CGMP Deviations: Active Pharmaceutical Ingredient (API) used for manufacture was stored in a non-GMP compliant warehouse at S.I.M.S., Italy.

Dec 4, 2014 Other Drugs Nationwide View Details →

Heart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads are not connected to the unit prior to use. The FR3, however, does not require pre-connection and the user may not discover the incompatibility issue until the

Nov 11, 2014 Diagnostic Equipment Nationwide View Details →

The storage conditions for indocyanine green (ICG) are not on the individual Fluorescence Imaging Procedure Kit box, the six-kit box, and the shipping box. If heat were to damage the material, the result would be a loss of fluorescence.

Nov 21, 2014 Diagnostic Equipment Nationwide View Details →

A design deficiency was discovered whereby the Drill/Awl Sleeve and Spring-Loaded Drill/Awl Sleeve instruments (the Sleeves) allow the mating Drill and Awl instruments to penetrate 1.0 mm beyond their intended depths during screw hole preparation.

Nov 11, 2014 Other Medical Devices Nationwide View Details →

A software anomaly was identified where the system may automatically transition from "Standby" to "Ready to Fire" state while in surgical use.

Nov 4, 2014 Surgical Instruments Nationwide View Details →

Getinge Disinfection AB received complaints regarding sediment residuals in the manifold of the washer and disinfector of Getinge 46-Series (CM302) with PACS 300/350 control. After performing detailed root cause analysis, the firm has identified the problem as insufficient water flow throughout the manifold due to formation of air pockets where sedimentation may occur.

Nov 26, 2014 Surgical Instruments Nationwide View Details →

The storage conditions for indocyanine green (ICG) are not on the individual Fluorescence Imaging Procedure Kit box, the six-kit box, and the shipping box. If heat were to damage the material, the result would be a loss of fluorescence.

Nov 21, 2014 Diagnostic Equipment Nationwide View Details →

Heart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads are not connected to the unit prior to use. The FR3, however, does not require pre-connection and the user may not discover the incompatibility issue until the

Nov 11, 2014 Diagnostic Equipment Nationwide View Details →

Two issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition workflow was linked to a wrong geometrical calibration file, resulting in an alignment offset for this specific mode of acquisition. 2. A gantry collision incident required the IBA service team to verify the gantry alignment and protons I x-ray beams colinearity, du

Nov 7, 2014 Other Medical Devices View Details →

Replacement LED Board kit of Natus neoBLUE2 Phototherapy system distributed after February 16, 2012 is subject of a Field Safety Notification because the system will exhibit a higher light intensity output than the original device. While some customers may prefer using these lights at the higher intensity levels, some are concerned that patients may receive more intensity than desired.

Nov 19, 2014 Other Medical Devices View Details →

Maquet has confirmed through a customer complaint that while the oxygenators and the pouches for a single lot of Quadrox-ID Pediatric Diffusion Membrane Oxygenator with Bioliine Coating are labeled correctly with the Bioline coating, the individual cartons and 4-pack shipping containers of this lot incorrectly bear Softline coating labels instead of the bioline coating labels.

Nov 5, 2014 Other Medical Devices View Details →