The recall was initiated because Beckman Coulter has identified a software error on AU5800 analyzers with the Automation Ready (AR) Sampler Kit. This software error results in the failure of the concentrated wash solution pump to stop, creating an overflow or leak on each analyzer unit.

Jun 3, 2013 Diagnostic Equipment Nationwide View Details →

Mesa Labs, based out of Bozeman, MT, informed SteriTec Products, Inc. of a product recall involving the EZTest Steam and Smart-Read Biological Indicators (BI) because the readout time is not in specification with the label claims.

Jun 10, 2013 Diagnostic Equipment Nationwide View Details →

Siemens initiate this recall due to a potential issue when using the sDIS matching rules of the syngo Dynamics Information System version 9.5 or VA10A with modality device configured to end study at association close. Images or DICOM SR objects sent from a modality device to syngo Dynamics may not be saved when the modality is configured at syngo Dynamics to end study at association close, and whe

Jun 10, 2013 Infusion Pumps View Details →

Mesa Labs, based out of Bozeman, MT, informed SteriTec Products, Inc. of a product recall involving the EZTest Steam and Smart-Read Biological Indicators (BI) because the readout time is not in specification with the label claims.

Jun 10, 2013 Diagnostic Equipment Nationwide View Details →

Malfunction of slap hammer.

Apr 25, 2013 Surgical Instruments View Details →

In vitro diagnostic test kit may exhibit invalid internal negative control results.

Jun 13, 2013 Diagnostic Equipment Nationwide View Details →

On March 28, 2013, Custom Spine, Inc. issued a minor removal of a lot of 18 screws that were found to be out of specification. The screw length was 5 mm shorter in length than the required specification for the screw size. The product is laser marked 8.5mm X 45mm, but its length measured 40mm.

Mar 28, 2013 Other Medical Devices Nationwide View Details →

A complaint was received from the field, that the drill guide was etched incorrectly, 12mm etch instead of 18mm drill guide. The firm identified that this constituted a field action and retrospectively reported this action.

Oct 12, 2012 Infusion Pumps View Details →

Stryker Osteosynthesis has become aware that LEFT GAMMA3 nails are marked as right nails.

May 21, 2013 Infusion Pumps View Details →

The Instruction for Use 1v9 Ref F and all previous versions specify incorrect dimensions for the Ready To Use (RTU) bottles from other suppliers than Leica. In case of a pressure failure of the device in combination with a bottle with incorrect dimensions, overflow can cause contamination of reagents. Additionally, the current instructions for use does not highlight the importance to check the filling levels of all reagent bottles (RTU and system bottles) prior to every process run. Without this important information, risk of tissue damage or loss is possible. Therefore the missing information will be distributed to all customers by a Field Service Notice.

Nov 13, 2012 Other Medical Devices Nationwide View Details →

GE Healthcare has recently become aware that during use of the Voluson E6 or E8 systems, that an error message Out of Memory can be displayed which could cause the system to freeze. This is an intermittent behavior that also requires a certain workflow. This issue can be resolved by rebooting the system.

Jul 13, 2012 Diagnostic Equipment Nationwide View Details →