DVOT SARS-CoV-2 ANTIGEN TEST KIT, Model Number FCY-AGT-01
FENG CHUN YUAN MEDICAL EQUIPMENT (SHENZHEN)CO.,LTD
The "DVOT brand" rapid test kits have not obtained the necessary authorization to be sold in the United States.
95,075 total recalls in our database
FENG CHUN YUAN MEDICAL EQUIPMENT (SHENZHEN)CO.,LTD
The "DVOT brand" rapid test kits have not obtained the necessary authorization to be sold in the United States.
FENG CHUN YUAN MEDICAL EQUIPMENT (SHENZHEN)CO.,LTD
The "DVOT brand" rapid test kits have not obtained the necessary authorization to be sold in the United States.
Mercury Enterprises, Inc.
Manual resuscitator circuit flow controller may come apart due to a missing manufacturing step.
Manufactured with a misaligned axis resulting in lenses with the incorrect power, may experience poor visual acuity
Philips North America
If the DIVA touch display generates touch inputs without user interaction, due to possibility of the erroneous acknowledgement of a patient alarm, there is a potential for a delay in the detection of a change in a patient condition which prompted the alarm to generate. Although unlikely, this could potentially result in a patient harm.
Tornier
Devices from one lot of Perform" Reversed Inserts may contain a locking ring that was assembled upside down which would prevent the humeral insert from securely seating in the stem or spacer.
Siemens Healthcare Diagnostics
Positive bias in patient results at and below the cut-off of 60 U/mL with the lots listed in the table below when compared to alternate lots on the Atellica IM analyzer.
Becton, Dickinson and Company, BD Bio Sciences
Sample preparation system with software may cause: 1) Multidispense feature enabled with 2-tube assay: Specimen volume dispensed into tube1 at least 3% higher than volume in tube2, or 2) Absolute counts calculated using Trucount Tubes: increased volume in tube1 may result in 20% higher absolute counts in tube 1 than tube 2; could result in additional blood collection, delayed results/patient therapy.
Siemens Healthcare Diagnostics
Positive bias in patient results at and below the cut-off of 60 U/mL with the lots listed in the table below when compared to alternate lots on the Atellica IM analyzer.
Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA.
Philips North America
The 866173 lntelliVue G7m Anesthesia Gas Modules could have a component issue that may trigger an interruption of the gas measurement. This issue concerns the G7m's sample gas pump which could stop working, producing a technical, INOP (inoperable) alarm. The pump may fail prematurely when there is friction caused by an abrasion. This recall is for an IMPORTANT PRODUCT NOTICE dated 1/18/23.
Medtronic MiniMed
Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA.
Pressure sensitive adhesive hooks attached to the sides of the anterior cervical orthosis collar panel can occasionally exhibit poor adhesion, which may lead to them gradually peeling off with repeated doffing, possibly leading to reduced immobilization of the cervical spine.
M & P Food Productions
Products contain undeclared sub ingredients: wheat and milk
Undeclared allergen (Fish) Ingredient
Skagit River Bakery & Cafe
Undeclared soy lecithin and soybean oil on label
AUTHENTIK FUDGE
Product may be contaminated with mold.
M & P Food Productions
Products contain undeclared sub ingredient: wheat.
M & P Food Productions
Products contain undeclared sub ingredient wheat.
M & P Food Productions
Products contain undeclared sub ingredients: wheat and milk.