Presence of Precipitate; white substance confirmed as Guaifenesin, an active ingredient was observed in some bottles. If the product is shaken or warmed the white particles goes into the solution.
All Recalls
95,123 total recalls in our database
Presence of Precipitate; white substance confirmed as Guaifenesin, an active ingredient was observed in some bottles. If the product is shaken or warmed the white particles goes into the solution.
ABILIFY (aripiprazole) Tablets, 30 mg, Rx only, Blister Pack of 100 tablets, Otsuka America Pharmaceutical Inc., Bristol-Myers Squibb NDC 59148-011-35
Bristol Myers Squibb Manufacturing Company
CGMP Deviations: A drum of Abilify 30 mg Tablets rejected during the compression stage was not segregated from the other portion of the lot and was inadvertently shipped, packaged and distributed.
0.9% Sodium Chloride Injection USP, 100 mL Hospira, Inc., Lake Forest, IL 60045 NDC 0409-7984-23,
Hospira
Lack of Assurance of Sterility: There is the potential for solution to leak from the administrative port to the fill tube seal.
Presence of Precipitate; white substance confirmed as Guaifenesin, an active ingredient was observed in some bottles. If the product is shaken or warmed the white particles goes into the solution.
Discoloration: Ethambutol Tablets USP 400 mg have tablets cores that may be discolored.
Chemical Contamination: The recall has been initiated based on multiple complaints received from pharmacists and consumers reporting that they detected an off-odor, described as moldy, musty or fishy in nature which has been identified as trace levels of Tribromoanisole (TBA) and Trichloroanisole (TCA).
Presence of Precipitate; white substance confirmed as Guaifenesin, an active ingredient was observed in some bottles. If the product is shaken or warmed the white particles goes into the solution.
Failed Dissolution Specification: This product recall is due to the product lot # 50077231 exceeding dissolution specifications. All other test specifications were met.
SIEMENS brand ARTISTE syngo RT Therapist (RTT) version 4.3, ARTISTE Model, Part No. 08162815, Digital linear accelerators, MOSAIQ OIS system, Distributed by SIEMENS Healthcare IM CR RO, CR Radiation Oncology, Manufactured by SIEMENS AG, MEDICAL SOLUTIONS, GERMANY.
Siemens Medical Solutions USA
Siemens became aware of a potential safety issue when using the "In-session resumption" feature of the syngo RT Therapist (RTT) 4.3 software. The RTT will send again the previously completed beam or segment for delivery to the control console, thus repeating the already delivered Monitor Units of the same beam or segment.
Products from lot no. 41865 may have been sent out with Incorrect Instructions for Use (IFU) labeling.
Bard Access Systems is recalling two lots of Hickman 9.0 French and one lot of Leonard 10.0 French Dual Lumen Catheters with SureCuff Tissue Ingrowth Cuff with Peel-Apart Percutaneous Introducer Systems due to the wrong size introducer being included with the kit.
The microelectrodes are not flush with the silastic substrate, may contain sharp edges capable of abrading the cortical surface, and may have remaining pieces which may fracture and remain in the cortical tissue with the electrode is explanted. Recall expanded on 5/2/2013 to include additional Macro Micro Subdural Electrodes.
Century Console Base, Model #043625051b. Intended for use as a platform for other ancillary equipment and disposable products in cardiopulmonary surgical procedures.
Salyer Prn Biomedical
Concerns that the addition of the UPS Battery Backup to the Century Console Base Shelf did not comply with the FDA premarket notification regulations.
The Membrane Switch Panel (referred to as "front panel" of the component part number A34-003-207, used on both the D3000 and D3000I Advantage Drive System, contained the incorrect symbology per third edition IEC60606-01 Medical Electrical Equipment: General requirements for basic safety and essential performance.
The Membrane Switch Panel (referred to as "front panel" of the component part number A34-003-207, used on both the D3000 and D3000I Advantage Drive System, contained the incorrect symbology per third edition IEC60606-01 Medical Electrical Equipment: General requirements for basic safety and essential performance.
BD is conducting a recall of the BD Vacutainer Plus Plastic Plasma Separator Tube (PST) with Light Green/BD Homogard Closure due to a deformed top of the tube reservoir reported in two customer complaints.
Synthes Multi-Vector Distractor - Multi-axial Mandible Distractor (Left), Part Number 487.967. For mandibular bone lengthening.
Synthes USA HQ
It was discovered that the sheer pin on the Multi-Vector Distractor may not be made according to product specifications.
Philips Digital Diagnost Stationary radiographic system
Philips Healthcare
When the operator for a wall stand view selects an "image rotation" different from default, or such image rotation is preset in the examination database, then the resulting image may be rotated in the wrong direction. When the operator manually rotates the image back, electronic side markers (if programmed) are rotated also and as a result may be placed in a wrong position inside the image.
Due to a software data storage limitation which only occurs under certain conditions, measurements of the Reagent Blank (RB) or Calibration (CAL) sample results are not saved and the software reverts to a previously stored, acceptable RB/CAL result.