Aesculap Implant Systems LLC is recalling the Minop Trocar due to the possibility it may have sharp edges on the distal end which may lead to the abrasion of the insulation when removing the electrode.
All Recalls
95,083 total recalls in our database
Volt Resistance Heated Slipper Gray (All sizes) Deep penetrating heat garment and wraps
2k Innovations
During an FDA inspection it was found that the firm claims unapproved therapy and medical claims on product information and website.
For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could cause an end user to use a non-sterile sheath as if it were sterile.
Unapproved change made by the supplier.
During an FDA inspection it was found that the firm claims unapproved therapy and medical claims on product information and website.
PRK Tray, part number DDS1044 Product packaged in a convenient manner for use in a general clinical procedure
Windstone Medical Packaging
AMS custom kits containing Medtronic Covidien Curity Eye Pad are recalled due to the potential for the sterile packaging of the eye pad to be compromised.
Vitrectomy Kit, part number CMP1016(A. Product packaged in a convenient manner for use in a general clinical procedure
Windstone Medical Packaging
AMS custom kits containing Medtronic Covidien Curity Eye Pad are recalled due to the potential for the sterile packaging of the eye pad to be compromised.
Volt Resistance Heated Slipper Black (All sizes) Deep penetrating heat garment and wraps
2k Innovations
During an FDA inspection it was found that the firm claims unapproved therapy and medical claims on product information and website.
Ophthalmic Pack, Part number WAL1107(A and WAL1190 Product packaged in a convenient manner for use in a general clinical procedure
Windstone Medical Packaging
AMS custom kits containing Medtronic Covidien Curity Eye Pad are recalled due to the potential for the sterile packaging of the eye pad to be compromised.
"Roche Diagnostics Corp. initiated a voluntary correction because a rack crash may occur on the Cobas 8100 bi-directional reformatter (BRF) and uni-directional reformatter (URF) modules with software version 02-xx, only when rack buffering is activated and the rack buffer is completely full. This issue can lead to sample spillage, posing a potential risk to operators / laboratory staff due to exposure to potentially infectious material, cross-contamination of samples in the affected racks and erroneous results due to sample carry-over."
GE Healthcare has become aware of an issue where a patient other than the intended is incorrectly selected by the operator in situations where DICOM Worklist search response time is slow. This issue is limited to certain Vivid Ultrasound systems and can result in the incorrect patient information showing on the screen during the exam. If not detected by the operator, images from the actual (intended) patient will be stored under the incorrect patient after the exam. This issue could lead to misdiagnosis. There have been no injuries reported because of this issue.
For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could cause an end user to use a non-sterile sheath as if it were sterile.
During an FDA inspection it was found that the firm claims unapproved therapy and medical claims on product information and website.
Combi Dockable Neurosurgery table for the Siemens MAGNETOM Aera (1.5T) and Skyra (3T) MRI Systems
Siemens Medical Solutions USA
Due to a potential manufacturing error, a rare possibility exists for a sudden drop in table height of the Combi Dockable Table Neurosurgery.
During an FDA inspection it was found that the firm claims unapproved therapy and medical claims on product information and website.
Bedside monitor CSM-1901(Life Scope G9) with main unit CU-192RA. The Problem only affects CSM-1901(Life Scope G9) communicating with CNS-6201A (PU-621RA) and CNS-9701A (MU-971RA).
Nihon Kohden America
The Pause function on central monitors will not automatically resume when connected to a Life Scope G9 patient monitor.
Vivid E95/E90/E80, H45581DC, H45581LB and H45581DA.
GE Healthcare
GE Healthcare has become aware of an issue where a patient other than the intended is incorrectly selected by the operator in situations where DICOM Worklist search response time is slow. This issue is limited to certain Vivid Ultrasound systems and can result in the incorrect patient information showing on the screen during the exam. If not detected by the operator, images from the actual (intended) patient will be stored under the incorrect patient after the exam. This issue could lead to misdiagnosis. There have been no injuries reported because of this issue.
Vitrectomy Pack, part number AMS4952 Product packaged in a convenient manner for use in a general clinical procedure
Windstone Medical Packaging
AMS custom kits containing Medtronic Covidien Curity Eye Pad are recalled due to the potential for the sterile packaging of the eye pad to be compromised.
Boston Scientific Imager II Angiographic Catheter, M001316581 (5-pack outer package UPN), M001316580, single unit, inner package UPN)
Boston Scientific
One lot of Imager II Angiographic Catheters is being recalled due to packaging mislabeling. The product may be labeled as Imager II Contra 2 curve catheters, when in fact the package contains Contra curve catheters.
A defective electrical component in the battery charging circuit in two lots of the ProBP 2400 may potentially cause an over-voltage battery charging condition when the device is connected to the external power supply/battery charger. This potential defect resides with the device, not the battery. The over-voltage condition can result in high NiMH battery temperatures that, in some cases, can reach levels sufficient to cause melting of the plastic battery door and other adjacent plastic and foam device components.