It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
All Recalls
92,900 total recalls in our database
medex TranStar DPT w/Two Stopcocks 10/EA, Product Code REF MX950X2SC
Smiths Medical ASD
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Medtronic Percept RC Implantable Neurostimulators (INS), Model B35300.
Medtronic Neuromodulation
A limited number of Percept RC Implantable Neurostimulators (INS), Model B35300 will not be able to communicate with the A610 DBS Clinician Programmer Application upon use and an "Invalid Device" message will appear.
Baxter Operating Table TruSystem 7000 (dV), Product code REF 1841049
Baxter Healthcare
The firm received customer reports stating that the batteries and their connectors experienced electrical short-circuits and/or emitted smoke.
Medline brand medical procedure convenience kits, labeled as: a) UNIVERSAL DRAPE PACK, REF DYNJ44114; b) DRAPE PACK, REF DYNJ82663B
MEDLINE INDUSTRIES, LP - Northfield
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
medex TranStar 84in (213cm) Single Monitoring Kit 10/EA, Product Code REF MX9505T
Smiths Medical ASD
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
AD-TECH Spencer Probe Depth Electrode, REF SD04R-AP58X-000 4-Contact 1.12mm (A style) Depth Electrode
Ad-Tech Medical Instrument
Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.
medex TranStar 60in (152cm) Single Monitoring Kit 10/EA, Product Code REF MX9504T
Smiths Medical ASD
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Product has been identified as having holes in the packaging.
medex TranStar SINGLE 84IN SET WITH ANTI-SHUNT STOPCOCK 10/EA, Product Code REF MXG300243R1
Smiths Medical ASD
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Baxter Operating table TruSystem 7000 U (MB), Product codes: a) REF 1604786; b) REF 1604788; c) REF 1723633
Baxter Healthcare
The firm received customer reports stating that the batteries and their connectors experienced electrical short-circuits and/or emitted smoke.
Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
Medline IV TEAM PICC, REF DYNJ48222; medical procedure convenience kit
MEDLINE INDUSTRIES, LP - Northfield
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
Medline brand medical procedure convenience kits, labeled as: a) ABDOMINAL HYSTERECTOMY TEGH-LF, REF DYNJ46730; b) HYSTEROSCOPY PACK, REF DYNJ49302A; c) PERI GYN PACK, REF DYNJ57526A; d) HYSTEROSCOPY TRAY ACS, REF DYNJ68750
MEDLINE INDUSTRIES, LP - Northfield
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
CADD Medication Cassette Reservoir, 100mL, FS, Product Code 21-7302-24
Smiths Medical ASD
Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
Baxter Operating Table TruSystem 7000 (MBW), Product code REF 1841048
Baxter Healthcare
The firm received customer reports stating that the batteries and their connectors experienced electrical short-circuits and/or emitted smoke.
CADD Medication Cassette Reservoir, 50mL, FS, Product Code 21-7301-24
Smiths Medical ASD
Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
medex TranStar Single ICP Set 1/EA, Product Code REF MX20897
Smiths Medical ASD
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.