140, 0224198334, 0224216499, 0224216563, 0224216564, 0224253258, 0224253259, 0224288640, 0224288641, 0224288642, 0224288643, 0224288644, 0224288686, 0224288687, 0224288688, 0224288689, 0224288690, 0224288691, 0224288692, 0224288693, 0224288694, 0224288695, 0224380514, 0224380515, 0224380516, 0224380517, 0224380518, 0224409896, 0224409897, 0224409899, 0224409900, 0224440199, 0224440200, 0224440201, 0224440202, 0224440203, 0224440204, 0224440205, 0224440866, 0224440867, 0224440868, 0224440869, 0224440870, 0224440871, 0224440873, 0224442313, 0224442314, 0224442315, 0224442316, 0224442317, 0224442318, 0224442319, 0224442320, 0224442321, 0224442322, 0224442323, 0224442324, 0224442325, 0224449126, 0224449127, 0224449128, 0224449129, 0224449130, 0224449131, 0224449132, 0224449133, 0224449134, 0224449135, 0224449136, 0224449137, 0224449138, 0224449139, 0224449140, 0224449141, 0224449142, 0224619284, 0224619285, 0224620226, 0224620227, 0224620228, 0224678399, 0224678400, 0224678401, 0224678564, 0224678565, 0224678906, 0224678908, 0224687409, 0224712020, 0224712021, 0224712022, 0224712023, 0224736512, 0224736513, 0224736514, 0224736515, 0224736516, 0224736517, 0224736837, 0224736838, 0224736839, 0224791803, 0224791804, 0224791805, 0224799391, 0224799392, 0224799393, 0224799394, 0224799395, 0224799396, 0224799397, 0224799398, 0224799399, 0224799400, 0224799401, 0224799402, 0224799403, 0224799806, 0224799807, 0224799808, 0224799809, 0224799810, 0224799811, 0224799812, 0224799813, 0224799814, 0224799815, 0224799816, 0224799817, 0224799818, 0224799819, 0224799820, 0224799821, 0224799822, 0224799823, 0224799824, 0224799825, 0224799826, 0224799827, 0224799828, 0224799830, 0224799831, 0224799834, 0224799835, 0224799836, 0224799837, 0224799838, 0224799839, 0225136017, 0225136018, 0225136930, 0225136931, 0225136932, 0225243969, 0225243970, 0225243973, 0225243974, 0225243975, 0225243976, 0225243977, 0225243978, 0225243979, 0225243980, 0225243981, 0225243982, 0225243983, 0225243984, 0225243985, 0225244366, 0225244367, 0225244368, 0225244369, 0225244370, 0225244371, 0225244372, 0225244373, 0225244374, 0227894725, 0227894827, 0227894828, 0227894829, 0227894832, 0227894833, 0227894834, 0227894835, 0227894836, 0227894842, 0227894843, 0227894844, 0227894845, 0227894846, 0227894847, 0227894848, 0227894849, 0227894851, 0227894852, 0227894853, 0227894854, 0227894856, 0227894857, 0227894858, 0227894859, 0227894861, 0227894863, 0227894864, 0227894865, 0227894866, 0227894867, 0227894868, 0227894869, 0227894872, 0227894873, 0227894874, 0227894875, 0227894876, 0227894877, 0227894878, 0227894879, 0228118366, 0228118367, 0228118368, 0228118370, 0228208483, 0228208484, 0228208485, 0228208846, 0228208847, 0228208848, 0228208849, 0228208850, 0228208851, 0228208855, 0228208856, 0228208857, 0228208858, 0228208859, 0228208860, 0228208861, 0228208862, 0228208863, 0228208868, 0228208869, 0228208870, 0228208872, 0228208873, 0228208874, 0228208875, 0228208876, 0228208877, 0228208878, 0228208879, 0228208880, 0228208881, 0228208883, 0228208884, 0228366377, 0228366378, 0228366379, 0228366380, 0228366381, 0228366382, 0228367327, 0228367328, 0228367329, 0228367330, 0228367331, 0228367332, 0228367333, 0228367334, 0228367335, 0228367336, 0228367337, 0228367338, 0228367339, 0228367340, 0228367341, 0228367342, 0228367343, 0228367344, 0228367349, 0228367350, 0228367351, 0228367352, 0228367353, 0228367355, 0228367356, 0228367357, 0228367358, 0228367359, 0228367362, 0228367363, 0228528286, 0228528287, 0228528288); 00763000111854 (0219505890, 0219505891, 0219505892, 0219505893, 0219505894, 0219533854, 0219875956, 0219988881, 0220138915, 0220248276, 0220248277, 0220253142, 0220277098, 0220368430, 0220369379, 0220369848, 0220394562, 0220584283, 0220584284, 0220584285, 0220627630, 0220632089, 0220643658, 0220780560, 0220785734, 0220868360, 0220925498, 0220936605, 0221343503, 0221972290, 0221972291, 0222138789, 0222230549, 0222230550, 0222230551, 0222230552, 0222622587, 0222622588, 0222622589, 0222638828, 0222645597, 0222667672, 0222685841, 0222742193, 0222742194, 0222742195, 0222819361, 0222819362, 0223008932, 0223008933, 0223008934, 0223008935, 0223008936, 0223316234, 0223449358, 0223456457, 0223568237, 0223568238, 0223568239, 0223568240, 0223852943, 0224187952, 0224187953, 0224187954, 0224187957, 0224409302, 0224440323, 0224440324, 0224518549, 0224518550, 0224518551, 0224518552, 0224518553, 0224518558, 0224518559, 0224518560, 0225136822, 0225136823, 0225244426, 0225244427, 0225244428, 0227894837, 0227894881, 0227894882, 0227969113, 0228208826, 0228208827, 0228208828, 0228208829, 0228365695, 0228365696); 00763000111861 (0222172842, 0222188636, 0222194734, 0222814174, 0222814175, 0222814176, 0222814177, 0222814178, 0222814180, 0222814181, 0222814182, 0222814183, 0222814184, 0223029972, 0223029973, 0223029974, 0223029975, 0223029976, 0223089378, 0223089379, 0223089380, 0223089381, 0223089382, 0223089383, 0223089384, 0223091453, 0223091455, 0223091607, 0223091608, 0223091609, 0223091610, 0223313914, 0223313915, 0223313916, 0223313917, 0223313918, 0223313919, 0223313920, 0223313921, 0223313922, 0223450743, 0223450744, 0223553145, 0223562226, 0223562227, 0223562228, 0223562229, 0223562230, 0223562231, 0223562232, 0223562233, 0223562234, 0223562235, 0223562236, 0223562238, 0223562239, 0223562240, 0223562241, 0223650153, 0223650154, 0223650155, 0223861159, 0223861544, 0223861780, 0223861881, 0223861998, 0223862583, 0223862584, 0223862600, 0223862602, 0223862608, 0224085308, 0224085309, 0224085310, 0224085311, 0224085312, 0224085313, 0224085314, 0224085315, 0224085316, 0224085317, 0224085318, 0224085319, 0224321608, 0224330255, 0224330256, 0224330257, 0224330258, 0224330259, 0224330260, 0224330261, 0224448536, 0224448537, 0224448538, 0224448539, 0224448540, 0224448541, 0224448542, 0224448543, 0224448544, 0224448545, 0224449106, 0224449107, 0224449108, 0224449109, 0224449110, 0224449111, 0224449112, 0224449113, 0224449114, 0224449115, 0224449116, 0224449117, 0224449118, 0224449119, 0224449120, 0224449121, 0224449122, 0224449123, 0224449124, 0225007573, 0225007574, 0225009606, 0225009607, 0225009608, 0225009609, 0225009610, 0225009611, 0225009612, 0225009613, 0225244394, 0225244395, 0225244396, 0225244397, 0225244398, 0225244401, 0225244402, 0225244403, 0225244404, 0225244405, 0225244406, 0225244407, 0225244408, 0225462856, 0225669344, 0225669345, 0225670466, 0227894884, 0227894885, 0227894926, 0227894927, 0227894929, 0227894933, 0227894934, 0227894935, 0228203474, 0228203475, 0228203476, 0228203477, 0228203478, 0228203479, 0228203480, 0228203481, 0228203482, 0228203483, 0228203484, 0228203485). 2. Model Number 64002, GTIN (Serial Numbers): 00613994544957 (N692230, N709765, N709766, N709769, N712972, N714659, N720090, N721922, N723172, N724723, N724752, N730633, N732361, N732366, N732368, N732369, N732369N01, N734116, N735283, N735294, N736643, N736644, N738447, N738450, N738459, N740632, N740717, N742744, N744753, N744754, N744756, N746628, N746635, N748190, N748193, N748193N01, N750163, N750166, N753183, N755406, N755416, N755435, N757678, N757684, N758562, N760449, N760457, N760465, N782607, N784725, N784740, N784754, N784755, N786840N01, N786849, N786915N01, N786917, N786917N01, N794131, N795243, N795246, N795247, N795271, N796405, N796416, N796419, N796432, N796441, N803281); 00613994604446 (0212745840, 0212749749, 0212754210, 0212761030, 0212761032, 0212761035, 0212761036, 0212761037, 0212761038, 0212774118, 0212774188, 0212774189, 0212774190, 0212774308, 0212782600, 0212782601, 0212782602, 0212784261, 0212785056, 0212785057, 0212787201, 0212787202, 0212797383, 0212797384, 0212798912, 0212804685, 0212804686, 0212815392, 0212815398, 0212846338, 0212847507, 0212847508, 0212847509, 0212850282, 0212850283, 0212850293, 0212850294, 0212852696, 0212852697, 0212856040, 0212862899, 0212863596, 0212864738, 0212875923, 0212878577, 0212878578, 0212881521, 0212885470, 0212885471, 0212885472, 0212885473, 0212886132, 0212894684, 0212894685, 0212894694, 0212894695, 0212894697, 0212895260, 0212895261, 0212895262, 0212895263, 0212895264, 0212895265, 0212900727, 0212901662, 0212905207, 0212905218, 0212905686, 0212906223, 0212922356, 0212922358, 0212922421, 0212926120, 0212926121, 0212928704, 0212929653, 0212929654, 0212929655, 0212929656, 0212929657, 0212933033, 0212933034, 0212933035, 0212933036, 0212933037, 0212933060, 0212933061, 0212933062, 0212933063, 0212933064, 0212933065, 0212933066, 0212933073, 0212935226, 0212940808, 0212940815, 0212956691, 0212963491, 0212963493, 0212963494, 0212963496, 0212963497, 0212965472, 0212965473, 0212965474, 0212965475, 0212965476, 0212965477, 0212965478, 0212965970, 0212966920, 0212966921, 0212970794, 0212970795, 0212971195, 0212971199, 0212971201, 0212971202, 0212971457, 0212980001, 0212983883, 0212983886, 0212996333, 0213001235, 0213001236, 0213013164, 0213017632, 0213017633, 0213017634, 0213017635, 0213021468, 0213021471, 0213021474, 0213023907, 0213037782, 0213037784, 0213037785, 0213039387, 0213071412, 0213071413, 0213071414, 0213071415, 0213071416, 0213071417, 0213071418, 0213071435, 0213071439, 0213071500, 0213071501, 0213071542, 0213071544, 0213071545, 0213071546, 0213071550, 0213071551, 0213072388, 0213072389, 0213072390, 0213079903, 0213079904, 0213080289, 0213080579, 0213085817, 0213085818, 0213085900, 0213085901, 0213085902, 0213085903, 0213091248, 0213094930, 0213094931, 0213095020, 0213095951, 0213097887, 0213097888, 0213097889, 0213097890, 0213103760, 0213103762, 0213103764, 0213103765, 0213103769, 0213115057, 0213115059, 0213115432, 0213127808, 0213127809, 0213127810, 0213127811, 0213127812, 0213127813, 0213127814, 0213127815, 0213127816, 0213127817, 0213127818, 0213127819, 0213127940, 0213132148, 0213136860, 0213136861, 0213136862, 0213136866, 0213136868, 0213147262, 0213147269, 0213147270, 0213147271, 0213147272, 0213147273, 0213147274, 0213174551, 0213182376, 0213182377, 0213182378, 0213191646, 0213191647, 0213192551, 0213192552, 0213192553, 0213237284, 0213237290, 0213413947, 0213413948, 0213413949, 0213413950, 0213413953, 0213413955, 0213413956, 0213413957, 0213413958, 0213413959, 0213417621, 0213417622, 0213429219, 0213429300, 0213429301, 0213433410, 0213445371, 0213445372, 0213452791, 0213452792, 0213452793, 0213456732, 0213456733, 0213456734, 0213456735, 0213456737, 0213462568, 0213462569, 0213462570, 0213462571, 0213462572, 0213462573, 0213462575, 0213466511, 0213467389, 0213473457, 0213473458, 0213578992, 0213578993, 0213583797, 0213583802, 0213584203, 0213588224, 0213588225, 0213588227, 0213605259, 0213632473, 0213632474, 0213632475, 0213632476, 0213632478, 0213632479, 0213634655, 0213634657, 0213634658, 0213634659, 0213649938, 0213649939, 0213650135, 0213650136, 0213650137, 0213650139, 0213650142, 0213650143, 0213650144, 0213650145, 0213650146, 0213650147, 0213650148, 0213650149, 0213650150, 0213650151, 0213650152, 0213650153, 0213650154, 0213650157, 0213650161, 0213650162, 0213650163, 0213650164, 0213650165, 0213650166, 0213650167, 0213650168, 0213650169, 0213650170, 0213650171, 0213650172, 0213650173, 0213650174, 0213650175, 0213650177, 0213650178, 0213650179, 0213650180, 0213650181, 0213650182, 0213650183, 0213650184, 0213650185, 0213650186, 0213650187, 0213650188, 0213650189, 0213650190, 0213650191, 0213650192, 0213650193, 0213650194, 0213650195, 0213650196, 0213650197, 0213650198, 0213650199, 0213650200, 0213650201, 0213650202, 0213650203, 0213650204, 0213650205, 0213650206, 0213650208, 0213650209, 0213650210, 0213650214, 0213650215, 0213650216, 0213650217, 0213650219, 0213650220, 0213650221, 0213650222, 0213650223, 0213650224, 0213650225, 0213650226, 0213650228, 0213650230, 0213650231, 0213650232, 0213650233, 0213650234, 0213650235, 0213650238, 0213650240, 0213650242, 0213650247, 0213650249, 0213650251, 0213650253, 0213650255, 0213650257, 0213650259, 0213650261, 0213650264, 0213650266, 0213650268, 0213650271, 0213650273, 0213650275, 0213650278, 0213650279, 0213650280, 0213650281, 0213650285, 0213650287, 0213650290, 0213650292, 0213650295, 0213650297, 0213650299, 0213650301, 0213650303, 0213650305, 0213650306, 0213650307, 0213650308, 0213650309, 0213650310, 0213650311, 0213650312, 0213650313, 0213650314, 0213650315, 0213650316, 0213650317, 0213650318, 0213650319, 0213650320, 0214103390, 0214103393, 0214468589, 0214468590, 0214468591, 0214468593, 0214468594, 0214468595, 0214468596, 0214468597, 0214468598, 0214468761, 0214468762, 0214468763, 0214468765, 0214468904, 0214468905, 0214468906, 0214468907, 0214468911, 0214469578, 0214469779, 0214470704, 0214470705, 0214475642, 0214485655, 0214485659, 0214485883, 0214486991, 0214496362, 0214496363, 0214496364, 0214518188, 0214518189, 0214519754, 0214528301, 0214528302, 0214528303, 0214528304, 0214528305, 0214528306, 0214531735, 0214531779, 0214540438, 0214540439, 0214554705, 0214554706, 0214557416, 0214557523, 0214566932, 0214566933, 0214566934, 0214573745, 0214573746, 0214573747, 0214573756, 0214574071, 0214578640, 0214585623, 0214590766, 0214603806, 0214603807, 0214609630, 0214609634, 0214610532, 0214610533, 0214611925, 0214616558, 0214616559, 0214620666, 0214621843, 0214621845, 0214621846, 0214624371, 0214624818, 0214624820, 0214625801, 0214625802, 0214647188, 0214647844, 0214648116, 0214648117, 0214649108, 0214649111, 0214649112, 0214653962, 0214653963, 0214657197, 0214692218, 0214692219, 0214697222, 0214700826, 0214701799, 0214701800, 0214712361, 0214712362, 0214808030, 0214816022, 0214816031, 0214816032, 0214824545, 0214837530, 0214842559, 0214865328, 0214865329, 0214865330, 0214865331, 0214865332, 0214874944, 0214882639, 0214882640, 0214884661, 0214887162, 0214887163, 0214899505, 0214915786, 0214923073, 0214932550, 0214934114, 0214935059, 0214935060, 0214935832, 0214941176, 0214947559, 0214947685, 0214947687, 0214947688, 0214954015, 0214959057, 0214974833, 0214974911, 0214974913, 0214975112, 0214975660, 0214975872, 0214975873, 0214975874, 0214975875, 0214975876, 0214975878, 0214975879, 0215009283, 0215031113, 0215031114, 0215033187, 0215035470, 0215035471, 0215066747, 0215066748, 0215066749, 0215076639, 0215076640, 0215076881, 0215076882, 0215076883, 0215099587, 0215099588, 0215099596, 0215099597, 0215099598, 0215100079, 0215100407, 0215112757, 0215112758, 0215112759, 0215112760, 0215112804, 0215112805, 0215116764, 0215116765, 0215116766, 0215116767, 0215116768, 0215116769, 0215116770, 0215126257, 0215126258, 0215132780, 0215141342, 0215141345, 0215146156, 0215146157, 0215156752, 0215156753, 0215167405, 0215167939, 0215167940, 0215170722, 0215170723, 0215170725, 0215170727, 0215171038, 0215172599, 0215176082, 0215198093, 0215198094, 0215198095, 0215201448, 0215201449, 0215201450, 0215226495, 0215226622, 0215226623, 0215226624, 0215226625, 0215227175, 0215233911, 0215233920, 0215234483, 0215245873, 0215253204, 0215253298, 0215255744, 0215255747, 0215259531, 0215259532, 0215268457, 0215268459, 0215271857, 0215271858, 0215271859, 0215271915, 0215272177, 0215272179, 0215273544, 0215273547, 0215273552, 0215273555, 0215280281, 0215280283, 0215280933, 0215282537, 0215283796, 0215283799, 0215283941, 0215288570, 0215288574, 0215288831, 0215356493, 0215356494, 0215356496, 0215356498, 0215356499, 0215356500, 0215358261, 0215358264, 0215358265, 0215358266, 0215358268, 0215358269, 0215358270, 0215358282, 0215358283, 0215358284, 0215358312, 0215358328, 0215358329, 0215358331, 0215358343, 0215358344, 0215358345, 0215358346, 0215358347, 0215358351, 0215358352, 0215358353, 0215358354, 0215358358, 0215359183, 0215359191, 0215359270, 0215359280, 0215361735, 0215361736, 0215372105, 0215385658, 0215385710, 0215385714, 0215385716, 0215385717, 0215385718, 0215385720, 0215385721, 0215385723, 0215385724, 0215385725, 0215385727, 0215385728, 0215385729, 0215385734, 0215385736, 0215385739, 0215385744, 0215385745, 0215385746, 0215385747, 0215385748, 0215385750, 0215385755, 0215385756, 0215385758, 0215385759, 0215385761, 0215385762, 0215385765, 0215385767, 0215385769, 0215385770, 0215385772, 0215385773, 0215385774, 0215385775, 0215385776, 0215385777, 0215385779, 0215385780, 0215385781, 0215385782, 0215385783, 0215385787, 0215385788, 0215385789, 0215385790, 0215385792, 0215392463, 0215392482, 0215392483, 0215392484, 0215392510, 0215392511, 0215392526, 0215392527, 0215392535, 0215392536, 0215392537, 0215392539, 0215392544, 0215472731, 0215472732, 0215472738, 0215472739, 0215472740, 0215474021, 0215474022, 0215474023, 0215474024, 0215474025, 0215475876, 0215477383, 0215477384, 0215477385, 0215499417, 0215501735, 0215501736, 0215522999, 0215523025, 0215523041, 0215523042, 0215523066, 0215589555, 0215589556, 0215596418, 0215606813, 0215606814, 0215614571, 0215657060, 0215700166, 0215700167, 0215700168, 0215819667, 0215819668, 0215934633, 0215934634, 0215936463, 0215942316, 0215942318, 0215954998, 0215954999, 0215955000, 0215964508, 0215972279, 0215978930, 0215988218, 0215988219, 0216004477, 0216004478, 0216004479, 0216004480, 0216016139, 0216018624, 0216018625, 0216065724, 0216111604, 0216156711, 0216171130, 0216171132, 0216171133, 0216171135, 0216171136, 0216171137, 0216206937, 0216206938, 0216206939, 0216206940, 0216207896, 0216207897, 0216207898, 0216208179, 0216208987, 0216209161, 0216209162, 0216462635, 0216547664, 0216550177, 0216561981, 0216579960, 0216591414, 0216592122, 0216624039, 0216624040, 0216640718, 0216640719, 0216640748, 0216644828, 0216644829, 0216658945, 0216661538, 0216661540, 0216668532, 0216668533, 0216670436, 0216670438, 0216683902, 0216683903, 0216683904, 0216683905, 0216683907, 0216683908, 0216683909, 0216691749, 0216701593, 0216707670, 0216730627, 0216748843, 0216748844, 0216748845, 0216748846, 0216766995, 0216766997, 0216777156, 0216781561, 0216782770, 0216789797, 0216790477, 0216812585, 0216833887, 0216866102, 0216866103, 0216880723, 0216888139, 0216888140, 0216903625, 0216903626, 0216934315, 0216943408, 0216943409, 0216970112, 0217000196, 0217000197, 0217014298, 0217014299, 0217039819, 0217046518, 0217101498, 0217101933, 0217101965, 0217103057, 0217134103, 0217134104, 0217134105, 0217148781, 0217157510, 0217178616, 0217178618, 0217178619, 0217187787, 0217220596, 0217221543, 0217244637, 0217249975, 0217249976, 0217287997, 0217307860, 0217307861, 0217311998, 0217311999, 0217324597, 0217335743, 0217335744, 0217342991, 0217342992, 0217346263, 0217346264, 0217348324, 0217348325, 0217375806, 0217375807, 0217375808, 0217375809, 0217375810, 0217375811, 0217375812, 0217375819, 0217399630, 0217399633, 0217399635, 0217399636, 0217403404, 0217412166, 0217422238, 0217422239, 0217425679, 0217440507, 0217440508, 0217440509, 0217440510, 0217483873, 0217484070, 0217486062, 0217486063, 0217486064, 0217497498, 0217497499, 0217497500, 0217534872, 0217560564, 0217560565, 0217566377, 0217600191, 0217600787, 0217612437, 0217635208, 0217635209, 0217635210, 0217664212, 0217727091, 0217759260, 0217778420, 0217782462, 0217795041, 0217809103, 0217820834, 0217821658, 0217830939, 0217843819, 0217863004, 0217873901, 0217874944, 0217881301, 0217911512, 0217911513, 0217933185, 0217939586, 0217977438, 0217986426, 0217992947, 0218003574, 0218003575, 0218011116, 0218019871, 0218022443, 0218028484, 0218040503, 0218043348, 0218043349, 0218061176, 0218086090, 0218086091, 0218104522, 0218122969, 0218156717, 0218156719, 0218156720, 0218156724, 0218156725, 0218156866, 0218156867, 0218178504, 0218282737, 0218297499, 0218297500, 0218297501, 0218297502, 0218299923, 0218299924, 0218299925, 0218441806, 0218441807, 0218513547, 0218513548, 0218513551, 0218513552, 0218513553, 0218523150, 0218775058, 0218803830, 0218815563, 0218977414, 0219008605, 0219014647, 0219048001, 0219048002, 0219048003, 0219049209, 0219050882, 0219065352, 0219065354, 0219065356, 0219065357, 0219093828, 0219144070, 0219176011, 0219191926, 0219199581, 0219201858, 0219203067, 0219203068, 0219226703, 0219248385, 0219248467, 0219248468, 0219248469, 0219248470, 0219248471, 0219248472, 0219288457, 0219289452, 0219289453, 0219289454, 0219289456, 0219289457, 0219370743, 0219370744, 0219374030, 0219375167, 0219375168, 0219393944, 0219403090, 0219423288, 0219441239, 0219465091, 0219465092, 0219465093, 0219465096, 0219468607, 0219468608, 0219468609, 0219472735, 0219472736, 0219472737, 0219535125, 0219541406, 0219541410, 0219541415, 0219573124, 0219605793, 0219627906, 0219627907, 0219687519, 0219697770, 0219724240, 0219724241, 0219724242, 0219726123, 0219726125, 0219737378, 0219899190, 0219899191, 0219916595, 0219916646, 0220037057, 0220037058, 0220044030, 0220044031, 0220044046, 0220184998, 0220184999, 0220185000, 0220185001, 0220185003, 0220185004, 0220185005, 0220187566, 0220187569, 0220187570, 0220187571, 0220187572, 0220187576, 0220187577, 0220187578, 0220187579, 0220187580, 0220187581, 0220187582, 0220187583, 0220187584, 0220187585, 0220187586, 0220187587, 0220622508, 0220629133, 0220629134, 0220629135, 0220629137, 0220629179, 0220629180, 0220629181, 0220629182, 0220629183, 0220629446, 0220629447, 0220629448, 0220629449, 0220629466, 0220629467, 0220629469, 0220629470, 0220629471, 0220629472, 0220629473, 0220629474, 0220629475, 0220629476, 0220629477, 0221034114, 0221034115, 0221034117, 0221034118, 0221034119, 0221053080, 0221316285, 0221610530, 0223192469, 0223192470, 0223192590, 0223193556, 0223200856); 00613994604453 (0212741247, 0212741248, 0212741249, 0212742419, 0212742483, 0212745770, 0212785058, 0212815602, 0212828591, 0212843763, 0212956684, 0212970796, 0213071419, 0213071540, 0213071541, 0213073473, 0213078456, 0213127807, 0213174305, 0213237292, 0213250893, 0213250894, 0213269072, 0213284733, 0213285393, 0213299619, 0213300272, 0213300273, 0213300279, 0213300280, 0213300281, 0213300282, 0213300283, 0213402619, 0213402861, 0213402865, 0213402866, 0213402867, 0213417326, 0213445379, 0213452794, 0213456725, 0213456726, 0213489219, 0213501865, 0213516313, 0213525256, 0213525259, 0213525300, 0213525307, 0213525308, 0213525310, 0213578174, 0213583801, 0213608287, 0213608288, 0213619023, 0213625015, 0213641249, 0213751757, 0213751758, 0213757301, 0213757302, 0213762800, 0213764861, 0213764862, 0213764863, 0213771966, 0213793175, 0213849987, 0214091533, 0214104013, 0214167671, 0214296666, 0214296667, 0214299340, 0214311302, 0214311303, 0214311304, 0214311307, 0214311308, 0214311309, 0214311318, 0214311319, 0214311341, 0214616550, 0214624816, 0214624817, 0214700827, 0214956585, 0215156747, 0215206629, 0215243796, 0216065728, 0216311195, 0216508909, 0216508910, 0216508911, 0216518391, 0216518392, 0217013121, 0217013122, 0217013123, 0217013124, 0217013125, 0217586255, 0217644986, 0217701988, 0217769612, 0217830725, 0217954517, 0218131774, 0218985898, 0219009806, 0219023827, 0219023828, 0219029040, 0219399824, 0219399825, 0219402886, 0219728126, 0220211195, 0220831736, 0221149643, 0221215835, 0221215836); 00613994604460 (0212803086, 0212894135, 0212894136, 0212894137, 0212894138, 021
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Medtronic Neuromodulation
- Reason for Recall:
- Patients implanted with a pocket adaptor (Model 64001 and/or 64002) are limited to HEAD ONLY MRI eligibility.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
A610 Clinician Programmer (CP) Software Application (app) with the following versions: Version 2.0.x (2.0.4584, 2.0.4594, 2.0.4605, 2.0.4630, and 2.0.4648), Version 3.0.x (3.0.1057, 3.0.1062, 3.0.1081, and 3.0.1098), Version 4.0.x (4.0.1052). Includes the below product descriptions: 1. ADAPTOR 64001 1X4 MVMT POCKET ADAPTOR, 2. ADAPTOR 64002 2X4 MVMT POCKET ADAPTOR, 3. INS 37602 Activa SC Neurostimulator, 4. INS 37603 Activa SCx Neurostimulator, 5. INS B35200 DBS PERCEPT PC, 6. INS B35300 PERCEPT RC V1
Product Codes/Lot Numbers:
Versions: Version 2.0.x (2.0.4584, 2.0.4594, 2.0.4605, 2.0.4630, and 2.0.4648), Version 3.0.x (3.0.1057, 3.0.1062, 3.0.1081, and 3.0.1098), Version 4.0.x (4.0.1052). 1. Model Number 64001, GTIN (Serial Numbers): 00613994544940 (N733965, N733969, N753252, N775354, N776291, N778120, N782773); 00613994604477 (0212774778, 0212775744, 0212775759, 0212776600, 0212793056, 0212804629, 0212805718, 0212864338, 0212874932, 0212890560, 0212890575, 0212895213, 0212895214, 0212895215, 0212907133, 0212932970, 0212932973, 0212932974, 0212935897, 0213003936, 0213011214, 0213037083, 0213103554, 0213103836, 0213103947, 0213119751, 0213126252, 0213126253, 0213126254, 0213126257, 0213126258, 0213126259, 0213145327, 0213238728, 0213238729, 0213238730, 0213238731, 0213238732, 0213238733, 0213296336, 0213296338, 0213296636, 0213336983, 0213336984, 0213379975, 0213391332, 0213403188, 0213410729, 0213447187, 0213453300, 0213453301, 0213462755, 0213462757, 0213494778, 0213503190, 0213512318, 0213512796, 0213513281, 0213526864, 0213526865, 0213526866, 0213526867, 0213526868, 0213526869, 0213526870, 0213526871, 0213526872, 0213557150, 0213557151, 0213557152, 0213566074, 0213575790, 0213589299, 0213612423, 0213612424, 0213612425, 0213627695, 0213632311, 0213632313, 0213643879, 0213643880, 0213649861, 0213649862, 0213649863, 0213649865, 0213649867, 0213649868, 0213649869, 0213649870, 0213649872, 0213649873, 0213649874, 0213649875, 0213649876, 0213662356, 0213917402, 0213938623, 0213938624, 0213938625, 0213938626, 0213938638, 0214153105, 0214153106, 0214220352, 0214220353, 0214220354, 0214220355, 0214220357, 0214221962, 0214221963, 0214447748, 0214447749, 0214490072, 0214494047, 0214494053, 0214494054, 0214494055, 0214494056, 0214503001, 0214528433, 0214537122, 0214537123, 0214539376, 0214574012, 0214574462, 0214617689, 0214617821, 0214694530, 0214694659, 0214734073, 0214734078, 0214734591, 0214735109, 0214735110, 0214827966, 0214892177, 0214892179, 0214899543, 0215002059, 0215039974, 0215068207, 0215069053, 0215157787, 0215170344, 0215172048, 0215189210, 0215191813, 0215204255, 0215546041, 0215546042, 0215546043, 0215546044, 0215624458, 0215631140, 0215732518, 0215744823, 0215774217, 0215781838, 0215781840, 0215782467, 0215782468, 0215782722, 0215796016, 0215796020, 0215803678, 0215803679, 0215857526, 0215857530, 0215898037, 0215899117, 0215899118, 0216018172, 0216171381, 0216171382, 0216207616, 0216346119, 0216377773, 0216403976, 0216419016, 0216499028, 0216499029, 0216499030, 0216499031, 0216658942, 0216658943, 0216664566, 0216759284, 0216774939, 0216774940, 0216810077, 0216810078, 0216810079, 0216810080, 0216810081, 0216919717, 0216919718, 0216970034, 0216977281, 0217000199, 0217046494, 0217057275, 0217067679, 0217067680, 0217067722, 0217068512, 0217068513, 0217068514, 0217170714, 0217170715, 0217170716, 0217170717, 0217170718, 0217170719, 0217170720, 0217179190, 0217179191, 0217179192, 0217179193, 0217179195, 0217179196, 0217179197, 0217179198, 0217179199, 0217306044, 0217306045, 0217309986, 0217309987, 0217309988, 0217309989, 0217309990, 0217309991, 0217309992, 0217333857, 0217333858, 0217333859, 0217333860, 0217333861, 0217333862, 0217333863, 0217440249, 0217596169, 0217596170, 0217619906, 0217664192, 0217671944, 0217671945, 0217750243, 0217927627, 0218001371, 0218173517, 0218318172, 0218318175, 0218318178); 00613994604484 (0213209850, 0213288462, 0213293456, 0213293457, 0213293458, 0213352261, 0213419249, 0213419340, 0213473603, 0213495294, 0213495295, 0213495296, 0213495297, 0213495298, 0213495299, 0213495340, 0213495341, 0213495342, 0213559327, 0213716708, 0213766926, 0213890732, 0213890733, 0214094542, 0214311406, 0214311407, 0214311408, 0214311414, 0214311415, 0214311416, 0214311419, 0214311474, 0214311475, 0214311476, 0214311477, 0214319850, 0214319853, 0214319967, 0214319968, 0214331598, 0214331599, 0216688332, 0216812514, 0217671913, 0217826925, 0217829246, 0217829247, 0217829248); 00613994604491 (0213925833, 0214282967, 0214282968, 0214301619, 0214403827, 0214404408, 0214404409, 0214448984, 0214448987, 0214451613, 0215280226); 00643169090866 (N693604, N695628, N697903, N697905, N707785, N707814, N707818, N709873, N711340, N712986, N716557, N724809, N724820, N724826, N724838, N726518, N728574, N730697, N742633, N742635, N744671, N744678, N746637, N746640, N748178, N748179, N751469, N751477, N753246, N753250, N753252, N755339, N755413, N755415, N758481, N769619, N769621, N771699, N779249, N781452, N782773, N786955, N786955N01, N796290, N796384, N808556, N810477, N810477N01, N818532, VA1UQ8W, VA1VSZ8, VA1VXY6, VA1W50Y, VA1XYNJ, VA1ZYXF, VA20GDV, VA214QW, VA21566, VA22BQP, VA239A9, VA239AC, VA25V40, VA283DR, VA28AQV, VA28SB9, VA29QGY, VA2AQ0X, VA2AQ9F, VA2BZYK, VA2E6W7, VA2E6W9, VA2ET07, VA2FMEF, VA2UEHE); 00643169936584 (VA22LW4, VA28SB9, VA29B21, VA2BK04, VA2BZYJ, VA2BZYK, VA2FMEF, VA2S9QU); 00763000111847 (0218354155, 0218354156, 0218354157, 0218354158, 0218354159, 0218354160, 0218354161, 0218366043, 0218366044, 0218366045, 0218370795, 0218370796, 0218370797, 0218425698, 0218425699, 0218426758, 0218605857, 0218616490, 0218669761, 0218717411, 0218720328, 0218720329, 0218774777, 0218780695, 0218780696, 0218786120, 0218803244, 0218803245, 0218821512, 0218904841, 0218952397, 0218952399, 0218952401, 0218960557, 0218960559, 0218960561, 0218960563, 0219134852, 0219143990, 0219160408, 0219185911, 0219201429, 0219201440, 0219206196, 0219214603, 0219245968, 0219305171, 0219305172, 0219305173, 0219305174, 0219337011, 0219349459, 0219375530, 0219375531, 0219377400, 0219390335, 0219398981, 0219422636, 0219425617, 0219433666, 0219497903, 0219518491, 0219587572, 0219587573, 0219612248, 0219612249, 0219620526, 0219621110, 0219715005, 0219717420, 0219718438, 0219718440, 0219765247, 0219765251, 0219868602, 0219890615, 0220071933, 0220075959, 0220100014, 0220256276, 0220260478, 0220263746, 0220276913, 0220297268, 0220297270, 0220316650, 0220368431, 0220368432, 0220375984, 0220428236, 0220430237, 0220584282, 0220590246, 0220590247, 0220590248, 0220590249, 0220590250, 0220590251, 0220594791, 0220594792, 0220614692, 0220648615, 0220668846, 0221036743, 0221053170, 0221184563, 0221186755, 0221227258, 0221264033, 0221264034, 0221264960, 0221272023, 0221281956, 0221281957, 0221323103, 0221430617, 0221444627, 0221491002, 0221514322, 0221514323, 0221514324, 0221514325, 0221546530, 0221566668, 0221577081, 0221594832, 0221622517, 0221649969, 0221690470, 0221741062, 0221741063, 0221755012, 0221765184, 0221802766, 0221803888, 0221803889, 0221803890, 0221827095, 0221839656, 0221848787, 0221849670, 0221849671, 0221849672, 0221850724, 0221881224, 0221894232, 0221894233, 0221972044, 0221972045, 0221972506, 0221979343, 0222006670, 0222006671, 0222025355, 0222038363, 0222054122, 0222139672, 0222139673, 0222139674, 0222139675, 0222139676, 0222139677, 0222139678, 0222139679, 0222154978, 0222154979, 0222154980, 0222154981, 0222395543, 0222395544, 0222395545, 0222396586, 0222396587, 0222396588, 0222396589, 0222396590, 0222396591, 0222396592, 0222396593, 0222396594, 0222396595, 0222396596, 0222396597, 0222396598, 0222396599, 0222396600, 0222396601, 0222412689, 0222412690, 0222418531, 0222418532, 0222418588, 0222418589, 0222418591, 0222426616, 0222427334, 0222427335, 0222427336, 0222429509, 0222580214, 0222621260, 0222621261, 0222621262, 0222621263, 0222622281, 0222629620, 0222630230, 0222630231, 0222630232, 0222646465, 0222652428, 0222653147, 0222653148, 0222658928, 0222660622, 0222660853, 0222677422, 0222684379, 0222687577, 0222687578, 0222689822, 0222689823, 0222741781, 0222741782, 0222741783, 0222742206, 0222818437, 0222818438, 0222818439, 0222818440, 0222832453, 0222845566, 0222845585, 0222845627, 0223009455, 0223009456, 0223009457, 0223009458, 0223019858, 0223023195, 0223023200, 0223023202, 0223037118, 0223037120, 0223083064, 0223083065, 0223090114, 0223091943, 0223091944, 0223091945, 0223095856, 0223095857, 0223099274, 0223121417, 0223130455, 0223257772, 0223259688, 0223259689, 0223259690, 0223259691, 0223259692, 0223259693, 0223259694, 0223259695, 0223259696, 0223259697, 0223259698, 0223259699, 0223259700, 0223259701, 0223315885, 0223316126, 0223316130, 0223316131, 0223316132, 0223316133, 0223316134, 0223316135, 0223316136, 0223316137, 0223316140, 0223316141, 0223316142, 0223316143, 0223316144, 0223316145, 0223316166, 0223316167, 0223316168, 0223316169, 0223316170, 0223316171, 0223316172, 0223354587, 0223450726, 0223450727, 0223450728, 0223450729, 0223450730, 0223450731, 0223450732, 0223450733, 0223450734, 0223450735, 0223450736, 0223450737, 0223450738, 0223450739, 0223450740, 0223450745, 0223450746, 0223450747, 0223450748, 0223450749, 0223456972, 0223456974, 0223456975, 0223456976, 0223562242, 0223562243, 0223562244, 0223562245, 0223562246, 0223562247, 0223562248, 0223562249, 0223562250, 0223562251, 0223562252, 0223562253, 0223562254, 0223562255, 0223562256, 0223562257, 0223562258, 0223562259, 0223562260, 0223562261, 0223852618, 0223852619, 0223852620, 0223852621, 0223852622, 0223852623, 0223852624, 0223859033, 0223861157, 0223861158, 0223861160, 0223861778, 0223861779, 0223861800, 0223861814, 0223861879, 0223861880, 0223861882, 0223861939, 0223861940, 0223861949, 0223861950, 0223862582, 0223862601, 0223862609, 0223862666, 0224025144, 0224034627, 0224034666, 0224034667, 0224082804, 0224082805, 0224085442, 0224085443, 0224085444, 0224085445, 0224085586, 0224085626, 0224085627, 0224085628, 0224085629, 0224085630, 0224107431, 0224107432, 0224107434, 0224107435, 0224107436, 0224107437, 0224107439, 0224107440, 0224107441, 0224107442, 0224107443, 0224107444, 0224107445, 0224107806, 0224107808, 0224134495, 0224134496, 0224134497, 0224159135, 0224159136, 0224159137, 0224159138, 0224159139, 0224159140, 0224198334, 0224216499, 0224216563, 0224216564, 0224253258, 0224253259, 0224288640, 0224288641, 0224288642, 0224288643, 0224288644, 0224288686, 0224288687, 0224288688, 0224288689, 0224288690, 0224288691, 0224288692, 0224288693, 0224288694, 0224288695, 0224380514, 0224380515, 0224380516, 0224380517, 0224380518, 0224409896, 0224409897, 0224409899, 0224409900, 0224440199, 0224440200, 0224440201, 0224440202, 0224440203, 0224440204, 0224440205, 0224440866, 0224440867, 0224440868, 0224440869, 0224440870, 0224440871, 0224440873, 0224442313, 0224442314, 0224442315, 0224442316, 0224442317, 0224442318, 0224442319, 0224442320, 0224442321, 0224442322, 0224442323, 0224442324, 0224442325, 0224449126, 0224449127, 0224449128, 0224449129, 0224449130, 0224449131, 0224449132, 0224449133, 0224449134, 0224449135, 0224449136, 0224449137, 0224449138, 0224449139, 0224449140, 0224449141, 0224449142, 0224619284, 0224619285, 0224620226, 0224620227, 0224620228, 0224678399, 0224678400, 0224678401, 0224678564, 0224678565, 0224678906, 0224678908, 0224687409, 0224712020, 0224712021, 0224712022, 0224712023, 0224736512, 0224736513, 0224736514, 0224736515, 0224736516, 0224736517, 0224736837, 0224736838, 0224736839, 0224791803, 0224791804, 0224791805, 0224799391, 0224799392, 0224799393, 0224799394, 0224799395, 0224799396, 0224799397, 0224799398, 0224799399, 0224799400, 0224799401, 0224799402, 0224799403, 0224799806, 0224799807, 0224799808, 0224799809, 0224799810, 0224799811, 0224799812, 0224799813, 0224799814, 0224799815, 0224799816, 0224799817, 0224799818, 0224799819, 0224799820, 0224799821, 0224799822, 0224799823, 0224799824, 0224799825, 0224799826, 0224799827, 0224799828, 0224799830, 0224799831, 0224799834, 0224799835, 0224799836, 0224799837, 0224799838, 0224799839, 0225136017, 0225136018, 0225136930, 0225136931, 0225136932, 0225243969, 0225243970, 0225243973, 0225243974, 0225243975, 0225243976, 0225243977, 0225243978, 0225243979, 0225243980, 0225243981, 0225243982, 0225243983, 0225243984, 0225243985, 0225244366, 0225244367, 0225244368, 0225244369, 0225244370, 0225244371, 0225244372, 0225244373, 0225244374, 0227894725, 0227894827, 0227894828, 0227894829, 0227894832, 0227894833, 0227894834, 0227894835, 0227894836, 0227894842, 0227894843, 0227894844, 0227894845, 0227894846, 0227894847, 0227894848, 0227894849, 0227894851, 0227894852, 0227894853, 0227894854, 0227894856, 0227894857, 0227894858, 0227894859, 0227894861, 0227894863, 0227894864, 0227894865, 0227894866, 0227894867, 0227894868, 0227894869, 0227894872, 0227894873, 0227894874, 0227894875, 0227894876, 0227894877, 0227894878, 0227894879, 0228118366, 0228118367, 0228118368, 0228118370, 0228208483, 0228208484, 0228208485, 0228208846, 0228208847, 0228208848, 0228208849, 0228208850, 0228208851, 0228208855, 0228208856, 0228208857, 0228208858, 0228208859, 0228208860, 0228208861, 0228208862, 0228208863, 0228208868, 0228208869, 0228208870, 0228208872, 0228208873, 0228208874, 0228208875, 0228208876, 0228208877, 0228208878, 0228208879, 0228208880, 0228208881, 0228208883, 0228208884, 0228366377, 0228366378, 0228366379, 0228366380, 0228366381, 0228366382, 0228367327, 0228367328, 0228367329, 0228367330, 0228367331, 0228367332, 0228367333, 0228367334, 0228367335, 0228367336, 0228367337, 0228367338, 0228367339, 0228367340, 0228367341, 0228367342, 0228367343, 0228367344, 0228367349, 0228367350, 0228367351, 0228367352, 0228367353, 0228367355, 0228367356, 0228367357, 0228367358, 0228367359, 0228367362, 0228367363, 0228528286, 0228528287, 0228528288); 00763000111854 (0219505890, 0219505891, 0219505892, 0219505893, 0219505894, 0219533854, 0219875956, 0219988881, 0220138915, 0220248276, 0220248277, 0220253142, 0220277098, 0220368430, 0220369379, 0220369848, 0220394562, 0220584283, 0220584284, 0220584285, 0220627630, 0220632089, 0220643658, 0220780560, 0220785734, 0220868360, 0220925498, 0220936605, 0221343503, 0221972290, 0221972291, 0222138789, 0222230549, 0222230550, 0222230551, 0222230552, 0222622587, 0222622588, 0222622589, 0222638828, 0222645597, 0222667672, 0222685841, 0222742193, 0222742194, 0222742195, 0222819361, 0222819362, 0223008932, 0223008933, 0223008934, 0223008935, 0223008936, 0223316234, 0223449358, 0223456457, 0223568237, 0223568238, 0223568239, 0223568240, 0223852943, 0224187952, 0224187953, 0224187954, 0224187957, 0224409302, 0224440323, 0224440324, 0224518549, 0224518550, 0224518551, 0224518552, 0224518553, 0224518558, 0224518559, 0224518560, 0225136822, 0225136823, 0225244426, 0225244427, 0225244428, 0227894837, 0227894881, 0227894882, 0227969113, 0228208826, 0228208827, 0228208828, 0228208829, 0228365695, 0228365696); 00763000111861 (0222172842, 0222188636, 0222194734, 0222814174, 0222814175, 0222814176, 0222814177, 0222814178, 0222814180, 0222814181, 0222814182, 0222814183, 0222814184, 0223029972, 0223029973, 0223029974, 0223029975, 0223029976, 0223089378, 0223089379, 0223089380, 0223089381, 0223089382, 0223089383, 0223089384, 0223091453, 0223091455, 0223091607, 0223091608, 0223091609, 0223091610, 0223313914, 0223313915, 0223313916, 0223313917, 0223313918, 0223313919, 0223313920, 0223313921, 0223313922, 0223450743, 0223450744, 0223553145, 0223562226, 0223562227, 0223562228, 0223562229, 0223562230, 0223562231, 0223562232, 0223562233, 0223562234, 0223562235, 0223562236, 0223562238, 0223562239, 0223562240, 0223562241, 0223650153, 0223650154, 0223650155, 0223861159, 0223861544, 0223861780, 0223861881, 0223861998, 0223862583, 0223862584, 0223862600, 0223862602, 0223862608, 0224085308, 0224085309, 0224085310, 0224085311, 0224085312, 0224085313, 0224085314, 0224085315, 0224085316, 0224085317, 0224085318, 0224085319, 0224321608, 0224330255, 0224330256, 0224330257, 0224330258, 0224330259, 0224330260, 0224330261, 0224448536, 0224448537, 0224448538, 0224448539, 0224448540, 0224448541, 0224448542, 0224448543, 0224448544, 0224448545, 0224449106, 0224449107, 0224449108, 0224449109, 0224449110, 0224449111, 0224449112, 0224449113, 0224449114, 0224449115, 0224449116, 0224449117, 0224449118, 0224449119, 0224449120, 0224449121, 0224449122, 0224449123, 0224449124, 0225007573, 0225007574, 0225009606, 0225009607, 0225009608, 0225009609, 0225009610, 0225009611, 0225009612, 0225009613, 0225244394, 0225244395, 0225244396, 0225244397, 0225244398, 0225244401, 0225244402, 0225244403, 0225244404, 0225244405, 0225244406, 0225244407, 0225244408, 0225462856, 0225669344, 0225669345, 0225670466, 0227894884, 0227894885, 0227894926, 0227894927, 0227894929, 0227894933, 0227894934, 0227894935, 0228203474, 0228203475, 0228203476, 0228203477, 0228203478, 0228203479, 0228203480, 0228203481, 0228203482, 0228203483, 0228203484, 0228203485). 2. Model Number 64002, GTIN (Serial Numbers): 00613994544957 (N692230, N709765, N709766, N709769, N712972, N714659, N720090, N721922, N723172, N724723, N724752, N730633, N732361, N732366, N732368, N732369, N732369N01, N734116, N735283, N735294, N736643, N736644, N738447, N738450, N738459, N740632, N740717, N742744, N744753, N744754, N744756, N746628, N746635, N748190, N748193, N748193N01, N750163, N750166, N753183, N755406, N755416, N755435, N757678, N757684, N758562, N760449, N760457, N760465, N782607, N784725, N784740, N784754, N784755, N786840N01, N786849, N786915N01, N786917, N786917N01, N794131, N795243, N795246, N795247, N795271, N796405, N796416, N796419, N796432, N796441, N803281); 00613994604446 (0212745840, 0212749749, 0212754210, 0212761030, 0212761032, 0212761035, 0212761036, 0212761037, 0212761038, 0212774118, 0212774188, 0212774189, 0212774190, 0212774308, 0212782600, 0212782601, 0212782602, 0212784261, 0212785056, 0212785057, 0212787201, 0212787202, 0212797383, 0212797384, 0212798912, 0212804685, 0212804686, 0212815392, 0212815398, 0212846338, 0212847507, 0212847508, 0212847509, 0212850282, 0212850283, 0212850293, 0212850294, 0212852696, 0212852697, 0212856040, 0212862899, 0212863596, 0212864738, 0212875923, 0212878577, 0212878578, 0212881521, 0212885470, 0212885471, 0212885472, 0212885473, 0212886132, 0212894684, 0212894685, 0212894694, 0212894695, 0212894697, 0212895260, 0212895261, 0212895262, 0212895263, 0212895264, 0212895265, 0212900727, 0212901662, 0212905207, 0212905218, 0212905686, 0212906223, 0212922356, 0212922358, 0212922421, 0212926120, 0212926121, 0212928704, 0212929653, 0212929654, 0212929655, 0212929656, 0212929657, 0212933033, 0212933034, 0212933035, 0212933036, 0212933037, 0212933060, 0212933061, 0212933062, 0212933063, 0212933064, 0212933065, 0212933066, 0212933073, 0212935226, 0212940808, 0212940815, 0212956691, 0212963491, 0212963493, 0212963494, 0212963496, 0212963497, 0212965472, 0212965473, 0212965474, 0212965475, 0212965476, 0212965477, 0212965478, 0212965970, 0212966920, 0212966921, 0212970794, 0212970795, 0212971195, 0212971199, 0212971201, 0212971202, 0212971457, 0212980001, 0212983883, 0212983886, 0212996333, 0213001235, 0213001236, 0213013164, 0213017632, 0213017633, 0213017634, 0213017635, 0213021468, 0213021471, 0213021474, 0213023907, 0213037782, 0213037784, 0213037785, 0213039387, 0213071412, 0213071413, 0213071414, 0213071415, 0213071416, 0213071417, 0213071418, 0213071435, 0213071439, 0213071500, 0213071501, 0213071542, 0213071544, 0213071545, 0213071546, 0213071550, 0213071551, 0213072388, 0213072389, 0213072390, 0213079903, 0213079904, 0213080289, 0213080579, 0213085817, 0213085818, 0213085900, 0213085901, 0213085902, 0213085903, 0213091248, 0213094930, 0213094931, 0213095020, 0213095951, 0213097887, 0213097888, 0213097889, 0213097890, 0213103760, 0213103762, 0213103764, 0213103765, 0213103769, 0213115057, 0213115059, 0213115432, 0213127808, 0213127809, 0213127810, 0213127811, 0213127812, 0213127813, 0213127814, 0213127815, 0213127816, 0213127817, 0213127818, 0213127819, 0213127940, 0213132148, 0213136860, 0213136861, 0213136862, 0213136866, 0213136868, 0213147262, 0213147269, 0213147270, 0213147271, 0213147272, 0213147273, 0213147274, 0213174551, 0213182376, 0213182377, 0213182378, 0213191646, 0213191647, 0213192551, 0213192552, 0213192553, 0213237284, 0213237290, 0213413947, 0213413948, 0213413949, 0213413950, 0213413953, 0213413955, 0213413956, 0213413957, 0213413958, 0213413959, 0213417621, 0213417622, 0213429219, 0213429300, 0213429301, 0213433410, 0213445371, 0213445372, 0213452791, 0213452792, 0213452793, 0213456732, 0213456733, 0213456734, 0213456735, 0213456737, 0213462568, 0213462569, 0213462570, 0213462571, 0213462572, 0213462573, 0213462575, 0213466511, 0213467389, 0213473457, 0213473458, 0213578992, 0213578993, 0213583797, 0213583802, 0213584203, 0213588224, 0213588225, 0213588227, 0213605259, 0213632473, 0213632474, 0213632475, 0213632476, 0213632478, 0213632479, 0213634655, 0213634657, 0213634658, 0213634659, 0213649938, 0213649939, 0213650135, 0213650136, 0213650137, 0213650139, 0213650142, 0213650143, 0213650144, 0213650145, 0213650146, 0213650147, 0213650148, 0213650149, 0213650150, 0213650151, 0213650152, 0213650153, 0213650154, 0213650157, 0213650161, 0213650162, 0213650163, 0213650164, 0213650165, 0213650166, 0213650167, 0213650168, 0213650169, 0213650170, 0213650171, 0213650172, 0213650173, 0213650174, 0213650175, 0213650177, 0213650178, 0213650179, 0213650180, 0213650181, 0213650182, 0213650183, 0213650184, 0213650185, 0213650186, 0213650187, 0213650188, 0213650189, 0213650190, 0213650191, 0213650192, 0213650193, 0213650194, 0213650195, 0213650196, 0213650197, 0213650198, 0213650199, 0213650200, 0213650201, 0213650202, 0213650203, 0213650204, 0213650205, 0213650206, 0213650208, 0213650209, 0213650210, 0213650214, 0213650215, 0213650216, 0213650217, 0213650219, 0213650220, 0213650221, 0213650222, 0213650223, 0213650224, 0213650225, 0213650226, 0213650228, 0213650230, 0213650231, 0213650232, 0213650233, 0213650234, 0213650235, 0213650238, 0213650240, 0213650242, 0213650247, 0213650249, 0213650251, 0213650253, 0213650255, 0213650257, 0213650259, 0213650261, 0213650264, 0213650266, 0213650268, 0213650271, 0213650273, 0213650275, 0213650278, 0213650279, 0213650280, 0213650281, 0213650285, 0213650287, 0213650290, 0213650292, 0213650295, 0213650297, 0213650299, 0213650301, 0213650303, 0213650305, 0213650306, 0213650307, 0213650308, 0213650309, 0213650310, 0213650311, 0213650312, 0213650313, 0213650314, 0213650315, 0213650316, 0213650317, 0213650318, 0213650319, 0213650320, 0214103390, 0214103393, 0214468589, 0214468590, 0214468591, 0214468593, 0214468594, 0214468595, 0214468596, 0214468597, 0214468598, 0214468761, 0214468762, 0214468763, 0214468765, 0214468904, 0214468905, 0214468906, 0214468907, 0214468911, 0214469578, 0214469779, 0214470704, 0214470705, 0214475642, 0214485655, 0214485659, 0214485883, 0214486991, 0214496362, 0214496363, 0214496364, 0214518188, 0214518189, 0214519754, 0214528301, 0214528302, 0214528303, 0214528304, 0214528305, 0214528306, 0214531735, 0214531779, 0214540438, 0214540439, 0214554705, 0214554706, 0214557416, 0214557523, 0214566932, 0214566933, 0214566934, 0214573745, 0214573746, 0214573747, 0214573756, 0214574071, 0214578640, 0214585623, 0214590766, 0214603806, 0214603807, 0214609630, 0214609634, 0214610532, 0214610533, 0214611925, 0214616558, 0214616559, 0214620666, 0214621843, 0214621845, 0214621846, 0214624371, 0214624818, 0214624820, 0214625801, 0214625802, 0214647188, 0214647844, 0214648116, 0214648117, 0214649108, 0214649111, 0214649112, 0214653962, 0214653963, 0214657197, 0214692218, 0214692219, 0214697222, 0214700826, 0214701799, 0214701800, 0214712361, 0214712362, 0214808030, 0214816022, 0214816031, 0214816032, 0214824545, 0214837530, 0214842559, 0214865328, 0214865329, 0214865330, 0214865331, 0214865332, 0214874944, 0214882639, 0214882640, 0214884661, 0214887162, 0214887163, 0214899505, 0214915786, 0214923073, 0214932550, 0214934114, 0214935059, 0214935060, 0214935832, 0214941176, 0214947559, 0214947685, 0214947687, 0214947688, 0214954015, 0214959057, 0214974833, 0214974911, 0214974913, 0214975112, 0214975660, 0214975872, 0214975873, 0214975874, 0214975875, 0214975876, 0214975878, 0214975879, 0215009283, 0215031113, 0215031114, 0215033187, 0215035470, 0215035471, 0215066747, 0215066748, 0215066749, 0215076639, 0215076640, 0215076881, 0215076882, 0215076883, 0215099587, 0215099588, 0215099596, 0215099597, 0215099598, 0215100079, 0215100407, 0215112757, 0215112758, 0215112759, 0215112760, 0215112804, 0215112805, 0215116764, 0215116765, 0215116766, 0215116767, 0215116768, 0215116769, 0215116770, 0215126257, 0215126258, 0215132780, 0215141342, 0215141345, 0215146156, 0215146157, 0215156752, 0215156753, 0215167405, 0215167939, 0215167940, 0215170722, 0215170723, 0215170725, 0215170727, 0215171038, 0215172599, 0215176082, 0215198093, 0215198094, 0215198095, 0215201448, 0215201449, 0215201450, 0215226495, 0215226622, 0215226623, 0215226624, 0215226625, 0215227175, 0215233911, 0215233920, 0215234483, 0215245873, 0215253204, 0215253298, 0215255744, 0215255747, 0215259531, 0215259532, 0215268457, 0215268459, 0215271857, 0215271858, 0215271859, 0215271915, 0215272177, 0215272179, 0215273544, 0215273547, 0215273552, 0215273555, 0215280281, 0215280283, 0215280933, 0215282537, 0215283796, 0215283799, 0215283941, 0215288570, 0215288574, 0215288831, 0215356493, 0215356494, 0215356496, 0215356498, 0215356499, 0215356500, 0215358261, 0215358264, 0215358265, 0215358266, 0215358268, 0215358269, 0215358270, 0215358282, 0215358283, 0215358284, 0215358312, 0215358328, 0215358329, 0215358331, 0215358343, 0215358344, 0215358345, 0215358346, 0215358347, 0215358351, 0215358352, 0215358353, 0215358354, 0215358358, 0215359183, 0215359191, 0215359270, 0215359280, 0215361735, 0215361736, 0215372105, 0215385658, 0215385710, 0215385714, 0215385716, 0215385717, 0215385718, 0215385720, 0215385721, 0215385723, 0215385724, 0215385725, 0215385727, 0215385728, 0215385729, 0215385734, 0215385736, 0215385739, 0215385744, 0215385745, 0215385746, 0215385747, 0215385748, 0215385750, 0215385755, 0215385756, 0215385758, 0215385759, 0215385761, 0215385762, 0215385765, 0215385767, 0215385769, 0215385770, 0215385772, 0215385773, 0215385774, 0215385775, 0215385776, 0215385777, 0215385779, 0215385780, 0215385781, 0215385782, 0215385783, 0215385787, 0215385788, 0215385789, 0215385790, 0215385792, 0215392463, 0215392482, 0215392483, 0215392484, 0215392510, 0215392511, 0215392526, 0215392527, 0215392535, 0215392536, 0215392537, 0215392539, 0215392544, 0215472731, 0215472732, 0215472738, 0215472739, 0215472740, 0215474021, 0215474022, 0215474023, 0215474024, 0215474025, 0215475876, 0215477383, 0215477384, 0215477385, 0215499417, 0215501735, 0215501736, 0215522999, 0215523025, 0215523041, 0215523042, 0215523066, 0215589555, 0215589556, 0215596418, 0215606813, 0215606814, 0215614571, 0215657060, 0215700166, 0215700167, 0215700168, 0215819667, 0215819668, 0215934633, 0215934634, 0215936463, 0215942316, 0215942318, 0215954998, 0215954999, 0215955000, 0215964508, 0215972279, 0215978930, 0215988218, 0215988219, 0216004477, 0216004478, 0216004479, 0216004480, 0216016139, 0216018624, 0216018625, 0216065724, 0216111604, 0216156711, 0216171130, 0216171132, 0216171133, 0216171135, 0216171136, 0216171137, 0216206937, 0216206938, 0216206939, 0216206940, 0216207896, 0216207897, 0216207898, 0216208179, 0216208987, 0216209161, 0216209162, 0216462635, 0216547664, 0216550177, 0216561981, 0216579960, 0216591414, 0216592122, 0216624039, 0216624040, 0216640718, 0216640719, 0216640748, 0216644828, 0216644829, 0216658945, 0216661538, 0216661540, 0216668532, 0216668533, 0216670436, 0216670438, 0216683902, 0216683903, 0216683904, 0216683905, 0216683907, 0216683908, 0216683909, 0216691749, 0216701593, 0216707670, 0216730627, 0216748843, 0216748844, 0216748845, 0216748846, 0216766995, 0216766997, 0216777156, 0216781561, 0216782770, 0216789797, 0216790477, 0216812585, 0216833887, 0216866102, 0216866103, 0216880723, 0216888139, 0216888140, 0216903625, 0216903626, 0216934315, 0216943408, 0216943409, 0216970112, 0217000196, 0217000197, 0217014298, 0217014299, 0217039819, 0217046518, 0217101498, 0217101933, 0217101965, 0217103057, 0217134103, 0217134104, 0217134105, 0217148781, 0217157510, 0217178616, 0217178618, 0217178619, 0217187787, 0217220596, 0217221543, 0217244637, 0217249975, 0217249976, 0217287997, 0217307860, 0217307861, 0217311998, 0217311999, 0217324597, 0217335743, 0217335744, 0217342991, 0217342992, 0217346263, 0217346264, 0217348324, 0217348325, 0217375806, 0217375807, 0217375808, 0217375809, 0217375810, 0217375811, 0217375812, 0217375819, 0217399630, 0217399633, 0217399635, 0217399636, 0217403404, 0217412166, 0217422238, 0217422239, 0217425679, 0217440507, 0217440508, 0217440509, 0217440510, 0217483873, 0217484070, 0217486062, 0217486063, 0217486064, 0217497498, 0217497499, 0217497500, 0217534872, 0217560564, 0217560565, 0217566377, 0217600191, 0217600787, 0217612437, 0217635208, 0217635209, 0217635210, 0217664212, 0217727091, 0217759260, 0217778420, 0217782462, 0217795041, 0217809103, 0217820834, 0217821658, 0217830939, 0217843819, 0217863004, 0217873901, 0217874944, 0217881301, 0217911512, 0217911513, 0217933185, 0217939586, 0217977438, 0217986426, 0217992947, 0218003574, 0218003575, 0218011116, 0218019871, 0218022443, 0218028484, 0218040503, 0218043348, 0218043349, 0218061176, 0218086090, 0218086091, 0218104522, 0218122969, 0218156717, 0218156719, 0218156720, 0218156724, 0218156725, 0218156866, 0218156867, 0218178504, 0218282737, 0218297499, 0218297500, 0218297501, 0218297502, 0218299923, 0218299924, 0218299925, 0218441806, 0218441807, 0218513547, 0218513548, 0218513551, 0218513552, 0218513553, 0218523150, 0218775058, 0218803830, 0218815563, 0218977414, 0219008605, 0219014647, 0219048001, 0219048002, 0219048003, 0219049209, 0219050882, 0219065352, 0219065354, 0219065356, 0219065357, 0219093828, 0219144070, 0219176011, 0219191926, 0219199581, 0219201858, 0219203067, 0219203068, 0219226703, 0219248385, 0219248467, 0219248468, 0219248469, 0219248470, 0219248471, 0219248472, 0219288457, 0219289452, 0219289453, 0219289454, 0219289456, 0219289457, 0219370743, 0219370744, 0219374030, 0219375167, 0219375168, 0219393944, 0219403090, 0219423288, 0219441239, 0219465091, 0219465092, 0219465093, 0219465096, 0219468607, 0219468608, 0219468609, 0219472735, 0219472736, 0219472737, 0219535125, 0219541406, 0219541410, 0219541415, 0219573124, 0219605793, 0219627906, 0219627907, 0219687519, 0219697770, 0219724240, 0219724241, 0219724242, 0219726123, 0219726125, 0219737378, 0219899190, 0219899191, 0219916595, 0219916646, 0220037057, 0220037058, 0220044030, 0220044031, 0220044046, 0220184998, 0220184999, 0220185000, 0220185001, 0220185003, 0220185004, 0220185005, 0220187566, 0220187569, 0220187570, 0220187571, 0220187572, 0220187576, 0220187577, 0220187578, 0220187579, 0220187580, 0220187581, 0220187582, 0220187583, 0220187584, 0220187585, 0220187586, 0220187587, 0220622508, 0220629133, 0220629134, 0220629135, 0220629137, 0220629179, 0220629180, 0220629181, 0220629182, 0220629183, 0220629446, 0220629447, 0220629448, 0220629449, 0220629466, 0220629467, 0220629469, 0220629470, 0220629471, 0220629472, 0220629473, 0220629474, 0220629475, 0220629476, 0220629477, 0221034114, 0221034115, 0221034117, 0221034118, 0221034119, 0221053080, 0221316285, 0221610530, 0223192469, 0223192470, 0223192590, 0223193556, 0223200856); 00613994604453 (0212741247, 0212741248, 0212741249, 0212742419, 0212742483, 0212745770, 0212785058, 0212815602, 0212828591, 0212843763, 0212956684, 0212970796, 0213071419, 0213071540, 0213071541, 0213073473, 0213078456, 0213127807, 0213174305, 0213237292, 0213250893, 0213250894, 0213269072, 0213284733, 0213285393, 0213299619, 0213300272, 0213300273, 0213300279, 0213300280, 0213300281, 0213300282, 0213300283, 0213402619, 0213402861, 0213402865, 0213402866, 0213402867, 0213417326, 0213445379, 0213452794, 0213456725, 0213456726, 0213489219, 0213501865, 0213516313, 0213525256, 0213525259, 0213525300, 0213525307, 0213525308, 0213525310, 0213578174, 0213583801, 0213608287, 0213608288, 0213619023, 0213625015, 0213641249, 0213751757, 0213751758, 0213757301, 0213757302, 0213762800, 0213764861, 0213764862, 0213764863, 0213771966, 0213793175, 0213849987, 0214091533, 0214104013, 0214167671, 0214296666, 0214296667, 0214299340, 0214311302, 0214311303, 0214311304, 0214311307, 0214311308, 0214311309, 0214311318, 0214311319, 0214311341, 0214616550, 0214624816, 0214624817, 0214700827, 0214956585, 0215156747, 0215206629, 0215243796, 0216065728, 0216311195, 0216508909, 0216508910, 0216508911, 0216518391, 0216518392, 0217013121, 0217013122, 0217013123, 0217013124, 0217013125, 0217586255, 0217644986, 0217701988, 0217769612, 0217830725, 0217954517, 0218131774, 0218985898, 0219009806, 0219023827, 0219023828, 0219029040, 0219399824, 0219399825, 0219402886, 0219728126, 0220211195, 0220831736, 0221149643, 0221215835, 0221215836); 00613994604460 (0212803086, 0212894135, 0212894136, 0212894137, 0212894138, 021
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2427-2024
Related Recalls
Medtronic 1x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64001; Stimulator, electrical, implanted, for parkinsonian tremor
Medtronic Neuromodulation
A limited number of Pocket Adaptor Kits, Model 64001 and 64002, are labeled with incorrect Use-By-Dates.
Software issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusion steps should automatically be arranged in consecutive order, from earliest to latest. However, in rare cases, the steps may be displayed out of order. If this occurs, the programmed Flex Infusion schedule will not be delivered at the intended time, if at all.
A71200 Vanta" Clinician Programmer Application (CP App) version 2.0.2465 and 2.0.2683, Product number A71200. Software is deployed and operated on Samsung tablets (CFN CT900).
Medtronic Neuromodulation
Complaints received that Vanta A71200 CP App does not function as intended during use. There is a potential for prolonged or postponed surgical procedure.