All Recalls

95,123 total recalls in our database

The Olympic Cool-Cap's Control Module has experienced a frozen screen during use. When this occurs, the on-screen information remains on display, but the system is no longer providing cooling treatment to the infants. The visible clock displayed at the upper right hand corner of the screen stops advancing.

May 9, 2012 Diagnostic Equipment Nationwide View Details β†’

The presence of plastic flash was identified at the tip of certain lots of Sarns Soft-Flow Aortic Cannula. The plastic flash has the potential to detach and contribute to an adverse patient effect. While this potential failure has not been reported from the field as an observed defect, it was determined that the product does not meet specifications and this condition could result in an unacceptabl

Nov 21, 2012 Other Medical Devices Nationwide View Details β†’

Spectrolyse PAI-1 activity assay, producing lower than expected absorbance values 0 IU/ml standard reducing the slope of the standard curve requiring more frequent sample dilutions.

Oct 24, 2012 Diagnostic Equipment Nationwide View Details β†’

The items in this lot are missing the ion implantation feature. The surface of the head may be softer and more prone to scratching, which in tum could potentially cause higher wear of the poly bearing. Increased wear could potentially lead to a shorter implant life.

Nov 1, 2012 Implants & Prosthetics Nationwide View Details β†’

The four screws that secure the patient table top to the carrier pulled out and caused the table top to detach. This issue may affect NeuViz Dual and NeuViz 16 scanners.

Oct 18, 2012 Diagnostic Equipment Nationwide View Details β†’

The presence of plastic flash was identified at the tip of certain lots of Sarns Soft-Flow Aortic Cannula. The plastic flash has the potential to detach and contribute to an adverse patient effect. While this potential failure has not been reported from the field as an observed defect, it was determined that the product does not meet specifications and this condition could result in an unacceptabl

Nov 21, 2012 Other Medical Devices Nationwide View Details β†’
πŸ₯ Medical Devices Class I - Dangerous

Firm has become aware of an unintended behavior when using syngo.plaza and issued an Update Instruction. The Update Instruction SY050/12/S provides a software update and addresses the following issues: - Patient Rename messages for the same patient are being continually processed for several days. This fills up the 'pcvdbsrv' log and potentially affects performance of the system. - Data could

Jun 20, 2012 Other Medical Devices Nationwide View Details β†’

Custom Procedural trays contain Stryker Hytrel Togas which are being recalled because the clear tape that is intended to aid in the prevention of patient's fluids from contacting the surgeon/surgical staff through the sleeve seam of the toga, was not applied during the manufacturing process.

Jun 8, 2012 Implants & Prosthetics Nationwide View Details β†’

Biomet has initiated this action following an investigation which identified that the thin shaft reamer may get stuck and not slide into the sleeve.

Nov 1, 2012 Implants & Prosthetics Nationwide View Details β†’

The presence of plastic flash was identified at the tip of certain lots of Sarns Soft-Flow Aortic Cannula. The plastic flash has the potential to detach and contribute to an adverse patient effect. While this potential failure has not been reported from the field as an observed defect, it was determined that the product does not meet specifications and this condition could result in an unacceptabl

Nov 21, 2012 Other Medical Devices Nationwide View Details β†’

On cobas c 701 and cobas c 702 modules of the Cobas 8000 series analyzer, a mismatch between the reagent probe positioning and the aspirated reagent volumes under certain conditions can lead to insufficient reagent pipetting. This may affect patient sample and controls recovery for the following tests: Alkaline Phosphatase (ALP2), Alanine Aminotransferase (ALT), Bicarbonate (CO2-L), Creatinine (C

Nov 1, 2012 Other Medical Devices Nationwide View Details β†’

Biomet has initiated this action following an investigation which identified that the thin shaft reamer may get stuck and not slide into the sleeve.

Nov 1, 2012 Implants & Prosthetics Nationwide View Details β†’

The presence of plastic flash was identified at the tip of certain lots of Sarns Soft-Flow Aortic Cannula. The plastic flash has the potential to detach and contribute to an adverse patient effect. While this potential failure has not been reported from the field as an observed defect, it was determined that the product does not meet specifications and this condition could result in an unacceptabl

Nov 21, 2012 Other Medical Devices Nationwide View Details β†’

Beginning May 24, 2011 Stryker Neptune 2 (N2) Waste Management Systems castors were replaced with a new castor design because investigation found the old design had the potential for the casters to bind compromising the Rover's ability to roll. This can result in additional strain on the operator, due to multiple exposures resulting in back pain.

May 24, 2011 Surgical Instruments Nationwide View Details β†’

The presence of plastic flash was identified at the tip of certain lots of Sarns Soft-Flow Aortic Cannula. The plastic flash has the potential to detach and contribute to an adverse patient effect. While this potential failure has not been reported from the field as an observed defect, it was determined that the product does not meet specifications and this condition could result in an unacceptabl

Nov 21, 2012 Other Medical Devices Nationwide View Details β†’

High-Powered Magnet desk toy

Global Goler Co. Limited, of Guangdong, China

🏠 Consumer Products Class I - Dangerous

When two or more magnets are swallowed, they can link together inside a child's intestines and clamp onto body tissues, causing intestinal obstructions, perforations, sepsis and death. Internal injury from magnets can pose serious lifelong health effects.

Dec 10, 2012 Furniture Nationwide View Details β†’