Cardiosave Intra-Aortic Balloon Pump (IABP) Pneumatic Interface Module (PIM) , Model Number 0997-00-1178
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial Numbers: MH353075E2 MH353057E2 MH353067E2 MH353077E2 MH356112G2 MH353056E2 MH356111G2 MH356119G2 MH353065E2 MH356117G2
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Datascope Corp.
- Reason for Recall:
- There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to slightly reduced IAB inflation during therapy or a reduction in the degree of augmentation provided to the patient during therapy.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Cardiosave Intra-Aortic Balloon Pump (IABP) Pneumatic Interface Module (PIM) , Model Number 0997-00-1178
Product Codes/Lot Numbers:
Serial Numbers: MH353075E2 MH353057E2 MH353067E2 MH353077E2 MH356112G2 MH353056E2 MH356111G2 MH356119G2 MH353065E2 MH356117G2
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1279-2023
Related Recalls
The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in the left battery bay limits full insertion of certain batteries and prevents proper charging. The right bay of the Battery Charging Station is not affected by this issue.
The firm has identified that the battery runtime and cycle specifications contained in the devices Instructions for Use (IFU) need to be updated.