Datascope

129 recalls Categories: Infusion Pumps, Other Medical Devices, Surgical Instruments, Diagnostic Equipment

Users were reporting that the device was not charging as expected. It was discovered that users were unaware that the Cardiosave console was not completely inserted into the hospital cart. If the console is not fully inserted back into the cart the battery(ies) will not charge.

Jul 31, 2023 Infusion Pumps Nationwide View Details →

Users were reporting instances of "Gas Loss in IAB Circuit" and "Gas Gain in IAB Circuit" alarms while providing therapy. An internal investigation of the complaints determined that there is a potential trigger for these alarms that was not listed in the IFU: Patient Movement (coughing, general movement, and swallowing). Blood in the circuit, kins and abrasions of the tubing/catheter and certain patient conditions could also trigger these alarms. The HHE also identified hardware issues related to these alarms.

Jul 31, 2023 Infusion Pumps Nationwide View Details →

Users reported "System Over Temperature" alarms associated with a loss of pumping and/or the Cardiosave system entering Standby mode.

Jul 31, 2023 Infusion Pumps Nationwide View Details →

Users were reporting instances of "Gas Loss in IAB Circuit" and "Gas Gain in IAB Circuit" alarms while providing therapy. An internal investigation of the complaints determined that there is a potential trigger for these alarms that was not listed in the IFU: Patient Movement (coughing, general movement, and swallowing). Blood in the circuit, kinks/abrasions of the tubing/catheter and certain patient conditions could also trigger these alarms. The firm's HHE also identified hardware issues related to these alarms.

Jul 31, 2023 Infusion Pumps Nationwide View Details →

Users were experiencing a failure in the IAB Fiber Optic Sensor input on the IABP when inserting the Intra-Aortic Balloon Fiber Optic connector.

Jul 31, 2023 Infusion Pumps Nationwide View Details →

LINEAR 7.5Fr.IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES D684-00-0479-01 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (APA) D684-00-0479-02 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (CHINA) D684-00-0479-01C LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (APA)(CHINA) D684-00-0479-02C LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (INDIA) D684-00-0479-07 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (APA) (INDIA) D684-00-0479-08 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (TURKEY) D684-00-0479-09 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES WITH APA (TURKEY) D684-00-0479-10 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES, US ONLY D684-00-0479-01U LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES, (APA) US ONLY D684-00-0479-02U LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES D684-00-0480-01 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (APA) D684-00-0480-02 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (CHINA) D684-00-0480-01C LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (APA)(CHINA) D684-00-0480-02C LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (INDIA) D684-00-0480-07 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (APA)(INDIA) D684-00-0480-08 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (TURKEY) D684-00-0480-09 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES WITH APA (TURKEY) D684-00-0480-10 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES, US ONLY D684-00-0480-01U LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES, (APA) US ONLY D684-00-0480-02U LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES D684-00-0478-01 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (APA) D684-00-0478-02 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (CHINA) D684-00-0478-01C LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (APA) (CHINA) D684-00-0478-02C LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (INDIA) D684-00-0478-07 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (APA) (INDIA) D684-00-0478-08 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (TURKEY) D684-00-0478-09 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES WITH APA (TURKEY) D684-00-0478-10 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES, US ONLY D684-00-0478-01U LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (APA) US ONLY D684-00-0478-02U PACKAGED INSERTION KIT - LINEAR 7.5 Fr. 25 cc IABs D884-00-0019-12 PACKAGED INSERTION KIT - LINEAR 7.5 Fr. 34/40 cc IABs D884-00-0019-13

Class I - Dangerous

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

Jun 23, 2023 Surgical Instruments Nationwide View Details →

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

Jun 23, 2023 Surgical Instruments Nationwide View Details →

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

Jun 23, 2023 Surgical Instruments Nationwide View Details →

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

Jun 23, 2023 Surgical Instruments Nationwide View Details →

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

Jun 23, 2023 Surgical Instruments Nationwide View Details →

MEGA 7.5Fr.IAB , Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES D684-00-0295-01 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES (APA) D684-00-0295-02 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES (W/O STATLOCK) D684-00-0295-03 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES (CHINA) D684-00-0295-05 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES AND APA (CHINA) D684-00-0295-06 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES W/O STATLOCK - INDIA D684-00-0295-07 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES W/O STATLOCK (APA) - INDIA D684-00-0295-08 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES (TURKEY) D684-00-0295-09 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES (APA) (TURKEY) D684-00-0295-10 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES, US ONLY D684-00-0295-01U MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES APA, US ONLY D684-00-0295-02U MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES D684-00-0294-01 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES (APA) D684-00-0294-02 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES (W/O STATLOCK) D684-00-0294-03 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES (CHINA) D684-00-0294-05 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES AND APA (CHINA) D684-00-0294-06 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES W/O STATLOCK - INDIA D684-00-0294-07 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES W/O STATLOCK (APA) INDIA D684-00-0294-08 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES (TURKEY) D684-00-0294-09 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES (APA) (TURKEY) D684-00-0294-10 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES, US ONLY D684-00-0294-01U MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES APA, US ONLY D684-00-0294-02U PACKAGED INSERTION KIT - MEGA 7.5 Fr. 30/40 cc IABs D884-00-0019-21

Class I - Dangerous

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

Jun 23, 2023 Surgical Instruments Nationwide View Details →

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

Jun 23, 2023 Surgical Instruments Nationwide View Details →

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

Jun 23, 2023 Surgical Instruments Nationwide View Details →

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

Jun 23, 2023 Surgical Instruments Nationwide View Details →

IABP may lose the ability to charge batteries in one or both bay slots. Therapy may be interrupted if batteries fail to charge and the device is disconnected from AC power. Low battery alarms may alert the User to the issue prior to interruption of therapy.

Jun 5, 2023 Infusion Pumps Nationwide View Details →

An unexpected shutdown of the IABP may occur due to failures of tantalum capacitors in the Power Management Board and/or Solenoid Board. This issue may lead to an unexpected interruption of therapy.

Jun 5, 2023 Infusion Pumps Nationwide View Details →

There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to slightly reduced IAB inflation during therapy or a reduction in the degree of augmentation provided to the patient during therapy.

Feb 28, 2023 Infusion Pumps View Details →

There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to slightly reduced IAB inflation during therapy or a reduction in the degree of augmentation provided to the patient during therapy.

Feb 28, 2023 Infusion Pumps View Details →

There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to slightly reduced IAB inflation during therapy or a reduction in the degree of augmentation provided to the patient during therapy.

Feb 28, 2023 Infusion Pumps View Details →

There have been reported failures of the high pressure helium regulator, which may cause a helium leak in the Cardiosave Hospital Cart. In instance of helium regulator failure, a Pump Console s internal reservoir of helium will not be replenished when docked into an impacted Hospital Cart. This may result in an insufficient amount of helium within the internal reservoir.

Feb 7, 2023 Infusion Pumps Nationwide View Details →

Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted.

Feb 7, 2023 Infusion Pumps Nationwide View Details →

Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted.

Feb 7, 2023 Other Medical Devices Nationwide View Details →

An unexpected shutdown of the IABP may occur due to loss of communication between the Executive Processor PCBA and the Video Generator PCBA. This may result in an interruption to therapy which may threaten the hemodynamic stability of the supported patient as the user is left unaware to the status of the Cardiosave IABP.

Feb 7, 2023 Other Medical Devices Nationwide View Details →

There have been reported failures of the high pressure helium regulator, which may cause a helium leak in the Cardiosave Hospital Cart. In instance of helium regulator failure, a Pump Console s internal reservoir of helium will not be replenished when docked into an impacted Hospital Cart. This may result in an insufficient amount of helium within the internal reservoir.

Feb 7, 2023 Infusion Pumps Nationwide View Details →

An unexpected shutdown of the IABP may occur due to loss of communication between the Executive Processor PCBA and the Video Generator PCBA. This may result in an interruption to therapy which may threaten the hemodynamic stability of the supported patient as the user is left unaware to the status of the Cardiosave IABP.

Feb 7, 2023 Other Medical Devices Nationwide View Details →

An unexpected shutdown of the IABP may occur due to a failure of the connection between the Coiled Cord cable and the Cable Assembly backplane to the Coiled Cord cable which provides the communication between the display head and base unit.

Feb 7, 2023 Infusion Pumps Nationwide View Details →

The Cardiosave IABP may shut down unexpectedly due to blood entering in the Cardiosave IABP when therapy is provided with a compromised intra-aortic balloon catheter.

Dec 19, 2022 Infusion Pumps Nationwide View Details →

Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic instability.

Nov 15, 2021 Infusion Pumps Nationwide View Details →

A Cardiosave Lithium-Ion Battery Pack used during protocol testing failed to meet the minimum runtime requirement per Getinge internal Product Specification. These nonconforming batteries were inadvertently released to customers.

Sep 21, 2021 Infusion Pumps Nationwide View Details →

The Cardiosave IABP may unexpectedly shut down when the device is running on AC power, only one battery is installed in the IABP, and the battery is physically removed while the battery is being charged. This occurs during a very specific set of conditions.

Sep 8, 2021 Infusion Pumps Nationwide View Details →

There are cybersecurity vulnerabilities in a widely used low-level TCP/IP software library that may result in a loss of communication to the Hospital Information System/Clinical Information System (HIS/CIS).

Mar 22, 2021 Other Medical Devices Nationwide View Details →

There are cybersecurity vulnerabilities in a widely used low-level TCP/IP software library that may result in a loss of communication to the Hospital Information System/Clinical Information System (HIS/CIS).

Mar 22, 2021 Other Medical Devices Nationwide View Details →

The device was distributed with the outer carton labeled with the manufacture date and the expiry date reversed. This showed the product as expired before it was manufactured.

May 28, 2019 Surgical Instruments View Details →

There is a potential for the over-pressure safety (OPS) valve in the tubing kits to not allow flow through.

Jul 27, 2017 Other Medical Devices View Details →

There is a potential for the over-pressure safety (OPS) valve in the tubing kits to not allow flow through.

Jul 27, 2017 Other Medical Devices View Details →

There is a potential for the over-pressure safety (OPS) valve in the tubing kits to not allow flow through.

Jul 27, 2017 Other Medical Devices View Details →

The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.

Oct 19, 2017 Other Medical Devices View Details →

The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.

Oct 19, 2017 Other Medical Devices View Details →

The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.

Oct 19, 2017 Other Medical Devices View Details →

The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.

Oct 19, 2017 Other Medical Devices View Details →

The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.

Oct 19, 2017 Other Medical Devices View Details →

The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.

Oct 19, 2017 Other Medical Devices View Details →