Subpotent Drug: Two of the active sunscreen ingredients, octinoxate and octisalate, are below the specifications for assay.
All Recalls
95,089 total recalls in our database
Subpotent Drug: Two of the active sunscreen ingredients, octinoxate and octisalate, are below the specifications for assay.
Subpotent Drug: Two of the active sunscreen ingredients, octinoxate and octisalate, are below the specifications for assay.
Subpotent Drug: Two of the active sunscreen ingredients, octinoxate and octisalate, are below the specifications for assay.
Subpotent Drug: Two of the active sunscreen ingredients, octinoxate and octisalate, are below the specifications for assay.
Subpotent Drug: Two of the active sunscreen ingredients, octinoxate and octisalate, are below the specifications for assay.
Sterility of device may be compromised due to sterile package breakage.
Beckman Coulter is recalling the UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System to reduce the potential for flammable liquids coming into contact with electrical components in the Stainer module possibly causing a fire.
The Centricity Laboratory removes free-text notes on a master panel when the user deletes the slave panel. The deletion of a slave panel will (auto slash) the corresponding test (item) results on the master panel. Additionally, the displayed and HL7 results show neither the appropriate abnormal flag value nor the reference range values, when using large_num.g >7 to report numeric results.
Software version 1.08 and 1.09 for the Specular Microscope CEM 530 included a change of analysis results feature that was not reviewed and approved by the FDA.
MAQUET Servo Humidifier 163, Inmed Mfg. Sdn. Bhd. Lot PT 2577, Jalan Perusahaan, 4, 34600 Kamunting, Perak, Malaysia.
Teleflex Medical
Cracking may occur at the 15mm and 22mm ISO female connector.
Interventional Radiology (IR) images are stored as JPEG2k Lossless in Merge Enterprise Archive (EA) and are not displaying correctly in RadSuite.
Through an error, the labeling failed to include the symbol or other text indicating that the diamond coated tips were single use only. A contributing factor to this error is that in other markets such as Japan, the diamond coated tips are multiple use and are not limited to single use only. The tips are stainless steel with a coating of fine diamond powder. The tips are sold non-sterile and are to be sterilized prior to first use. The tips were cleared as single use only. This problem was discovered during an ongoing, proactive, quality system improvement plan that was initiated by Nakanishi in October 2014. Because the correct instructions for sterilization are not available, it is possible that the devices might not be sterile prior to use. The absence of the single use statement introduces a risk of cross-contamination if the tips are reprocessed and reused.
INTEGRA Spinal Tray, Catalogue No: 3404875, Contents: 10 Each per Case, Sterile.
Integra LifeSciences Corp. d.b.a. Integra Pain Management
Integra LifeSciences is recalling Spinal Trays, Cat No. 3404875, because of the affected recalled Hospira 5% Lidocaine HCL that is a component of some lots of the Spinal Trays.
Siemens identified an issue with the ratio equation provided on the ADVIA Chemistry XPT System Software Test Definition (TDef) Version 1.0 disks (SMN: 11127343 and 11222123). This issue affects only the ADVIA Chemistry Hemoglobin A1c_3 Automated Pretreatment (A1c_3) results when HbA1c values are reported in International Federation of Clinical Chemistry (IFCC) equivalent units (HbA1cR). The ratio equation used to calculate ADVIA Chemistry Hemoglobin A1c_3 Automated Pretreatment (A1c_3) results in IFCC units contains an error. The error results in falsely depressed HbA1cR results.
Field failures were reported related to the da Vinci Xi Surgical System EndoWrist Stapler 45 and 30 instruments having an interruption in shaft rotation and / or instrument jaws not opening.
Potential for a leak to occur on the back-up O2 and air e-size cylinder yokes on the A7 Anesthesia Delivery System.
PCR Compano Basic Reader using 100-127V as mains voltage. Model numbers: 732027, 732028, and 732040. Radiological image processing system (Computed Radiography).
Philips Electronics North America
A defect component of the Compano reader power supply can cause the power supply to burn and some smoke to leave the housing.
Terumo Custom Cardiovascular Procedure Kits are labeled as Xcoating surface coated tubing, but tubing is non-coated
SMR allen wrench 5 mm, product code 9013.50.210
Limacorporate S.p.A
Product found to be dimensionally non-conforming (slightly over dimensioned on the diameter of the Zimmer connection) and non-functional for connecting with some T-handles.