All Recalls

95,018 total recalls in our database

🏥 Medical Devices Class I - Dangerous

During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen freeze), and/or printer disconnections was identified. The consequence of these issues would be a delay in reporting results due to the required system shutdowns to recover the system. This delay in reporting results could occur during any assay, including which the speed of the result is deemed critical or STAT and could result in a delay of treatment in a patient. Due to the release strategy of this software version, the user could not update the system but a field engineer had to manually install each affected system.

Oct 10, 2023 Other Medical Devices Nationwide View Details →

STERILE PVP SOLUTION, REF DYNDA1907

MEDLINE INDUSTRIES, LP - Northfield

🏥 Medical Devices Class I - Dangerous

The product description on the labeling includes the word "Sterile" or STRL", however the contents of the solution inside the foil sachet is non-sterile.

Nov 3, 2023 Other Medical Devices Nationwide View Details →

Use of a 1.4 drill in hard bone conditions caused increased insertion forces leading to subsequent failures of the inserter, when the inserters bend, it may also move off the drill hole location leading to damage of the suture construct during insertion. Suture damage may lead to suture breaks during use.

Oct 25, 2023 Implants & Prosthetics Nationwide View Details →