Software defect that impacts image alignment when using the automatic stitching option for long length images. The defect could cause stitching inconsistencies which may go undetected on a radiograph.

Aug 7, 2015 Diagnostic Equipment Nationwide View Details →

The cannulation of the Concave and Convex Reamer Heads may be too small to allow the Guide Pin to freely pass through.

Jul 20, 2015 Infusion Pumps Nationwide View Details →

Labeling Revision: Additional instructions for the use of peritoneal dialysis transfer sets including the sanitization the connection, healthcare provider monitoring of thyroid function, caution against reuse or reprocessing of the devices, warning against use if tip protectors are not in place, and advisement that the product does not contain natural rubber latex.

Feb 25, 2013 Surgical Instruments Nationwide View Details →

Sterilization tray lids for the Tiger and Tiger Headless Cannulated Screw Systems were reported to occasionally be difficult to remove from the sterilization tray.

Jul 20, 2015 Implants & Prosthetics Nationwide View Details →

Medtronic MiniMed is recalling the Guardian Real-Time Continuous Glucose Monitoring System because it has a language translation error that impacts the Finnish language. Specifically, the Monitor has a Finnish translation error in Predictive Alerts setting: the on-screen Finnish translation for LOW/HIGH is reversed as HIGH/LOW.

Jul 22, 2015 Diagnostic Equipment View Details →

Labeling Revision: Additional instructions for the use of peritoneal dialysis transfer sets including the sanitization the connection, healthcare provider monitoring of thyroid function, caution against reuse or reprocessing of the devices, warning against use if tip protectors are not in place, and advisement that the product does not contain natural rubber latex.

Feb 25, 2013 Other Medical Devices Nationwide View Details →

Low PO2 results. QC failures of the PO2 parameter, affecting primarily Levels 1 and 2, caused by a calibration issue with the PO2 parameter. This issue may not be detected since QC results can be below mean values, but still within 2 standard deviations (SD) limits. Potential for erroneously low PO2 results in patient samples, especially in blood samples with PO2 values below 50 mmHg

Jul 14, 2015 Diagnostic Equipment Nationwide View Details →

Hardy Diagnostics is recalling HardyCHROm MRSA due to a report of the potential for breakthrough of the methicillin-sensitive Staphylococcus aureus (MSSA) QC strain, ATCC 29213 on the identified lot.

Jul 9, 2015 Diagnostic Equipment View Details →

Alcon is initiating a voluntary recall of all LX3 Microscope Floor Stands after receiving reports concerning loss of functionality due to issues with the cabling between the power, host and illumination modules, potentially causing the system to display a system message, and/or resulting in illumination failure.

Aug 11, 2015 Infusion Pumps Nationwide View Details →

Images from the Centricity PACS-IW with Universal Viewer and Centricity Universal Viewer may be missing when a system parameter MapRoute is set to a value greater than 1.

Aug 11, 2015 Diagnostic Equipment Nationwide View Details →

Low PO2 results. QC failures of the PO2 parameter, affecting primarily Levels 1 and 2, caused by a calibration issue with the PO2 parameter. This issue may not be detected since QC results can be below mean values, but still within 2 standard deviations (SD) limits. Potential for erroneously low PO2 results in patient samples, especially in blood samples with PO2 values below 50 mmHg

Jul 14, 2015 Diagnostic Equipment Nationwide View Details →

The cannulation of the 3.0/4.0 Cannulated Drill Bit inner diamter is too small to allow for the guide wire to pass completely through.

Jul 20, 2015 Implants & Prosthetics Nationwide View Details →

The markings on both versions of the OFFSET FEMORAL ALIGNMENT GUIDES are incorrect. Specifically, the markings on the alignment guides are reversed from the intended position that should provide proper alignment for the offset revision stems when used with the femur.

Jun 9, 2015 Implants & Prosthetics Nationwide View Details →