Failed Impurities/Degradation Specifications: Fentanyl-n-Oxide (FNO) degradant exceeding the specification limits
All Recalls
95,075 total recalls in our database
There is a possibility that specific lots of catheters that have recently been distributed may be at risk for an increased potential for kinking at the proximal end where the catheter connects to the drug infusion pump.
Medtronic Custom Perfusion Systems, indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures.
Medtronic Perfusion Systems
Medtronic custom perfusion tubing packs contain Terumo OPS valves that were recalled due to a lack of flow through the valve.
Medtronic is voluntarily recalling two production lots of its Covidien Vistec" x-ray detectable sponges de to possible contamination with human blood caused by handling of the product during the manufacturing process and prior to final packaging and Ethylene Oxide sterilization.
Philips Brilliance 64 computed tomography x-ray system
Philips Medical Systems (Cleveland)
Numerous issues related to software Brilliance iCT 4.1.6 software version.
Ingenuity TF PET/CT - a integrated diagnostic X-ray Computed Tomography (CT) and Positron Emission Tomography (PET) system
Philips Medical Systems (Cleveland)
One of the two wire channels mounted inside the Gantry Separation Unit (GSU) that is also used to secure the top two GSU covers, was missing mounting hardware.
Risk of unintended motion while the AIRO system is in transport mode.
Implant Direct, INTERACTIVE HEALING COLLAR, PART NUMBER 6530-15, 5.0mmL: 3.0mmD Platform
Implant Direct Sybron Manufacturing
InterActive Healing Collar, Lot Number 104203, labeled as sterile with distributed prior to being sterilized.
AArtis Q/ Q.Zen biplane Artis zee biplane Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Siemens Medical Solutions USA
Within a specific number of Artis zee biplane and Artis Q biplane systems, a connection in the ceiling mount consisting of 12 screws may not have been attached and fastened according to quality standards. A quality inspection will be initiated to ensure all screws in the rotation unit meet quality standards.
Safety Deluxe Bone Marrow Biopsy/Aspiration Tray, Catalog # 986 Product Usage: Surgical procedure
Robert Busse & Co. Inc. d.b.a. Busse Hospital Disposables
BD SafetyGlide Needle 22GA 1-1/2 inch may contain the presence of loose polyprophylene foreign matter above release specification.
The design verification for the previous packaging configuration G928 does not cover the 70cm wires. A design verification has been completed to move the 70cm guide wires to a new packaging configuration. As a result, the products packaged in the previous packaging configuration are being removed.
Cardiovascular Systems, Inc. (CSI) is removing the products because it was identified that the labeling for two lots of product s were potentially inadvertently switched during production.
The design verification for the previous packaging configuration G928 does not cover the 70cm wires. A design verification has been completed to move the 70cm guide wires to a new packaging configuration. As a result, the products packaged in the previous packaging configuration are being removed.
When a prior study is being replaced in the workflow step, in certain scenarios (based on the Display Protocol configuration) the prior study will only be replaced in the active Workflow Step and not show in all other workflow steps. The other workflow steps will continue to show the initially loaded study.
PC-1000, PC-1000/Laser1000, Model Numbers 800724-2 & 800724-3
Panoramic Rental
The lift motor used to raise and lower the overhead chassis will wear if not properly maintained according to the maintenance schedule.
PowerPICC Provena Catheters, as 3 Fr. Single lumen and 4 Fr. Dual Lumen. Packaged in a poly tray with Tyvek lid within a header bag configuration or a breather bag configuration.
Bard Access Systems
Possible leaks on the 3 Fr. Single lumen and 4 Fr. Dual lumen PowerPICC catheters at the luer hub extension leg junction.
The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.
The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.
The heat-shrink insulation tube may split during autoclave sterilization/re-sterilization. If a split occurs, the heat-shrink could fall off during a surgical procedure and if cautery is needed, there is a potential for burn due to insulation failure.
Sterile packaging of the disposable light handle cover may contain gaps in the seal of the pouch.